The institutional handout Measurement of Emf Using a Potentiometer, Oklahoma Christian University laboratory (Experimental Activity: calibration procedure, measuring a test cell, and measuring internal resistance of a cell) compares unloaded and loaded battery readings. Its apparatus uses a slidewire potentiometer, calibrated resistance coils, a reference cell, a galvanometer, a nominal flashlight cell, a voltmeter, and a switched variable resistor branch with an ammeter. Calibration uses the reference cell's tagged electromotive force; zero galvanometer deflection establishes a null measurement. With the load branch open, the null reading draws no test-cell electric current. The protocol compares newer and older cells using both instruments, then records simultaneous loaded terminal electric potential difference and electric current. It prescribes brief measurements, opening the load switch between readings, and periodic calibration rechecks. These are documented procedures, not run data.[ref]
For the battery internal resistance model, . The handout's analysis estimates electromotive force from the intercept and from the negative slope, then compares the intercept with the independent null reading.[ref]
Additional replication controls are to repeat the open-load reading after the sweep, return to an earlier load setting to detect drift, and record temperature, elapsed time, instrument resolution and measurement error. Include ammeter and contact electrical resistance in the external path; allow for voltmeter loading if a voltmeter replaces the null instrument. The Thévenin measurement experiment gives the corresponding meter-loading correction. Curvature or time-dependent readings can reject the constant- approximation, but a terminal curve alone does not identify which internal transport or electrode process caused the deviation. These extra controls and interpretations are model-based extensions, not reported outcomes of the handout.
An electrical calorimetry experiment asks whether a measured electrical work input accounts for a measured warming. The University of Iowa's demonstration uses an insulated flask containing water, an immersed heater, a thermometer, electrical meters and a timer. Its procedure compares the duration, electrical readings and temperature rise. The following low-power measurement design and synthetic readout develop that idea; they are not measurements reported by the demonstration. [ref]
Use a covered calorimeter with weighed water, an immersed resistor, a calibrated thermometer and gentle stirring. Put an ammeter in series with the heater branch and a high-input-electrical resistance voltmeter across the immersed resistor's terminals. Record , , temperature and room temperature before, during and after switching the heater on. Measure the electric potential difference at the heater, rather than at the supply: lead and contact electrical resistance outside the receiving boundary can consume power that never reaches the water. Account for voltmeter loading if the ammeter also includes the sensing branch's electric current.[ref]
The electrical readout gives operating electrical resistance when , and power . To test Ohm's law, obtain several pairs at the same temperature; one ratio alone cannot establish a linear response. For small electrical resistance, use four-terminal sensing to exclude lead drops and reverse electric current to test stable offsets from the thermoelectric effect. Use brief low-power resistance measurements if the intended result is the unheated value: the test itself causes Joule heating.
For a covered assembly at approximately constant pressure, with negligible evaporation, negligible endpoint mechanical energy changes and no phase transition, take
Here is the water's specific heat capacity at constant pressure, is the participating vessel, heater and sensor heat capacity, is signed outward heat leakage, and is stirring work entering the assembly. The storage term approximates its enthalpy change; for a rigid fixed-volume receiver use the corresponding heat capacity at constant volume. Calibrate independently if the purpose is to test electrical input against thermal storage. Using the same heating run to define and then claiming it verifies the equality would be circular.
Figure 1.
An immersed resistor connects electrical readings to a corrected calorimeter warming
. Original apparatus schematic and synthetic temperature trace, not Joule's original apparatus or a measured dataset. Separate sensing leads indicate the heater-terminal electric potential difference; the ammeter is in the heater branch. Shading marks of input at . The dashed trace suppresses room heat leakage but keeps the same heat capacity. Details.
For the synthetic trace, stipulate , , , , room temperature , and loss coefficient , with negligible stirring input. The heater runs for , beginning at room temperature. The calorimeter heat-loss correction model is
It predicts , , storage and outward loss . Neglecting leakage undercounts deposited energy by about ; assuming no loss predicts . After switching off, the excess temperature undergoes exponential decay with time constant . A cooling slope can estimate only with known and a credible single-temperature model.
Controls separate causes of a temperature rise. Check the electrical meters' zeros and thermometer calibration; keep water mass, immersion, lid and stirring fixed between runs. Record a heater-off stirred baseline to estimate sensor drift and stirring input. Repeat cooling runs over the same temperature range to estimate environmental heat transfer; a baseline taken only at room temperature cannot measure warming-dependent leakage. Vary electric current while keeping and initial thermal conditions comparable: corrected deposited energy should scale as for equal durations. An independent calorimeter calibration or known reference heat capacity makes this a quantitative test rather than a qualitative observation.
Finite sensor response, incomplete thermal equilibrium, changing or heat capacity, evaporation, and imperfect loss corrections limit the inference. A final temperature alone cannot distinguish smaller electrical input from larger heat leakage or a larger participating heat capacity. Agreement tests the stated assembly and its energy accounting; it does not show that all electrical work in arbitrary devices becomes heat.
Historically, the Royal Society's archive description of James Prescott Joule's manuscript dated 13 October 1840 records a coil of wire in a jar of water, a thermometer and a galvanometer, and summarizes his inference that warming depends on electrical resistance and the square of electric current intensity. This electrical experiment precedes Joule's paddle-wheel experiment; the input here is electrical work, whereas the paddle supplies mechanical work. The catalogue, rather than the inaccessible manuscript images, supports this account. [ref]
Electric potential Created 2026-09-24 Updated 2026-10-14
In electrostatics, the electric potential tells us the interaction potential energy per unit test electric charge in the electric field of prescribed sources. Its most directly useful quantity is an electric potential difference: how much work per unit electric charge an external agent must supply to move that test electric charge slowly between two points. This replaces the bookkeeping of a vector force along a path with the difference of two scalar values. The test electric charge must be small enough not to rearrange the sources appreciably.
For motion from to , write . With fixed sources and only the electric force doing work,
Here is interaction potential energy, is kinetic energy, and the line integral uses the electric field produced by the other electric charges. For slow transport with unchanged kinetic energy, the external work is , so for nonzero . This external work has the opposite sign to the work done by the electric field. The SI unit is the volt,
A volt measures energy per unit electric charge, not energy itself. For example, crossing an electric potential difference of changes the potential energy of by , but changes that of by . Both test electric charges encounter the same electric potential difference; their opposite signs reverse the energy change.[ref]
The electrostatic potential is a scalar function of position. The electric field is a conservative vector field, and all three components follow from the negative gradient of the electric potential:
The electric field points toward decreasing electric potential. A positive test electric charge feels electric force in that direction; a negative test electric charge feels electric force in the opposite direction, since . A freely accelerated negative test electric charge therefore moves toward higher electric potential while lowering its potential energy .[ref]
Only differences fix physical electrostatic potentials: adding the same constant everywhere, , changes neither the gradient nor any electric potential difference. Choosing an electrical reference potential is this elementary gauge transformation. Setting zero at infinity is convenient for finitely many localized point charges; it is not available for an ideal uniform electric field extending to infinity. There we can choose a finite reference point. Assigning zero at ground is a convention, whereas physically grounding an electrical conductor also allows exchange of electric charge to impose a boundary value. Zero electric potential at a point never by itself implies zero electric field.[ref]
For a fixed point charge at the origin in vacuum, choose zero at infinity. Coulomb's law gives
where is the vacuum permittivity. Use the stipulated approximation and . At ,
A test electric charge has and radial electric force , directed inward. Moving it inward from to raises the electric potential from to but lowers its potential energy:
This is the Coulomb potential energy calculation using electric potentials. The source's singular self-field at is excluded; when finding the force on a source point charge, use only the electric field of the other sources.[ref]
For a uniform electric field , choose . Then
Moving from to gives . For the change in potential energy is and the electric field does of work. For , the same prescribed displacement gives : the external agent must supply that work for slow transport. This uniform electric field approximates the interior of a parallel-plate capacitor sufficiently far from the edges; the fringing field of a capacitor limits the approximation.[ref]
An equipotential is a level set . On a smooth equipotential with nonzero gradient, every tangent displacement has , so the electric field is normal to the equipotential. Thus moving a test electric charge along an equipotential requires no net electrical work. Around a point charge, the equipotentials are spheres; the circles in the diagram are their sections, not an assumption of two-dimensional Coulomb's law. A uniform electric field has parallel plane equipotentials. For contours separated by the same small electric potential difference, smaller normal separation means a stronger electric field, approximately . At a point where , the electric field is zero and has no direction to call normal.[ref]
The superposition principle adds electric potentials as signed scalars. For finitely many stationary point charges at in vacuum,
For example, the electric dipole with at and at has zero electric potential everywhere in the midplane , yet its electric field at the midpoint is . With and , this is . At , the same model gives and . Cancellation of electric potentials is quite different from cancellation of vector electric fields.
Figure 2.
Computed dipole equipotentials and electric-field directions: zero potential on the vertical midline still permits a nonzero field
. Details.
A motivating measurement is conductive-paper electric potential mapping. A documented teaching experiment uses slightly conductive paper on an insulating board, two circular electrodes in highly conducting ink, a steady supply, and a digital voltmeter. Connect the supply to the two electrodes, keep the voltmeter common lead on the negative electrode, and use its other lead as a movable probe. Adjust the supply using the measured electric potential difference between electrodes; the published procedure uses . The observable is the voltmeter reading at the probe position . Locate several points reading , record their positions on a separate map, and join them into an equipotential; repeat at other readings. Lift the probe between contacts so it does not damage the paper. This procedure makes the electrical reference potential concrete: every reading compares two contacts.[ref]
From neighboring readings infer an electric field component by a finite difference, . Reconstruct two in-plane components with displacements along two independent directions. Alternatively, draw the electric field normal to mapped equipotentials, toward lower electric potential. The second diagram illustrates the same measurement principle with bar electrodes spanning a uniform sheet, insulating top and bottom edges, and a synthetic model . The model readings at and at imply . These values are calculated illustrations, not reported experimental data. The documented circular-electrode experiment similarly estimates the electric field from readings at successive centimetre marks.[ref]
Useful controls include checking the voltmeter zero, checking that the electric potential difference along each painted electrode is negligible, and repeating reference readings to detect supply drift. Reversing the supply should reverse the signs relative to the same reference contact. Contact resistance, leakage electric currents, and touching the sheet can distort the map. A voltmeter has finite input electrical resistance and draws a small electric current; voltmeter loading changes the quantity being measured if that electrical resistance is insufficiently large relative to the sheet's effective source electrical resistance.[ref][ref]
The paper carries a steady electric current, so this is a planar analogue rather than a direct measurement of isolated point charges in three-dimensional vacuum. For constant electrical conductivity, Ohm's law and steady charge conservation give and , hence the two-dimensional Laplace equation away from electrodes. Insulating sheet edges impose zero normal electric current, while finite electrodes fix boundary values. This derivation explains both the analogy and its limits: inhomogeneous paper, finite boundaries, and finite electrodes can change the contours. One should not fit the three-dimensional point charge formula to a planar map without testing that model. Small probe spacing resolves spatial variation better, but amplifies uncertainty in the inferred finite difference. The published laboratory instructions explicitly ask students to assess whether their proposed field fit is appropriate.[ref][ref]
The scalar description has a crucial limit. Faraday's law of induction for a stationary closed path says
where is magnetic flux. Nonzero changing magnetic flux gives nonzero circulation, incompatible with a single-valued electric potential alone. More generally, a scalar electric potential and a magnetic vector potential together give
Consequently, in the presence of induction, two voltmeter lead routes between the same contacts can produce different readings if the enclosed magnetic flux changes. This follows by subtracting the two path line integrals; it is a deduction from Faraday's law of induction, not a claimed outcome of the paper experiment. The steady setup and negligible induction justify interpreting its readings as electric potential differences. Merely introducing a scalar electric potential that depends on time cannot supply a nonzero curl of the electric field.[ref]
A short mathematical justification connects the statements above. Along a smooth path, the chain rule gives . Integrating and using yields the endpoint formula by the fundamental theorem for line integrals, proving path independence. Conversely, vanishing circulation around every closed path lets us define independently of the route, with . Zero curl is a local test; for a continuously differentiable curl-free vector field, a simply connected open region suffices for a global potential. The explicit finite-source formula already gives a global electric potential away from its source points.[ref]
Loaded voltage divider 2026-10-14
Connect a passive electrical load of electrical resistance from the junction to the reference. In the steady direct current model, and share the same electric potential difference and are connected in parallel. Their equivalent electrical resistance is
This is the replacement required when a voltmeter is connected across . The inspected teaching guide explicitly models the meter this way.[ref]
The Kirchhoff's current law proof makes the changed assumption visible:
The upper resistor carries the sum of the lower and load electric currents, so the two original resistors no longer have equal electric currents. In the stipulated example, , the upper current is , and each lower branch carries .
For a digital multimeter used as a voltmeter, let its input electrical resistance on the selected voltage range be . With no other load, set ; if an intended load is also present, use in place of . Two instruments connected simultaneously add two branches. Their ideal indicated value equals the voltage of this altered circuit; voltmeter loading can exist even with perfect measurement calibration. Real readout additionally has measurement uncertainty, offsets, and finite resolution. Consult the actual range and input specifications rather than assigning every digital multimeter the same .
For an oscilloscope, the resistance-only model applies to settled direct current with a compatible input coupling setting. The oscilloscope input and cable also have capacitance; changing signals can therefore have loading that depends on frequency. A probe changes the impedance seen at its tip and the relation between tip voltage and displayed voltage; use the probe-and-input combination's specification and attenuation setting. NJIT's guide models the parallel electrical resistance and capacitance and adjusts probe compensation with a square-wave readout.[ref]
The Field Mapper Kit (PK-9023), Product Manual 012-04346D by PASCO scientific (p. 3, Parallel plate capacitor) proposes a direct investigation of the plate-length-to-gap ratio using conductive-paper electric potential mapping. Its finite straight electrodes represent a cross-section of the plates. The apparatus is a gridded resistive sheet on a corkboard, silver-ink electrodes, pushpin electrical contacts, a – direct supply capable of , and a voltmeter with input electrical resistance at least . The manual specifies waiting 20 minutes for the ink to reach its maximum electrical conductivity, then checking each electrode: its internal voltage drop should be no more than of the applied electric potential difference.[ref]
For the readout, keep one voltmeter lead on the reference electrode and scan the other across the sheet. Mark points having the same reading to trace an equipotential. A second method holds the two probe tips at a fixed separation and pivots them to find the largest electric potential difference; a suggested separation is one centimetre. Avoid printed grid marks when making contact.[ref]
The inference follows from the measured electric potential, rather than assuming straight electric field lines. With probe separation and direction , a local finite difference gives
An arrow toward rising electric potential follows , so it must be reversed to show . To investigate the fringing field of a capacitor, compare contours well inside the gap with contours near and beyond the ends, then repeat at a changed gap while keeping the sheet boundary and plate length fixed. Curving contours and an outside gradient establish nonuniformity in this planar setup. Compare at recorded positions to distinguish simple voltage rescaling from geometric changes; no measured values are supplied here.
Additional useful controls are repeat reference readings to detect drift, reversal of supply polarity, and repeating selected readings with reduced probe spacing. Keep contact quality and sheet edges fixed during the aspect-ratio comparison. Finite probe separation averages the gradient; reducing it improves spatial resolution but magnifies the effect of voltage uncertainty. Voltmeter loading, imperfect ink electrodes, and inhomogeneous sheet electrical conductivity can mimic distortions. These proposed checks supplement the manufacturer's electrode check, rather than claiming a completed experiment.
The sheet carries a steady electric current. For uniform electrical conductivity, Ohm's law and electric charge conservation give the two-dimensional Laplace equation away from electrodes, as explained in conductive-paper electric potential mapping. Uncontacted sheet edges have zero normal electric current, whereas the diagram model fixes zero outer electric potential. These distinct boundary conditions must be matched before a quantitative comparison. The apparatus reveals a planar electric potential map; it does not directly measure the free-space three-dimensional capacitance of metal plates.
Open-circuit voltage 2026-10-14
The terminal electric potential difference with zero external electric current, when a finite value exists. A reading from a finite-input-electrical resistance voltmeter approximates it subject to voltmeter loading.
With two shared leads of electrical resistances , Ohm's law gives
Using the Pt100 local linear model, positive lead electrical resistance appears as a warmer temperature. At true , stipulated leads of each change into , producing : a calculated bias of about . Subtracting a measured loop correction works only while that loop electrical resistance remains appropriate; cable temperature changes can invalidate it.
Figure 1.
Two shared leads include cable voltage drops; four leads sense at the platinum element
. Red paths carry electric current; blue paths connect an ideal high-input-electrical resistance voltmeter. In the two-wire circuit the sensing terminals span both leads and the element. In four-terminal sensing they span the element alone. The stated lead values and temperature bias are illustrative calculations. Source. Details.
Four-terminal sensing separates the electric current and electric potential difference leads. Negligible sensing electric current suppresses sensing-lead drops; force-lead drops lie outside the interval. Real voltmeter loading, connections, and leakage still need checking.[ref]
For a stable offset from the thermoelectric effect, two signed electric potential differences obey and . The current-reversal calculation in four-terminal sensing yields
At stipulated , and , and recover exactly in this model. If the offset or sample temperature changes between readings, cancellation is incomplete. Reversal also leaves Joule heating unchanged because it depends on .
Resistor tolerance Created 2026-10-14 Updated 2026-10-25
Resistor tolerance specifies the permissible manufacturing deviation of a resistor's electrical resistance from its nominal value under stated conditions. For nominal electrical resistance and fractional resistor tolerance , the specified range is . This differs from measurement uncertainty: resistor tolerance describes a component specification, while measurement uncertainty describes knowledge of a measured value; the tolerance alone is not a probability distribution for that value.
Two nominally equal resistors can have different actual electrical resistances, changing their voltage divider ratio even without an external electrical load. In a loaded voltage divider, use the actual electrical resistances to calculate both the unloaded output and the Thévenin resistance governing voltage divider loading error. A mismatch from resistor tolerance and a change from voltmeter loading are distinct effects that can occur together.
Thermopile series voltage 2026-10-18
Use the A-positive terminal order of Seebeck effect for each thermocouple. Its thermocouple measuring junction has temperature and its reference connections have temperature . Define the difference of Seebeck coefficients as . Connect the B reference end of one thermocouple to the A reference end of the next at a common local temperature, so the terminal electric potential differences add:
The first A reference terminal is the positive meter input; the final B reference terminal is the negative input. Following this connected path and adding the individual Seebeck effect integrals proves the result. Matched meter leads with common endpoint temperatures cancel as in an isothermal terminal block. The manufacturer's sum formula also allows unequal individual contributions.[ref]
Figure 1. . Original idealized wiring, showing three of thermocouples. Each A leg rises from reference temperature to absorber temperature , and each B leg returns. B-to-A connections at join adjacent pairs; the thermally shared absorber and support do not short the electrically distinct junctions. A-positive and B-negative meter connections match the Seebeck effect convention. Arrows indicate absorbed electromagnetic radiation and escaping heat, not electric current. Details.
If all pairs share and the same nearly constant difference of Seebeck coefficients throughout that temperature interval, then
This is the pair's relative Seebeck coefficient, not either material's individual coefficient or a universal device constant. As an original constructed example, choose , and . Each thermocouple contributes and
Reversing the meter connections changes the sign. Exchanging and reverses the integrals and hence the electromotive force; this reversible sign response does not make a detector with heat conduction a reversible thermodynamic process. Reversing the wiring of one pair instead gives , because one contribution is subtracted rather than added. For this example the result is , and at equal junction temperatures every contribution vanishes.
The open-circuit condition matters. If the array's total electrical resistance is and the meter input has electrical resistance , the elementary voltage divider model gives . Thus is needed to suppress voltmeter loading. Appreciable electric current also permits Joule heating and the thermoelectric heat transfer to disturb the assumed junction temperatures.
Thévenin load response 2026-10-14
A positive resistive electrical load obeys Ohm's law, . Its resistive load line intersects the Thévenin terminal response at
provided . For the Thévenin divider example, gives and ; gives and . At , the voltage is , half the open-circuit voltage.
Figure 1.
Load-line intersections and load voltage for the 12 V divider
. Left: the Thévenin terminal response falls from at zero delivered electric current to zero voltage at ; three positive resistive load lines select operating points. Right: both the original loaded voltage divider calculation and its equivalent give the same voltage curve, approaching as increases. Calculated values only. Source. Details.
For this positive , increasing reduces the drawn electric current and raises toward . This explains voltmeter loading: a voltmeter itself is an electrical load. The equivalence also applies to a nonlinear electrical load through intersection with its terminal relation, but an arbitrary load need not give a unique operating point. If negative cancels positive , the displayed formula is singular; a real active circuit's limits and dynamics must then be considered.
An authored teaching experiment uses an adjustable steady-voltage supply, a digital multimeter, fixed resistors, and a potentiometer or resistance decade box. Its original circuit uses a supply and , , and resistors; the student compares and electrical loads. The readout is load voltage, recorded against prediction and deviation. The student then removes the electrical load to measure the open-circuit voltage, disconnects the supply and substitutes a wire to measure the Thévenin resistance with the multimeter's resistance mode, and restores the supply for an independent method: adjust the variable electrical load until the voltmeter reads half the unloaded voltage. Finally the student builds the equivalent and repeats both load measurements. Its tables are blank recording forms, not reported observations. [ref]
The half-voltage inference assumes negligible voltmeter loading and positive . A finite voltmeter input electrical resistance actually gives an apparent unloaded reading
When the same voltmeter remains connected across a variable electrical load , the network sees . Setting its reading to , rather than to the true , gives
Thus the uncorrected half-reading method underestimates positive . In the original example, a stipulated gives and half-reading load , instead of . These are model predictions for a deliberately loading meter, not the specification or observed readout of the source experiment's instrument.
Finite readings can also estimate the terminal line without a physical short. For two distinct load settings with corrected delivered electric currents and terminal voltages ,
If the voltmeter remains across each electrical load, the delivered electric current is , including the meter branch. Omitting it biases the inferred slope.
Useful additional controls are to measure each resistor separately before assembly, verify terminal polarity, repeat an earlier load setting to detect supply drift, and compare a third setting with the inferred Thévenin terminal response. These are proposed controls, not reported outcomes of the source experiment. Closely spaced electric currents make the slope estimate sensitive to voltage resolution and measurement error; widely separated settings may trigger source current limiting or Joule heating that changes electrical resistance. Lead and contact electrical resistance, meter input uncertainty and rounding limit the inference. An ohmmeter injects its own test electric current; its reading on a powered network is not the suppressed-network ratio. Measurements with independent sources physically removed do not characterize an active dependent electrical source whose control or supply has also been disconnected.
Southern Illinois University's ET 304A guide, Lab 4: Voltage Measurements and Meter Loading Effects, documents an instructive test: build equal-arm voltage dividers on a supply, then measure across the lower arm with a digital multimeter, analog multimeter, and oscilloscope. Starting with two resistors, repeat with equal pairs from to . It requires separate recorded readings, unloaded predictions, and predictions including each instrument's input electrical resistance. The guide obtains meter resistances from specification sheets and scope resistance from the input-panel marking. Equal ratios control the unloaded target while changing susceptibility to voltmeter loading. The provided tables are blank: the guide specifies a procedure, not measured outcomes.[ref]
Preparation and accuracy checks are made concrete by Berkeley's Lab 1: Introductory Experiments and Linear Circuits I. Its different divider uses nominal and resistors and a supply on a breadboard. Measure isolated resistors first, record actual supply voltage, recalculate with these measured values, then record the junction voltage with the digital multimeter. It explicitly compares actual values with resistor tolerance and the final voltage with a range-dependent accuracy bound from the instrument manual. This is an uncertainty check, not an independent measurement calibration.[ref]
For a controlled comparison, additionally record the selected instrument range, probe factor, and input coupling; connect each instrument separately to the same electrical reference potential, and monitor the input rather than treating the supply label as an exact voltage. These are proposed controls for interpreting the documented procedures. A voltage indication and a resistance indication from the same digital multimeter do not independently validate its measurement calibration; a traceable reference comparison would be needed for that. Resistor tolerance describes component variation, while measurement uncertainty describes what is known about measured values. Using nominal arms alone can confuse mismatch with loading: equal nominal arms each within one-percent tolerance can give unloaded ratios from to .
The inference to test is specific: once actual arm values, input voltage, and meter input electrical resistance are accounted for, the readout should approach the loaded voltage divider prediction within the remaining measurement uncertainty. A discrepancy can instead reflect contacts, changing temperature, source droop, or an input that is not a fixed passive resistor. A source with positive series electrical resistance can be modeled by replacing with only when denotes its ideal internal voltage; when the actual voltage across the divider is measured, that terminal value is already . During settling, or with appreciable input capacitance and changing signals, the steady direct current calculation is insufficient. Neither institutional procedure supplies measured values for the original example above.
Voltmeter loading 2026-10-14
Voltmeter loading is the change in an electric potential difference caused by the electric current drawn by the measuring voltmeter.
For a loaded voltage divider in steady direct current, a voltmeter connected across the lower resistor places its finite input electrical resistance in parallel with the lower arm's electrical resistance and any intended electrical load of electrical resistance . With positive, fixed electrical resistances, an upper resistor of electrical resistance , and an ideal voltage source supplying electric potential difference , Kirchhoff's current law and Ohm's law give
Here is the connected circuit's electric potential difference across the voltmeter; omit if there is no intended electrical load. The parallel connection and need for large input electrical resistance are standard voltmeter operating principles.[ref]
Voltmeter loading is a bias relative to the target electrical circuit with the voltmeter absent and any intended electrical load still present. Voltage divider loading error instead compares the attached-load output with the wholly unloaded output. An accurate readout of , supported by measurement calibration, can still differ from either target: calibration accuracy concerns the indication of the connected electric potential difference, whereas voltmeter loading changes that electric potential difference itself.
A real voltmeter has finite input electrical resistance ; a galvanometer also has a finite winding electrical resistance. Away from balance, its electric current changes both intermediate electric potentials. This is voltmeter loading when the detector is a voltmeter. Model the detector as a passive resistor, with no offset or injected electric current. Thévenin's theorem supplies the full response without pretending the connected bridge remains two independent voltage dividers.
With the detector removed, the equivalent source is . Replace the ideal supply by a wire to calculate the source electrical resistance seen between : nodes merge, making parallel and parallel. Consequently
Since the attenuation factor is strictly positive for , the null remains . A smaller detector electrical resistance reduces the measured electric potential difference off balance but does not shift the ideal null. Its electric current sensitivity instead depends on ; detector gain and resolution also matter in deciding how closely a null can be located.
Figure 1.
Finite detector resistance reduces bridge voltage sensitivity while preserving the null
. Calculated curves use and . All cross zero at ; the inset resolves their different slopes. These curves are a model, not measured data. Source. Details.
In the plotted example, gives and . A detector with instead reads . Using that loaded reading in the unloaded voltage divider formula would underestimate the deviation of from . Balancing the bridge avoids this particular systematic measurement error.
The local derivative makes the sensitivity explicit. For and , the first-order Taylor expansion is
At the plotted null, the open-circuit slope is , compared with for . Increasing the supply improves this ideal sensitivity but also increases Joule heating. A finite source electrical resistance requires recomputing ; the displayed parallel-pair expression assumes an ideal supply.