An interlaboratory comparison organizes measurements of the same or similar items by at least two laboratories under agreed conditions. Its purpose may be to compare measurement reproducibility, evaluate a measurement procedure, or assess participants through proficiency testing. The last purpose evaluates performance against criteria specified beforehand; it is one use of interlaboratory comparison, not a synonym for every such study.[ref]
Selection, Use and Interpretation of Proficiency Testing (PT) Schemes by Brian Brookman and Ian Mann. §§2–3, printed pp. 3–4: scope and definitions
A comparison begins with its experimental design. When the aim is to compare laboratory analysis, the organizer needs evidence of reference material homogeneity and reference material stability: unequal or changing test items would otherwise be confounded with differences between laboratories. If collection is part of the measurement procedure, the design must instead include analytical sampling explicitly. In a 2002 collaborative study, several samplers collected wheat and green coffee from bulk lots. The reported between-sampler variation was significant for two wheat analytes; for coffee, a significant repeatability contribution was detected without a significant sampling-reproducibility contribution. Thus even the importance of collection differences depended on the material and analyte.[ref]
Collaborative trials of the sampling of two foodstuffs, wheat and green coffee by Michael Thompson and others. Abstract: comparison of sampling reproducibility and repeatability for wheat and green coffee
For a common proficiency-test z-score, let laboratory report , let the assigned value in proficiency testing be , and let the standard deviation for proficiency assessment be . ThenThe denominator describes the scheme's assessment scale. It is not automatically the laboratory's own measurement uncertainty or the observed standard deviation of participants. The guide distinguishes this proficiency-test z-score from scores incorporating uncertainty, and discusses choosing the assessment scale to match the intended use.[ref]
An original hypothetical example shows why the assigned value matters. Suppose three laboratories report , and in the same units, with a stipulated assessment scale . Using their arithmetic mean as an interlaboratory consensus value gives and scores . An independently established assigned value of , with the same assessment scale, instead gives . Close agreement between participants is compatible with a shared displacement from an independent reference.
The identifiability issue is visible in a simple model,where is the reference quantity value, is a common systematic error, and describes laboratory-specific deviations. Pairwise differences eliminate both and . Moreover, replacing by leaves every predicted unchanged. Consequently agreement alone cannot identify a common bias. Evidence from an independently characterized certified reference material or an independent measurement calibration can help resolve this ambiguity. Its own measurement uncertainty and any shared dependencies must then enter the comparison model; one should not silently interpret the simple proficiency-test z-score as a full uncertainty calculation. This is the practical importance of distinguishing measurement precision, measurement trueness and shared bias in an interlaboratory consensus.