John D. Ramboz is the first author, with Oskars Petersons, of NIST's 1991 account of a laboratory current transformer calibration service. The report describes ratio and phase measurements, their uncertainties and quality-control procedures for electrical measurement work.[ref]
A Calibration Service for Current Transformers by John D. Ramboz and Oskars Petersons. Title page and Abstract
It is a primary technical record of the service and its measurement conventions.The separation of magnitude and phase has a direct consequence for electric power measurements. Consider an original sinusoidal model with root-mean-square electric potential difference , secondary current magnitude , true current ratio and primary-current lag . True active power isIf the secondary leads the primary by , simply multiplying the secondary signal by nominal ratio givesprovided . This uses the lead-positive current-transformer phase displacement convention stated in the report.[ref]
With exact magnitude conversion and small in radians, the relative power error is approximately . For stipulated rad and , that term is about . At , its first-order value vanishes. Thus the same phase error can have different effects at different power factors.
This derivation is an explanatory example, not a reported performance test or an attribution of the calculation to Ramboz. It explains why a calibration report needs operating conditions, a sign convention and measurement uncertainty, in addition to a nominal transformer ratio. Precision in the current magnitude does not remove phase-dependent power error.
Oskars Petersons worked on electrical metrology at the National Institute of Standards and Technology. With John D. Ramboz, Petersons coauthored the 1991 report describing NIST's laboratory current transformer calibration service. The report treats current ratio and current-transformer phase displacement as distinct measurement results.[ref]
A Calibration Service for Current Transformers by John D. Ramboz and Oskars Petersons. Title page; Abstract; §1.3, Terms Related to Transformer Calibration
A current transformer allows a secondary electric current to represent a larger or smaller primary current. Its marked ratio is a nominal conversion, whereas the measured ratio can differ. The report expresses the measured ratio as the marked ratio multiplied by the current-transformer ratio correction factor, and uses a secondary-leading sign convention for the reported phase displacement.[ref]
A Calibration Service for Current Transformers by John D. Ramboz and Oskars Petersons. §1.3, printed pp.5–6; Figure 3
An original calculation illustrates the distinction. A marked ratio of A: A gives . If the current-transformer ratio correction factor is , the measured magnitude ratio is . A secondary reading of 5 A then represents 200.04 A under those stated conditions, rather than exactly 200 A.
Magnitude alone is not a full alternating-current description. If the secondary phasor leads the primary by and their magnitude ratio is , a consistent convention isThe magnitude correction and phase rotation address different errors. A measurement system can reproduce one accurately while requiring correction of the other. The joint report's calibration circuits, including compensated current comparators, give these distinctions an apparatus-level setting rather than leaving them as ideal-transformer identities.[ref]
A Calibration Service for Current Transformers by John D. Ramboz and Oskars Petersons. §2.1, Calibration Circuits, Figures 8–9
Coauthor of Maintenance and Validation of Liquid-in-Glass Thermometers by Christina D. Cross and others.
Identity: www.nist.gov/people/paul-williams
Identity: www.nist.gov/people/joshua-hadler
Identity: www.nist.gov/people/john-h-lehman
David J. Livigni coauthored a comparison of two cryogenic radiometers at NIST. Comparing independently realized optical-power measurements tests consistency and helps identify systematic contributions to measurement uncertainty.[ref]
Comparison of Two Cryogenic Radiometers at NIST by Jeanne M. Houston and others. author list and the described study’s abstract or introductory account
Homepage: www.nist.gov/people/gregory-f-strouse
- A New NIST Automated Calibration System for Industrial-Grade Platinum Resistance Thermometers by Gregory F. Strouse and others
- Effects of different methods of preparation of ice mantles of triple point of water cells on the temporal behaviour of the triple-point temperatures by G. T. Furukawa and others
- A New NIST Automated Calibration System for Industrial-Grade Platinum Resistance Thermometers by Gregory F. Strouse and others
- Effects of different methods of preparation of ice mantles of triple point of water cells on the temporal behaviour of the triple-point temperatures by G. T. Furukawa and others
Jonathan Pearce is a metrologist whose research develops practical temperature measurement with contact sensors. His NPL work includes improved thermocouples, models of impurity effects during metal solidification, and methods for detecting or reducing sensor calibration drift. These connect the physics of materials with the reliability of measurements in industrial conditions.[ref]
https://www.npl.co.uk/our-people/jonathan-pearce. Research highlights: thermocouples, high-temperature standards, impurity modelling and self-validating sensors
In the coauthored Guide on Secondary Thermometry: Industrial Platinum Resistance Thermometers by Jonathan Pearce and others, an important measurement problem is resistance thermometer self-heating: the electrical measuring current heats the sensor whose temperature is being inferred. The guide describes estimating the zero-current resistance from measurements at different currents, while accounting for the surrounding medium. Its treatment of electrical resistance in the leads also explains why four-terminal sensing differs from two-wire measurements. These details show how measurement uncertainty arises from the complete sensor-and-readout arrangement.[ref]
Homepage: www.npl.co.uk/our-people/jonathan-pearce
Portrait of Jonathan Pearce
. Portrait identified on the National Physical Laboratory profile.Homepage: www.npl.co.uk/our-people/richard-rusby
Homepage: www.nist.gov/publications/guide-secondary-thermometry-industrial-platinum-resistance-thermometers
Homepage: www.inrim.it/en/contacts/luigi-iacomini
Homepage: www.nist.gov/publications/guide-secondary-thermometry-industrial-platinum-resistance-thermometers
Randall M. Schoonover investigated precision measurements of mass and mass density. A coauthored study used an electronic force balance to weigh a silicon crystal in air and water, applying buoyancy and measurement calibration corrections. The reported crystal density agreed with its accepted value within 2.4 parts per million.[ref]
The Electronic Balance and Some Gravimetric Applications (The Density of Solids and Liquids, Pycnometry and Mass) by Randall M. Schoonover and others. Printed pp. 299–300, abstract and hydrostatic-weighing principles
The method obtains mass density from the two buoyant-force readings and known fluid densities; it does not require an independent measurement of the crystal's volume.[ref]
James Taylor coauthored research on electronic-balance measurements of solid and liquid mass density, mass, and vessel volume. The study explained how an electromagnetic restoring force returns a loaded balance to a sensed null position, then related the readings to buoyancy and measurement calibration. It also treated pycnometer density measurement, where temperature-dependent vessel capacity must be distinguished from the balance reading.[ref]
The Electronic Balance and Some Gravimetric Applications (The Density of Solids and Liquids, Pycnometry and Mass) by Randall M. Schoonover and others. Printed p. 299 byline; pp. 302–305, balance operation; pp. 313–316, pycnometer calibration and application