Current clamp prescribes injected electric current while recording membrane potential. A constant command does not force membrane potential to remain constant: a neuron can produce action potentials or synaptic potentials while that command is unchanged. At zero commanded electric current, the amplifier still measures membrane potential; intrinsic ionic electric currents continue. The name concerns current supplied by the instrument, not the cancellation of all currents in the biological cell.
In whole-cell recording, the measured pipette voltage during a current pulse includes the patch-clamp series resistance drop. A modern authored methods article explicitly describes current-pulse protocols and this distortion.[ref] Check the access contribution with small pulses and distinguish the immediate pipette-voltage jump from the slower membrane capacitance response. A zero-input interval measures a baseline, not proof that the instrument has no voltage offset. The passive example below explains these distinctions without asserting any new physiological measurements.
Model a spatially uniform cell membrane as a capacitor of membrane capacitance in parallel with membrane electrical resistance and a fixed leak reversal potential . Exclude active ion channels, seal leak current and pipette capacitance. For electric current injected into the biological cell, electric charge conservation yields
Starting at and applying constant gives the ordinary differential equation solution
The steady state rise is ; the time constant is . With , , and , the rise is and . At one time constant the rise is . The figure's uncompensated pipette voltage adds for . The membrane voltage is continuous at step onset in this model, while the pipette voltage immediately acquires that access drop.
Checking the units is instructive: electrical resistance times capacitance has units , and electric current times electrical resistance has units of voltage. Unlike the passive voltage clamp transient under patch-clamp series resistance, this ideal current clamp membrane time constant is set by , not . Strong current steps can recruit voltage-gated ion channels; the passive model then fails and cannot generate an action potential. Small pulses of both signs, stable access and repeated baseline measurements help test whether the passive approximation is appropriate.