Magnitude zero point 2026-10-05
A magnitude zero point fixes the reference measurement in a specified photometric passband. Relative to a zero-magnitude reference, the apparent magnitude is . Both measurements must use the same passband and calibration convention.
Optical spectrum 2026-10-05
An optical spectrum describes the distribution of optical radiation over wavelength or frequency. A spectrograph records it by separating wavelengths, while a photometric passband integrates a chosen part.
The traditional infrared photometric bands extending the optical photometric system into this range are approximately
Their photometric passbands lie mainly in atmospheric windows between strong molecular absorption bands. This makes ground-based observations practical. The exact centers depend on the system: is near , and shortened and bands reduce unwanted background or absorption. At the longest wavelengths, atmospheric and instrumental thermal radiation make sky brightness especially important. A modern band near is an additional choice, rather than one of the traditional extensions.
The magnitude zero point is defined by the reference response . With zero magnitude assigned to that reference, the apparent magnitude is
The two responses must use the same photometric passband and instrumental weighting. If the reference has assigned magnitude instead, add to the right-hand side.
Spectral flux density 2026-10-05
Spectral flux density is radiant power received per unit area and per unit wavelength or frequency. The two conventions satisfy under , hence when the arguments refer to the same photons. Integrating over a photometric passband gives its measured flux.