Threshold timing sensitivity of an RC step response measures how a crossing time changes when the absolute threshold or signal height changes. For a relaxed step response , with , the rising crossing is
Holding the other inputs fixed gives
A threshold approaching the step height encounters a flatter part of the exponential response, making the first sensitivity large.
For an original timing estimate, choose , , and effective small threshold-noise standard deviation . At , local linear propagation gives timing standard deviation about . At , it gives about . The latter threshold also needs a longer input pulse to be reached at all, as shown in threshold-crossing time of a finite rectangular pulse through an RC load.
The two voltage sensitivities can cancel under a different measurement calibration model. If the threshold is set to a fixed fraction of the same varying height, and , their contributions sum to zero. The crossing becomes , independent of amplitude in this ideal model. Such correlated threshold and amplitude errors in RC timing must retain covariance rather than add two independent voltage uncertainties.[ref]
The local noise rule also follows from dividing a small voltage perturbation by the crossing slope . It assumes a well-defined nearby crossing and sufficiently small perturbations; it does not describe multiple noise crossings, comparator switching dynamics, or a pulse ending before the threshold. A bench test can compare controlled threshold shifts with oscilloscope crossing times while holding the actual step height and initial state fixed.
The time above threshold for an RC-filtered rectangular pulse is the separation of its rising and falling threshold crossings, rather than the applied pulse duration. In the relaxed first-order model, put with and . The parent crossing calculation gives
This excludes comparator propagation dynamics and uses the same threshold on both edges. A Schmitt trigger with unequal switching thresholds needs a different formula.
The positive-width relation can be inverted exactly:
Its derivative is . Thus a measured positive determines a unique input duration under the stated model. A zero-width observation does not: all rejected durations have that same outcome.
For an original example, take , and . The output remains above threshold for about , longer than the input pulse because of its decaying tail. Substitution into the inverse relation recovers .
For long pulses, tends to . It is positive below a half-height threshold and negative above one. At exactly half height, for every finite accepted pulse, with the difference tending to zero for long pulses. Consequently a measured above-threshold width cannot be equated to the input width without the threshold and filter model.
A pulse-generator/oscilloscope comparison can record both widths while varying threshold. The measured RC decay should be checked independently, using the natural-response method in the primary MIT laboratory.[ref] Initial residual charge, hysteresis, and different rising/falling comparator delays alter the observation; the algebra above only inverts the ideal capacitor waveform.
A timing-skew bound for a ratiometric measurement limits the resistance-ratio error caused by sampling the signal and reference at different times. It converts a bound on the fractional rate of change of the shared excitation into an acquisition-timing requirement. This is useful even when the two analog filters are exactly matched.
Use the model in sampling skew after matched ratiometric filters: the common filtered excitation is , the true resistance ratio is , and the reference is sampled after the signal. Then . Suppose is continuously differentiable and its logarithmic derivative obeys throughout the entire interval between the samples. The fundamental theorem of calculus gives
The same proof works for either sign of . Writing , exponentiation yields the exact interval
This is a deterministic bound under the stated slope assumption, rather than an expansion truncated at first order.
For an original timing budget, suppose and require at most fractional resistance error from skew alone. It suffices to require
The slightly smaller-than- value accounts for the exponential upper bound. An excitation proportional to reaches one side of the logarithmic bound exactly, so the bound is attainable in this model.
A sufficient way to establish is to verify and , giving . A filter cutoff by itself supplies neither a positive excitation floor nor this fractional-slope bound. Near an excitation zero, a small absolute change can produce a large resistance-ratio error. The primary filter-design report establishes the need to match signal and reference dynamics and to allow filter settling; the inequality above adds an original unequal-time error budget.[ref]
A proposed bench check uses a stable sensor substitute and reference resistor, records the filtered excitation with an oscilloscope, and deliberately changes the acquisition delay. Compare each observed ratio with the interval predicted from a verified excitation floor and slope bound. This isolates timing effects from a fixed reference-resistor error in ratiometric resistance measurement.
An unobserved delayed-pulse counterexample to a fitted tail bound constructs two admissible optical inputs giving the same finite measurement record but different missed energies. The future-domain validity of a fitted power-law optical envelope article gives a nonnegative, continuously differentiable delayed pulse with adjustable energy. The counterexample persists through a calibrated causal detector, not just an ideal instantaneous power readout.
Let , where the additional pulse has support after the record ends at . A linear receiver with impulse response and responsivity measures . Causality means for . Therefore, for every recorded time ,
since every argument is negative. This causal indistinguishability of delayed optical inputs on a finite record holds even when the impulse response is known exactly and the recorded data are noiseless.
For an original example, record through and place the smooth sine-squared pulse of the parent article at , with width and peak . Its extra optical energy is . A chosen causal exponential receiver with decay time produces no difference in the initial record, although it will respond when the later pulse arrives. If its impulse response has unit area, the complete later output-area difference is in the receiver's output-area units.
Any rule using only the common initial record must return the same proposed bound for both inputs. Without a justified restriction on later illumination, the delayed pulse energy can be chosen larger than any proposed finite guarantee. More accurate calibration of that same initial record cannot resolve this information gap.
NIST characterized its detector, cable and oscilloscope together with a short known optical input before correcting pulse measurements.[ref] Such response calibration addresses receiver distortion. Source shutoff, exclusion of later pulses or a future-valid rate envelope supplies the additional physical information needed for a missed-energy guarantee. The delayed pulse here is a mathematical construction, not a reported defect of that apparatus.
Voltage divider Created 2026-10-13 Updated 2026-10-14
A voltage divider uses two resistors connected in series to obtain a fraction of an applied electric potential difference. Let the upper electrical resistance be , the lower one , and the input be relative to a common electrical reference potential. The output is the junction's electric potential difference relative to that same reference, across . The familiar ratio assumes zero output electric current: no external electrical load draws from the junction. In a steady direct current model with fixed resistors obeying Ohm's law and an ideal voltage source,
Here is the open-circuit voltage; correspond to in the header formula. With no junction branch, Kirchhoff's current law makes the same electric current pass through both resistors, and Ohm's law gives the two voltage drops. For , division gives ; if , the predicted output is zero but this ratio is undefined.
An original stipulated example has and . With zero output electric current, and . An attached electrical load instead gives ; a load gives . These are exact model predictions, not measured readings or specifications of a particular multimeter or oscilloscope.
Figure 1.
The same voltage divider has three different predicted outputs as the attached load changes
. Original circuit and readout comparison for ideal input and two resistors. Each finite connects across ; the displayed numbers are model predictions with ideal readout, not experimental data. Source. Details.
The divider is useful for scaling a voltage or converting a change in a sensor's electrical resistance into an electric potential difference. Its beauty is also its trap: multiplying both electrical resistances by the same positive factor preserves the unloaded ratio, but changes the effect of any fixed electrical load. A divider therefore supplies a predictable ratio under specified loading conditions; it is not a source that maintains that voltage for arbitrary output electric currents. A Wheatstone bridge compares two divider junctions, while Thévenin's theorem describes the divider's response to many different loads.
Southern Illinois University's ET 304A guide, Lab 4: Voltage Measurements and Meter Loading Effects, documents an instructive test: build equal-arm voltage dividers on a supply, then measure across the lower arm with a digital multimeter, analog multimeter, and oscilloscope. Starting with two resistors, repeat with equal pairs from to . It requires separate recorded readings, unloaded predictions, and predictions including each instrument's input electrical resistance. The guide obtains meter resistances from specification sheets and scope resistance from the input-panel marking. Equal ratios control the unloaded target while changing susceptibility to voltmeter loading. The provided tables are blank: the guide specifies a procedure, not measured outcomes.[ref]
Preparation and accuracy checks are made concrete by Berkeley's Lab 1: Introductory Experiments and Linear Circuits I. Its different divider uses nominal and resistors and a supply on a breadboard. Measure isolated resistors first, record actual supply voltage, recalculate with these measured values, then record the junction voltage with the digital multimeter. It explicitly compares actual values with resistor tolerance and the final voltage with a range-dependent accuracy bound from the instrument manual. This is an uncertainty check, not an independent measurement calibration.[ref]
For a controlled comparison, additionally record the selected instrument range, probe factor, and input coupling; connect each instrument separately to the same electrical reference potential, and monitor the input rather than treating the supply label as an exact voltage. These are proposed controls for interpreting the documented procedures. A voltage indication and a resistance indication from the same digital multimeter do not independently validate its measurement calibration; a traceable reference comparison would be needed for that. Resistor tolerance describes component variation, while measurement uncertainty describes what is known about measured values. Using nominal arms alone can confuse mismatch with loading: equal nominal arms each within one-percent tolerance can give unloaded ratios from to .
The inference to test is specific: once actual arm values, input voltage, and meter input electrical resistance are accounted for, the readout should approach the loaded voltage divider prediction within the remaining measurement uncertainty. A discrepancy can instead reflect contacts, changing temperature, source droop, or an input that is not a fixed passive resistor. A source with positive series electrical resistance can be modeled by replacing with only when denotes its ideal internal voltage; when the actual voltage across the divider is measured, that terminal value is already . During settling, or with appreciable input capacitance and changing signals, the steady direct current calculation is insufficient. Neither institutional procedure supplies measured values for the original example above.