In their 1946 letter, Felix Bloch, William Webster Hansen, and Martin Packard reported water at room temperature. A steady magnetic field along and an oscillating field along produced a signal in a pickup coil along , displayed on an oscilloscope. Some samples included paramagnetic additives to accelerate thermal equilibrium. Resonance fields at two frequencies gave the same field-to-frequency ratio, consistent with the proton value.[ref]
The detailed paper describes glass water bulbs inside tuned receiving coils. Transmitter coils, shields and a conducting paddle controlled leakage. Field modulation at 60 Hz drove the horizontal oscilloscope sweep. A diode and audio amplification supplied the vertical trace. Vibration changed coupling: their largest difficulty. Coil tuning capacitors formed tank circuits. The actual receiver omitted the conceptual diagram's radiofrequency amplifier.[ref]
The orthogonal coil axes suppress direct transmitter pickup, but do not eliminate it. A field-dependent signal at the nuclear resonance must be distinguished from leakage and noise; merely detecting a radiofrequency voltage is insufficient. This is why the field sweep and the independently checked field-to-frequency ratio matter.
Awarded jointly to Felix Bloch and Edward Mills Purcell for new methods of precise nuclear magnetic measurement and related discoveries. The nuclear magnetic resonance article explains their distinct experiments.[ref]
In 1938, Isidor Isaac Rabi, Jerrold Zacharias, Sidney Millman and Polykarp Kusch introduced molecular-beam magnetic resonance. Their 1939 account describes a heated oven, vacuum slits, inhomogeneous A/B magnetic fields with opposite gradients, and a uniform static C magnetic field. Conductors supplied transverse oscillatory radiofrequency (RF) fields. Transitions spoiled refocusing, reducing the tungsten-filament detector signal. Alignment compared A-only and B-only attenuation; C-field sweeps at fixed RF frequency limited oscillator pickup. Matching frequency/field ratios across compounds checked nuclear assignments. Magnetic field calibration limited absolute magnetic moments; weak intervening fields caused unwanted transitions.[ref]
Figure 1.
Functional beam selector: unchanged states refocus; RF transitions cause detector loss
. Original functional diagram of the Rabi molecular beam magnetic resonance experiment; paths are schematic and not to scale. Details.
The readout has a useful logical structure: A and the apertures admit trajectories correlated with a magnetic moment projection; B tests whether that projection has survived C. The two deflectors compensate for unchanged states, whereas a changed projection experiences a different force in B. A dip followed by recovery as the static magnetic field passes through resonance is therefore a population-sensitive signal. A comparison with RF absent or far from resonance is the natural baseline in this functional model; RF pickup, drift or background loss must not be mistaken for a state transition. A dip alone does not establish a pure initial quantum state or resolve an entire time-domain Rabi cycle.
This beam method measures transmitted particles after state-dependent deflection. The later bulk nuclear magnetic resonance methods of Felix Bloch and Edward Mills Purcell instead use a macroscopic sample and detect nuclear induction or RF absorption; NMR spectroscopy builds on that bulk setting. The common resonant coupling does not make their preparations or readouts interchangeable.[ref][ref]