Electrical substitution radiometer Created 2026-10-14 Updated 2026-10-14
An electrical substitution radiometer measures radiant flux by replacing absorbed optical power with calibrated Joule heating. At fixed temperature, feedback control holds the receiver equally warm with the beam blocked or admitted: after corrections, absorbed optical power equals the reduction in supplied electrical heating. Incident radiant flux additionally requires radiometric absorptance and entrance-path corrections. This connects radiometry to electrical standards.[ref]
Spectroradiometric Detector Measurements: Ultraviolet, Visible, and Near-Infrared Detectors for Spectral Power by Thomas C. Larason and others. §5.2, printed p. 30: electrical substitution and electrical standards
Picture a small absorbing cavity connected to a colder reference block by a radiometer thermal link. A heater supplies Joule heating, a thermometer senses the receiver, and feedback control adjusts the heater. When light supplies some of the heating, the controller supplies less. The readout is the change in heater power, rather than the beam's brightness or its number of photons. NIST's institutional description gives a concrete receiver with a conical interior, absorbing coating and weak Kapton support, operated with a receiver heater and thermometer.[ref]
Pico-Watt ACR by National Institute of Standards and Technology. Description, Design Parameters for Improved ACR Sensitivity; Key Design Features: conical receiver, coating, constant-temperature loop and thin-walled Kapton link
Cavity, radiometer thermal link and heater readout in an electrical substitution radiometer
. Original functional schematic, not a drawing of a particular laboratory instrument. The shutter defines dark and illuminated states; incident radiant flux is specified at the receiver entrance. The cavity absorbs a fraction , while the remainder escapes. The reference block and receiver have distinct controlled temperatures. The thermometer signal drives feedback control, and heater voltage and electric current provide the electrical power readout. Lead heating, stray radiation and spatial heating differences require separate checks. Details. Photometry (optics) 2026-10-13
Photometry weights visible electromagnetic radiation by standardized visual response functions; radiometry measures its energy and power without visual weighting.[ref]
The question is how much light a source supplies under a stated visual weighting. A beam containing at the defining frequency has luminous flux . Moving the same radiant flux to a region of smaller spectral luminous efficacy reduces its luminous flux. Changing the optical spectrum changes this conversion even when its total power stays fixed. This makes photometry useful for comparing lighting, while radiometry remains necessary for an energy budget. The lumen counts luminous flux; the candela additionally specifies its distribution over solid angle.
Two measurement routes implement the weighting: measure an optical spectrum and evaluate a weighted integral, or use a photometer whose spectral responsivity approximates the prescribed weighting. Both require a specified collection geometry. Neither is a complete prediction of an individual's perception of a scene.[ref]
Radiometry Created 2026-10-13 Updated 2026-10-14
Radiometry measures the energy and power carried by electromagnetic radiation, including light. It asks how much radiation a source supplies, where it travels, and how much reaches a receiving surface. The measurements are physical rather than weighted by the sensitivity of the human eye: equal radiant fluxes at different wavelengths need not look equally bright. [ref].
The five quantities below answer different questions about the same radiation. The International System of Units makes the distinctions visible: a watt is a joule per second, while a steradian measures solid angle. [ref]
https://www.nist.gov/pml/special-publication-811/nist-guide-si-chapter-4-two-classes-si-units-and-si-prefixes. NIST Guide to the SI, Chapter 4: derived units
.| Quantity | What is resolved? | Units |
|---|---|---|
| Radiant flux | Total radiation power in the specified beam or collection | |
| Radiant intensity | Power per outgoing solid angle | |
| Radiance | Power per projected area and per solid angle | |
| Irradiance | Incident power per receiving area | |
| Radiant exitance | Outward emitted power per emitting area |
Resolving wavelength or frequency adds a further “per unit interval”, as in spectral radiance. The notation is consistent with NIST's table of spectral quantities. In particular, radiant flux is power, whereas the existing astronomical radiative flux is power per area. [ref]
https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=901354. NISTIR 6657, A Rational Methodology for Estimating the Luminous Flux Based upon Color Primaries from Digital Projection Displays, 2009, Table 1
.This bookkeeping makes an optical camera's collected signal calculable from a source's radiance and the geometry of the receiving aperture. It also lets radiation measurements from different instruments be compared. For example, NIST describes separate calibrations of spectral radiance and spectral irradiance for instruments used to observe Earth. [ref]
https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=104550. Johnson, Brown and Rice, Metrology for remote sensing radiometry, §§3.4.1–3.4.2
. Spectral radiance 2026-10-14
Spectral radiance is radiance per unit frequency or wavelength interval. Its astronomical synonym specific intensity is usually written or ; this is a spectral radiance, distinct from the radiant intensity in radiometry. For positive small intervals,The two expressions refer to the same spectral slice, with the positive widths related through frequency and wavelength . For frequency measured in hertz, , the units are . For wavelength measured in metres, . Integrating over a spectral band gives its radiance, and then multiplying by the collecting projected area and accepted solid angle gives a small uniform beam's radiant flux. A spectral measurement therefore retains information that the band-integrated power alone loses.
For an ideal blackbody, Planck's law gives and at temperature , independent of outgoing direction. Its existing frequency and wavelength peaks of Planck's law illustrate why the coordinate matters. The simpler flat-band calculation below isolates that same change of variables without the thermal physics.