Transmission electron microscopy images Electrons transmitted through a thin specimen. Electron diffraction from a selected region supplies complementary information about its atomic arrangement. A useful transmitted signal requires suitable specimen thickness.[ref]
Transmission Electron Microscope (TEM) by JEOL. JEOL, Another characteristic electron diffraction: thin specimens, images and diffraction patterns
The image also depends on which outgoing Electron directions reach the detector. Transmission electron microscopy contrast makes this selection explicit: changing the objective aperture can change an image without changing the specimen. Interpreting the result requires electron diffraction, a specified detector response and measurement calibration.
Transmission electron microscopy contrast is a difference in recorded signal between specimen regions under specified illumination, angular acceptance and detector settings. For a positive reference signal , one possible signed contrast convention isNegative means darker than the reference; it does not specify a surface height or chemical composition. The definition must accompany a numerical contrast value, since other conventions exist. A conspicuous feature need not establish fine spatial resolution: detecting a small signal difference and distinguishing two nearby features are different measurement questions.
An objective aperture selects outgoing Electron directions in the objective lens's back focal plane. Passing the central transmitted beam gives bright-field transmission electron microscopy; selecting a separated diffracted beam gives dark-field transmission electron microscopy.[ref]
In a paraxial approximation, an outgoing angle maps approximately to position in this focal plane, where is the objective focal length. A centred opening of radius therefore admits . Selecting a direction selects a component of the outgoing wavefunction; the subsequent image still assigns intensity to specimen positions. The opening is not a hole cut into a real-space specimen image. Its angular selection must be distinguished from selecting a specimen region for electron diffraction.
Objective aperture selection for bright-field transmission electron microscopy and dark-field transmission electron microscopy
. Original functional diagram. Lower relay boxes denote image formation, not literal ray paths. Scales and beam angles are illustrative. Source. Details.Bright-field transmission electron microscopy selects the central transmitted beam; deflection into excluded electron diffraction beams can make a region dark.[ref]
Bright-field image by JEOL. Definition: transmitted beam and dark diffracting regions
Dark-field transmission electron microscopy selects a diffracted beam while excluding the central beam. Regions contributing to that selected electron diffraction beam can appear bright.[ref]
Dark-field image by JEOL. Definition: selected diffracted beam and bright contributing regions
Diffraction contrast is image-intensity variation caused by spatially varying electron diffraction conditions. JEOL documents a silicon semiconductor-interconnect specimen at : its differently oriented crystals are imaged in bright-field transmission electron microscopy and same-area dark-field transmission electron microscopy, alongside an electron diffraction pattern. Selecting the marked diffracted beam makes the corresponding crystal bright after it was dark. The preparation route, thickness and exposure are not supplied. The changed objective aperture selection provides a comparison within one specimen; the readout supports orientation-dependent contrast rather than a map of surface height or chemical composition.[ref]
Diffraction contrast by JEOL. Definition and silicon example, panels a–c: 200 kV, same-area comparison and selected diffraction spot
This comparison is useful precisely because the material has not been replaced between the two images. It does not prove that thickness and chemical composition never influence transmission electron microscopy contrast. Instead, it supplies a concrete counterexample to assigning either quantity uniquely from brightness. A proposed extension is to repeat the first aperture setting after the second: recovery of the original contrast tests whether specimen motion or irradiation changed the comparison.
Consider an original illustrative model with exactly two outgoing scattering channels: a central beam and one diffracted beam . Let be specimen thickness and a constant coupling with units of inverse length. Normalize the probability amplitudes to incident probability current, assume equal longitudinal speeds, and specifyThese ordinary differential equations define a lossless toy model, not a fit to the silicon example or a complete theory of crystal scattering. Differentiating once more gives , soThe identity expresses conservation within these two scattering channels. Alternatively, substituting the ordinary differential equations into makes the two cross terms cancel. Here ideal bright-field transmission electron microscopy accepts fraction and rejects fraction ; ideal dark-field transmission electron microscopy accepts precisely the rejected channel. This complementarity requires the declared two-channel completeness and absence of loss.
Even with known , thickness is ambiguous: and both give . Near , ; extending that approximation through a full oscillation would erase the ambiguity incorrectly. The model isolates reversible transfer between probability amplitudes; it does not say that every thicker specimen must look darker.
Real dark-field transmission electron microscopy usually collects only a subset of the directions rejected in bright-field transmission electron microscopy. Other elastic scattering directions, inelastic scattering and detector losses prevent a general two-image conservation rule. To keep loss and background separate, extend the counting model explicitly. Let be the fraction absent from both recorded beam channels, describe the remaining fraction, be incident count per exposure, the detector efficiency and its mean additive background. Then the expected value of recorded count isFor separately acquired equal-illumination exposures with known , the two corrected mean counts sum to , not necessarily . Removal from these channels is not synonymous with physical absorption: an Electron can leave the chosen acceptance or change kinetic energy without being stopped in the specimen. Background subtraction removes a mean offset, but its counting variance remains.
JEOL's minimum-dose method focuses and corrects astigmatism away from the recording position before exposing it, reducing irradiation of the chosen field.[ref] A downstream objective aperture cannot undo irradiation that has already reached the specimen: rejecting more outgoing Electrons does not itself reduce specimen exposure.
For an original dose budget, define as incident Electron count per specimen area, rather than absorbed energy per mass. A uniform beam of electric current magnitude , illuminating area for time , giveswhere is the elementary charge. Choosing , and gives approximately incident Electrons, or . This is an invented exposure calculation, not a safe-dose recommendation. Focusing and searching exposures must also enter the accumulated budget.
If detected counts follow a Poisson distribution, a mean count has standard deviation . Ignoring background, accepted counts therefore have relative counting uncertainty, while have . Reducing accepted counts can reveal useful diffraction contrast while increasing counting uncertainty. It is useful to specify both contrast and count budget when discussing electron microscope resolution.
A proposed measurement calibration sequence records a beam-blocked detector background, an illuminated empty support region, and a specimen image with fixed voltage, exposure and detector gain. Test detector linearity with exposure time before comparing corrected intensities. Calibrate the image scale against a known spacing, as in microscope scale calibration, and independently calibrate electron diffraction spacing at its own lens settings; one calibration does not supply the other. Register compared regions, record objective aperture selection, specimen tilt and accumulated exposure, then repeat the first setting to test drift and irradiation changes. These are proposed controls, not undocumented steps attributed to JEOL's silicon example.
Articles by others on the same topic
Transmission Electron Microscope by LD SEF (2019)
Source. Images some gold nanopraticles 5-10 nm. You can also get crystallographic information directly on the same machine.