A chemical shift in photoelectron spectroscopy is a difference in binding energy for the same kind of core electron in different chemical surroundings, using comparable energy references. Identifying a chemical element assigns its family of lines; resolving chemical shifts within that family asks how its chemical atoms are bonded. For example, distinguishing a metal from its chemically modified surface requires more than detecting that element in both places.
The removal energy compares the initial electronic state with the state left after emission. Redistribution of the remaining Electrons changes that final state's energy; a measured shift therefore includes relaxation as well as initial bonding effects.[ref]
This makes chemical shift in photoelectron spectroscopy useful but conditional. Compare the same electron shell, the same photoelectron binding energy reference, and compatible sample conditions. Agreement with a chemically appropriate reference supports an assignment; a line position by itself need not uniquely determine an oxidation state. An unresolved sum of two chemical components can even have a centroid where neither component lies. The following Gaussian mixture distribution shows exactly how that happens without claiming a material identification.
Let a conducting sample and conducting analyzer be in good electrical contact, unbiased, at equilibrium, and without appreciable charging during illumination. Their Fermi levels then coincide. Choose that common Fermi level as zero. Define as the binding energy relative to this zero: removing an Electron into the Electron reservoir at the Fermi level costs . It is not the cost of removing the Electron all the way to the sample's vacuum level.
Write the sample and analyzer work functions as and . Their local vacuum levels lie at and above the common Fermi level. Conservation of energy for an Electron emitted without additional losses gives total outgoing energy , where is the Planck constant and the photon frequency. Immediately outside the sample its kinetic energy is . The contact-related electric potential changes this to
Thus the sample work function cancels when the measured kinetic energy is referenced to the analyzer. Substituting for in the final conversion would mix two different energy references. An instrument's calibrated binding energy scale can incorporate its effective analyzer work function, so the correction need not appear as a separate operator-entered number.
As an original numerical example, stipulate , and a line at . The predicted analyzer kinetic energy is . If , the surface value is instead ; the contact-related increase is . A second line at gives at the analyzer. A chemical shift in photoelectron spectroscopy therefore appears as a shift in measured kinetic energy. These are chosen values, not a measured carbon spectrum.
Charging breaks the assumed reference. If a uniformly charging sample rises by electric potential relative to the analyzer, an escaping Electron loses an additional energy , with the positive elementary charge. Applying the unbiased conversion consequently returns
Two features from the same stable potential retain their separation. Features from regions at different potentials acquire different offsets, which can mimic chemical differences or broaden a line. A conducting holder alone does not establish electrical contact through an insulating surface layer. Compare repeat scans, electrical contact and charge-compensation settings before interpreting an offset as a change in chemical bonds.
Measurement calibration is a separate issue: a known reference line tests the scale, while charge compensation controls the specimen's electric potential. In an original offset model , one known line determines . An unknown scale factor in requires two distinct known lines to determine both and . Re-reading a reference after a sample checks drift. Treating an unidentified contamination peak as a universally fixed reference would assume the chemical identity and electrical behavior that the measurement should test.
Kai Siegbahn's early instrument used iron-free magnetic double focusing and a thin-window Geiger-Müller counter. His later instrument, compatible with ultra-high vacuum, used a spherical electrostatic electron energy analyzer and monochromatic aluminum X-rays. Argon mixed with methane supplied simultaneous reference lines.[ref]
Two sulfur lines in sodium thiosulfate distinguished chemical environments within one compound. A separate gold comparison showed a fingerprint obscuring substrate lines.[ref]
For an unknown sample, first document its preparation and transfer, then compare survey and detailed scans under reproducible illumination, contact and acquisition conditions. Repeat a region to test change during exposure. If a cleaning treatment is used, compare before and after rather than assuming that cleaning preserves the original chemical bonds. Report the photoelectron binding energy reference, reference lines, charge-compensation conditions and instrumental settings with the inferred chemical shift in photoelectron spectroscopy. A chemical interpretation should survive these checks as well as explain the line separation.