A certified reference material is a reference material supplied with authoritative documentation assigning specified property values and their associated uncertainties and traceabilities through valid procedures. The certification applies to the stated properties and conditions. It does not make every conceivable property of the material a certified value.[ref]
A usable certificate identifies the material and relevant property, its assigned value, the uncertainty convention, conditions of use and the evidence needed for metrological traceability. Both reference material homogeneity and reference material stability matter when a portion is used later. The VIM distinguishes reference materials suitable for precision control from those with assigned values suitable for measurement calibration or measurement trueness control.[ref]
For a physical example, the archived NIST SRM1922 mineral-oil certificate gives refractive index at and , with expanded uncertainty and coverage factor .[ref] Its corresponding standard uncertainty is .
For an original comparison under those same conditions, suppose a refractometer gives with independent standard uncertainty . The discrepancy is , and its combined standard uncertainty is
Thus the discrepancy is . This calculation is a diagnostic comparison, not an automatic pass/fail rule. Correlated contributions, a different temperature or an inappropriate sample invalidate the simple comparison model. The certificate supplies a documented reference and its limits; the user's complete measurement result still includes the measuring instrument and procedure.
Coverage factor 2026-11-05
A coverage factor multiplies combined standard uncertainty to give expanded uncertainty, . It is dimensionless. Choosing it requires a target coverage and assumptions about the result's distribution and the reliability of the estimated standard deviation.
For a normal distribution with known standard deviation, gives a central interval of about 95 percent probability. With a small sample and an estimated variance, an appropriate Student's t-distribution quantile can be larger. A stated factor therefore communicates a method, not merely an error-bar width.[ref]
Expanded uncertainty 2026-11-05
An expanded uncertainty is , where is the combined standard uncertainty and a stated coverage factor. A result written aims to describe an interval with a specified coverage interpretation.
If a normal approximation is appropriate and the standard uncertainty is well determined, gives approximately 95 percent coverage. This interpretation requires those assumptions; multiplying by two alone does not prove a coverage probability. Report the factor and the basis for its choice.[ref]
Measurement result 2026-11-05
A measurement result comprises the values attributed to a measurand together with the information needed to interpret those values. It commonly takes the form of a measured quantity value and its measurement uncertainty, but the result can contain a fuller probability distribution. A displayed number alone need not supply this information.[ref]
For an original electrical resistance example, report a corrected value at , with standard uncertainty . Specify that the uncertainty is a standard uncertainty, identify the resistor and the temperature condition, and retain the measurement procedure and measurement calibration records. Writing only “100.020” omits both the unit and the meaning of the last digits. Writing without stating what the second number represents is also incomplete.
An alternative report might use expanded uncertainty with coverage factor . This is a different uncertainty convention for the same central estimate; it is not a newly observed change in the resistor. A probability interpretation requires the distributional assumptions accompanying that convention.
An instrument's indication, a corrected measured quantity value and the complete measurement result therefore play different roles. Corrections from the measurement model, the specified measurand, environmental conditions and metrological traceability can all matter when comparing results. Extra displayed digits cannot replace the omitted measurement uncertainty.
Measurement uncertainty Created 2026-10-14 Updated 2026-11-05
Measurement uncertainty describes the spread of values reasonably attributable to a specified measurand, using the available information. It differs from measurement error, which is a difference from a reference value: the uncertainty can be quantified even when that difference is unknown. A useful result therefore states what was measured, the unit of measurement, the value, and how its uncertainty was evaluated.[ref]
For a wire's electrical resistance obtained from , uncertainty in the voltage and electric current propagates through that measurement model. Shared measurement calibration effects can make the inputs correlated, so adding percentages or treating every contribution as independent can be wrong. The measurement uncertainty budget records those assumptions. Report a standard uncertainty when a standard-deviation scale is intended, or an expanded uncertainty with its stated coverage factor; a bare plus-or-minus number does not specify which was used.
Reference material 2026-11-05
A reference material is established as suitable for a specified measurement use because the relevant properties are sufficiently homogeneous and stable. The requirement concerns particular properties and an intended use, rather than every possible attribute of the material.[ref] Reference material homogeneity concerns variation within or between the portions used; reference material stability concerns maintaining the relevant properties over time and under stated conditions.
Reference materials can support measurement precision control even without an assigned quantity value. An assigned value is needed when the material serves in measurement calibration or a check of measurement trueness.[ref] A certified reference material adds authoritative documentation specifying property values with associated uncertainties and traceabilities obtained through valid procedures.[ref]
A concrete physical example is the archived 2009 certificate for National Institute of Standards and Technology SRM 1922, a mineral oil characterized for refractive index. Its characterization used a precision spectrometer and goniometer arrangement with the minimum deviation method, oil in a hollow optical prism, and water-bath temperature control. Measurements covered several wavelengths and temperatures; multiple portions were examined to assess homogeneity.[ref]
The certificate's Table 1 reports, at wavelength 589.3 nm and temperature , refractive index 1.46945 with expanded uncertainty and coverage factor 2. This wavelength's value was obtained through a dispersion fit. The listed temperature coefficient is .[ref]
An original calculation using that linear coefficient predicts
The predicted change, in magnitude, is larger than the quoted expanded uncertainty at the reference conditions. Temperature therefore belongs with the assigned optical value; it is not an incidental label. This calculation illustrates the certificate's model, rather than supplying a new certified value or uncertainty.
Record the material identity, certificate version, property, wavelength, temperature and handling history with a measurement result. The archived certificate supplies a documented example of those links, rather than a claim about a currently available product or the validity of a particular bottle today.