In NIST's 2001 comparison, the Laser-Optimized Cryogenic Radiometer held its cavity near and controlled the heat sink separately. Opening the beam reduced heater power at fixed cavity temperature; an AC bridge read the temperature. Tests used stabilized, spatially filtered lasers at , and , with horizontal input and a movable photodiode transfer detector in the cavity's optical plane. The paper contrasts this with the High Accuracy Cryogenic Radiometer's optical heating followed by two electrical heating cycles and interpolation.[ref]
The two laboratories compared calibrations of the same transfer detectors, and checked their spatial response. Alignment differences and transported dust limited that comparison; its agreement did not prove universal equivalence. Its correction tables retain window transmission, cavity radiometric absorptance, alignment, electrical readout and optical-electrical heating non-equivalence. The heat sink was controlled, but §3 does not specify its numerical setpoint: the reference in the preceding example is stipulated.[ref]
For a reproducible substitution measurement, record the beam's wavelength and entrance boundary, receiver and reference temperatures, heater terminal voltage and electric current, dark readings before and after illumination, and the settling rule. Check beam collection at the aperture, window transmission, radiometric absorptance, deposited lead heating, and sensitivity to beam position. Repeatability alone misses a stable correction error. Photometric calibration can subsequently use this radiant flux scale, but photometry still needs the appropriate spectral visual weighting; the cold receiver does not itself measure luminous flux.
Cryogenic radiometer Created 2026-10-14 Updated 2026-10-14
A cryogenic radiometer uses low-temperature operation to improve electrical substitution radiometry through particular material and thermal properties. Lower heat capacity of a copper receiver permits a substantial cavity without proportionately slow relaxation; cold shields reduce unwanted thermal radiation, and heater leads with superconductivity can suppress resistive lead heating. These features do not make radiometric absorptance exactly one or abolish optical-electrical heating non-equivalence.[ref]
Cold surroundings matter as much as a cold receiver. In a simplified gray-surface model, net outward thermal radiation is , using emissivity , area , shield temperature and the Stefan-Boltzmann constant . Cold shields reduce the incoming term as well as the receiver's emission; cooling the receiver alone while leaving it exposed to warm surroundings does not remove incoming radiation. An entrance window still provides an optical and thermal path. Vacuum reduces gas-mediated heat transfer, while an intentional radiometer thermal link remains necessary. This model explains specific losses, rather than promising a universal improvement factor or perfect accuracy.
An explicit lumped model makes the cancellation understandable. Treat the receiver as having uniform temperature , heat capacity , heater power deposited in the receiver, net parasitic input , and outward heat transfer to a reference at temperature . With incident radiant flux and radiometric absorptance , conservation of energy gives
The dark state has no admitted beam. Compare settled dark and light states, denoted and ; each has . Subtracting their balances gives
where and . At identical and , with unchanged surroundings and parasitic heating, both correction differences vanish. Thus , and . Notice the signs: extra stray thermal radiation admitted with the beam also reduces heater power, so it must be subtracted from that reduction before attributing the result to the intended beam.
Measure the heater's own terminal voltage and electric current : for steady electrical heating, , where is its electrical resistance. A supply voltage that includes wire voltage drops counts Joule heating in the leads too; only the portion deposited in the receiver belongs in this balance. Changes in that deposited lead heat belong in . Optical-electrical heating non-equivalence adds a further instrument-specific correction when the uniform-temperature model is insufficient.[ref]
Optical-electrical heating non-equivalence means that equal absorbed optical and electrical powers do not necessarily give identical thermometer responses. Light and the heater may deposit heat in different places, producing different temperature gradients and different routes of thermal conduction to the support. Holding one thermometer reading fixed need not hold every part of the cavity at the same temperature. This differs from incomplete radiometric absorptance: one concerns where absorbed heat goes, the other how much incident light becomes absorbed heat.
A useful accounting convention is
where the signed correction is determined by characterization of the instrument and represents the residual departure from the lumped electrical substitution heat balance. It is not a universal extra source of energy. Varying the beam location or heater location, comparing equivalent temperature excursions, and checking repeatable dark returns can test the assumptions; no single small correction proves that every other correction vanishes.