SEC light-scattering detection adds information from scattering of light to the separation. Static scattering with a concentration readout can infer a weight-average molar mass for material in an eluting slice without the standard-family retention calibration. Dynamic scattering estimates a diffusion coefficient and hence a hydrodynamic radius. “Absolute” here means independent of a retention-to-molar mass standard curve; it still needs calibrated detectors, optical response parameters, and a valid sample model. Coeluting species produce a mixture-weighted result, rather than individually identified molecules.
A concrete manufacturer test used a Bio SEC-3 column, with particles, an Agilent 1260 Infinity G5611A quaternary pump and G5667A autosampler, a diode-array flow cell, and dual-angle/dynamic scattering detectors. A buffer solution at pH flowed at ; column and scattering detector were at . Absorbance at supplied a concentration-sensitive readout. Dilution and loading comparisons tested response: seventeen ovalbumin injections spanning – gave a reported average with relative standard deviation. The note reports poorer separation of thyroglobulin from aggregates and an elevated mass estimate, illustrating the coelution limit. Its sample filtration also limits interpretation of larger protein aggregation products removed before injection.[ref]