A ring laser is a laser whose optical cavity follows a closed circulating path through a laser gain medium. Mirrors can fold the path into a triangle or another polygon; guided structures can implement the same circulating feedback principle. A travelling optical wave can return to its starting position and direction after one full circuit. The laser threshold requires round-trip gain to balance the complete losses; a closed path alone is not enough to produce laser oscillation.[ref]
Figure 1.
Original schematic of a three-mirror ring laser, a laser gain medium, two possible circulation directions and one output port
. Idealized optical-path diagram rather than a reconstruction of the McNamara experiment. The arrows indicate possible travelling waves; their simultaneous stable lasing is not assumed. Details.
The phase condition explains the allowed normal modes. Neglect dispersion and take a constant round-trip reflection phase . If the circuit's optical path length is , resonance requires
Subtracting the equations for consecutive modes gives the free spectral range
More generally the spacing is approximately the inverse round-trip group delay. A vacuum circuit of full geometric perimeter has a round-trip time about and mode spacing about . There is no additional factor of two: already counts the whole returning route.[ref]
Gain and phase impose different conditions. In an original startup model let the three mirror intensity reflectivities be , and , and let the intensity gain accumulated during one complete circuit be , with other losses neglected. The round-trip intensity multiplier is
The modeled field can grow from a seed. Reducing to gives , so it decays. These are intensity quantities, not amplitude reflectivities. Saturation and competition between normal modes must determine real steady operation.
A reciprocal ideal ring supports possible clockwise and counterclockwise waves. Their simultaneous stable emission is a dynamical question because both can interact with the same laser gain medium. A ring laser gyroscope uses rotation-dependent differences between the two circulation directions through the Sagnac effect; ring geometry by itself does not guarantee a useful rotation measurement.
A different motivating experiment used the directional dynamics as a nonlinear sensor. In 1988 Bruce McNamara, Kurt Wiesenfeld and Rajarshi Roy periodically modulated the asymmetry between counter-rotating lasing modes and added controlled external noise. Their measured signal-to-noise ratio improved for injected noise relative to no external noise, demonstrating stochastic resonance. This observation concerns a particular modulated dynamical response, not a claim that adding noise improves every optical measurement.[ref]
A ring laser gyroscope measures a component of angular velocity from the frequency difference between two counter-circulating waves of a ring laser. Rotation changes their round-trip propagation times through the Sagnac effect. The active optical cavity converts this propagation difference into different resonant frequencies; combining the output waves gives an electrical beat signal through optical heterodyne detection.
For a planar evacuated ideal ring, let be its full perimeter, its oriented vector area, its angular velocity and the nominal vacuum wavelength. At small rotation speed the round-trip time difference has magnitude . With and nominal optical frequency , the resonant-frequency difference has magnitude . Choosing the output sign consistently with the area orientation gives
This is an ideal first-order scale law, with reciprocal paths and resolved counter-circulating modes. Optical dispersion and actual apparatus behaviour require a fuller response model. The underlying circulation-time difference is the one measured by a Sagnac interferometer.[ref]
For an original square ring with sides, and . At and normal-axis rotation ,
Tilting the ring changes the measured projection, not the full rotation magnitude. Reconstructing a general rotation vector requires independent directional information. The kilohertz-scale difference is also distinct from the optical carrier frequency and the cavity's free spectral range.
A historical experiment made the frequency comparison concrete. W. M. Macek and D. T. M. Davis, Jr. reported rotation sensing in 1963 using a continuous-wave helium-neon ring laser on the line. Their metre-square apparatus used four gas tubes with Brewster windows and four external corner mirrors. The first-page diagram directs the two beams toward combining optics and a photocathode, linking the geometric circulation difference to a detectable beat.[ref]
For two coherent output powers , the ideal combined signal includes
One such intensity record supplies a beat-frequency magnitude, not automatically the rotation sign, since changing the sign can be absorbed into the unknown phase. A directional readout needs additional phase information or a suitable reference. Mode coupling, frequency bias and detector response must also be tested rather than treating a closed optical path as a complete calibrated sensor.