Model a spatially uniform cell membrane as a capacitor of membrane capacitance in parallel with membrane electrical resistance and a fixed leak reversal potential . Exclude active ion channels, seal leak current and pipette capacitance. For electric current injected into the biological cell, electric charge conservation yields
Starting at and applying constant gives the ordinary differential equation solution
The steady state rise is ; the time constant is . With , , and , the rise is and . At one time constant the rise is . The figure's uncompensated pipette voltage adds for . The membrane voltage is continuous at step onset in this model, while the pipette voltage immediately acquires that access drop.
Checking the units is instructive: electrical resistance times capacitance has units , and electric current times electrical resistance has units of voltage. Unlike the passive voltage clamp transient under patch-clamp series resistance, this ideal current clamp membrane time constant is set by , not . Strong current steps can recruit voltage-gated ion channels; the passive model then fails and cannot generate an action potential. Small pulses of both signs, stable access and repeated baseline measurements help test whether the passive approximation is appropriate.
Patch clamp 2026-10-13
Patch clamp connects an electrode to a small area of cell membrane through a solution-filled patch pipette. A gigaohm seal around the rim makes tiny electric currents through ion channels distinguishable from seal leak current. The beautiful change of scale is from a population's average response to individual opening and closing events in single-channel recording.
Separate two questions. The configuration determines which cell membrane is accessible: an intact patch in cell-attached recording, an excised inside-out patch or outside-out patch, or intracellular access in whole-cell recording. The recording mode determines what is commanded and what is observed: voltage clamp commands an electric potential difference and measures electric current; current clamp commands electric current and measures membrane potential. Whole-cell recording supports both modes. A patch can contain several ion channels; a small recording area does not establish that only one is present.
Figure 1.
Patch clamp configurations and synthetic single-channel recording
. Original schematic of cell-attached recording and whole-cell recording, with synthetic electric current traces and a single-channel current-voltage relation. The prescribed opening times are identical at and solely to compare amplitudes. These are not experimental data or a model of opening probability. Membrane potential means the actual voltage across the patch, not automatically the pipette command. Details.
The 1976 article Single-channel currents recorded from membrane of denervated frog muscle fibres describes an extracellular fire-polished pipette with a – opening and – electrical resistance. It contained saline and a channel-activating compound. Two separate intracellular electrodes supplied local voltage clamp of the muscle fibre; the patch amplifier measured local electric current. Cooling to – and hyperpolarization helped resolve elementary events. Figure 2 shows step-like pulses with a filter at and . Discrete amplitudes and overlapping events supported individual ion channel openings rather than a smoothly varying population response. This was neither whole-cell recording through the patch pipette nor current clamp. The specialized conditions improved visibility but limit generalization to warmer preparations; filtering can conceal brief openings.[ref]
For a new single-channel recording experiment, useful controls include a matched recording without the activating compound, controlled membrane potential, repeated seal tests, and an explicitly stated filter frequency response. These are proposed controls, not claims that every one was performed in that 1976 figure. Changing membrane potential tests the single-channel current-voltage relation; changing the recording filter tests whether a short pulse's apparent amplitude is an instrument artifact. A flat baseline establishes neither an absence of ion channels nor an absence of very brief events.
Patch-clamp series resistance, also called patch-clamp access resistance, is the effective electrical resistance between the pipette's sensed voltage and the cell interior. It includes tip access and any remaining obstruction after break-in. For injected electric current into the biological cell, an ideal resistive access path gives
Thus through produces a difference: a substantial possible voltage clamp error. The electric current in this equation is the access-path current, excluding any bypass seal leak current. In current clamp, the same access drop contaminates the pipette voltage during injection. Bridge compensation subtracts an estimated ; an estimate that changes during recording leaves an error.
An independent passive voltage clamp calculation makes both speed and error visible. Write and , with fixed leak reversal potential , constant membrane electrical resistance and membrane capacitance . Assume an ideal command at the pipette node and omit seal leak current and pipette capacitance. Electric charge conservation gives
For a step from , the ordinary differential equation gives
With the figure's parameters and , the time constant is , the final membrane rise is , and access-path electric current falls from to . Only when does the time constant approach and the final voltage approach the command. Actual voltage clamp also depends on amplifier feedback control and the space clamp problem.
Seal leak current 2026-10-14
Seal leak current bypasses the membrane patch through an imperfect seal. For an ideal ohmic seal to a zero-voltage bath, , positive from pipette to bath. At and its magnitude is , far larger than the synthetic single-channel step. A stable baseline can be subtracted, but changing leakage can mimic a biological response. Intrinsic membrane leak through ion channels is distinct from this bypass: interpreting both as the same electrical conductance obscures whether the biological cell or the seal changed.
Whole-cell recording 2026-10-13
Whole-cell recording establishes intracellular electrical access by rupturing the membrane under a sealed patch pipette while retaining the gigaohm seal. Under adequate voltage clamp, the measured electric current combines contributions from accessible ion channels, membrane capacitance and seal leak current. Current clamp instead follows membrane potential while supplying a chosen electric current. Whole-cell recording is useful for linking a biological cell's total electrical response to its ion channels, but neither mode automatically controls distant dendrites or preserves the original cytosol.
In the 1981 methods demonstration, a bovine chromaffin biological cell was tested before and after patch rupture. Small voltage clamp steps produced a much larger capacitive transient after break-in. Integrating it gave for a cell; its decay indicated patch-clamp series resistance below . Depolarizing commands then revealed ionic electric currents. Separate current clamp recordings showed spontaneous action potentials. The comparison established access to the cell interior, rather than a claim that every pipette command equals membrane potential. Slower transients sometimes indicated incomplete rupture, and these small-cell results do not establish spatial control of a branched neuron.[ref]
Figure 1.
Synthetic whole-cell recording: patch-clamp series resistance and passive responses
. An original equivalent electrical circuit uses , and . The current clamp step and voltage clamp step are separate synthetic experiments. The passive model omits active ion channels, seal leak current, pipette capacitance and amplifier dynamics. Curves are calculations, not experimental measurements. Details.