In NIST's 2001 comparison, the Laser-Optimized Cryogenic Radiometer held its cavity near and controlled the heat sink separately. Opening the beam reduced heater power at fixed cavity temperature; an AC bridge read the temperature. Tests used stabilized, spatially filtered lasers at , and , with horizontal input and a movable photodiode transfer detector in the cavity's optical plane. The paper contrasts this with the High Accuracy Cryogenic Radiometer's optical heating followed by two electrical heating cycles and interpolation.[ref]
Comparison of Two Cryogenic Radiometers at NIST by Jeanne M. Houston and others. §§2–3, printed pp. 642–643: apparatus, wavelengths, readout and the two substitution protocols
The two laboratories compared calibrations of the same transfer detectors, and checked their spatial response. Alignment differences and transported dust limited that comparison; its agreement did not prove universal equivalence. Its correction tables retain window transmission, cavity radiometric absorptance, alignment, electrical readout and optical-electrical heating non-equivalence. The heat sink was controlled, but §3 does not specify its numerical setpoint: the reference in the preceding example is stipulated.[ref]
Comparison of Two Cryogenic Radiometers at NIST by Jeanne M. Houston and others. §§3–5, printed pp. 643–646, especially Tables 2–3 and spatial-uniformity checks
For a reproducible substitution measurement, record the beam's wavelength and entrance boundary, receiver and reference temperatures, heater terminal voltage and electric current, dark readings before and after illumination, and the settling rule. Check beam collection at the aperture, window transmission, radiometric absorptance, deposited lead heating, and sensitivity to beam position. Repeatability alone misses a stable correction error. Photometric calibration can subsequently use this radiant flux scale, but photometry still needs the appropriate spectral visual weighting; the cold receiver does not itself measure luminous flux.
David J. Livigni Created 2026-10-14 Updated 2026-11-05
David J. Livigni coauthored a comparison of two cryogenic radiometers at NIST. Comparing independently realized optical-power measurements tests consistency and helps identify systematic contributions to measurement uncertainty.[ref]
Comparison of Two Cryogenic Radiometers at NIST by Jeanne M. Houston and others. author list and the described study’s abstract or introductory account
Jeanne M. Houston Created 2026-10-13 Updated 2026-11-05
A researcher in optical metrology who compared two NIST cryogenic radiometers using common photodiode trap detectors. Each radiometer independently calibrated the detectors’ spectral responsivity, allowing consistency to be tested without assuming that either instrument was error-free. The comparison illustrates electrical substitution radiometry: absorbed optical power is related to the electrical heating needed to produce the same thermal response.[ref]
Comparison of Two Cryogenic Radiometers at NIST by Jeanne M. Houston and others. Abstract and §§2–5: trap-detector transfer, substitution protocols and uncertainty comparison