Oscilloscope 2026-10-14
An oscilloscope displays an electrical signal versus time. In an acoustic wavelength measurement, the microphone trace supplies the tone's frequency and oscillation amplitude as the detector changes position.
Oscilloscope probe 2026-10-18
A connection that couples a measured electric potential difference into an oscilloscope. Its electrical resistance and capacitance contribute to instrument loading.
RC low-pass laboratory 2026-10-18
The University of Colorado Boulder's Lab 3: Filters, §§7.1–7.4, specifies measuring the resistor and capacitor, then driving the assembled RC low-pass filter with a function generator. A two-channel oscilloscope reads actual input and output using high electrical impedance. Sweep the frequency, record the voltage amplitude ratio, and locate . A control compares a cable with a oscilloscope probe on a resistive voltage divider, isolating measurement capacitance before using the probe for the filter. A slow square wave then lets discharge timing estimate the time constant independently. Agreement supports the RC low-pass frequency response; source output impedance, probe loading, and instrument signal bandwidth limit it. These are prescribed measurements, with no measured dataset supplied here.[ref]
The two-channel setup can also test the predicted phase difference directly: for positive frequency , measure the output's time lag between corresponding features of the two sinusoids. For this single-stage model the phase lies between and , so choose the lag within a quarter cycle and calculate . At the cutoff frequency the predicted lag is one eighth of a cycle. This tests the timing part of the frequency response as well as the voltage ratio.
A real electrical load can draw electric current from the output. For a load resistor across the capacitor, Kirchhoff's current law instead gives
Therefore the loaded continuous-time frequency response is
Loading reduces the eventual gain and shortens the time constant. With the constructed values and , , , and the corner relative to the loaded low-frequency gain moves to . A step now approaches . A quicker rise alone would conceal the error in the final voltage.
With the ideal input voltage source set to zero, the output impedance seen by the electrical load is
This follows from Thévenin's theorem: the resistor and capacitor are then in parallel. A sufficient small-loading condition is over the frequencies of interest. An oscilloscope is itself an electrical load: its input electrical resistance and cable or probe capacitance can both change the response. Pure extra parallel capacitance increases and lowers the corner, whereas a finite parallel electrical resistance gives the different gain-and-corner changes above.
The definition of input also matters. If voltage is prescribed inside a source with series electrical resistance , the unloaded response to that internal source uses . If the actual voltage at the filter input terminal is measured, the ratio of output to that measured input still uses . Measuring both terminals avoids mistaking a source setting for the voltage delivered to the filter.
An actual teaching experiment in LM-0300 Fabry Perot Spectrum Analyser by Walter Luhs, §§3.1.1–3.1.2, prepares a green laser beam by aligning it near and far along a rail, then centring back-reflected light from adjustable curved mirrors. An actuator using piezoelectricity moves one mirror. A screen first displays interference fringes; a focusing optical lens and photodiode then replace the viewing arrangement. The oscilloscope records the photodiode signal and the piezo drive reference on separate channels, using the latter as its trigger. Alignment and mirror separation are adjusted to suppress extra transverse normal modes; source temperature and drive electric current are controlled. Repeated peaks calibrate the scan using the appropriate free spectral range, and their spacing divided by their full width at half maximum estimates finesse.[ref]
This experiment uses a confocal curved-mirror cavity, whose repeated peak families can have spacing in air; it must not be calibrated blindly with the parallel-mirror . Extra transverse normal modes can mimic source features. A finite measured peak width bounds source sharpness only after instrumental broadening is considered; it does not determine the source's intrinsic width by itself.[ref]
For an independent ideal scan calculation, hold a narrow laser wavelength fixed and vary the parallel-mirror gap at normal incidence. Advancing one resonance order requires . If changes uniformly over the measured interval, the time between repeated peaks divided by their full width at half maximum in time equals finesse. The cancellation of scan rate follows from : both peak spacing and peak width are divided by the same . It fails if the scan rate changes appreciably over that interval. A practical control in the University of Washington's motor-driven flat-mirror experiment is to check that successive peaks are equally spaced before using the time axis this way; its handout also warns that table strain and aggressive low-pass filtering alter the trace.[ref]
Thermal noise measurement 2026-10-18
MIT's laboratory guide mounts resistors on an aluminum box beneath a metal beaker shield, with short twisted leads. Its SR560 amplifier uses nominal gain before a Krohn-Hite – band-pass filter; an oscilloscope reads root mean square voltage. Calibration sweeps an attenuated sinusoid and measures the actual gain . Students alternate the resistor and a short circuit, subtracting squared readings , and disconnect the ohmmeter during noise acquisition. Lead and input capacitance enter the gain-squared integral. With settings unchanged, varying electrical resistance or temperature tests the predicted mean square. The guide prescribes repeated readings and warns against averaging noise waveforms to zero. This is a documented teaching protocol, not a supplied dataset of measured outcomes.[ref]
The subtraction requires care: shorting the input changes the source electrical impedance. An amplifier's input current noise can therefore produce different voltage noise in the two configurations; a shorted baseline alone does not remove that contribution. Gain drift, interference and insufficient observation time also limit the inference. These are limitations of interpreting the measurement, not corrections that can be inferred from the nominal signal bandwidth alone. In an independent analysis one would compare the measured difference with
where represents the measured shunt capacitance and excludes that separately modeled shunt. If calibration already includes the same shunt response, applying the denominator again counts the attenuation twice. The thermal noise worked example and both original figures illustrate ideal models; they claim no measured agreement or measured value of the Boltzmann constant.
Voltage divider Created 2026-10-13 Updated 2026-10-14
A voltage divider uses two resistors connected in series to obtain a fraction of an applied electric potential difference. Let the upper electrical resistance be , the lower one , and the input be relative to a common electrical reference potential. The output is the junction's electric potential difference relative to that same reference, across . The familiar ratio assumes zero output electric current: no external electrical load draws from the junction. In a steady direct current model with fixed resistors obeying Ohm's law and an ideal voltage source,
Here is the open-circuit voltage; correspond to in the header formula. With no junction branch, Kirchhoff's current law makes the same electric current pass through both resistors, and Ohm's law gives the two voltage drops. For , division gives ; if , the predicted output is zero but this ratio is undefined.
An original stipulated example has and . With zero output electric current, and . An attached electrical load instead gives ; a load gives . These are exact model predictions, not measured readings or specifications of a particular multimeter or oscilloscope.
Figure 1.
The same voltage divider has three different predicted outputs as the attached load changes
. Original circuit and readout comparison for ideal input and two resistors. Each finite connects across ; the displayed numbers are model predictions with ideal readout, not experimental data. Source. Details.
The divider is useful for scaling a voltage or converting a change in a sensor's electrical resistance into an electric potential difference. Its beauty is also its trap: multiplying both electrical resistances by the same positive factor preserves the unloaded ratio, but changes the effect of any fixed electrical load. A divider therefore supplies a predictable ratio under specified loading conditions; it is not a source that maintains that voltage for arbitrary output electric currents. A Wheatstone bridge compares two divider junctions, while Thévenin's theorem describes the divider's response to many different loads.
Southern Illinois University's ET 304A guide, Lab 4: Voltage Measurements and Meter Loading Effects, documents an instructive test: build equal-arm voltage dividers on a supply, then measure across the lower arm with a digital multimeter, analog multimeter, and oscilloscope. Starting with two resistors, repeat with equal pairs from to . It requires separate recorded readings, unloaded predictions, and predictions including each instrument's input electrical resistance. The guide obtains meter resistances from specification sheets and scope resistance from the input-panel marking. Equal ratios control the unloaded target while changing susceptibility to voltmeter loading. The provided tables are blank: the guide specifies a procedure, not measured outcomes.[ref]
Preparation and accuracy checks are made concrete by Berkeley's Lab 1: Introductory Experiments and Linear Circuits I. Its different divider uses nominal and resistors and a supply on a breadboard. Measure isolated resistors first, record actual supply voltage, recalculate with these measured values, then record the junction voltage with the digital multimeter. It explicitly compares actual values with resistor tolerance and the final voltage with a range-dependent accuracy bound from the instrument manual. This is an uncertainty check, not an independent measurement calibration.[ref]
For a controlled comparison, additionally record the selected instrument range, probe factor, and input coupling; connect each instrument separately to the same electrical reference potential, and monitor the input rather than treating the supply label as an exact voltage. These are proposed controls for interpreting the documented procedures. A voltage indication and a resistance indication from the same digital multimeter do not independently validate its measurement calibration; a traceable reference comparison would be needed for that. Resistor tolerance describes component variation, while measurement uncertainty describes what is known about measured values. Using nominal arms alone can confuse mismatch with loading: equal nominal arms each within one-percent tolerance can give unloaded ratios from to .
The inference to test is specific: once actual arm values, input voltage, and meter input electrical resistance are accounted for, the readout should approach the loaded voltage divider prediction within the remaining measurement uncertainty. A discrepancy can instead reflect contacts, changing temperature, source droop, or an input that is not a fixed passive resistor. A source with positive series electrical resistance can be modeled by replacing with only when denotes its ideal internal voltage; when the actual voltage across the divider is measured, that terminal value is already . During settling, or with appreciable input capacitance and changing signals, the steady direct current calculation is insufficient. Neither institutional procedure supplies measured values for the original example above.