Burette Created 2026-10-14 Updated 2026-10-18
A burette delivers an adjustable measured volume of a chemical solution; the change in its reading gives the delivered volume in an acid-base titration.
Its downward-increasing scale measures a change in reading, not the initial liquid inventory. For initial reading and final reading without refilling, the nominal delivered volume is . An illustrative to reading therefore represents . There is no requirement to start at zero. An air bubble disappearing from the tip uses part of the apparent delivery to fill the tip, and a hanging drop has not yet entered the sample; both break the intended volume account.
Figure 1.
A pipette selects the sample and a burette measures added titrant by two liquid meniscus readings
. Original schematic, not a photograph or experimental readout. The apparatus shows a clamped burette, mixed sample, and optional glass electrode beside the stirring bar. Enlarged illustrative scale readings increase downward; the bottom of the curved liquid meniscus gives and . Their difference is the added volume. Details.
For transparent aqueous samples, read the bottom of the liquid meniscus at eye level; viewing from above or below shifts its apparent alignment. The mark is a comparison between a curved liquid surface and a scale, not a direct measure of amount of substance. Eye-level reading and the distinction between contained and delivered volumes are part of solution preparation and volumetric calibration.[ref]
Both readings contribute measurement uncertainty. If each has independent standard deviation , their difference has standard deviation ; a common additive reading offset cancels, while a scale-factor error remains. For the inventory , uncertainty in titrant concentration remains even with perfect readings. This is why repeat delivery, independent volume calibration, and titrant standardization answer different questions.
Columbia University's institutional laboratory manual, Experiment 1, Glassware Calibration, Primary and Secondary Standards, and Manual Titrations, Part 3, specifies two weighed – portions of potassium hydrogen phthalate, transferred completely into separate Erlenmeyer flasks. Dissolve each in approximately distilled water; any warmed sample cools to room temperature before titration. Add two or three drops of phenolphthalein. Rinse and condition a clamped burette with the diluted sodium hydroxide, then titrate while swirling to faint pink lasting about seconds. Calculate concentration from each portion's own mass and delivered titre; repeat if results scatter. Part 1 identifies a Mettler balance, and Part 2 includes paired weighings of exposed sodium hydroxide and potassium hydrogen phthalate. These are prescribed procedures, not reported measurements.[ref]
The same manual directs keeping the diluted sodium hydroxide covered and limiting air exposure after titrant standardization.[ref] The MIT handout explains the associated carbon dioxide problem: carbonate contamination of sodium hydroxide is controlled during preparation, bottles are promptly capped, and atmospheric carbon dioxide makes the pink phenolphthalein endpoint fade.[ref] Consequently, a standardized concentration applies to the stored chemical solution under its handling conditions; continued uptake of carbon dioxide can change the reacting inventory per delivered volume.
The color is the readout, and stoichiometry supplies the inference. Phenolphthalein indicates a pH range, so its first reproducible pale pink is an operational titration endpoint, rather than a direct observation of equal reacting inventories. A potentiometric titration can locate an endpoint from a pH versus volume curve, but glass electrode calibration alone still does not determine sodium hydroxide concentration. Repeat portions test scatter; incomplete dissolution, an unmatched titration blank, or a shared endpoint error can bias all repeats together.
The apparatus and motivating measurement are described in permanganate-iron titration; the following is an original account of its possible errors. Assume correct titrant standardization, the acidic permanganate-iron balance, quantitative conversion of the sampled , and additive contributions expressed as equivalent titrant volumes. Write
Here is the sample titre, the titration blank, background reagent consumption, and signed offsets of the observed titration endpoints from the respective stoichiometry conditions after background consumption. The subscripts distinguish sample and blank. is additional sample-specific permanganate demand, absent from the blank; is apparent delivery used to fill an initially air-containing burette tip rather than enter the flask. Assume the blank's tip is already filled; the sample should also have after proper preparation. These terms are a diagnostic model, not measured components of the stipulated blank.
Figure 1.
Burette delivery, titration endpoint and titration blank describe different parts of the inference
. Original schematic and additive volume model, not a photograph or measured titration curve. The numerical decomposition assumes matched background and endpoint contributions and no tip-air or sample-specific demand. Boxes are not to scale; the blank's reagent consumption and endpoint offset are not separately specified. The exact equivalence point has no guaranteed visible colour flag. Details.
Subtracting the two accounts proves the error relation
The simple inference is valid when this residual is negligible. Matching water and sulfuric acid tests the background terms, but does not by itself establish : flask composition and viewing conditions can differ. A coloured sample can mask residual permanganate, changing the assigned titration endpoint without a corresponding redox consumption. A sample-specific reducing agent instead contributes positive by consuming permanganate. A reagent titration blank cannot measure either effect just by omitting the sample.
A transient purple streak before mixing is not the mixed-flask titration endpoint. Conversely, the equivalence point is the five-to-one inventory match, without a promised visible flag. In the ideal selective model, a persistent excess colour requires some permanganate beyond that match. No particular drop size, waiting time or endpoint offset is stipulated here; the same stated colour criterion must be applied to sample and blank, and its transferability checked.
The model also predicts error direction. If a tip bubble disappears during sample delivery, makes the inferred amount of substance too high because final minus initial burette reading includes liquid that stayed in the tip. Conditioning the tip before the first reading prevents this particular discrepancy; subtracting a separate titration blank is not an instrument repair. Repeated aliquots test scatter, a known added inventory tests the five-to-one response, and an independent titrant standardization tests . These are the proposed controls from permanganate-iron titration, with distinct diagnostic roles; none alone proves absence of all sample interference.
Primary standard 2026-10-18
A primary standard supplies a reliably known reacting inventory from which a titrant concentration can be determined.
The defining advantage is known high purity: the weighed portion can be assigned a known amount of substance. A titrant whose concentration is established against that inventory is a secondary standard solution in IUPAC terminology.[ref]
An accurate balance determines total mass, including any unwanted water or other material. To infer reacting inventory, one also needs known composition, stability during storage and weighing, complete transfer and dissolution, and a sufficiently complete chemical reaction of known stoichiometry. A finely resolved balance cannot establish these chemical facts. Potassium hydrogen phthalate is useful because a suitably prepared portion provides a known acid inventory for sodium hydroxide titrant standardization; the KHP standardization protocol below shows how that claim is put into practice.
Preparation belongs to the particular material and method. For example, the MIT handout specifies drying potassium hydrogen phthalate at for hours or overnight, cooling in a desiccator, and weighing by difference; its footnote identifies removal of superficial moisture as the purpose. This is a condition of that potassium hydrogen phthalate procedure, not a universal instruction to heat every primary standard in the same way.[ref]
Redox titration Created 2026-10-13 Updated 2026-10-18
A redox titration adds a measured amount of substance of one reactant to determine another through a known redox chemical equation. The burette reading becomes an amount through concentration; the balanced chemical equation supplies the amount ratio. A colour endpoint additionally needs a chemical explanation, as in the permanganate-iron titration.
The useful bridge is an Electron inventory. If each sample entity supplies Electrons and each titrant entity accepts , matching the formal exchanged amount of substance of Electrons gives
Here is the titrant molar concentration and the volume attributable to the sample at the equivalence point. These are net half-reaction counts, not a claim that free Electrons accumulate in the flask. The redox conversion must be sufficiently complete and selective for the stated stoichiometry to identify .
This is why a small visible signal can reveal a much larger invisible inventory: the signal locates a stopping point, while titrant standardization and stoichiometry turn delivery into amount of substance. The titration endpoint estimates the equivalence point; repeatable colour alone establishes neither correct titrant concentration nor chemical selectivity. The permanganate-iron endpoint and blank model separates these assumptions.
How can an invisible dissolved inventory be measured? The unsigned laboratory handout Solubility Product Constant, Ksp, for Calcium Hydroxide by Acid-Base Titration, Procedure, starts with saturated calcium hydroxide, supplied by the teacher or prepared from calcium and water. Its apparatus includes filter paper, a funnel, stoppered flasks, a volumetric pipette, a volumetric flask and a burette. Two successive rounds of filtration remove particles. The filtrate is stoppered to reduce carbon dioxide exposure. A aliquot is diluted with deionized water and titrated with standardized hydrochloric acid; bromothymol blue gives a pale-green titration endpoint. The readout is the difference between final and initial burette readings; four consistent trials are required.[ref]
Unlike AB, calcium hydroxide has one-to-two dissolution stoichiometry:
For molar solubility in the ideal model with no background calcium or hydroxide, no additional chemical reactions, and negligible self-ionization of water, and . Hence , with units ; it is not the AB constant used in the figures. The neutralization chemical equation is . Let be the known titrant molar concentration, its delivered volume at the equivalence point, and the original aliquot volume. Then the inventory calculation is
Dilution of the aliquot changes its concentration during the acid-base titration, but preserves its initial amount of substance of hydroxide. Therefore , rather than the diluted flask volume, belongs in the denominator.
The handout identifies two biases: carried-over calcium hydroxide solid dissolves during acid-base titration, consuming extra titrant; carbon dioxide consumes hydroxide and forms calcium carbonate. Filtration and stoppering address these effects, while repeat trials test consistency. Its review questions also ask about dissolved ion pairs, which filtration cannot remove.[ref]
Further controls would compare a water blank, independently verify titrant standardization, and record the saturation temperature; these are proposed checks, not reported measurements. A titration endpoint estimates the equivalence point and may introduce a systematic offset. Even perfect inventory measurement does not determine activity coefficients. Recovering the dimensionless thermodynamic solubility product additionally requires
This explains both the power and the limit of the experiment: acid-base titration measures a dissolved inventory, while the link from that inventory to a thermodynamic equilibrium constant still depends on a chemical model.
Titrant standardization 2026-10-18
Titrant standardization determines the actual titrant concentration by reaction with a known inventory, often a primary standard. It is distinct from calibrating the delivered volume or a glass electrode's response. For a known monoprotic acid inventory and a suitable blank-corrected endpoint,
An original illustrative standard of requiring , with matched blank , gives . The measured quantities here are stipulated examples. Independent solution preparation and volumetric calibration tests the pipette, burette, and volumetric flask; titrant standardization tests the reacting inventory per delivered volume.
The useful chain is known mass, known composition, known amount of substance, and then a measured titre through a specified stoichiometry. If each standard unit consumes units of titrant, the general inventory equation is , where is the corrected titre. The one-to-one formula above has . A titration endpoint supplies an observable stopping rule; whether it represents the equivalence point accurately remains a separate question. Standardization endpoint and blank bias makes the direction of the resulting error explicit.