Bourdon tube 2026-11-05
A Bourdon tube is a curved, hollow elastic element that converts a pressure difference into movement of its free end. A common form has an oval tube cross-section and a roughly C-shaped centreline. Fluid pressure is applied through the fixed end, while the other end is closed and free to move. WIKA describes the cross-section tending toward a more circular shape under internal pressure and the resulting end motion being transmitted to a pointer.[ref]
In a Bourdon tube pressure gauge, the tube, linkage, gearing and pointer form a mechanical measuring chain. The sensing element undergoes elastic deformation; the movement converts a small displacement into a readable measurement indication. A steady pressure reading does not require continuous flow through the closed tube.
For an original small-range model, let free-end displacement be , with gauge pressure and stipulated . Let the linkage produce pointer angle with . At ,
The combined response is . These are illustrative coefficients, not a specification for a WIKA instrument. The actual response depends on tube geometry, material and movement.
The reference pressure belongs in the interpretation. With the outside of the element exposed to local atmosphere, the response measures gauge pressure, rather than absolute pressure. WIKA separately describes sealed reference arrangements for absolute-pressure instruments.[ref]
For instrument calibration, connect the gauge and a suitable measurement standard to a common pressure source, allow equilibration, and compare several increasing and decreasing pressures within the intended range. Record the relative height of fluid-filled pressure ports because hydrostatic pressure can add a difference between them. Differences between increasing and decreasing readings can reveal hysteresis or backlash; a zero shift after loading can reveal a changed mechanical state. The original linear model alone does not establish the gauge's measurement uncertainty.
A Bourdon tube pressure gauge combines a pressure-sensitive Bourdon tube with a motion-transmission mechanism, pointer and calibrated scale. Pressure-driven elastic deformation moves the free end of the tube; a linkage and geared movement convert that motion into an indication on the dial. WIKA's technical diagram identifies the tube, end piece, link, movement, pointer and pressure connection as distinct parts.[ref]
The complete gauge is a measuring system. Its measurement indication depends on the sensing tube and the motion transmission, including possible backlash and hysteresis. For a vented case, the tube responds to pressure relative to the surrounding atmosphere, so the intended quantity is gauge pressure. The dial alone does not define an absolute pressure reference.
For an original instrument calibration comparison, connect the gauge and a suitable pressure reference to a common fluid source. Apply several increasing pressure steps, wait for stable readings, and return through the same pressures. Record each discrepancy and distinguish the two loading directions. This tests the assembled readout rather than only the elastic element.
The locations of the pressure ports can matter as much as their shared supply. If the test gauge's port is below the reference port in a stationary water-filled connection, take and . Hydrostatic pressure gives
Without that correction, a correct lower gauge can appear to overread the upper reference. These are original apparatus values, not manufacturer test data.
Read the error specification for the actual model and size. WIKA's June2025 US sheet, for example, states an accuracy of of span for the specified four- and six-inch versions, while listing a different specification for the smaller version.[ref] A percentage of span is not the same percentage of the instantaneous reading. Retain range, conditions, corrections and measurement uncertainty with the reported result.
Caliper 2026-11-05
A caliper measures a dimension by relating the separation of contacting measuring faces to a scale or readout. A sliding-jaw caliper has a fixed jaw and a movable jaw; appropriate faces can measure outside dimensions, inside dimensions or steps, while a depth bar can supply a separate depth measurement. These modes use different contact geometries even when they share one slider.[ref]
A Vernier scale, dial or electronic readout converts slider position into a measurement indication. It does not remove the need to align the contacts with the desired length. For an outside measurement, the opposed faces must meet the intended surfaces without dirt, excessive force or tilt. An inside measurement uses its designated faces and can require geometry-specific corrections; it is not merely the outside-jaw reading with a different label.
An original small-angle calculation shows why jaw geometry matters. Suppose the displacement reference and contact point are separated by , and a moving-jaw tilt produces angle . The resulting offset contribution has magnitude approximately
It exceeds a reading increment. These are stipulated dimensions and tilt, not a measured error for a particular caliper. The primary guide gives the same lever-arm mechanism for caliper jaw tilt error and identifies contact force, wear and thermal differences as physical error sources.[ref]
A useful check closes clean outside faces under consistent light force, then compares readings at several separations with suitable calibrated artefacts such as gauge blocks. Repeating the comparison near different contact positions can reveal a geometry-dependent error that a single zero check misses. Apply a Vernier zero correction when using that readout, while retaining uncertainty from the reference, alignment and repeatability.
The measurement result concerns the specified object dimension under stated conditions, not just slider displacement. Thermal expansion of either the object or instrument, deformation under contact force and the selected faces belong in the measurement model. More displayed digits or a finer scale cannot by themselves establish better measurement accuracy.
Double substitution weighing determines the difference between an unknown mass and a reference through repeated substitutions on the same comparator, with a small known sensitivity mass used to determine response under load. NIST's SOP4 compares standard and unknown twice, rather than relying on the balance to establish their entire masses from its zero.[ref]
For the source's sequence A, let S denote the standard, X the unknown and sw the added sensitivity mass. With optional tare masses omitted, the four pan loads are
producing measurement indications . The standard remains the reference; the small addition probes transducer sensitivity close to the actual working load.[ref]
An original simplified model neglects buoyancy, tare differences and drift, and assumes a common linear response coefficient . Write and for the added mass. Relative to the standard load, the readings are
The two comparison differences estimate the same quantity; averaging them gives . The sensitivity step is , so
For original and readings , this gives . The offset and the unknown response coefficient cancel under the model.
Practical weighing records timing and environmental conditions, keeps settings fixed through the comparison, and controls handling and equilibration. The full NIST procedure supplies buoyancy corrections, uncertainty contributions and a comparison using a check standard.[ref] For equally spaced readings, a linear additive drift cancels from , but can still affect the sensitivity step . Neither the simplified equation nor the method's name guarantees immunity to drift, contamination, nonlinear response or uncertainty in the measurement standard.
Gain ranging 2026-11-05
Gain ranging changes the sensitivity of a measuring signal path to keep its output within a useful interval as the input changes. A high-sensitivity setting resolves smaller inputs; a lower-sensitivity setting accommodates larger ones before amplifier clipping. Combining the settings can extend the instrument's usable dynamic range, provided their input intervals overlap and their measurement calibration is consistent.
A concrete apparatus is a switchable transimpedance amplifier converting electric current into an output electric potential difference. The DLPCA-200 manufacturer diagram separates a current-to-voltage stage from programmable amplification, and its performance tables give gain-dependent noise, input limits and bandwidth. The settings therefore change more than the decimal point of a display.[ref]
For an original ideal calculation, suppose two settings follow , where is input current, the offset and zero-mean output noise. Stipulate a usable offset-corrected output from to in either setting. With , the input interval is to . With , it is to . Each setting spans a ratio 200, but their overlapping union spans a current ratio 20000. These are stipulated intervals, not specifications for the cited amplifier.
Range-overlap calibration checks that both settings estimate the same current in their common – interval. A range-switching hysteresis rule can switch down in sensitivity above and back up below , reducing repeated switching near one boundary. The output must settle before it is accepted as a new measurement indication.
An analog-to-digital converter that merely changes its printed units cannot recover an already clipped upstream signal. Likewise, increasing amplification does not automatically improve input-referred noise; the new setting may have different noise sources, bandwidth and drift. A meaningful combined range includes these effects and the required measurement uncertainty, rather than multiplying nominal gain settings without checking their usable intervals.
Measured quantity value 2026-11-05
A measured quantity value is a value used to represent a measurement result. It may be obtained from one observation or calculated from replicate observations, for example as an arithmetic mean. The VIM distinguishes this value from a raw measurement indication and from the fuller result with its measurement uncertainty.[ref]
For an original electrical resistance example, suppose three instrument indications are , and . A measurement calibration gives additive indication offset , so the corrected values are , and . Their arithmetic mean is the representative measured quantity value
This value is a numerical estimate of the specified measurand, such as the resistor's resistance at a stated temperature. It is not established as an exact true value merely by correcting the indications.
The three values have sample standard deviation . If repeat deviations are independent, their contribution to the mean's standard uncertainty is . Suppose the common correction has independent standard uncertainty . First-order propagation of uncertainty gives
Repeating the readings reduces the first term under these assumptions; it does not repeatedly remeasure and average away the same common correction. These are illustrative uncertainty inputs, not observed equipment data.
Reporting the measured quantity value therefore requires its unit and measurand definition. Reporting the complete measurement result also identifies the associated measurement uncertainty, its convention and relevant conditions. A change from ohms to milliohms changes the numerical expression, while leaving the represented physical value unchanged.
Measurement indication 2026-11-05
A measurement indication is the value supplied by a measurement instrument or measuring system. It may appear as a pointer position, displayed number, printed output or transmitted code. The indication and the corresponding measurand value need not be values of the same kind of quantity.[ref]
For an original mechanical readout, suppose the indicated pointer position is a length , related to input pressure by
The measurement indication then corresponds to under this measurement model. The position is observable; the pressure value is inferred using the response relation. A printed pressure scale could carry out part of that conversion physically without changing the underlying distinction.
If the pointer position is uncertain by , that contribution corresponds to in this stipulated linear model. Uncertainty in the zero , transducer sensitivity , temperature response and instrumental drift can add further contributions. Closely spaced scale markings therefore do not establish the total measurement uncertainty.
Record the original indication and instrument settings when applying corrections. A corrected measured quantity value can change after improved measurement calibration even though the recorded indication remains the same. For example, revising the assumed zero from to changes the inferred pressure from to . The complete measurement result accompanies the inferred value with its uncertainty and relevant conditions; the raw measurement indication alone cannot supply that information.
Range-switching hysteresis uses different thresholds for changing from a more sensitive instrument range to a less sensitive one and for changing back. Within the intervening interval, the current range is retained. The selected range therefore depends on the previous state as well as the present measurement indication. This prevents small fluctuations near a single boundary from repeatedly triggering gain ranging.
A concrete electronic analogue is an electronic comparator with two switching thresholds. The primary circuit example specifies upper and lower thresholds of and and explains why one threshold can produce unwanted transitions with a noisy or slowly changing input. Those are comparator circuit values, not specifications for an autoranging measurement instrument.[ref]
For an original two-range current measurement, let denote high sensitivity and low sensitivity. Use an input-current estimate , expressed on the same calibrated scale in either range. Switch when , switch when , and otherwise keep the previous state. The hysteresis interval has width .
Starting in , the indication sequence selects . A single threshold at would instead select and switch at every step after the first. At a steady underlying with bounded readout fluctuations of , neither boundary is crossed; either previous range remains selected.
The thresholds should lie within the usable overlap established by range-overlap calibration. If the two ranges give inconsistent current estimates, a switch can itself move the indication across the opposite threshold and create a cycle. The comparison must therefore use the corrected measurand estimate, not identical raw output voltages from different gains.
Hysteresis is not temporal averaging and does not ensure that the new reading has settled. Reserve headroom for a rapidly rising input, reject readings during the required settling time, and verify that the less sensitive range still meets its noise-limited detection threshold. An input that jumps beyond the current range can cause amplifier clipping before the controller reacts, even when its two-threshold rule is correct.
WIKA 2026-11-05
WIKA is a manufacturer of measuring instruments whose technical publications describe pressure measurements using elastic sensing elements. Its technical information distinguishes pressure-driven deformation from the linkage and pointer mechanism that turns it into a measurement indication.[ref] Its documentation is useful for connecting a physical measurement principle to the specifications of a complete instrument.
A concrete specification also illustrates the difference between an error limit and measurement uncertainty. WIKA's June2025 US sheet for specified four- and six-inch gauges gives a base limit of of span and an additional temperature error of of span per away from the reference.[ref]
For an original instrument example with span , the base bound is . A temperature departure contributes a further bound . Adding those magnitudes gives a conservative combined bound under this simplified interpretation; it is not a statistical root-sum-square uncertainty calculation. At a reading of , even the base bound is of the reading, despite being of span.
The example concerns the stated version and conditions, not a brand-wide accuracy claim. Use the actual model, range and pressure gauge temperature error specification together with instrument calibration evidence when interpreting a particular Bourdon tube pressure gauge.