An electrical substitution radiometer measures radiant flux by replacing absorbed optical power with calibrated Joule heating. At fixed temperature, feedback control holds the receiver equally warm with the beam blocked or admitted: after corrections, absorbed optical power equals the reduction in supplied electrical heating. Incident radiant flux additionally requires radiometric absorptance and entrance-path corrections. This connects radiometry to electrical standards.[ref]
Picture a small absorbing cavity connected to a colder reference block by a radiometer thermal link. A heater supplies Joule heating, a thermometer senses the receiver, and feedback control adjusts the heater. When light supplies some of the heating, the controller supplies less. The readout is the change in heater power, rather than the beam's brightness or its number of photons. NIST's institutional description gives a concrete receiver with a conical interior, absorbing coating and weak Kapton support, operated with a receiver heater and thermometer.[ref]
Figure 1. . Original functional schematic, not a drawing of a particular laboratory instrument. The shutter defines dark and illuminated states; incident radiant flux is specified at the receiver entrance. The cavity absorbs a fraction , while the remainder escapes. The reference block and receiver have distinct controlled temperatures. The thermometer signal drives feedback control, and heater voltage and electric current provide the electrical power readout. Lead heating, stray radiation and spatial heating differences require separate checks. Details.
An explicit lumped model makes the cancellation understandable. Treat the receiver as having uniform temperature , heat capacity , heater power deposited in the receiver, net parasitic input , and outward heat transfer to a reference at temperature . With incident radiant flux and radiometric absorptance , conservation of energy gives
The dark state has no admitted beam. Compare settled dark and light states, denoted and ; each has . Subtracting their balances gives
where and . At identical and , with unchanged surroundings and parasitic heating, both correction differences vanish. Thus , and . Notice the signs: extra stray thermal radiation admitted with the beam also reduces heater power, so it must be subtracted from that reduction before attributing the result to the intended beam.
Measure the heater's own terminal voltage and electric current : for steady electrical heating, , where is its electrical resistance. A supply voltage that includes wire voltage drops counts Joule heating in the leads too; only the portion deposited in the receiver belongs in this balance. Changes in that deposited lead heat belong in . Optical-electrical heating non-equivalence adds a further instrument-specific correction when the uniform-temperature model is insufficient.[ref]
A radiometer thermal link is the intentional path for thermal conduction from receiver to reference block. Near a chosen operating point, write with local conductance in . A weaker link increases the steady temperature rise per added power, , but also increases the uncontrolled relaxation time when the link dominates other heat transfer. These are local linear approximations: actual thermal conductivity depends on temperature, and active feedback control introduces its own response. NIST explicitly describes the tradeoff between weaker links and smaller receiver heat capacity.[ref]
Stipulate a model receiver held at with a reference at . Choose , so outward thermal conduction is . Ignore other heat transfer, take in both states, and stipulate perfect optical/electrical thermal equivalence. These values specify an original example; they are not measurements or claimed performance of a NIST instrument.
With the beam blocked, supply . With light admitted, the required heater power is . For stipulated radiometric absorptance at the chosen single wavelength,
The illuminated receiver still gets of heating. The remaining of incident radiant flux escapes without heating it. Calling the measured heater reduction “incident power” would underestimate the stipulated incident radiant flux by .
Figure 1.
The same receiver temperature with different electrical and optical heating
. Original stipulated steady-state example, not experimental data. Both columns total of deposited heating and balance the same outward thermal conduction. Incident radiant flux is , of which is absorbed; the that escapes is outside the heating column. No response time or measured uncertainty is implied. Details.
Two variations expose different pitfalls. First, if admitting the beam also adds of stray heating while the same is absorbed from the intended beam, the heater falls to . The apparent reduction is now ; subtracting recovers the beam's absorbed power. Second, keep the original heater readings but stipulate a light-state receiver offset at fixed reference temperature. The link carries an extra ; the inferred absorbed power becomes , and incident radiant flux becomes . This second variation deliberately violates exact fixed-temperature substitution.
The heater can supply only nonnegative power in this model. Consequently an absorbed beam above the available dark heater power cannot be compensated at the chosen setpoint merely by reducing the heater. The operating point, link, or admitted radiant flux must change; zero heater demand is a range boundary, not evidence of greater radiant flux sensitivity.
Optical-electrical heating non-equivalence means that equal absorbed optical and electrical powers do not necessarily give identical thermometer responses. Light and the heater may deposit heat in different places, producing different temperature gradients and different routes of thermal conduction to the support. Holding one thermometer reading fixed need not hold every part of the cavity at the same temperature. This differs from incomplete radiometric absorptance: one concerns where absorbed heat goes, the other how much incident light becomes absorbed heat.
A useful accounting convention is
where the signed correction is determined by characterization of the instrument and represents the residual departure from the lumped electrical substitution heat balance. It is not a universal extra source of energy. Varying the beam location or heater location, comparing equivalent temperature excursions, and checking repeatable dark returns can test the assumptions; no single small correction proves that every other correction vanishes.
A cryogenic radiometer uses low-temperature operation to improve electrical substitution radiometry through particular material and thermal properties. Lower heat capacity of a copper receiver permits a substantial cavity without proportionately slow relaxation; cold shields reduce unwanted thermal radiation, and heater leads with superconductivity can suppress resistive lead heating. These features do not make radiometric absorptance exactly one or abolish optical-electrical heating non-equivalence.[ref]
Cold surroundings matter as much as a cold receiver. In a simplified gray-surface model, net outward thermal radiation is , using emissivity , area , shield temperature and the Stefan-Boltzmann constant . Cold shields reduce the incoming term as well as the receiver's emission; cooling the receiver alone while leaving it exposed to warm surroundings does not remove incoming radiation. An entrance window still provides an optical and thermal path. Vacuum reduces gas-mediated heat transfer, while an intentional radiometer thermal link remains necessary. This model explains specific losses, rather than promising a universal improvement factor or perfect accuracy.
In NIST's 2001 comparison, the Laser-Optimized Cryogenic Radiometer held its cavity near and controlled the heat sink separately. Opening the beam reduced heater power at fixed cavity temperature; an AC bridge read the temperature. Tests used stabilized, spatially filtered lasers at , and , with horizontal input and a movable photodiode transfer detector in the cavity's optical plane. The paper contrasts this with the High Accuracy Cryogenic Radiometer's optical heating followed by two electrical heating cycles and interpolation.[ref]
The two laboratories compared calibrations of the same transfer detectors, and checked their spatial response. Alignment differences and transported dust limited that comparison; its agreement did not prove universal equivalence. Its correction tables retain window transmission, cavity radiometric absorptance, alignment, electrical readout and optical-electrical heating non-equivalence. The heat sink was controlled, but §3 does not specify its numerical setpoint: the reference in the preceding example is stipulated.[ref]
For a reproducible substitution measurement, record the beam's wavelength and entrance boundary, receiver and reference temperatures, heater terminal voltage and electric current, dark readings before and after illumination, and the settling rule. Check beam collection at the aperture, window transmission, radiometric absorptance, deposited lead heating, and sensitivity to beam position. Repeatability alone misses a stable correction error. Photometric calibration can subsequently use this radiant flux scale, but photometry still needs the appropriate spectral visual weighting; the cold receiver does not itself measure luminous flux.