A cesium chloride density gradient forms when the salt redistributes in a rotating chemical solution. It can supply matching-density positions for DNA during isopycnic centrifugation.[ref]
https://www.beckman.com/resources/technologies/centrifugation/density-gradient-centrifugation/. Types of Density Gradients and Types of Gradient Material: self-forming CsCl gradients and nucleic-acid separation
Density-gradient centrifugation 2026-10-13
Density-gradient centrifugation uses a liquid whose density varies along the separation direction. A band does not by itself establish which mechanism produced it: rate-zonal centrifugation separates by migration rate before equilibrium, whereas isopycnic centrifugation uses equilibrium positions set by buoyant density.
Isopycnic centrifugation 2026-10-13
In isopycnic centrifugation, a species concentrates near the position where the liquid's density matches its buoyant density. In a density gradient increasing outward, material displaced inward encounters less-dense liquid and tends to move outward; material displaced outward encounters denser liquid and tends to move inward. Diffusion broadens the resulting band. If the density gradient never reaches the species' buoyant density, a matching equilibrium position need not exist within the vessel.[ref]
https://www.beckman.com/resources/technologies/centrifugation/ultracentrifugation/. An Introduction to Ultracentrifugation, Isopycnic Centrifugation: matching buoyant density and diffusion in equilibrium gradients
The 1958 paper The replication of DNA in Escherichia coli by Matthew Meselson and Franklin W. Stahl (pp. 672–676, Figures 1–4) makes the distinction concrete. Its separation apparatus was a Spinco Model E ultracentrifuge with an Escherichia coli lysate in a cesium chloride density gradient. Ultraviolet wave absorption photographs of the rotating sample were scanned with a recording microdensitometer: peak position reported buoyant density, while signal above baseline reported DNA concentration.[ref]
Mixed lysates labeled fully with the nitrogen isotope and unlabeled lysates supplied reference bands; a mixture including intermediate material checked its position relative to those references. Figure 1 followed band formation over time, helping distinguish a migrating distribution from an approximately stationary one. These controls address resolution, calibration and approach to equilibrium. They do not make every ultraviolet-absorbing component identifiable as DNA, or turn a band into a direct image of individual strands. The DNA replication article develops the copying inference; here the key result is a separation by buoyant density, not a claim that heavier-labeled DNA is physically larger.[ref]
The replication of DNA in Escherichia coli by Matthew Meselson and Franklin W. Stahl. pp. 672–673, Figures 1–2; p. 675, Figure 4 density references and microdensitometer baseline; p. 676, Experimental: Spinco Model E and ultraviolet photographs