Hydrodynamic radius 2026-10-14
The hydrodynamic radius is the radius of a hypothetical no-slip sphere with the same translational diffusion coefficient as the dissolved species. In the dilute continuum regime the Stokes–Einstein relation giveswhere is the Boltzmann constant, the absolute temperature, the dynamic viscosity of the solvent, and the translational diffusion coefficient. This effective radius need not equal the radius of gyration or a geometric molecular radius. It is a useful descriptor for SEC molar mass calibration, rather than a complete specification of how every shape enters every pore.
Polystyrene 2026-10-14
Polystyrene is a polymer used as a calibration-standard family in organic-solvent size-exclusion chromatography. Its hydrodynamic size in a specified solvent supplies a conditional SEC molar mass calibration for other dissolved chains.
SEC molar mass calibration 2026-10-14
SEC molar mass calibration measures standards under the same column, solvent, and temperature conditions as the unknown. Over a suitable fractionation interval, a common empirical approximation iswhere specifies the molar mass unit and has inverse-volume units. Interpolation is conditional on the standards' size–molar mass relationship. SEC measures access associated with hydrodynamic size, so a branched polymer, an extended chain, and a compact protein need not share that relationship.
In this model the , , and standards have hydrodynamic radii , , and and elute at , , and . An unknown at would be assigned by this calibration. But a species that expands from to elutes at and is assigned without gaining any mass. The numbers demonstrate ambiguity, rather than predict a real unfolding transition.
Changes in conformational isomerism, protein folding, protein denaturation, and protein aggregation can therefore shift peaks. Association usually increases size; compaction can reduce size at unchanged molar mass. Adsorption creates a different ambiguity: delayed elution need not indicate a smaller species. A different solvent can change both polymer coil dimensions and interactions with the packing. The SEC protein conformation experiment tests these alternatives with additional detectors.
Size-exclusion chromatography Created 2026-10-13 Updated 2026-10-14
Size-exclusion chromatography separates dissolved molecules by their hydrodynamic size in the chosen solvent. A larger species that cannot enter the packing's pores travels with the liquid between particles and elutes earlier. Smaller species explore more internal liquid and elute later. The column is not a sieve that permanently traps the large species: exclusion makes them leave first. “Gel filtration” is common for aqueous separations; “gel permeation chromatography” is common for organic-solvent analysis of polymers. SEC molar mass calibration translates size-dependent elution into a standard-relative estimate of molar mass, rather than a universal absolute measurement.
This makes SEC useful for separating proteins from small dissolved species, comparing protein aggregation, and characterizing dissolved polymer chains. The residual-group monitoring of polyurethane formation experiment already illustrates polymer calibration: its reported molar masses are relative to polystyrene standards. Size-exclusion characterization of chain polymers explains why an elution trace alone cannot count chains or establish their growth mechanism.
Two-population molar mass example 2026-10-14
Suppose a synthetic polymer population contains three chains of molar mass for every one chain of molar mass . Multiplying both chain counts by any common factor leaves the averages unchanged. Substitution into the existing number-average molar mass and weight-average molar mass definitions givesThe larger chains are only one quarter of the chain count but supply of the mass. Thus a weighted mean that emphasizes mass gives a larger result than one that weights each molecule equally. Total mass divided by total chain amount of substance gives , while is not the mean mass of a randomly chosen chain. Inferring chain counts from size-exclusion chromatography still requires the existing SEC detector response and SEC molar mass calibration assumptions.