Semiconductor Created 2026-10-05 Updated 2026-10-20
A semiconductor has an electronic band gap and a conductivity strongly affected by temperature, impurities and carrier injection. Optical absorption can create mobile electrons and holes, enabling a photodiode.
The useful feature of a semiconductor is that its charge carrier population can be controlled. Semiconductor doping changes the supply of mobile electrons or electron holes; carrier injection changes their population away from thermal equilibrium. A transistor turns that control into an electronic amplifier: a small input signal controls a larger flow of supplied energy.
For a uniform semiconductor in a weak electric field, a simple drift model writes its electrical conductivity asHere is the elementary charge, are the number densities of mobile electrons and electron holes, and their positive charge carrier mobilities. The two contributions add because opposite electric charges drift in opposite directions and therefore carry electric current in the same direction. This model assumes fixed temperature and mobilities and omits diffusion from density gradients.
An original dimensionless example isolates the effect of carrier injection. Suppose initially and . Adding equal excess populations gives , even though the added electron holes remain a minority carrier population. A semiconductor can respond strongly to a small injected population without changing its underlying band gap.
Single-molecule absorption in a molecular crystal Created 2026-10-14 Updated 2026-10-20
In 1989, William E. Moerner and Lothar Kador reported individual pentacene absorbers in para-terphenyl in Optical detection and spectroscopy of single molecules in a solid by W. E. Moerner and L. Kador (pp. 2535–2538). Different local environments spread photon absorption frequencies; tuning into the sparse wings selected few resonant molecules despite many molecules in the illuminated volume.
Their tunable dye laser focused through a numerical aperture lens onto a – crystal immersed in superfluid helium at . Frequency modulation accompanied either a electric field between coated sapphire disks, detected at , or a quartz shear transducer. An avalanche photodiode measured transmitted-light photon absorption, not fluorescence. Double modulation suppressed residual laser amplitude background.[ref]
Optical detection and spectroscopy of single molecules in a solid by W. E. Moerner and L. Kador. pp. 2535–2537: sample, lens, dye laser, FM/Stark and FM/ultrasound apparatus and detection
Repeated scans gave the predicted modulated line shape. Blank crystals and off-resonance scans gave background; changing modulation frequency expanded and contracted the feature as predicted. Agreement between both perturbation methods supported the individual-absorber inference. The authors acknowledged out-of-focus molecules and the unlikely alternative of multiple absorbers with exactly coincident frequencies.[ref]
Optical detection and spectroscopy of single molecules in a solid by W. E. Moerner and L. Kador. Fig. 1(d–g), Fig. 2(b–g), and pp. 2537–2538: blanks, modulation-frequency changes, background and coincident-absorber qualification
This is evidence for a selected absorber in a prepared, known chemical substance and host. It does not establish arbitrary unknown molecule identity from one spectral feature, provide an atom-resolved photograph, or count every molecule in the crystal. A measurement sensitive to only a narrow excitation range cannot detect a molecule whose transition lies outside that range. This distinction separates the total particle inventory from the number responding to a particular probe.
Single-photon avalanche diode 2026-10-25
A single-photon detector based on a reverse-biased photodiode operated above its breakdown voltage. A carrier generated by photon absorption can trigger a self-sustaining impact ionization avalanche. Quenching lowers the bias to stop that avalanche; recovery creates detector dead time. Random internally generated carriers can also trigger indistinguishable pulses, producing dark counts. Single-photon avalanche diodes ordinarily detect an event rather than resolve the incident photon number. [ref]
https://hub.hamamatsu.com/us/en/ask-an-engineer/detectors/sipm-mppc-spad-questions-and-answers.html. Hamamatsu Photonics, What is SPAD? and How does a SPAD work?: above-breakdown bias and quenching
Solar limb-darkening drift-scan experiment 2026-10-14
A motivating measurement used the McMath solar optical telescope at Kitt Peak. The Sun's image drifted across a -high, -wide optical slit; a double-pass spectrograph selected wavelength intervals from wide at to at . An EMI 9750 QB photodetector or silicon photodiode supplied a digitized signal: samples were summed into 4096 positions per crossing, including dark and off-limb sky readings. [ref]
Solar limb darkening 1986–1990 (λλ303 to 1099 nm) by Heinz Neckel and Dietrich Labs. §2, p. 93, instrumental compilation: slit dimensions, double pass, detectors, sampling and drift-curve layout; p. 113, acknowledgements identifying the telescope
Controls included daily realignment of optical telescope mirrors, wavelength settings between spectral lines insensitive to the rotational Doppler shift, selection of detectors without detectable hysteresis, and measured electronic nonlinearity corrections below of maximum signal. Limb-position calibration combined centres of gravity of the edge derivative profiles with almanac semidiameters; the preliminary edge shifts were consistent with astronomical seeing of approximately arcseconds full width at half maximum. [ref]
The decreasing centre-to-limb readout gives a specific intensity profile, constraining stratification of the photosphere under a radiative transfer model. Astronomical seeing, instrumental scattering of light, and solar granulation limit this inference; brightness alone does not uniquely determine temperature versus geometric height. These observations need not follow the stipulated linear limb-darkening law.
Spectral mismatch correction 2026-10-14
For collected spectral radiant flux and a linear photodiode with spectral responsivity , the dark-subtracted electric current is . A reference optical spectrum sets the conversion from electric current to collected luminous flux. Applying it to another optical spectrum givesAssume nonzero calibration and test signals. Then is the corrected luminous flux. Multiplying either optical spectrum by an overall positive scale leaves unchanged. If is proportional to , for every optical spectrum. Uniformly overfilled aperture measurements use wavelength-resolved irradiance instead, multiplying the electric current expression by aperture area; the same ratio cancellation gives the illuminance correction. This is the algebra behind source-dependent calibration, rather than a new measured correction curve. [ref]
Spectral radiant flux 2026-10-14
Spectral radiant flux resolves radiant flux by wavelength or frequency:Its units are per metre of wavelength or per hertz of frequency. A density quoted per nanometre must be integrated with wavelength intervals in nanometres; its numerical value is times the per-metre density. The frequency and wavelength spectral radiance conversion applies equally to spectral radiant flux, through the change of variables formula:with the speed of light in vacuum and expressed per metre. The Jacobian determinant preserves integrated power rather than the height or location of an optical spectrum's peak.
For two explicitly synthetic optical spectra, let and quote the following densities in :The triangle under and the rectangle under both have area : . Their band radiant fluxes differ: – contains of , while – contains of . These are specified functions, not measured source spectra.
The dark-subtracted electric current of a linear photodiode isusing its spectral responsivity. The figure's stipulated gives and despite equal total radiant flux. Symmetry places 's mean wavelength at and 's at ; a linear averages to its value at that mean. A single electric current therefore cannot determine an unknown broadband radiant flux without additional spectral information or a sufficiently flat calibrated response.
Spectral responsivity Created 2026-10-14 Updated 2026-10-18
For a photodiode in its linear regime, spectral responsivity is the dark-subtracted electric current per incident radiant flux in a narrow band at wavelength , measured in . More generally a photodetector's readout can be electrical voltage or another quantity, with corresponding units. A broad-band readout weights the incoming optical spectrum by the detector's spectral responsivity; it is not the unweighted total incident power. [ref]
https://www.nist.gov/programs-projects/spectral-responsivity-measurements-and-spatial-spectral-power-responsivity. NIST, Spectral responsivity measurements: wavelength-dependent sensitivity and spatial characterization
. Ultraviolet Created 2026-10-05 Updated 2026-10-20
Ultraviolet is electromagnetic radiation with wavelengths shorter than the violet end of ordinarily visible light. The CIE convention uses –, subdivided into UV-A at –, UV-B at – and UV-C at –. These divisions name intervals; they do not mark universal changes in every material's response. The boundary with visible light is not sharp, and other fields use near-, far-, vacuum- and extreme-ultraviolet divisions. [ref]
https://cie.co.at/eilvterm/17-21-008. CIE, International Lighting Vocabulary (2020), ultraviolet radiation: ranges and boundary caveats
. For example, NASA's Solar Dynamics Observatory observes extreme-ultraviolet bands below ; the CIE interval is therefore not the full extent of astronomical ultraviolet usage. [ref]https://sdo.gsfc.nasa.gov/data/channels.php. NASA, Solar Dynamics Observatory HMI and AIA channels: the Atmospheric Imaging Assembly's ultraviolet and extreme-ultraviolet bands
.The CIE ultraviolet interval corresponds to frequencies about – and photon energies –, using vacuum wavelength and with Planck constant and speed of light . Shorter wavelength means larger energy per photon; larger irradiance means more incident power per area. These are different comparisons. At the same incident power, a ultraviolet beam has fewer photons per second than a infrared beam, even though each ultraviolet photon has more energy.
A common ultraviolet mechanism is photon absorption promoting an electron from a ground state to an excited state of a chemical atom or molecule. By contrast, many infrared absorption bands involve stretching or bending chemical bonds. For either mechanism, both the energy difference and the coupling to the incident electromagnetic field matter. One concrete example is the absorption discussed for butadiene: an electron moves between molecular orbitals, and the position of the absorption reveals their energy separation. Ultraviolet absorption can therefore probe molecular orbitals, while infrared spectroscopy probes many molecular motions. [ref]
https://openstax.org/books/organic-chemistry/pages/14-7-ultraviolet-spectroscopy. John McMurry, Organic Chemistry, §14.7: ultraviolet excitation and the 217 nm butadiene example
.Subsequent chemical reactions depend on the absorbing chemical substance and its environment; the label ultraviolet alone does not predict a reaction rate. At sufficiently short wavelength, a photon can cause photoionization if its energy exceeds the relevant ionization energy. There is no single wavelength that separates all ionizing from all nonionizing behavior for every chemical substance. Likewise, eventual heating after photon absorption is possible in both ultraviolet and infrared. The distinction is useful because different optical spectra expose different processes, not because the two ranges obey different laws of electromagnetism.
Ritter's 1801 observation used a chemical readout rather than a thermometer. In Caltech/IPAC's account, sunlight passed through a glass optical prism and silver chloride was placed in different parts of the optical spectrum. Darkening was weak toward red, stronger toward violet, and pronounced beyond the visible violet edge. The measured evidence was a change in the chemical substance's appearance at a spatial location with no visible color, establishing radiation beyond the eye's ordinary range. It was not a calibrated measurement of a universal ultraviolet optical spectrum. [ref]
https://coolcosmos.ipac.caltech.edu/page/lesson_ritter_experiment. Caltech/IPAC Cool Cosmos, The Ritter Experiment (2022): the historical silver-chloride observation
.A controlled modern interpretation of such a chemical test would compare equal exposure durations and matched fresh samples, keep an unexposed sample in darkness, monitor temperature, and distinguish scattered visible light from the intended short-wavelength beam with characterized filtering. These are proposed controls, not an assertion that Ritter performed them. Unequal sample preparation or heating can change a reaction rate. The optical prism, terrestrial atmosphere and sample response restrict which wavelengths the test detects. Darkening is a detector-weighted response, so its spatial maximum is not automatically the maximum of incoming radiant flux or of photon energy.
Thomas Larason and Jeanne Houston's NIST report documents a quantitative ultraviolet detector comparison. An argon source feeds a double-diffraction grating spectrometer; a shutter and mirrors deliver the selected band to a working-standard or test photodiode, successively at the same position. A beam splitter supplies a monitor photodiode. Amplifiers and voltmeters record detector and monitor simultaneously; shutter-closed readings provide dark baselines. The – scan, typically with bandpass, measures spectral responsivity. Monitor normalization suppresses source drift; known spectral lines calibrate wavelength. The report assesses temperature effects and unwanted light outside the band, weighted by detector response. The latter need not cancel between different spectral responsivities. [ref].
This determines electrical response without visual perception. Bandpass and spatial nonuniformity limit the calibration; aging and changed temperature can affect later use. The scan includes visible light and ultraviolet; no response below follows from it.
For a narrow beam, the idealized substitution calculation isHere is spectral responsivity; and are dark-subtracted detector and monitor electric currents. This follows by writing for each detector and using the monitor to track changes in radiant flux. It assumes linear response, unchanged beam splitting and collection geometry, and a sufficiently narrow band. For an illustrative standard with , ratios and give . These invented arithmetic inputs illustrate the method; they are not NIST measurements.
An independent photon-counting model connects spectral responsivity to quantum efficiency. If is the number of collected electrons per incident photon, there is no internal multiplication, and is the elementary charge, thenAt illustrative constant and , this gives about at and at . The ultraviolet readout is smaller in this model because there are fewer incident photons, not because their individual energies are smaller. Real quantum efficiency depends on wavelength, windows, absorption and carrier collection; a detector's band gap and construction determine its usable range. Constant across these two wavelengths is an analytical comparison, not a claimed specification of a real photodiode.
Photon energy across conventional ultraviolet and infrared ranges, with a schematic detector-substitution experiment
. Details. The curve is the calculated relation , and the shaded ranges use the stated CIE conventions. The lower diagram abstracts the detector comparison rather than reproducing an instrument drawing. Together with the infrared heat-balance illustration, it shows why a calibrated detector response, rather than what looks bright to an observer, is the bridge between invisible radiation and a quantitative measurement.

