Rank deficiency from deterministic spectral upsampling occurs because a deterministic map cannot manufacture independent measurements. For and input covariance matrix ,When the output has more channels than the input, its covariance matrix is necessarily singular. Covariance of count-conserving rebinned Raman channels gives the conserved-count example and its full propagated uncertainty.[ref]
Using that parent's two-to-three channel map, the output always satisfies . If the input readings are and , the rebinned vector is . Recovering the original readings as and shows that the third displayed number is not a third independent observation.
For an original fitting example, suppose both input means equal an unknown and their independent measurement-error variances are known to be . Fitting in the original channels gives , with variance . A correct generalized least squares fit after rebinning gives the same result and uncertainty. It must use the singular covariance on its supported range, for example through the Moore-Penrose inverse, together with the exact linear constraints.
The range restriction in singular-covariance spectral fitting matters if a proposed output model fails those constraints. A residual along cannot arise from this rebinning model at all. Merely inserting a pseudoinverse into a quadratic objective would assign zero penalty to that impossible direction. Either enforce the range condition or supply a physically justified additional noise model; adding a small diagonal term solely to make the matrix invertible changes the statistical assumptions.
The motivating Raman apparatus measures detector-array spectra before any subsequent numerical grid change.[ref]
Femtosecond Broadband Stimulated Raman: A New Approach for High-Performance Vibrational Spectroscopy by David W. McCamant and others. Instrumentation, array detection and spectral calibration; Data Collection, equation 4
Working on the original grid often avoids this singular fitting problem while retaining the actual measurement information. Rank-one covariance spike Created 2026-10-06 Updated 2026-11-05
For a unit vector and , a rank-one covariance spike is the covariance matrix . Its eigenvalue is in the direction and one on the orthogonal complement. It arises by adding an independent scalar normal distribution signal to a standard Gaussian random vector.
Ratiometric cancellation of RC timing gain errors follows when the same positive gain multiplies every voltage level relevant to a threshold ratio. With zero initial baseline, fixed time constant , high level and threshold ,Replacing by with leaves and the time unchanged. This is an exact invariance, not merely the first-order cancellation of two uncertainty terms. It is a ratiometric measurement property of the specified model.
For an original example, choose , and . The crossing is approximately . Gains and both give the same result when they scale high level and threshold together. A shared readout gain can cancel from the calculated time in the same way; whether a reference also changes the physical circuit must be stated separately.
The mismatch remaining after common RC gain cancellation depends on the relative gains, not their common magnitude. Let the high level use gain and the threshold gain , with . Provided ,In the example, a one-percent relative mismatch delays the crossing by about , compared with from the first-order expression. Thresholds nearer the high level amplify the residual.
A nonzero initial baseline must share the gain too: is unchanged by multiplying all three levels by . Scaling only the high level and threshold against a fixed nonzero baseline is another experiment. Likewise, additive errors follow the conditions in common offset in RC timing, not the two-level scaling rule.
Uncertainty in the time constant, independent threshold error, baseline memory, and failure to reach the threshold remain even when one common gain cancels. Their covariance structure must follow the complete measurement model, as in correlated threshold and amplitude errors in RC timing and NIST's propagation law.[ref]
Ratiometric measurement 2026-11-05
A ratiometric measurement infers a quantity from a ratio of two measured signals chosen to share an unwanted multiplicative variation. The useful question is which inputs actually share that variation: a reference resistor error or a change in the sensor itself generally survives the cancellation.
In a shared-excitation resistance ratio measurement, a current source passes the same electric current through a sensor of electrical resistance and a reference resistor . Their electric potential differences are and . An ideal normalized analog-to-digital converter reading is thereforeThe current source need not have a perfectly constant output for this cancellation, provided both measurements sample the same electric current, the denominator is nonzero, and the converter stays within its operating ranges. This is a standard arrangement for a resistance thermometer.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. §2.1.3, pp. 10–11, shared-current resistance measurement
For an original ideal example, take , and . Then , and . Increasing by gives and : the ratio remains . If the converter instead uses an independent fixed reference and the calculation assumes the original , the inferred electrical resistance becomes .
Shared excitation cancels in a resistance ratio; a fixed independent reference retains the excitation error. Original ideal calculation, not measured data.
Details. The reference-resistor error in ratiometric resistance measurement remains even with perfect shared excitation. If the actual reference resistor is but the calculation uses its nominal value, thenThus a reference error makes the nominal estimate of a true sensor about . Reference accuracy and its variation with temperature still matter.[ref]
Nor does the ratio remove sensor self-heating in a ratiometric resistance measurement. Joule heating dissipates , which is in the example. A temperature rise changes the actual electrical resistance of a resistance thermometer; the ratio then measures that changed resistance. Reducing the electric current or comparing several currents can reveal this physical effect.[ref]
The law of propagation of uncertainty makes the role of covariance explicit. For small perturbations of nonzero mean signals, the first-order measurement uncertainty of obeysThe contribution from a shared excitation fluctuation cancels through the covariance term. Treating the two signals as independent random variables would discard that cancellation. Other errors need their own inputs in the complete model.[ref]
For time-varying excitation, ratiometric cancellation with matched signal and reference filters requires matching dynamics as well as shared wiring. With constant resistances and identical linear time-invariant systems of impulse response , the filtered signals are and , so their ratio is still wherever defined. Here denotes convolution. Different filters or different sampling times can leave an excitation-dependent error.[ref]
RTD Ratiometric Measurements and Filtering Using the ADS1148 and ADS1248 Family of Devices by Luis Chioye. First-page abstract and low-pass filter design discussion, matching input and reference filters
A simple proposed bench test first replaces the sensor by a stable calibrated resistor, sweeps the electric current within the instrument limits, and records both electric potential differences and their ratio. That separates electrical cancellation from changing sensor temperature. Repeating with a resistance thermometer in a stable bath then tests Joule heating; changing the reference resistance or one filter tests the predicted failure modes. These are proposed checks of the model, rather than reported experimental results.
Rayleigh radius from Gaussian components 2026-10-18
The joint probability density of isIn polar coordinates, , , and the absolute Jacobian determinant is . Thus the joint density of radius and angle factors asIntegrating over angle gives the Rayleigh distribution. The same factorization shows that the angle has uniform distribution and is independent of the radius. This is a short change of variables argument: an annulus of width has area to first order.
The assumptions are substantive. Unequal component variances give elliptical rather than circular density contours; a nonzero mean displaces the center. Even zero covariance and equal normal distributions do not suffice without independence. For an original counterexample, let and let be an independent Rademacher random variable. Put . Then has the same normal distribution as , and . Nevertheless, all points lie on or , and has densitywhich is positive at zero, unlike the Rayleigh distribution. This pair is not a Gaussian random vector.
Receiver-latency uncertainty in the inferred input-time horizon propagates uncertainty in timing calibration into the latest input time that could have influenced a stored output record. For one channel, the horizon in input-time horizon of a delayed optical receiver record is , with last output time and effective input-to-output timing offset . HenceThis linear variance identity retains shared timing influences.[ref]
A common calibration offset can cancel when it enters both quantities. In an original model, let errors be and , with independent components of standard deviations , and . Then and . Discarding their common-offset covariance would instead give .
The shared component must come from the actual calibration model. Here is the effective timing offset relating the input and output coordinates, so an output-clock alignment error can enter both and . A physical transport-delay error is not automatically correlated with the last-sample timestamp.
For hard timing limits, propagate validated bounds rather than reinterpret standard uncertainty as a guarantee. If is known and , then . Claiming the optimistic endpoint as though it held for every allowed delay would overstate the input-time reach.
With several channels, . Channel switching under uncertainty in a combined receiver horizon makes this a nonsmooth function when the maximizing channel is uncertain. Propagate the justified joint timing set or distribution; a derivative of only one nominally best channel can miss such switching.
NIST's primary pulse calibration measured the detector, cable and oscilloscope together using a short known optical input.[ref]
Calibration Service for Low-level Pulsed-Laser Radiometers at 1.06 µm: Pulse Energy and Peak Power by Rodney W. Leonhardt. Appendix C.1–C.2, printed pp. 36–37, Figure C1
The numerical timing errors above are chosen examples, not values from that apparatus. A timing horizon remains a necessary information limit; it does not guarantee invertibility of the receiver or bound future optical illumination. Recovery-estimate covariance 2026-11-05
Recovery-estimate covariance describes joint variation between an estimated extraction recovery and another input used with it, such as the measured recovered amount. It matters when a recovery correction divides one estimate by the other: shared data can make their uncertainties correlated.
For corrected amount , first-order uncertainty propagation givesThe sign of the covariance contribution follows from the opposite signs of the quotient's sensitivity coefficients. A positive covariance reduces the propagated variance relative to a calculation assuming independence; a negative covariance increases it. Neither sign is a discretionary choice.[ref]
Evaluation of measurement data — Guide to the expression of uncertainty in measurement by JCGM. §5.2, correlated input quantities and covariance terms.
An original shared-data example sets , where independent measured amounts have nominal values and , with standard uncertainties and . Then andSubstitution gives , hence . This agrees with the exact algebraic cancellation . Ignoring the shared numerator would instead give .
The cancellation also reveals a validation trap: using the same recovered amount both to estimate its recovery and to correct itself does not independently demonstrate accurate recovery of an unknown sample. Build the joint measurement model from the underlying observations. Shared calibration standards, preparations or instrument drift may create other covariance structures; matched replicate data or an explicit common-input model can estimate them. Do not assume independence merely because a table lists two separate estimates.
Reference-assisted spectral offset separation uses an additional reference observation to distinguish a physical spectral change from an instrumental offset. The reference must respond to the offset in a known way while excluding, or separately modelling, the sample’s physical change. Merely collecting another signal does not guarantee spectral offset identifiability.
For a local linear measurement model, writewhere is the desired change, the shared offset and . With known sensitivities and zero-mean residual errors,removes the offset contribution. This is a subtraction of modelled responses, not a claim that the uncorrected sample and reference frequencies are physically identical.
An original example uses , and , giving . A reference may have a much larger offset response than the sample; the response ratio supplies the necessary conversion.
The uncertainty of a reference-corrected spectral change depends on joint errors. For fixed ,With , and correlation , this gives approximately ; assuming independence instead gives . The primary GUM supplies the correlated-input propagation rule used in this calculation.[ref]
If is itself estimated, its uncertainty also matters. Under an additional independent-input first-order approximation, the contribution is . Large uncertainties, correlations with the fitted ratio or nonlinear responses require a more complete analysis.
A controlled experiment records stable reference features through the same acquisition and processing chain while deliberately perturbing the instrumental offset. This can estimate and test whether the response is linear. Sample/reference timing must also match the assumed offset: a drifting offset at different acquisition times can leave an uncorrected residual. Repeated paired data, rather than unrelated error estimates, are needed to characterize the relevant covariance.
Reference identity in algebraic uncertainty cancellation asks whether two appearances of a reference are the same measured quantity or merely readings given the same descriptive name. If a transfer is but a correction subtracts , thenThe last difference vanishes identically only when the same reference value is reused. Separate acquisitions may be correlated, independent, or affected by drift; sharing a label does not decide their covariance.
For independent and , also independent of the reference pair, propagation givesWith equal reference standard uncertainties and correlation coefficient , the residual reference contribution is . It is zero for identical reused values, for independent readings, and is not generally zero just because the displayed numbers agree. The GUM requires significant correlations from common inputs to be included.[ref]
An original acquisition model separates what cancels:Here is a common uncertain calibration contribution, while are independent repeat errors. Their difference removes both and , but leaves . If mg and each repeat error has standard deviation 2 mg, each reading has variance and their covariance is . The difference therefore has variance , not zero.
Adding independent mg and mg gives . Reusing exactly one recorded reference value in both places instead gives . These constructed cases have different data-reduction histories despite similar-looking equations.
Reference lineage in covariance models records acquisition identifiers, shared calibration inputs and coefficients in each derived quantity. A calibration change or time-dependent drift must appear in that model if significant. Algebraic simplification is valuable only when it preserves the identity of the underlying inputs; renaming two acquisitions with the same symbol does not make their errors cancel.
Reference-resistor error in ratiometric resistance measurement is the error left when an instrument converts a measured resistance ratio using an inaccurate value for its reference resistor. Shared excitation removes one nuisance input; it does not establish the denominator's true electrical resistance.
Let be the actual reference resistor value and the value used in calculation. For an ideal simultaneous reading , the inferred sensor resistance isWriting gives the exact fractional error . The familiar is its first-order approximation. If a deterministic bound is , the inferred fractional error lies inA symmetric reference tolerance therefore produces a slightly asymmetric resistance-error interval.
For an original example, suppose the instrument uses while the actual reference resistor is . A true sensor is reported as , an error near . Multiplying the shared excitation by any common positive number leaves this error unchanged in the ideal model. Using a measured and corrected instead of its nominal marking can remove the known part, while the corrected value retains its own measurement uncertainty.
Temperature variation can produce the remaining error after measurement calibration. Locally, , where is the temperature coefficient of resistance. For example, a -parts-per-million-per-kelvin coefficient over contributes roughly parts per million; it survives excitation cancellation. The primary circuit guide explicitly requires reference accuracy and low drift, and separates these requirements from shared-current cancellation.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. §2.1.3, printed pp. 10–11, reference accuracy and drift
For small random errors, applying the law of propagation of uncertainty to givesThe covariance term must follow how the reference value and reading were obtained; it cannot simply be assumed zero when inputs are shared.[ref]
Reference-state transfer in overlap normalization uses a known optical state to estimate a state-dependent instrumental offset, then transfers that correction to the specimen measurement. Its central assumption is that the relevant state-gain ratio remains applicable between the reference and specimen. Repeatability of the reference alone does not establish transfer.
Let , andSuppose specimen and reference observations satisfyWith known reference state , the corrected specimen log-ratio isThe residual is exactly the failure of the instrumental offset to transfer. Acquisition order can address time drift, but cannot by itself prove .
An original matched-state example has and . If the specimen's state-gain ratio differs from the reference's by , the corrected ratio still contains 0.02. Ignoring that residual gives an inferred radius ratio approximately 0.9917 instead of one. Repeating both measurements can reduce independent scatter while preserving that systematic offset.
Reference uncertainty must also propagate. Near a matched reference, . For , its magnitude is 2.4. A standard uncertainty 0.01 in the reference log-radius therefore contributes 0.024 to the corrected log-ratio before combining other errors. If the same reference estimate corrects many specimens, this uncertainty is shared; it does not become independent for each specimen. The GUM explicitly retains covariance from common references.[ref]
Operating-point transfer of an overlap reference requires checking the variables capable of changing the state response: optical power, detector bias, electronic range, frequency and illumination. A useful proposed test brackets specimen readings with reference readings and repeats the reference at several operating powers. A trend then tests a transfer assumption; an unchanged reference at only one power leaves power-dependent transfer untested.
NIST's receiver apparatus controls spatial mode and polarization with a polarizer and single-mode-fibre coupling, while separately calibrating the electrical response.[ref]
Frequency response metrology for high-speed optical receivers by Paul D. Hale and others. PDF p. 1, Heterodyne measurements
A reference must preserve the relevant optical and electrical conditions, or the analysis must model their differences. The transfer calculation above is an original diagnostic construction, not a claimed NIST protocol.Reference-versus-sample allocation in repeated isotope-ratio measurements distributes a finite measurement budget between sample readings and the reference used to calibrate them. Suppose independent sample and reference readings have first-order gain-error variances and . Averaging sample readings and reference readings gives log stiffness-ratio variancewhere is the variance of a differential calibration error shared by all ratio estimates. The shared-reference floor in averaged isotope stiffness ratios explains why that last component does not average down. Covariance propagation must reflect which observations and calibration influences are actually shared.[ref]
Let one sample reading cost budget units and one reference reading cost , with . Treating counts continuously for the initial optimization, a constraint multiplier givesThe counts are proportional to standard uncertainty divided by the square root of cost, rather than simply to each variance. A budget-independent common component changes the attainable floor but not this optimum ratio under the stated model.
For an original example, take equal costs, , and 40 total readings. The optimum is 30 sample and ten reference readings. Without a common component, the predicted log-ratio standard uncertainty is about , compared with for twenty of each. A common standard uncertainty raises those values to about and , respectively.
An integer allocation near a continuous isotope-reference optimum must compare feasible neighboring count choices and any acquisition-order constraints. More reference readings help only to the extent that their errors are independent; repeatedly measuring an unchanged unknown calibration bias does not create independent references. If reference drift depends on elapsed time, a single constant cost and common-variance term may be inadequate.
In practice, record which reference observations calibrate which sample spectra and account for any correlated peak fitting, isotope-composition uncertainty and spectral offsets. The variance and cost model above is a chosen allocation problem, not an experimentally established precision specification for an isotope measurement system.
For air approximated by index one, , so . Taking the variance gives the displayed first-order error formula, with angles and their errors in radians. Independence removes the covariance term; hard error bounds require a different combination rule. Common angular-zero errors can produce covariance and should not automatically be added in quadrature.
Regurgitant fraction 2026-11-05
The regurgitant fraction is the proportion of a valve-related total ejected blood volume that returns through the leaking valve. For the left ventricular balance used to assess mitral regurgitation, let be total ventricular stroke volume, the matched forward aortic volume and the regurgitant volume. With no other abnormal communication or confounding valve leak,The imaging recommendations use complementary volume and flow measurements to quantify these quantities.[ref]
Recommendations for Noninvasive Evaluation of Native Valvular Regurgitation by William A. Zoghbi and others. Section II.B.6, volumetric methods; Figure 8, p.316
An original example takes mL and mL. Then mL and RF is 30%. The quotient instead expresses leakage relative to forward delivery; it is not the same fraction. The denominator is part of the definition.
Equal fractions also need not indicate equal transported amounts. A 30 mL leak out of 100 mL and a 60 mL leak out of 200 mL both have RF 30%. Conversely, an unchanged 30 mL regurgitant volume gives RF 50% if total ejection falls to 60 mL. These stipulated cases isolate the normalization rather than assigning a clinical severity category.
The two measurements in a derived RF may share errors. For , first-order propagation givesThe covariance term should follow the actual acquisition and data reduction, rather than assuming independence because forward and total volume have different names.[ref]
Relative calibration error in an isotope-corrected stiffness ratio is the fractional error remaining when spectral calibration affects the two compared bands differently. Write the model of isotope-corrected Raman stiffness ratio as , where and the positive denote vibrational wavenumbers. For small errors,The mass correction does not remove differential spectral calibration error.
A useful calibration model is , with common and separate gain errors and offsets . Its spectral contribution becomeswith the mass term added if uncertain. A single multiplicative scale cancels exactly when no additive error is present. Separate gains do not: comparing two spectra calibrated independently can lose that cancellation.
For an original example, take and . A differential gain creates about relative error in . A common offset instead contributes about . If the separate gains each have independent standard uncertainty, their combined contribution to the relative standard uncertainty of is .
The separate spectral-calibration gains in isotope stiffness inference must therefore be distinguished from a shared reference error. Covariance propagation keeps precisely those shared and separate influences.[ref] Recording both samples with the same validated spectral scale can help, but drift or changed dispersion can still produce differential errors.
The primary graphene isotope experiment measured Raman bands over isotope concentrations and laser excitation energies and found the harmonic-oscillator mass dependence.[ref]
Probing carbon isotope effects on the Raman spectra of graphene with different ¹³C concentrations by Bruno R. Carvalho and others. Abstract, isotope-mass dependence of Raman frequencies
Small residual stiffness changes require the additional calibration and mode-model checks above; the numerical errors here are chosen examples, not quoted experimental uncertainties. Release-expansion thermometry 2026-10-18
A spatial image after a chosen dark interval is a different readout from continuously recording arrivals at a fixed probe. Here is an original idealised fluorescence imaging model: release a dilute cloud, let it move freely, then apply a short uniform probe pulse. Neglect collisions, residual forces, attenuation and motion during the pulse. Use each image's cloud centre to remove mean drift. Uniform gravity shifts that centre without changing the centred variance.
For each atom let along a horizontal imaging axis. Expanding the square about the expected values gives exactlyNo Gaussian distribution is needed for this variance identity. A jointly Gaussian distribution of initial positions and velocities does make the later density Gaussian. For a separable three-dimensional cloud viewed along , integration along the viewing direction gives the column densityHere is atom number. Integrating this image along leaves a one-dimensional Gaussian distribution with standard deviation . This is a projected density measurement, not a three-dimensional radius or a direct velocity histogram. Under the uniform, optically thin probe assumptions, expected fluorescence is proportional to this column density. A known, independent point spread function with finite variance adds its own variance to the measured width; probe duration and intensity gradients need separate treatment.
For a Maxwell-Boltzmann distribution, , where is atomic mass and the Boltzmann constant. For a nonthermal distribution this defines a directional second-moment temperature, which need not describe every feature of the distribution. A covariance between position and velocity can initially narrow the cloud even though is positive. The physical covariance matrix requires .
Choose , and , giving correlation coefficient . With in milliseconds, the numerical variance in square millimetres is . Its minimum occurs at , with standard deviation . These chosen values satisfy the covariance matrix constraint; early contraction is compatible with free motion.
Generate thirteen synthetic variance observations at equally spaced times from to , adding independent Gaussian noise of standard deviation . Linear least squares fits all three coefficients in . The example yields , and . With the chosen approximate sodium mass , the true is and the fitted value is . This is a fit to generated width data, not historical observations.
If the same data are fitted to , forcing , the inferred falls to . The omitted term is partly absorbed into both fitted coefficients. In noiseless linear least squares with a free intercept, its precise effect on the quadratic coefficient iswhere the subscript denotes equal-weight moments over the chosen observation times, using one consistent unit system. Since the ratio is positive for distinct nonnegative times, negative biases this estimate downwards. An incorrect release-time origin can also generate a linear term, so fitting it does not by itself identify a physical initial covariance. Timing calibration, spatial resolution, residual magnetic fields and probe-induced recoil energy remain separate controls. Neither a cold-looking image nor a single long expansion time identifies all three coefficients.
Residual-offset uncertainty in nail thickness estimates is the uncertainty contributed by an imperfectly known additive background after correction of spectral signals. It differs from a known uncorrected offset's bias: subtraction can remove the best estimate of a background while leaving uncertainty about that estimate.
For an original reduced optical model, let two measured normalized signals be and subtract the same background estimate . Define and suppose the attenuation contrast is known. The inferred nail plate thickness isIts sensitivity coefficients areWith mutually independent raw signal estimates and background estimate, first-order uncertainty propagation givesThis follows the sensitivity-and-covariance approach in the measurement guide.[ref]
Guide to the expression of uncertainty in measurement by JCGM. §§5.1–5.2, propagation of input uncertainties and correlated quantities
For stipulated , , , and , the estimated thickness is approximately 0.5108 mm and its combined standard uncertainty is 0.0177 mm. The common-background contribution alone is 0.0133 mm. Averaging 100 independent repetitions of each raw signal while reusing the same background estimate reduces the total only to about 0.0134 mm; it does not reduce that shared contribution by ten.
An equivalent calculation treats corrected signals as correlated: under the stated independence assumptions. Omitting this covariance would describe a different experiment with independent background errors. Near zero corrected signal, the logarithm also becomes strongly nonlinear, so this local approximation needs further checking.
The primary nail experiment included a fluorescence correction in its optical transmission measurement.[ref]
The transmission of optical radiation through human nails by S. G. Parker and B. L. Diffey. Abstract, fluorescence correction
Background emission and dark current must be characterized separately when they have different spectral responses. The equations above are an illustrative uncertainty budget, not a validated thickness calibration or reported precision from that experiment.A response-normalized tolerance for baseline-feasibility residuals measures a pulse estimator's constraint errors against physically chosen response limits. For residual , assign positive tolerances in the corresponding response units and calculateThe criterion means every specified response limit is satisfied. Numerical feasibility residual for baseline constraints explains why a raw unscaled norm alone has no universal physical interpretation.
These tolerances can enforce a deterministic bias budget. If signal and background coefficients are bounded by and , thenThe second inequality follows when the normalized test passes. The allocation of a deterministic bias budget among baseline constraints therefore connects a numerical acceptance criterion to a stated physical error allowance.
For an original example, bound the signal amplitude by , constant background by and drift rate by . Choose signal-response tolerance , constant-response tolerance and linear-time-response tolerance . Their maximum bias contributions are respectively , and , totaling . If all actual residuals are half these tolerances, and the corresponding bias bound is .
The normalized test can be invariant to constraint units. Multiplying one equation and its required response by a nonzero factor multiplies its residual by ; replacing its tolerance by keeps the ratio unchanged. The associated coefficient bound changes inversely, so the physical product is unchanged as well.
NIST's before/after pulse baselines supply a measurement setting where residual offset and drift responses matter.[ref]
Calibration Service for Low-level Pulsed-Laser Radiometers at 1.06 µm: Pulse Energy and Peak Power by Rodney W. Leonhardt. §3.5, printed p. 8; §7.2, Leakage Effect
The example bounds and tolerances are stipulated, not specifications of that apparatus. This deterministic leakage test is separate from noise variance, which needs the actual measurement covariance.[ref]Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results by Barry N. Taylor and Chris E. Kuyatt. Appendix A, equation A-3
Backgrounds outside the assumed family also need separate treatment.Free bending has two zero-energy filament modes, so it does not define an unrestricted normalizable thermal height distribution. Fix the mean displacement and linear tilt, or project out that affine kernel. On the positive complement, the equipartition theorem gives independent Gaussian mode variances for normalized modes. The resulting covariance is the bending operator pseudoinverse multiplied by . It depends on how the rigid motion is fixed.