An actual teaching experiment in LM-0300 Fabry Perot Spectrum Analyser by Walter Luhs, §§3.1.1–3.1.2, prepares a green laser beam by aligning it near and far along a rail, then centring back-reflected light from adjustable curved mirrors. An actuator using piezoelectricity moves one mirror. A screen first displays interference fringes; a focusing optical lens and photodiode then replace the viewing arrangement. The oscilloscope records the photodiode signal and the piezo drive reference on separate channels, using the latter as its trigger. Alignment and mirror separation are adjusted to suppress extra transverse normal modes; source temperature and drive electric current are controlled. Repeated peaks calibrate the scan using the appropriate free spectral range, and their spacing divided by their full width at half maximum estimates finesse.[ref]
This experiment uses a confocal curved-mirror cavity, whose repeated peak families can have spacing in air; it must not be calibrated blindly with the parallel-mirror . Extra transverse normal modes can mimic source features. A finite measured peak width bounds source sharpness only after instrumental broadening is considered; it does not determine the source's intrinsic width by itself.[ref]
For an independent ideal scan calculation, hold a narrow laser wavelength fixed and vary the parallel-mirror gap at normal incidence. Advancing one resonance order requires . If changes uniformly over the measured interval, the time between repeated peaks divided by their full width at half maximum in time equals finesse. The cancellation of scan rate follows from : both peak spacing and peak width are divided by the same . It fails if the scan rate changes appreciably over that interval. A practical control in the University of Washington's motor-driven flat-mirror experiment is to check that successive peaks are equally spaced before using the time axis this way; its handout also warns that table strain and aggressive low-pass filtering alter the trace.[ref]
Single-slit diffraction Created 2026-10-14 Updated 2026-10-18
Single-slit diffraction is the spreading of light transmitted by one slit. Consider an ideal rectangular slit of width , illuminated normally by a uniform plane wave with a single wavelength in the surrounding medium. This monochromatic wave has uniform wave amplitude and equal incident wave phases across the slit, providing the required optical coherence. In the scalar Fraunhofer diffraction model, the normalized angular wave intensity is
Here is measured from the forward normal in the plane across the slit width. The sinc function uses the unnormalized convention . The limit makes the central value one. This ideal envelope neglects additional direction-dependent propagation and detector factors; near the forward direction it describes the normalized screen irradiance. [ref].
The noncentral zeros satisfy
The restriction follows from the range of the sine of a real angle. The first minima have when . If , this model has no noncentral zero in the forward hemisphere; equality puts the first zeros at grazing directions, outside the small-angle approximation and off any finite perpendicular screen. [ref].
For , define the central width as the separation between the two first minima. Its angular width and its width on a perpendicular screen are
The angular approximation uses radians and requires ; the screen approximation uses . The tangent relation is exact straight-line geometry, even though the underlying Fraunhofer diffraction field is approximate. This width is distinct from the full width at half maximum.
For , and , the positive first minimum is
The dimensionless far-field parameter is . Doubling the slit width to gives and parameter ; both curves below use the Fraunhofer diffraction approximation. Choosing halves those parameters and doubles both screen widths.
Figure 1.
Normalized single-slit diffraction on a screen: a narrower slit produces a wider central maximum
. Details.
Each curve is normalized to its own ; the arrows measure between first minima. The calculation uses before evaluating the sinc function, rather than replacing by . The lower diagram shows the exact screen geometry with the vertical scale enlarged. Normalization removes the change in peak brightness when the slit width changes.
The wave interference explanation is especially simple: divide the slit into equal narrow strips and add their complex amplitudes, rather than their wave intensities. With wavenumber , the continuous slit gives
Integrating the complex exponential gives the endpoint difference; its continuous value at zero is one. Taking the squared complex modulus gives the stated wave intensity. This is the Fourier transform of an interval indicator expressed in angular coordinates. Feynman's continuous-source argument obtains the same pattern by letting the number of sources tend to infinity. [ref].
At the positive first minimum, the wave phases span exactly across the slit. Every strip in one half has a partner away, with equal wave amplitude and phase difference , so the pairs cancel. At , all the wave phases align instead. [ref].
Figure 2.
Complex amplitudes across a slit: aligned contributions at broadside and cancellation at the first minimum
. Details.
The drawing uses equally weighted midpoint samples , with . At the first minimum their complex amplitudes are a common rotation of the eight roots of unity, so their head-to-tail polygon closes exactly. This finite sum illustrates cancellation; the continuous integral, rather than eight point sources, defines the slit pattern at general angles. The circle shows the continuous range of wave phases; coloured arrows show the eight samples.
A narrower aperture requires a larger angle to accumulate the cancelling phase difference, so narrowing the slit widens the beam. This is a useful limit on geometrical optics: confining light more tightly at an opening increases its subsequent angular spread. At fixed incident irradiance and slit height, reducing also reduces transmitted power, exposing a practical balance between diffraction-limited resolution and light throughput. In a diffraction grating, this single-slit diffraction envelope modulates the wave interference between different slits.
A motivating measurement used the McMath solar optical telescope at Kitt Peak. The Sun's image drifted across a -high, -wide optical slit; a double-pass spectrograph selected wavelength intervals from wide at to at . An EMI 9750 QB photodetector or silicon photodiode supplied a digitized signal: samples were summed into 4096 positions per crossing, including dark and off-limb sky readings. [ref]
Controls included daily realignment of optical telescope mirrors, wavelength settings between spectral lines insensitive to the rotational Doppler shift, selection of detectors without detectable hysteresis, and measured electronic nonlinearity corrections below of maximum signal. Limb-position calibration combined centres of gravity of the edge derivative profiles with almanac semidiameters; the preliminary edge shifts were consistent with astronomical seeing of approximately arcseconds full width at half maximum. [ref]
The decreasing centre-to-limb readout gives a specific intensity profile, constraining stratification of the photosphere under a radiative transfer model. Astronomical seeing, instrumental scattering of light, and solar granulation limit this inference; brightness alone does not uniquely determine temperature versus geometric height. These observations need not follow the stipulated linear limb-darkening law.
For , the Fourier transform of is . A weak interaction between energy levels of gap gives final probability amplitude , where is the transition matrix element. Only when does the single resonant lobe give an approximately squared Lorentzian line shape excitation profile. Its full width at half maximum is in angular frequency.
Temporal optical coherence Created 2026-10-14 Updated 2026-10-18
Temporal optical coherence compares a field at one position with a delayed copy. In a Michelson interferometer, unequal round-trip lengths supply this delay. Even a source with rapidly changing wave phase can give high interferometric visibility close to zero delay, because both arms then deliver nearly the same source history. A source narrowed in optical spectrum generally retains correlation over larger delays; a single width does not describe sources with several separated lines. [ref].
For a stationary scalar field, choose the carrier convention , with frequency in cycles per second. Let be its spectral density per unit frequency, with . Its normalized field correlation is the Fourier transform
The optical spectrum weights the rotating contributions; a broad spread of frequencies allows them to cancel at short delays. This is the stationary field correlation–spectrum relation. The relevant spectrum is that of the optical field in the detected mode, rather than the low-frequency fluctuations of a detector's light intensity. [ref].
Here are two stipulated line profiles, each centered at with full width at half maximum . Write and use normalized profiles on the whole detuning line:
The first is a Lorentzian, mathematically a Cauchy distribution; the second is a Gaussian function. Taking their Fourier transforms gives, with the carrier factor omitted from the magnitudes,
These are calculated model results, not measured line profiles. The detuning approximation treats the narrow line as extending to arbitrarily large positive and negative ; for optical applications require . Exact spectra of the complex positive-frequency field have no negative absolute frequencies. Changing to angular frequency replaces by ; it must not silently preserve the same numerical constant.
Figure 1.
Michelson interferometer geometry and calculated interferometric visibility versus delay for stipulated line profiles
. Details.
Equal detected single-arm light intensities give interferometric visibility equal to the correlation magnitude. Doubling the Gaussian full width at half maximum from to halves every fixed-contrast delay. At the model gives Gaussian interferometric visibility approximately at and at ; the Lorentzian gives . The dashed curve instead keeps the Gaussian correlation and changes to , multiplying all interferometric visibilities by . Its lower contrast at zero delay therefore comes entirely from intensity imbalance.
Time-bandwidth product Created 2026-10-18 Updated 2026-10-20
The product of a specified temporal width and spectral frequency width. State whether each width refers to power or field amplitude and whether it is a full width at half maximum or a standard deviation.
T. J. Spencer is a coauthor of the Central Laser Facility's1999 report on a single-shot instrument for third-order intensity autocorrelation and optical pulse temporal contrast measurements.[ref]
An original calculation illustrates a pulse-characterization issue relevant to that work. Let a Gaussian pulse have intensity
Completing the square in its third-order intensity autocorrelation gives
The pulse's intensity full width at half maximum is , while the correlation width is . Their ratio is therefore
A measured correlation width cannot simply be reported as the original pulse width. The conversion depends on the assumed pulse shape and the correlation being measured. This Gaussian example explains that distinction; it is not a reconstruction of Spencer's individual experimental tasks or a fit to the1999 data.
Choose a closed two-level system, equal stimulated absorption and emission strengths, a monochromatic pump, and no extra dephasing. Let be the lower and upper populations, and let spontaneous emission remove upper population at rate . In this convention the natural linewidth is full width at half maximum in angular frequency, its half width at half maximum is , and its full width at half maximum in hertz is . All below are angular detunings.
Let be the on-resonance saturation parameter. An effective stimulated transition rate that reproduces the monochromatic steady response is
Here describes each stimulated direction, while describes irreversible spontaneous emission. The rate equation gives the steady populations; it is not a description of transient Rabi oscillations. Setting gives
and the population difference relevant to net weak-probe photon absorption is
The pump depletes the lower population by , but reduces the population difference by : stimulated emission contributes as well as missing ground-state absorbers. On resonance, gives , , ; gives , , . As , the two populations approach , rather than producing a population inversion. Far off resonance, and .
For moving atoms use in these pump-prepared populations. A weak counterpropagating probe interrogates them near . A simple population-only approximation to its absorption coefficient averages the weak-probe Lorentzian line shape times over the Maxwell-Boltzmann velocity distribution. It neglects interference between the beams, changes of velocity during illumination, and multilevel optical pumping. Those restrictions matter for real alkali chemical atoms.
An unresolved Gaussian pulse-width estimate occurs when the measured correlation width and characterized response do not establish a positive pulse width with useful precision under the assumed Gaussian model. It is a limitation of the measurement and inference, not evidence that the pulse has zero duration.
In Gaussian pulse-width deconvolution uncertainty, the inferred full width at half maximum is
For , a positive central value exists. Whether the data resolve it depends on the uncertainty in , the response model and their covariance, not just on the sign of the central radicand.
Take original central values and . They give and . Independent standard uncertainties give, to first order,
This uncertainty is almost three times the central difference. Propagating once more through the square root gives ; that symmetric approximation is poorly suited to the boundary . The GUM requires attention to higher-order terms when nonlinearity invalidates the first-order approximation.[ref]
A simple range calculation makes the difficulty visible without assigning a probability. Suppose only that plausible widths lie in and . These ranges include equal widths, hence , while their maximum positive radicand is , giving . These stipulated ranges are not a calibrated confidence interval; they demonstrate sensitivity to response uncertainty.
A negative observed need not imply an imaginary physical duration: noise, incorrect response characterization or a non-Gaussian pulse can invalidate the naive inversion. Conversely, dropping all negative simulated radicands and averaging the rest conditions on a selected subset and biases the inference. A Monte Carlo method must implement a stated input or likelihood model and a justified treatment of the physical boundary.
Report the measured and response widths, their conventions, uncertainties and dependence. An upper limit is meaningful only with its statistical construction and coverage assumptions. Improved response characterization or an independent temporal measurement can resolve the ambiguity; quoting more digits of the positive square root cannot.