A crystal X-ray spectrometer selects a narrow wavelength interval by diffraction at known interplanar spacing and Bragg angle . For fixed diffraction order , a known lets an angular measurement determine . This reverses the crystallography problem of finding from a known wavelength.
Siegbahn's laboratory developed X-ray tubes and evacuated spectrographs for longer wavelengths. Rotating the crystal selected reflections; photographic plates recorded positions. Opposite-side exposures measured angles on both sides of the direct beam. Later tube instruments mounted a beam-defining slit opposite the plate holder. A separate double instrument compared two reflected beams with two ionization chambers. These were different instruments, not one universal assembly.[ref]
The Nobel biography describes constant-temperature spectrographs and a precision angular scale read by diametrically opposed microscopes.[ref] Temperature control matters because thermal expansion changes ; an accurately read angle cannot compensate for an unknown interplanar spacing.
Figure 1. . Original functional schematic, not a reconstruction of a particular Siegbahn instrument. Two slits illustrate collimation; the selected incident direction, rotatable crystal and movable readout define the geometry. The Bragg angle is measured from the lattice planes; the beam deflection is . The photographic plate and scanning ionization chamber are alternative readouts. Component sizes, angles and positions have no historical scale. Details.
The schematic separates preparation, selection and readout. An X-ray tube target supplies Characteristic X-rays with a possible bremsstrahlung background. A collimator limits incident directions; evacuation reduces air X-ray attenuation. Rotating the crystal brings different wavelengths into reflection. A photographic plate stores line positions over an exposure, whereas an ionization chamber can measure collected electric charge at successive angles. Neither readout alone establishes absolute wavelength: an angular zero, angular-scale calibration, diffraction order, and any required refraction correction enter the inference. Narrower slits reduce accepted angular spread but also reduce signal; a weak line can become harder to locate despite a narrower instrumental response.
A control reported in Siegbahn's lecture photographed a line system with copper-sulphate and calcium-sulphate crystals. Irregular relative shifts around copper's characteristic frequency supported anomalous dispersion; the arrangement excluded differing absorption across that frequency as the explanation.[ref] A finer angular scale alone cannot remove this physical refraction correction.
Precision X-ray emission spectroscopy turns closely spaced spectral lines into quantitative constraints on atomic energy levels. Bragg's law makes a crystal an angular ruler for wavelength: resolving nearby directions can reveal components hidden inside an apparently single line. Accurate line positions and the ability to separate two lines are related, but distinct, experimental achievements.
Karl Manne Georg Siegbahn received the 1924 Nobel Prize in Physics; his lecture was delivered on 11 December 1925.[ref]
The Barkla characteristic X-ray experiment distinguished element-dependent K and L groups through absorption and excitation. Moseley's law related corresponding line frequencies across chemical elements to atomic number. Measuring the components within those groups asks a further question: which distinct atomic energy differences produce the neighbouring spectral lines?
Siegbahn's lecture organizes measured spectral lines into energy levels and resolved double levels.[ref] The payoff is a map of atomic energy differences as well as an element's position in the periodic table.