Electron microscope Created 2026-10-14 Updated 2026-10-20
An electron microscope forms a spatially enlarged representation of a specimen using electrons, electron lenses, and a detector. Its appeal is access to structure too fine for an ordinary optical microscope, but a short de Broglie wavelength alone does not guarantee fine electron microscope resolution. Increasing magnification can enlarge an existing blur without revealing any additional structure.
In transmission electron microscopy, an electron gun illuminates a thin specimen; magnetic electron lenses form and enlarge its transmitted image on a fluorescent screen or detector. A vacuum reduces collisions with gas. Lens electric currents control focusing and magnification.[ref]
Transmission Electron Microscope (TEM) by JEOL. JEOL, The difference between Electron Beam and Light: gun, vacuum, magnetic lenses and screen
Electron microscope resolution Created 2026-10-18 Updated 2026-10-20
Electron microscope resolution concerns distinguishable specimen detail under a specified contrast criterion. It depends on diffraction, lens defects, electron energy spread, specimen motion and detection. A smaller de Broglie wavelength permits finer detail in principle, but neither wavelength nor magnification is the instrument's measured resolution.
For an electron accelerated from negligible initial kinetic energy through a positive electric potential difference , combine , the relativistic energy-momentum relation and the de Broglie wavelength:Here is the Planck constant, the electron mass, the elementary charge, and the speed of light. This includes the relativistic correction omitted by the low-voltage electron accelerating-voltage wavelength approximation. With and SI defining values of , calculated wavelengths are approximately , and at , and respectively. The rounded mass agrees with the recommended value.[ref]
For a circular aperture admitting a small semi-angle , the Rayleigh criterion suggests a diffraction-limited resolution scale . At and , this gives , already much larger than the wavelength. This is an ideal aperture estimate. Transmission electron microscopy with coherent illumination and a particular specimen requires its own image-contrast analysis; this scale alone is not a universal two-feature resolution test.
An original illustrative blur model adds competing effects:Define as a characteristic transverse blur from the electron-lens analogue of spherical aberration, and as one from energy-dependent focusing, the analogue of chromatic aberration. The coefficients have units of length, is a chosen characteristic kinetic energy spread, and is a fixed residual blur. Adding their squares is a declared heuristic, not a measured point spread function or a rigorous resolution theorem. The model omits contrast transfer, specimen scattering and signal statistics, so its minimum is only an estimate within its assumptions.
Choose , and at . At , the four contributions are about , , and , giving . At , the diffraction contribution falls to but the assumed spherical aberration contribution grows to . Thus opening the aperture indefinitely worsens this model's image. All these values are calculations with chosen parameters, not observations of any microscope.
Electron microscopy Created 2026-10-14 Updated 2026-10-20
Electron microscopy is the practice of preparing specimens, producing images with an electron microscope, and interpreting their contrast. A dark region can reflect electron deflection out of the admitted beam rather than complete absorption. A useful image therefore needs a stated preparation, detection signal and length calibration; brightness by itself does not identify a material. Transmission electron microscopy probes thin specimens, whereas a scanning electron microscope can examine the exposed surface of a bulk specimen.[ref]
Scanning Electron Microscopes (SEM) by JEOL. JEOL, Observation of the micro world using electrons and Various Function of SEM: transmitted versus surface-generated signals
Focused ion beam milling 2026-10-20
Removal of material with a focused ion beam, for example to prepare a thin transmission electron microscopy specimen. Implantation, amorphization and redeposition can require a low-energy finishing step.
Freeze-etching 2026-11-05
Freeze-etching removes surface ice from a frozen specimen by sublimation under vacuum, often after freeze-fracture microscopy has exposed a new face. Removing that ice can reveal structures initially hidden beneath it. Fracturing opens the specimen mechanically; etching subsequently changes the exposed surface by removing material.[ref]
Brief Introduction to Freeze Fracture and Etching by Gisela Höflinger. Opening definition; Sublimation and condensation of water, freeze etching and contamination
A physical preparation holds the specimen cold in a controlled vacuum, permits a chosen amount of ice removal, and then applies the coating appropriate for electron microscopy. The exposed block face can be examined with a scanning electron microscope, or a supported replica can be examined by transmission electron microscopy. Coating and transfer preserve information about a prepared face rather than making the original bulk material transparent.[ref]
Brief Introduction to Freeze Fracture and Etching by Gisela Höflinger. Etching and coating discussion; SEM block-face and TEM replica workflows
For an own geometrical model, suppose a flat ice surface recedes uniformly at while an embedded feature does not sublime. A feature initially below the surface is first exposed atAfter the ice has receded , leaving the feature protruding in this model. The numbers are stipulated, not a measured etching rate or a recommended preparation setting.
A useful control compares matched specimens after different etching durations while recording specimen temperature, transfer conditions and coating geometry. Progressive exposure would be consistent with ice removal, but a different appearance can also arise from changed preparation or contamination. Condensation deposits material rather than removing it, so contamination must not be mistaken for an etched feature.
The ideal recession calculation assumes a uniform rate and an unchanged embedded feature. Real local composition and heat transfer can violate both assumptions. Interpret image contrast together with etching and coating history rather than assigning every relief feature directly to the initial specimen.
Freeze-fracture electron microscopy 2026-11-05
Freeze-fracture electron microscopy reveals structures along a fracture through a frozen specimen, commonly within a biological membrane. The exposed surface is shadowed with a thin platinum-carbon deposit; supporting carbon stabilizes the replica, which is cleaned and examined by transmission electron microscopy. The image is of this replica, not a direct view through the original thick tissue.[ref]
Freeze-fracture electron microscopy by Nicholas J. Severs. Abstract and protocol, fracture, metal-carbon replica and TEM examination
A practical preparation rapidly freezes tissue, fractures it in a cold vacuum apparatus, deposits the replica and transfers the cleaned film to a microscope grid. Membranes often split through their hydrophobic interior, exposing particles and complementary depressions associated with membrane proteins. The fracture plane is not generally the original outer cell surface.
This method is useful because it displays lateral membrane organization that a thin cross-section can hide. However, freezing artifacts, shadowing direction and fracture-path selection affect appearances. Combining replica structure with immunoelectron microscopy can help identify selected proteins, while an unlabeled intramembrane particle should not be assigned a molecular identity from shape alone.
Freeze-fracture microscopy 2026-11-05
Freeze-fracture microscopy exposes internal structure by breaking a frozen specimen and examining the revealed face with electron microscopy. Freezing makes it possible to prepare a surface through material that would otherwise change substantially during ordinary handling in a vacuum. Physical applications include inspecting internal interfaces in frozen mixtures; the observed face is a fracture-selected surface rather than an arbitrary transparent section.[ref]
Brief Introduction to Freeze Fracture and Etching by Gisela Höflinger. Opening definition and physical preparation discussion, frozen fracture and material interfaces
A preparation workflow freezes the specimen, fractures it while cold, and maintains protected low-temperature transfer. Optional freeze-etching removes surface ice by sublimation under vacuum. A thin oblique metal deposit supplies topographic contrast, with carbon providing support where a replica is required. The prepared block face can be examined with a scanning electron microscope, or a replica can be examined by transmission electron microscopy.[ref]
Brief Introduction to Freeze Fracture and Etching by Gisela Höflinger. Sublimation and condensation discussion; freeze-fracture coating and SEM block-face versus TEM replica routes
A geometrical illustration explains directional shadowing. Model deposition as parallel trajectories making angle above a flat surface. A protrusion of height casts a nominal shadow lengthFor an illustrative , gives , whereas gives about . These are ideal geometric predictions, not measured features or a calibration of a coating instrument. Finite coating thickness, trajectory spread and surface shape affect real contrast.
A useful preparation-control experiment compares otherwise matched frozen samples fractured and coated at different orientations, recording temperature history, etching time and deposition direction. Features that change with coating direction may reflect shadow geometry; features that change with freezing or etching conditions may reflect preparation. The fracture path can favor weak interfaces, and deposited material can obscure small structures, so apparent topography is not automatically an unbiased reconstruction of the original specimen.
The freeze-fracture electron microscopy article describes the replica route in more detail. Here the physical preparation and block-face comparison explain why the microscope image must be interpreted together with the specimen's preparation history.
Gisela Höflinger 2026-11-05
Gisela Höflinger is an author on electron microscopy specimen preparation. Her manufacturer-hosted profile identifies a background in industrial engineering and responsibility for a coating-instrument line at Leica Microsystems. Her 2014 tutorial Brief Introduction to Freeze Fracture and Etching by Gisela Höflinger explains the physical preparation steps behind images of frozen specimens.[ref]
https://www.leica-microsystems.com/science-lab/authors/detail/gisela-hoeflinger/. Originating author profile: identity, industrial-engineering background and EM-sample-preparation role
The tutorial distinguishes freeze-fracture microscopy, which exposes an internal face by breaking a frozen specimen, from freeze-etching, which removes surface ice by sublimation. It describes angled heavy-metal deposition for topographic contrast, followed by a supporting carbon layer, and distinguishes viewing a block face by scanning electron microscopy from viewing a replica by transmission electron microscopy.[ref]
Brief Introduction to Freeze Fracture and Etching by Gisela Höflinger. Definition; Sublimation and condensation of water; Freeze fracture to generate images.
A geometrical example explains why coating angle matters. For a narrow deposition beam at angle above a flat surface, an ideal step of height casts a shadow of length . With and , that length is about 23 nm. This is a simplified shadow construction, not a reconstruction of data from the tutorial. It connects Höflinger's preparation topic to a practical problem: the recorded contrast depends on both specimen structure and the way its surface was coated.
Microtome 2026-11-05
A microtome is an instrument for microtomy: a supported specimen and a knife move relative to one another to separate a thin physical section. The apparatus combines a specimen holder, a knife holder, a cutting motion and a controlled feed that presents new material for the next cut. An ultramicrotome is specialized for very thin sections, including sections for transmission electron microscopy.
A concrete materials-research installation at Yale uses a Leica Ultra UC7, with a trimming mill and glass knife maker. Its listed specimen materials include polymers, material fibres and soft metals.[ref]
Microtome at the Yale Institute for Nanoscience and Quantum Engineering. About the instrument, accessories and materials list
For the thin-section cutting cycle, JEOL describes advancing the specimen by a preset feed and moving it against a diamond knife.[ref]Ultramicrotomy by JEOL. Glossary, specimen advance and downward cutting stroke
The supports must hold the specimen and knife securely enough that commanded motion is meaningfully related to the material removed.An original ideal feed-drive calculation illustrates the distinction between mechanism and outcome. Suppose a hypothetical drive advances a lead screw by pitch per revolution, has reduction ratio motor revolutions per screw revolution, and receives equally spaced command steps per motor revolution. Its nominal advance per step isFor chosen , and , this gives . Four steps command a nominal section feed of . This is an illustrative mechanism, not a claim about the UC7's internal drive. Motor motion does not independently establish the physical section thickness: backlash, elastic motion, thermal expansion and section compression can intervene.
A bounded apparatus experiment would cut a stable polymer reference at several feed settings, measure the recovered physical section thickness independently, and retain the cutting order and operating conditions. Agreement of motor counts with the command tests the drive readout; agreement of measured sections with the intended thickness tests more of the complete process. The distinction parallels the need for an independent readout when testing positioning repeatability.
Microtomy 2026-11-05
Microtomy prepares thin physical sections with a microtome, exposing interior structure for microscopy. The knife separates a layer from a supported specimen while a controlled feed determines the nominal depth increment. Physical section thickness, section orientation and cutting-induced deformation all enter the interpretation of an image. Ultramicrotomy is the specialized thin-section technique used in applications such as transmission electron microscopy.
The preparation method also serves materials research. Yale's instrument description lists a Leica Ultra UC7 with a trimming mill and glass knife maker, and identifies polymers, material fibres and soft metals among suitable specimen materials.[ref]
Microtome at the Yale Institute for Nanoscience and Quantum Engineering. About the instrument, accessories and material list
Manufacturer guidance discusses polymer blends and composite materials, along with the possibility of section compression and cutting artifacts.[ref]Ultramicrotomy Techniques for Materials Sectioning by Leica Microsystems. Opening materials discussion; How to prepare high-quality sections of polymers
A useful teaching experiment would compare cuts through a straight, known circular material fibre at different section orientations. A normal vector to the cut aligned with the fibre axis gives a circular profile. An oblique cut gives an ellipse even when the fibre has suffered no deformation. Image the recovered sections with an optical microscope, measure profile dimensions with an established image stack axis calibration, and record the angle between the section's normal vector and the fibre axis. This tests the geometry of the section rather than assuming that every measured elongation is a physical distortion.
For an original ideal calculation, model the fibre boundary as an infinite circular cylinder of radius , with axis along the unit vector , where . Parallel section planes have equation . The squared Euclidean distance to the axis is , obtained by subtracting the squared axial orthogonal projection from . At the section plane, the boundary therefore satisfiesThis oblique section of a circular cylinder is an ellipse with semi-major and semi-minor axes and areaThe area of an ellipse gives the last expression. For chosen and , the semi-major and semi-minor axes are approximately and . The profile area is times the perpendicular-profile area, an increase of about 41.4 percent entirely explained by section orientation.
The center of an ellipse in this model has coordinate . In an ordered image stack, two sampled section planes separated by consequently have lateral centre displacement . At the chosen and , this displacement is . Here is the spacing between sampled planes, which can differ from the physical section thickness if some sections are not imaged.
The calculation assumes a circular, straight fibre, known section geometry and negligible physical section thickness relative to resolved features. One elliptical profile alone cannot distinguish an oblique circular fibre from a differently shaped fibre or an unknown deformation. Comparing known reference features and following their positions through an image stack provides additional evidence. Image registration can align images, but an inferred shift may contain real specimen geometry as well as instrument motion; removing every shift would erase the tilt information in this example.
Nicholas J. Severs 2026-11-05
Nicholas J. Severs is a researcher in cellular ultrastructure whose 2007 freeze-fracture electron microscopy protocol lists an affiliation with the National Heart and Lung Institute at Imperial College London. His methodological work explains how fracture planes reveal the internal organization of cell membranes.[ref]
Freeze-fracture electron microscopy by Nicholas J. Severs. Author affiliation and abstract, Nature Protocols 2, pp.547–576
The protocol freezes a biological specimen, fractures it, deposits platinum and carbon to form a replica, and cleans that replica for transmission electron microscopy. The microscope thus examines a durable record of the exposed surface. Optional removal of ice by sublimation exposes additional relief. This makes preparation a scientific part of the measurement: the observed surface depends on where the specimen fractured and which material the procedure exposed.[ref]
Freeze-fracture electron microscopy by Nicholas J. Severs. Abstract, four preparation stages, etching and planar membrane views
Transmission electron microscopy accelerates electrons through a thin sample, then uses electron lenses to form an image or diffraction pattern. Elastic scattering encodes projected structure and potential; inelastic scattering carries excitation and chemical information. Thin specimens reduce unwanted multiple scattering and absorption, although coherent dynamical scattering can still matter in a crystal. High-resolution phase-contrast images depend on defocus and the transfer function as well as the specimen; they are not direct height maps.
The relativistic wavelength is . A shorter wavelength permits fine resolution but does not guarantee it. A schematic probe/image blur budget contains diffraction , spherical aberration and chromatic aberration , plus source size, instabilities, specimen motion and radiation damage. Optimizing the aperture angle balances these terms; increasing accelerating voltage can also alter damage and contrast.
Bright-field imaging accepts the central transmitted beam or its neighborhood. Regions that scatter electrons outside that acceptance can appear dark. Conventional dark-field imaging selects a diffracted beam. Annular dark-field electron imaging in a scanning TEM instead collects a range of scattered angles while a probe is scanned. A sufficiently high-angle detector often gives strong atomic-number sensitivity and more nearly incoherent contrast; thickness, channeling and detector geometry prevent a universal conversion of intensity to composition.
Electron energy loss spectroscopy measures energy lost by transmitted electrons. Core-ionization thresholds identify elements, while near-edge fine structure probes bonding and unoccupied states. At successive probe positions, one can record a spectrum image or form a map from a selected edge after background subtraction. Plural scattering, edge overlap, delocalization, specimen thickness and counting statistics limit the interpretation. Chemical specificity comes from excitation energies and spectral structure, not from brightness alone.
A development already established before this 2012 paper is aberration-corrected transmission electron microscopy: multipole correctors compensate leading lens aberrations, allowing useful larger angular acceptance and finer probes. Correction does not remove every chromatic, stability or specimen limit. A concrete application was atom-by-atom annular-dark-field analysis of monolayer boron nitride with substitutional defects, resolving chemical differences between sites in a nonuniform lattice.[ref]
Atom-by-atom structural and chemical analysis by annular dark-field electron microscopy by O. L. Krivanek and others. 2010, low-voltage aberration-corrected STEM of monolayer boron nitride
Combining structural maps with spatially resolved core-loss spectra supplies another route to distinguish composition and bonding at interfaces. Both exploit a calibrated imaging or spectroscopy response rather than assuming every visible spot is chemically self-identifying. Past exam of the materials science course of the University of Cambridge 2019 Paper 1 2 a Solution 2026-10-20
Features only a few tens of nanometres wide lie below the ordinary visible-light diffraction limit, approximately . The microscope cannot resolve individual rods, and unpolarized illumination does not directly reveal the orientation-dependent optical response. Scanning electron microscopy and atomic force microscopy can resolve the surface pattern. Transmission electron microscopy is another possible choice if a suitably thin specimen can be prepared.
In transmission electron microscopy, the electron objective lens forms an intermediate real-space image and intermediate/projector lenses enlarge it onto a screen or camera. The total optical magnification is the product of the relevant stage magnifications. The user changes imaging-lens excitations to select magnification, while separately controlling illumination and focus. Display resizing introduces an additional enlargement, so the recorded pixel calibration matters.
In scanning transmission electron microscopy, a focused probe is rastered across the specimen and a detector signal is assigned to each probe position. The displayed magnification is display width divided by scanned specimen width. The user increases magnification by reducing raster-deflection amplitude, independently of choosing a smaller incident probe with the probe-forming lenses. TEM changes image-forming lens magnification; STEM changes the raster field of view. A smaller field or more pixels does not create finer physical resolution than the probe, specimen scattering and detection permit. A calibrated real-space scale is more transferable than the apparent size on a monitor.[ref]
Scanning Electron Microscope A To Z by JEOL. printed p.7, raster-width/display-width magnification; this geometric relation also applies to scanned transmitted-electron imaging
Disperse a very small amount of the dry powder in a clean volatile solvent compatible with aluminium oxide, use brief gentle sonication to separate loose agglomerates, and place a dilute drop on a holey carbon-coated transmission electron microscopy grid. Wick away excess liquid and dry thoroughly. Look first at low magnification and select isolated particles or thin edges over holes, rather than an electron-opaque pile. Record a solvent-only grid to identify contamination and compare several independently prepared areas.
This is quick and preserves individual particles without mechanical thinning. However, spray pyrolysis can make hollow or porous particles that may fracture during sonication; sedimentation and drying can select sizes and aggregate particles. aluminium oxide is insulating, so charging or beam-induced changes can complicate images, and the carbon support can overlap weak contrast. Vary dose and dwell time to test stability. A projected image of one selected thin particle is not an unbiased three-dimensional size or porosity distribution of the whole powder.
Use a low-dose, dose-efficient acquisition rather than assuming that a lower accelerating voltage is always gentler. Focus and align on a sacrificial neighbouring region, blank the beam between exposures, illuminate only when recording and use an efficient detector. A broad, low-current-density transmission electron microscopy exposure can avoid a high instantaneous local dose; scanning can also work if probe current, dwell time and repeated visits are controlled.
Radiolysis starts with inelastic scattering and ionization. For an appropriate thin specimen, increasing the electron energy can reduce the inelastic interaction probability and energy deposition per incident electron, so a relatively high voltage can help. However, increasing voltage can also activate knock-on displacement. The optimum therefore depends on the material and the amount of useful information obtained per damaging event; alone does not determine it. Cooling may reduce diffusion-driven chemical changes but does not eliminate primary ionization. An experimental study of nonbiological layered polymer films explicitly found that the voltage choice must balance damage with image signal.[ref]
Optimal acceleration voltage for near-atomic resolution imaging of layer-stacked 2D polymer thin films by Baokun Liang and others. 2022, accelerating-voltage comparison and discussion of the inelastic scattering cross-section
Moiré fringes arise when two overlapping periodic structures contribute closely spaced reciprocal lattice vectors to the transmission electron microscopy image. Their relative phase varies slowly across the image, producing a coarse modulation with a period much larger than either atomic spacing. The structures may have different lattice spacings, a small relative rotation, or both. The modulation is an interference or beat phenomenon; its spacing is not itself a new atomic-plane spacing. Its orientation is perpendicular to the difference vector of the two contributing projected reciprocal-lattice vectors.
Reducing void swelling means delaying stable cavity growth and reducing its rate after incubation. Several approaches act on different parts of that mechanism.
Choose an alloy and initial microstructure with favorable defect sink populations. Grain boundaries, stable interfaces, precipitates and a controlled dislocation structure can remove mobile defects and promote recombination before they feed voids. The relevant quantity is sink efficiency and bias, rather than the number of visible microstructural features alone.
Cold working can increase the starting dislocation density and prolong incubation in suitable austenitic steels. Irradiation and thermal recovery can later change that structure, so an initial benefit need not persist indefinitely. Solute additions and controlled precipitation can also change vacancy mobility, defect trapping and cavity nucleation. Their effectiveness must be assessed together with corrosion, ductility and irradiation-assisted changes in composition.
Where other constraints permit, choose a more swelling-resistant material family, such as an appropriate ferritic-martensitic steel, rather than assuming every stainless steel has the response of an austenitic grade. This introduces its own tradeoffs in toughness, temperature window and compatibility.
Manage the operating temperature and accumulated dose to avoid the most unfavorable swelling regime, and account for transmutation-gas production. Neither simply minimizing temperature nor adding gas traps guarantees an improved component: the former can increase hardening and embrittlement, and gas-stabilized cavities can become nuclei for later swelling.
Useful comparisons require specimens with controlled composition and heat treatment, measured dose and temperature histories, and both dimensional measurements and transmission electron microscopy. Matching final dose alone does not establish that two irradiation conditions produce the same swelling.
Serial-section electron microscopy 2026-11-05
Serial-section electron microscopy reconstructs a volume from an ordered image stack of thin physical slices. In the transmission electron microscopy implementation, an ultramicrotome cuts successive slices, which are collected on a support and imaged through the specimen. Image registration aligns corresponding features; image segmentation identifies structures to follow between slices. JEOL's glossary describes sections of 40–100 nm and illustrates the separate registration and segmentation stages.[ref]
The motivation is geometrical: one isolated slice cannot show whether a feature continues, divides, or ends outside that slice. An ordered image stack provides observations along the third coordinate, although finite section-thickness anisotropy and missing material limit what can be inferred. A change between neighboring images may come from specimen structure, a slice deformation, or imperfect image registration.
For a chosen numerical example, suppose each of consecutive sections has uniform thickness , with lateral pixel pitch . Ignoring lost material and thickness variation, the sections coverThe voxel sampling is 12.5 times coarser along the cutting direction. These are chosen dimensions, not measurements from either cited experiment. The first and last section centers are only apart: center-to-center extent differs from total represented thickness. Neither spacing is automatically the achieved spatial resolution.
Image resampling onto a cubic grid can make an image stack easier to display, but interpolation supplies no additional independent observations. Likewise, smooth image registration cannot resolve missing-section ambiguity in volume reconstruction: a small connecting feature could lie wholly inside lost material. Checking a reconstructed connection therefore needs evidence in the acquired slices, not just a visually smooth rendered surface.
A historically important automation change illustrates the distinction between section and surface imaging. In 2004, Winfried Denk and Heinz Horstmann placed an ultramicrotome with a diamond knife inside a scanning electron microscope. Their serial block-face scanning electron microscopy imaged the exposed block surface between cuts instead of collecting the removed sections. They reported stacks several hundred layers deep, 50–70 nm sectioning, and typically less than 10 nm lateral position jitter. These observations tested an apparatus for acquiring closely aligned serial images; they did not remove the finite spacing between observations.[ref]
Serial Block-Face Scanning Electron Microscopy to Reconstruct Three-Dimensional Tissue Nanostructure by Winfried Denk and Heinz Horstmann. Abstract; Results, lateral-position stability; Materials and Methods, Microtome
A mechanical detail makes the measurement chain concrete. Their specimen-advance lever reduced the motorized micrometer's motion by 0.11. Thus its stated position measurement uncertainty of 50 nm contributed after that reduction; other sources of section-thickness error still required assessment.[ref]
Serial Block-Face Scanning Electron Microscopy to Reconstruct Three-Dimensional Tissue Nanostructure by Winfried Denk and Heinz Horstmann. Materials and Methods, Microtome, specimen-advance mechanism
Transmission electron microscopy Created 2026-10-18 Updated 2026-10-20
Transmission electron microscopy images electrons transmitted through a thin specimen. Electron diffraction from a selected region supplies complementary information about its atomic arrangement. A useful transmitted signal requires suitable specimen thickness.[ref]
Transmission Electron Microscope (TEM) by JEOL. JEOL, Another characteristic electron diffraction: thin specimens, images and diffraction patterns
The image also depends on which outgoing electron directions reach the detector. Transmission electron microscopy contrast makes this selection explicit: changing the objective aperture can change an image without changing the specimen. Interpreting the result requires electron diffraction, a specified detector response and measurement calibration.