Choose two hypothetical environments A and B of one chemical element, with line centers and . Model each line by a normalized Gaussian function with common standard deviation , and stipulate integrated count weights , . On the binding energy axis , the normalized count density is
All energies in this expression use electronvolts. Each component's integrated area equals its count weight, and . There is no background, statistical counting fluctuation, or additional emission feature in this stipulated model. These choices deliberately isolate photoelectron peak overlap and resolution.
Figure 1.
Two chemical components can blend into one maximum
. Original arbitrary normalized data, not a measured spectrum. The chosen Gaussian mixture distribution has centers and and integrated count weights and . Changing the common standard deviation from to changes the number of maxima from two to one, while leaving the centroid at . These count weights are not mole fractions; a chemical abundance inference requires cross section and analyzer-response calibration. Details.
Direct integration, or the law of total expectation applied to the component label, gives the centroid
The centroid lies between the chemical environments and is not either component's binding energy. Expanding within each component, the odd cross term integrates to zero. The resulting variance is
For the variance is and the two maxima are separated. For it is and there is one maximum. Each Gaussian's full width at half maximum is , approximately and respectively. The chemical separation remains in both cases. The number of visible maxima therefore does not equal the number of chemical components.
One interpretation of this change is additional independent Gaussian instrument blur. The convolution of probability densities adds the variances, so broadening to requires response standard deviation . A normalized response with zero mean preserves the centroid and integrated counts. Real line shapes and backgrounds can depart from this model, and separating hidden components is an inverse problem; a plausible fit does not alone establish uniqueness.
Finally, count weights describe detection. In a simple response model, let denote the number of emitting sites sampled in environment , its effective photoionization cross section, and the probability that its emitted Electron contributes to the selected measured line. Then
The common factor includes exposure. Here includes escape from the sample, analyzer transmission and detection. Count weights equal site fractions only if the relevant response factors are equal or corrected. They are not automatically molecular mole fractions; different molecules can supply different numbers of the selected emitting sites. In an original example, equal site numbers with give the plotted count ratio despite a site ratio.
X-ray photoelectron spectroscopy uses X-ray excitation to investigate Electron removal energies, including those of core electrons. The common name electron spectroscopy for chemical analysis emphasizes its use in distinguishing chemical surroundings. The whole recorded spectrum can also contain Electrons produced by processes other than direct photoionization; a peak assignment must identify the emission process as well as the chemical element.
In a modern implementation, a quartz crystal selects aluminum characteristic X-rays according to Bragg's law and focuses them onto the sample. Vacuum reduces gas collisions and contamination; a separately pumped entry chamber limits exposure to air. Lenses collect photoelectrons; an electrostatic electron energy analyzer transmits a narrow kinetic energy band. Stepping that band and counting detector pulses produces the spectrum. An electron flood source can compensate positive electric charge on insulating samples. Ion cleaning can alter vulnerable samples.[ref]
For a scan with count and dwell time at setting , compare rates , rather than raw counts when dwell times differ. Under an independent Poisson distribution model, the count standard deviation is approximately for a sufficiently populated bin. Longer acquisition reduces this relative counting measurement uncertainty; it does not automatically improve separation of overlapping peaks. Photoelectron peak overlap and resolution gives an example in which the same chemical separation produces either two maxima or one.
Figure 1.
Photoelectron energy selection and the analyzer energy reference
. Original functional diagram for an unbiased electrical conductor in electrical contact with an electron energy analyzer. The lower panel uses a common Fermi level; the two vacuum levels differ because the work functions differ. Vertical distances are schematic. The outgoing Electron's total energy is measured above the analyzer's vacuum level to obtain its kinetic energy. This is an explanatory model, not a reconstruction of one historical instrument. Details.