The International Temperature Scale of 1990 is a practical scale for reproducible temperature measurements. Its assigned reference points and interpolation procedures approximate thermodynamic temperature. A calibrated standard platinum resistance thermometer transfers this scale over specified intervals; it does not define the modern kelvin.[ref]
A standard platinum resistance thermometer uses very pure, annealed platinum with strain-limiting construction. An industrial platinum resistance thermometer commonly uses supported wire or a film. Their response curves differ.[ref]
The word standard describes the instrument's metrological role. Calibrated SPRTs realize specified interpolation intervals of the International Temperature Scale of 1990 and transfer reference temperatures to other thermometers.[ref] They are primary standards for realizing that practical scale, but their electrical resistance alone does not determine absolute thermodynamic temperature without calibration. Primary thermometry such as the corrected gas thermometer obtains thermodynamic temperature from a physical law. A reference in a comparison bath may be an SPRT; an individually calibrated industrial platinum resistance thermometer can also serve as a working reference with an appropriate uncertainty.
Thermocouple voltage calibration relates an actual assembled sensor's electromotive force to independently established junction temperatures. It must be distinguished from selecting a nominal letter-type conversion curve. In Thermocouples Calibrations Services by National Institute of Standards and Technology (opening method description and bath-calibration table), NIST describes comparison with a maintained reference thermocouple, comparison with a standard platinum resistance thermometer in controlled heat baths, and fixed-point calibration at the freezing points of zinc, aluminum, silver and gold on the International Temperature Scale of 1990. These are independent thermal references, unlike a test that checks only a voltage-input circuit.[ref]
For a concrete electrical check, the SCC-TC guide describes shorting the input after 30 minutes of powered operation and adjusting its offset to a zero-voltage readout. Its separate calibrator procedure requires equal connection temperatures at the calibrator and module, measures the supplied voltage, converts it to temperature, and compares with the calibrator setting. Such an injected-signal test checks the readout chain; it does not expose a real measuring junction to an independently known bath temperature.[ref]
Useful additional checks follow from the model: log the raw voltage, the reference temperature, and the final inferred temperature separately; hold stable while changing and test whether cold-junction compensation preserves the inferred ; compare against an independent thermometer at several stable heat bath settings. Check immersion and repeatability instead of assuming a steady display establishes thermal equilibrium. These are proposed diagnostic checks, not new experimental results. The equal-temperature and polarity controls are already described under Seebeck effect.
Limits arise separately from the thermocouple reference function, sensor-wire changes, thermal contact, terminal temperature differences, and the voltage-input circuit. In particular, the SCC-TC guide warns of connector-to-module temperature gradients and voltage drops caused by the fault-detection electric current flowing through lead electrical resistance.[ref] A fitted curve or compensation algorithm cannot make all these errors vanish. No single accuracy is assigned to every thermocouple.
Triple-point cell 2026-10-14
A triple-point cell turns phase coexistence into a reproducible reference for thermometer calibration. The NIST water-cell demonstration uses a sealed glass vessel containing water of specified isotopic composition and an open central thermometer well. Cooling this well with solid carbon dioxide and alcohol freezes an ice mantle while leaving liquid water and its vapor. The coolant is outside the sealed sample; the sample's vapor space does not communicate with room air.[ref]
In the actual NIST cell-comparison experiment, newly frozen mantles were compared with an aged reference mantle in a controlled maintenance bath. A warm aluminium rod melted a thin inner layer around the well, allowing the mantle to rotate freely and establishing an ice–water interface close to the sensor. A standard platinum resistance thermometer and a resistance bridge measured the electrical resistance ratio to a calibrated, temperature-controlled reference resistor. Readings at two currents permitted extrapolation to zero electrical power, correcting self-heating; measured immersion depths supplied hydrostatic pressure corrections. Comparing preparations over time tested stabilization rather than assuming freshly frozen ice was ready.[ref]
The inference is a reproducible fixed-point response for the thermometer, not a direct reading of three chemical potentials. The reference cell anchors the comparison; a water point alone does not determine a thermometer's response at every other temperature. The ideal argument excludes impurity and isotope variation, finite interface effects and unintended heat transfer; real cells also require adequate immersion and stable electrical references. Repeatability is evidence that these effects are controlled, not a reason to omit their uncertainty.
Since 20 May 2019 the kelvin has been defined through a fixed numerical value of the Boltzmann constant. The thermodynamic temperature of the water triple point is consequently a measured quantity with uncertainty, rather than an exact value imposed by the kelvin definition. Its continued use as an assigned reference on the International Temperature Scale of 1990 is a separate convention.[ref][ref]