A fluorescence excitation spectrum scans excitation wavelength while detecting fluorescence in a fixed emission band. A fluorescence emission spectrum does the converse. The inspected spectrofluorometer example uses an excitation monochromator scan with the emission monochromator fixed, and reverses their roles for emission.[ref]
An operational model explains why this need not reproduce absorption. For one dilute, unsaturated emitting species, write for absorbed fraction, for incident photon rate, for fluorescence quantum yield, and for the fraction of emitted photons counted in the selected band. The background-subtracted count rate is
Only after correcting incident photon rate, and with constant yield and emission-band collection, is this proportional to absorbed fraction. A nonfluorescent absorber contributes to absorption but not this readout. Thick samples, multiple emitting species, and reabsorption of fluorescence complicate the model.
Fluorescence lifetime 2026-10-14
For the constant-rate model in fluorescence quantum yield, fluorescence lifetime is the mean residence time in the emitting excited state and the timescale of its exponential decay. The detected fluorescent photon rate after an ideal instantaneous pulse is for . Its amplitude and its decay timescale are separate quantities.
For a stipulated example, rates give and . Rates give the same but . Conversely, doubling doubles the ideal amplitude without changing . Equal fluorescence lifetimes need not imply equal brightness or equal fluorescence quantum yields.
Different environments can supply several populations, giving rather than one exponential decay. A fitted single timescale then depends on the selected model and detection band. FRET donor lifetime uses the change in a donor's decay when transfer introduces an additional loss route; the FRET branch develops that interpretation.
Fluorescence quantum yield counts emitted fluorescent photons per absorbed excitation photon. It differs from quantum efficiency, which describes the photodetector.[ref]
For an original constant-rate model, let be the radiative decay rate and the sum of competing rates without fluorescent photon emission. An initial population in the emitting excited state obeys the rate equation
Integrating over all time gives
Here every absorbed excitation is assumed to reach this emitting population; additional losses before that stage require another factor. Under steady weak excitation, absorbed photon rate and total detection fraction give . The count rate therefore mixes excitation, yield, and detection, whereas specifies a decay timescale.
Fluorophore 2026-10-14
A fluorophore is a fluorescent molecule or fluorescent part of a molecule. Its usefulness as a label depends on how efficiently it absorbs the chosen excitation light, its fluorescence quantum yield, and its resistance to photobleaching. A large fluorescence quantum yield alone does not guarantee a bright image if few excitation photons are absorbed or collected.