Apical impulse 2026-11-05
The apical impulse is a localized movement of the chest wall associated with the heart's mechanical activity, assessed by inspection or palpation. It is a surface observation rather than the anatomical cardiac apex itself. Its position, area, duration and timing within the cardiac cycle describe different features of the observation.
A primary study compared physical examination with two-dimensional echocardiography in 41 patients. The investigators found that impulse location relative to selected chest landmarks did not reliably identify increased left ventricular end-diastolic volume.[ref] The observation illustrates why an external landmark should be tested against an independent measurement instead of being treated as a direct cavity-volume reading. It does not establish a universal diagnostic rule for all patients.
A simple original measurement model explains one source of ambiguity. Let represent an internal mechanical displacement and let the measured local surface displacement be
The coefficient summarizes an idealized mechanical coupling and viewing direction; represents other contributions. If the same internal motion mm is observed with or and zero background, the surface amplitudes are 1 mm and 0.5 mm. Changing the coupling can change the observed amplitude without changing the internal motion.
Real chest wall mechanics are more complicated than a single gain. The model nevertheless explains why body position, contact and the chosen observation site belong to the measurement description. Neither a surface displacement amplitude nor an apparent lateral position alone specifies ejection fraction or ventricular volume.
Cardiac image foreshortening occurs when a cardiac view represents a shorter apparent base-to-apex extent than the intended anatomical chamber length, often because the imaging plane misses the true cardiac apex. It can create an apparently thick false apex and affect derived ejection fraction and deformation measurements.
A primary echocardiography study acquired optimal and deliberately foreshortened views in 72 participants using a GE Vivid E9 system. The altered acquisition window was one rib-space above the optimal window; two analysis systems were used to compare derived quantities.[ref] This tests acquisition geometry while keeping the participant the same, rather than interpreting every shorter-looking chamber as a biological change.
An original ellipsoid model distinguishes a missed apex from a simple projected length. Let
with the true cardiac long axis along . The central long-axis plane includes both axial tips and has full length . A parallel offset plane instead intersects an ellipse whose apparent axial extent is
At , . A 10 cm true extent therefore appears as 8 cm in this model even though the slice was translated rather than rotated. Its apparent tips are not the original chamber tips.
This model is a geometrical explanation, not a clinical correction formula. A shorter outline can alter measured volume differently at end-diastolic volume and end-systolic volume; the ejection fraction error need not follow the length error by one fixed percentage. Consistent anatomical landmarks and independent views help distinguish actual chamber shape from the chosen section. Improving the repeatability of an outline inside an incorrect plane does not recover anatomy that was never acquired.
Figure 1.
A parallel cardiac slice can miss the true apex
. Original sections of an ideal ellipsoid with axial semiaxis 5 cm and transverse semiaxis 2 cm. The central plane has full length 10 cm; a parallel plane offset by 0.6 times the transverse semiaxis has length 8 cm. Dashed outline shows the central section for comparison. Synthetic geometry, not patient data. Details.
Cardiac short-axis stack 2026-11-05
A cardiac short-axis stack is a series of images across a cardiac chamber at successive positions approximately transverse to its cardiac long axis. In ventricular volumetry, the series must cover the intended cavity from its basal boundary toward the cardiac apex. A primary cardiovascular magnetic resonance imaging protocol specifies linked long-axis planning and short-axis coverage.[ref]
The stack supplies a spatial sum rather than a single diameter. Let be the segmented cavity area in slice at phase , and let be the represented distance along the stack normal. A basic approximation is
All areas in a given sum must refer to a compatible phase of the cardiac cycle. Image segmentation, slice coverage and phase matching are separate inputs to the volume estimate.
An original example uses four representative areas of 2, 6, 10 and 6 cm, each assigned 1 cm of axial extent. Their represented volume is 24 cm, or 24 mL. Omitting the basal 6 cm contribution reduces the result to 18 mL without changing the tracing quality of the other slices. If slice thickness is 0.8 cm but center spacing is 1 cm, the numerical integration convention must also account for the intervening extent; treating thickness and represented spacing as interchangeable changes the estimate.
Phase mixing causes a different error. An area from one slice near end-filling and another near end-ejection do not generally describe a cavity that existed at one time. A stack can therefore be complete spatially yet inconsistent temporally. With correctly matched phases, the derived end-diastolic volume and end-systolic volume can be used for stroke volume and ejection fraction; those formulas do not repair missing anatomy or mismatched acquisition phases.
Dilated cardiomyopathy 2026-11-05
Dilated cardiomyopathy is a form of cardiomyopathy involving enlargement of a ventricular chamber together with impaired pumping by the heart muscle. The left ventricle is a central site of the disorder, and other chambers can also be affected. Enlargement and pumping function are related but distinct observations; chamber size alone does not determine how much blood is expelled.[ref]
A simple volume example makes that distinction quantitative. Let end-diastolic volume be and end-systolic volume be . Then
Here SV is stroke volume and EF is ejection fraction. A stipulated chamber changing from to ejects with EF . Another changing from to ejects with EF . A larger filled volume need not produce a larger stroke volume. These are constructed mechanical examples, not diagnostic cutoffs or reference ranges.
Imaging can test several parts of this mechanism. Echocardiography follows chamber dimensions and contraction; cine magnetic resonance imaging can quantify volumes and longitudinal motion. In a primary study, Marcus Carlsson and colleagues compared twelve patients with dilated cardiomyopathy, twelve controls and twelve elite triathletes. Their analysis combined atrioventricular-plane displacement with anatomical area to investigate the longitudinal contribution to ventricular pumping.[ref]
This experiment illustrates why the condition cannot be represented by one chamber diameter alone: geometry, motion and transported volume supply different information. Doppler echocardiography and its velocity-time integral provide complementary flow observations, while a clinical diagnosis also requires the surrounding evidence and cause of the myocardial abnormality.
Ejection fraction 2026-11-05
Ejection fraction is the fraction of a ventricle's end-diastolic volume lost from its cavity during ejection:
The numerator is volumetric stroke volume; is end-systolic volume. The chamber must be specified: right ventricular ejection fraction concerns the right ventricle, whereas a calculation for the left ventricle concerns its own pair of volumes.
Time-resolved magnetic resonance imaging can estimate the two volumes from a consistently defined cavity. A primary right-ventricular study used cine images and three-dimensional modelling in 120 subjects.[ref] The ejection fraction is a derived quantity, not an additional independent volume observation.
An original example illustrates its scale dependence. Volumes of 120 and 60 mL give and stroke volume 60 mL. Volumes of 240 and 120 mL give the same ejection fraction but twice the stroke volume. Consequently, ejection fraction alone does not determine cardiac output; heart rate and an appropriate forward-flow interpretation are also needed.
Its response to shared measurement errors is instructive. Multiplying both measured volumes by the same positive scale leaves the ejection fraction unchanged. Adding the same offset does not:
For the 120/60 mL example, and mL change the calculated fraction from to , even though the volume difference remains 60 mL. This is an error-model example, not a universal correction for image segmentation. Likewise, cavity volume loss can include backward flow through a heart valve; the formula alone does not establish net useful delivery.
End-diastolic volume 2026-11-05
End-diastolic volume is the volume of blood in a specified ventricle at the end of filling, before the subsequent ejection. It is the starting volume in the usual volumetric stroke volume calculation and the denominator of ejection fraction:
The right ventricle and left ventricle each have their own end-diastolic volume. The chamber, phase and measurement convention must be identified before two values are compared.
For a constructed example, a chamber filling to 120 mL and ending ejection at 50 mL has a stroke volume of 70 mL and ejection fraction about . Another chamber filling to 180 mL and ending at 110 mL has the same 70 mL stroke volume, but ejection fraction about . Equal ejected volumes do not imply equal relative emptying.
The quantity can be obtained from time-resolved images by choosing the filling phase and identifying the complete cavity. A primary right-ventricular magnetic resonance imaging study generated volumes from cine images with three-dimensional modelling; its reference analysis also considered age, sex and body size.[ref] These factors help explain why an isolated volume should not be treated as a universal reference value.
For a simple sum-of-slices model, , where is the measured cavity area in a slice and its represented thickness. Omitting a slice with area 10 cm and represented thickness 0.8 cm omits about 8 mL. This example connects image segmentation and coverage to a concrete volume error; it does not require a spherical ventricle.
End-systolic volume 2026-11-05
End-systolic volume is the volume of blood remaining in a specified ventricle at the end of its ejection phase. It must be associated with a particular chamber and cardiac cycle. Subtracting it from the preceding end-diastolic volume gives the chamber's volumetric stroke volume:
For example, mL and mL give mL. The residual 60 mL is not a separate compartment of permanently trapped blood; it is a chamber-volume measurement at one phase of a recurring cycle.
Measuring end-systolic volume requires both a phase choice and a cavity boundary. A primary magnetic resonance imaging study used cine images and three-dimensional modelling to measure right ventricular volumes and function in 120 healthy subjects.[ref] An image's smallest-looking cavity in one view is not automatically the smallest three-dimensional volume.
An original sensitivity calculation makes the boundary issue tangible. If a selected cavity volume mL has surface area cm, a hypothetical uniform outward boundary shift of cm gives a first-order volume error mL. This geometrical approximation does not prescribe how a real image should be segmented, but shows how many small boundary displacements can accumulate into a volume error.
At fixed end-diastolic volume, increasing end-systolic volume reduces both stroke volume and ejection fraction. That algebra does not identify the physiological cause: changing contraction, loading and valve behavior can affect the observed volumes. A single residual-volume number is therefore a measurement of state, not a complete mechanism.
Marcus Carlsson 2026-11-05
Marcus Carlsson investigates cardiac pumping using magnetic resonance imaging. His 2007 primary study measured longitudinal atrioventricular-plane displacement and related it to stroke volume in twelve controls, twelve elite triathletes and twelve patients with dilated cardiomyopathy.[ref]
The study combined a longitudinal displacement with a short-axis area to estimate the longitudinal contribution to ejected volume.[ref] A shorter displacement need not imply a proportionally smaller contribution when the chamber's transverse dimensions differ.
An original constant-area illustration gives
With stipulated and cm, the modeled contribution is 40 mL. Doubling area to while halving displacement to 0.5 cm gives the same 40 mL. If total stroke volume is 80 mL in both examples, both longitudinal fractions are one half. These are geometric examples, not group values from the study.
This explains why comparing displacement alone and comparing its contribution to stroke volume are different questions. The experimental paper supplies actual images and a specified geometric estimator, while the toy model isolates the multiplication responsible for that distinction. Neither a displacement nor its volume fraction alone determines every aspect of ventricular function; ejection fraction also requires the corresponding chamber volume.
Figure 1.
Marcus Carlsson, named portrait linked from his Karolinska Institutet profile
.
Regurgitant fraction 2026-11-05
The regurgitant fraction is the proportion of a valve-related total ejected blood volume that returns through the leaking valve. For the left ventricular balance used to assess mitral regurgitation, let be total ventricular stroke volume, the matched forward aortic volume and the regurgitant volume. With no other abnormal communication or confounding valve leak,
The imaging recommendations use complementary volume and flow measurements to quantify these quantities.[ref]
An original example takes mL and mL. Then mL and RF is 30%. The quotient instead expresses leakage relative to forward delivery; it is not the same fraction. The denominator is part of the definition.
Equal fractions also need not indicate equal transported amounts. A 30 mL leak out of 100 mL and a 60 mL leak out of 200 mL both have RF 30%. Conversely, an unchanged 30 mL regurgitant volume gives RF 50% if total ejection falls to 60 mL. These stipulated cases isolate the normalization rather than assigning a clinical severity category.
The two measurements in a derived RF may share errors. For , first-order propagation gives
The covariance term should follow the actual acquisition and data reduction, rather than assuming independence because forward and total volume have different names.[ref]
RF is therefore a normalized transport quantity. A peak jet velocity, a valve image and ejection fraction provide other information; none supplies RF without the matching flow or volume model. Specify the valve, flow convention and acquisition interval when comparing measurements.
Serkan Ünlü 2026-11-05
Serkan Ünlü investigates cardiac imaging and the dependence of measured heart function on image acquisition. His official institutional profile identifies cardiology as his unit and lists cardiac and cardiovascular research areas.[ref]
Ünlü is first author of a primary study testing cardiac image foreshortening in echocardiography. The investigators imaged 72 participants using a GE Vivid E9 system and deliberately collected altered apical views from a window one rib-space above the optimal position. They compared ejection fraction and longitudinal deformation measurements using two analysis systems.[ref] This is an experimental acquisition comparison: the view changes while the patient's chamber is not assumed to have become anatomically shorter.
An original geometric example shows why a constant scale error and a phase-dependent distortion have different consequences. Let a modeled chamber length change from to cm. Its fractional shortening is
Multiplying both lengths by 0.9 leaves the ratio unchanged. If instead the measured lengths become 9 and 7 cm because the same 1 cm terminal segment is missed in each phase, the apparent fractional shortening is . Neither model is a complete echocardiography reconstruction; they demonstrate why the error mechanism matters even in a normalized measurement.
The primary study found that deliberately foreshortened views changed measured deformation, with the effect differing by myocardial layer and analysis system.[ref] Ünlü's documented contribution is to test that dependence experimentally. Repeatable software output does not establish that the acquired image represents the intended anatomical view.