Thermopile Created 2026-10-14 Updated 2026-10-18
A thermopile connects multiple thermocouples in series to turn a temperature difference into a larger electrical signal. Absorbed thermal radiation can heat its sensing junctions while reference junctions remain cooler; the signal depends on the thermal and optical design.
A thermopile is an array of thermocouples connected as a series circuit with their electromotive forces adding in the chosen terminal order. It measures a collective temperature difference through the Seebeck effect. An open-circuit electric potential difference requires no externally supplied electric current through the array: a high-input-electrical resistance voltmeter reads the signal. The heat that maintains the junction temperature difference still comes from somewhere, such as absorbed electromagnetic radiation.
In a membrane detector, an absorbing film heats the measuring junctions while reference junctions lie on the surrounding support. Actual junction temperatures can differ across the membrane.[ref]
Thermopile detectors by Hamamatsu Photonics K.K.. p. 2, §2 Structure, Figs. 2-1–2-2: absorbing film, membrane, junction locations and nonuniform temperature
This converts difficult-to-observe heat transfer into an accessible electrical signal; the series connection combines many small contributions without requiring every thermocouple to have a separate meter. Thermopile absorber 2026-10-18
A thermopile absorber converts incident electromagnetic radiation into heat near the thermocouple measuring junctions. Electrically separate junctions can share its temperature through thermal contact. The reference support provides a path for heat transfer and a comparison temperature; “cold” denotes this reference role, even if it is warmer than the absorber.
Distinguish incident power from absorbed power. A window, the illuminated area, and the absorption of the coating determine their relation. A detector's measured voltage sensitivity can depend on where a small illuminated spot lies.[ref]
Thermopile detectors by Hamamatsu Photonics K.K.. p. 3, §3-1 Sensitivity, Eq. (3-1) and following paragraph: uniform illumination versus a localized spot
The model in thermopile thermal response takes absorbed power as its input, so an optical calibration is a separate step. Thermopile series voltage 2026-10-18
Use the A-positive terminal order of Seebeck effect for each thermocouple. Its thermocouple measuring junction has temperature and its reference connections have temperature . Define the difference of Seebeck coefficients as . Connect the B reference end of one thermocouple to the A reference end of the next at a common local temperature, so the terminal electric potential differences add:The first A reference terminal is the positive meter input; the final B reference terminal is the negative input. Following this connected path and adding the individual Seebeck effect integrals proves the result. Matched meter leads with common endpoint temperatures cancel as in an isothermal terminal block. The manufacturer's sum formula also allows unequal individual contributions.[ref]
Thermopile detectors by Hamamatsu Photonics K.K.. p. 2, Eq. (2-1): total output is the sum of individual thermocouple outputs
If all pairs share and the same nearly constant difference of Seebeck coefficients throughout that temperature interval, thenThis is the pair's relative Seebeck coefficient, not either material's individual coefficient or a universal device constant. As an original constructed example, choose , and . Each thermocouple contributes andReversing the meter connections changes the sign. Exchanging and reverses the integrals and hence the electromotive force; this reversible sign response does not make a detector with heat conduction a reversible thermodynamic process. Reversing the wiring of one pair instead gives , because one contribution is subtracted rather than added. For this example the result is , and at equal junction temperatures every contribution vanishes.
The open-circuit condition matters. If the array's total electrical resistance is and the meter input has electrical resistance , the elementary voltage divider model gives . Thus is needed to suppress voltmeter loading. Appreciable electric current also permits Joule heating and the thermoelectric heat transfer to disturb the assumed junction temperatures.
Thermopile thermal response 2026-10-18
A useful original model treats the thermopile absorber as one approximately uniform temperature with heat capacity , connected to a fixed reference temperature by thermal conductance . Both parameters are positive constants here. Let be an added absorbed power above the initial background. Linearize the outgoing heat transfer about that background. With measured from an initial state having , conservation of energy gives the first-order linear differential equationThe left side is the rate of stored energy; the two terms on the right are added power and outgoing heat transfer. The effective thermal conductance can include several paths, rather than just heat conduction along the wires. The model assumes a stable reference, negligible electrical loading, a common absorber temperature, and changes small enough for constant parameters.
For a step from zero to constant at , subtract the steady solution . The difference obeys exponential decay, so the initial condition fixesThe time constant has units . With the same constructed and , choose , and . ThenAt one time constant, the response is of its final value: . Reaching requires . A voltmeter read too soon therefore underestimates the steady signal even when its electrical calibration is perfect.
The steady ratio is in this constructed model. Halving thermal conductance doubles this ratio but also doubles the time constant if heat capacity stays fixed. Reducing heat capacity alone speeds the response without changing its steady value. Increasing gives a larger signal only at fixed individual temperature differences: adding legs can also change thermal conductance, heat capacity and electrical resistance. The algebra identifies coupled design choices, rather than predicting a particular manufactured detector.
A finite electrical resistance produces Thermal electrical noise. For nearly uniform absolute temperature and effective measurement signal bandwidth , its root mean square open-circuit voltage isHere is the Boltzmann constant. Thermal electrical noise in the sensing element is an important contribution, and the signal bandwidth enters its voltage variance.[ref]
Thermopile detectors by Hamamatsu Photonics K.K.. p. 3, §3-2 Noise, Eq. (3-2): element resistance, absolute temperature and measurement bandwidth
If identical pairs have fixed electrical resistance and fixed , then : the signal grows as while this noise grows as . The resulting signal-to-noise ratio grows only as under those assumptions. Reference temperature drift, environmental electromagnetic radiation, and meter noise remain separate limits; reducing signal bandwidth trades fast information for less integrated electrical noise.