Derek Jones (magnetic resonance researcher) 2026-11-05
Derek Jones is a physicist at Cardiff University whose research develops quantitative magnetic resonance imaging, especially diffusion tensor imaging. He trained in physics at Nottingham, medical physics at Leeds and completed a Leicester doctorate on diffusion tensor imaging. His work asks how acquisition design changes what can be inferred from a measured signal.[ref]
Derek Jones profile at Cardiff University. Overview, Research and Biography: research focus and1993/1995/1998 education
Derek Jones, from his labeled Cardiff University staff-profile portrait
. Jones investigated how diffusion MRI gradient sampling affects the measurement uncertainty of inferred diffusion tensors. In 2004, Jones used Monte Carlo simulations to compare sampling schemes at a diffusion weighting b-value of , tracking tensor orientation, trace and fractional anisotropy. It showed why a minimal set of directions need not give robust estimates; its numerical findings belong to the tested models and noise conditions.[ref]
The effect of gradient sampling schemes on measures derived from diffusion tensor MRI: A Monte Carlo study by Derek K. Jones. Original abstract: controlled b-value, tensor-derived outcomes and comparison of angular sampling schemes
An earlier conference report made the comparison with a fixed observation budget: one baseline and 60 diffusion-weighted observations. Six unique orientations repeated 10 times and 30 orientations repeated twice therefore used the same number of diffusion-weighted observations. Varying angular coverage while controlling that budget separates a direction-design question from simply acquiring more data.[ref]
The Effect of Gradient Sampling Scheme on Estimates of Fiber Orientation: Implications for Fiber Tractography by D. K. Jones. Methods and Table1:61 total rows, unique orientations and repetition counts
A noise-free calculation illustrates diffusion tensor identifiability. With separately known baseline , fixed and a unit direction , the Gaussian diffusion tensor model givesThe symmetric matrix has six independent coefficients. Its quadratic form produces a row in a linear least squares design. Directions give six observations but only matrix rank three: opposite directions repeat a row and the off-diagonal coefficients remain undetermined.
Adding the normalized mixed directions , and to the three positive coordinate axes instead gives matrix rank six. For example,This establishes uniqueness in the stated noise-free model. It does not establish small measurement uncertainty or an optimal experimental design.
Original noise-free diffusion-tensor design matrices: six antipodal directions give rank three, while axes plus three mixed directions give rank six
. Details. Jones also interviewed John E. Tanner about the physical origins of pulsed field gradient NMR. The interview preserved concrete details of the apparatus, including Tanner's carefully spaced mica stack and the practical difficulties of rapidly switching magnetic field gradients.[ref]
John Tanner, the return to Germelshausen, and the work that nearly never happened by Derek Jones. pp.6–7: restricted-diffusion apparatus and switching problems
Edward O. Stejskal 2026-11-05
Edward O. Stejskal was a physical chemistry researcher known for nuclear magnetic resonance methods. His Illinois doctoral work used nuclear spin relaxation to investigate molecular motion in solids. At Wisconsin he developed pulsed field gradient NMR with his first doctoral student, John E. Tanner; later he worked at Monsanto and became a professor at North Carolina State University.[ref]
Obituary: Edward Stejskal by the University of Illinois Department of Chemistry. Graduate research, Wisconsin collaboration, Monsanto and NC State appointments
Their 1965 apparatus measured proton spin echoes at using opposed gradient coils. Keeping the applied magnetic field gradient small during the radio frequency pulses and echo detection removed two limitations of continuously strong gradients: difficult excitation across a broad resonance and an echo too narrow to measure easily. They tested the method on dry glycerol and reported a diffusion coefficient of at . This is their reported measurement, rather than a general reference value for all glycerol samples.[ref]
Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient by E. O. Stejskal and J. E. Tanner. pp.288–289: measurement motivation; pp.291–292: 20MHz apparatus and glycerol result
The Stejskal-Tanner equation gives a useful ideal prediction. For unrestricted diffusion, two equal rectangular gradient pulses of duration and magnitude , with starting times separated by , giveHere is the diffusion coefficient, the diffusion weighting b-value, and the angular gyromagnetic ratio; is the otherwise matched echo without the applied diffusion-encoding gradients. An intervening spin echo pulse reverses their effective encoding sign, although the two physical gradient pulses have the same sign. The accumulated encoding returns to zero for stationary atomic nuclei, while random displacements attenuate the ensemble signal.[ref]
Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient by E. O. Stejskal and J. E. Tanner. pp.289–290: Theory and rectangular-gradient case
An original idealized gradient-pair diagram: identical physical pulses, refocusing of the stationary position tag, and the time integral of squared encoding that determines diffusion attenuation
. Details. The finite-duration correction can be checked directly. In this ideal diagram, rises linearly to , remains there for , then falls linearly to zero. Each ramp contributes to , giving . This integral explains why replacing the pulses by instantaneous position tags loses a correction when is appreciable.
At Monsanto, Stejskal and Jacob Schaefer combined cross-polarization (nuclear magnetic resonance) with magic angle spinning to obtain high-resolution solid-state nuclear magnetic resonance. In cross-polarization magic-angle spinning, polarization transfer strengthens weak nuclear responses while sample rotation averages appropriate anisotropic interactions; the two operations address different experimental obstacles.[ref]
Obituary: Edward Stejskal by the University of Illinois Department of Chemistry. Monsanto collaboration with Jacob Schaefer and combined cross-polarization/magic-angle spinning
Restricted diffusion 2026-11-05
Restricted diffusion is diffusion whose displacement statistics are constrained by boundaries that particles cannot cross. A reflecting boundary condition models an impermeable wall: particles continue moving, but there is no normal diffusive flux through it. Finite permeability would require a different boundary model.
A physical demonstration uses a liquid confined between solid lamellae. In his first-person account, John E. Tanner described constructing 99 mica layers separated by aluminium-foil strips, removing wrinkles and burrs to obtain regular spacing; Edward O. Stejskal inspected the stack with a stereomicroscope. Pulsed magnetic-field gradients encode displacements in the nuclear magnetic resonance signal, allowing motion normal to the barriers to be compared with less confined directions.[ref]
John Tanner, the return to Germelshausen, and the work that nearly never happened by Derek Jones. Printed p.6, mica-stack account and Figure3
This is a material confinement experiment, not a change in the liquid's chemistry.For an original one-dimensional model, let particles diffuse in with constant diffusion coefficient :The wall conditions are homogeneous Neumann boundary conditions for the diffusion equation. Start in equilibrium, so has the uniform distribution on . Its variance isAt long time separation, the positions decorrelate. Consequently the mean squared displacement approachesAt short times relative to , most displacements have not encountered a wall, and the leading scale is the free-diffusion result . For a hypothetical gap , the long-time plateau is and its root is about .
Equilibrium restricted diffusion saturates in a reflecting interval
. Original Neumann-eigenmode calculation with uniform initial positions; time is and mean squared displacement is divided by . Perfectly reflecting walls give a plateau , while free diffusion continues as . These are model curves, not measurements of the mica stack. Details.The initial condition matters: fixing instead gives the long-time value . Likewise, echo attenuation in a confined sample must not automatically be converted into a free-diffusion coefficient. Confinement geometry, timing and the selected boundary model all affect the inference.



