A force-microscopy mode keeping the probe in contact while monitoring cantilever deflection. Constant-deflection feedback records the vertical actuator motion needed to maintain an approximate normal-force setpoint; finite probe geometry and specimen deformation still affect the image.
Friction force microscopy 2026-10-20
Measurement of lateral tip–surface forces from torsion of an atomic-force-microscope cantilever.
Jump-to-contact 2026-10-20
An instability in a compliant force probe when the attractive force gradient exceeds the restoring stiffness of the cantilever. The tip then jumps from a noncontact branch into contact; the missing branch cannot be followed quasistatically with that stiffness.
Atomic force microscopy senses tip–sample interactions through a compliant cantilever. Contact deflection, oscillation amplitude, phase or resonance frequency can be converted into a feedback signal. Chemical bonding and short-range repulsion, van der Waals forces, electrostatic forces and, where relevant, magnetic or capillary interactions produce contrast. Apparent height at a fixed force or frequency setpoint can therefore contain interaction contrast as well as geometry. Lateral resolution depends on the effective tip apex and interaction range; finite tip radius broadens features. Vertical sensitivity also depends on thermal and detector noise, cantilever calibration, vibration, drift and feedback bandwidth. Strongly localized short-range forces favor atomic resolution, whereas extended interactions average over more of the tip and surface.
For small oscillations about equilibrium separation , write and linearize . The constant force is absorbed into the equilibrium displacement. The cantilever equation becomes
Consequently the small-amplitude AFM frequency shift is
The linear expression is the small-force-gradient approximation, , in addition to the small-oscillation assumption. For an attractive force , and the resonance softens. A finite oscillation amplitude samples a weighted range of forces rather than only the gradient at one separation, and destroys local stability in this simple static model.
Figure 1.
AFM frequency shift: exact local linearized oscillator and the weak-gradient approximation
. Details.
In a scanning electron microscope, a focused electron beam scans the surface and a detector records an emitted signal at each beam position. Low-energy secondary electrons escape from near the surface; changes in local slope, edges and detector-facing geometry give strong secondary-electron topographic contrast. Backscattered electrons result from elastic deflection of primary electrons and generally have a larger yield for higher atomic number. Their backscattered-electron atomic-number contrast therefore helps separate compositional regions. Detector geometry, crystal orientation and topography can also affect this signal, so brightness is not a unique elemental identification.
An environmental scanning electron microscope admits gas around the specimen while pressure-limiting apertures and differential pumping preserve the electron-optical column's vacuum. Electron collisions ionize the gas. Positive ions can neutralize negative charging of a poorly conducting specimen, enabling observation without a conductive coating. A gaseous secondary electron detector uses gas amplification to collect secondary-electron signals under these conditions. Gas pressure, path length and scattering must be controlled because they also affect the probe and background.[ref]
Finally, acquire an energy-dispersive X-ray spectroscopy spectrum at each pixel to add elemental specificity. The beam creates inner-shell vacancies, and characteristic X-rays emitted as they fill identify chemical elements. Selecting fitted characteristic-line intensities yields elemental maps that can be compared with the electron image. Peak overlap, detector response, absorption and the X-ray-generating interaction volume limit identification, quantification and spatial resolution. This gives chemical-element contrast; resolving different bonds or oxidation states generally needs more detailed spectral information than one broad elemental map.
The fixed support supplies the end shear force and the largest bending moment, . In elementary beam bending, the greatest longitudinal tensile and compressive stresses lie on opposite outer faces of the section at the fixed end, with zero longitudinal bending stress on the neutral axis. For a downward tip load on a horizontal cantilever, the upper face is in tension and the lower face in compression. Transverse shear stress has a different distribution, largest near the neutral axis for a rectangular section; “highest force” should therefore be interpreted with the stress component specified.
With fixed, bending strength requires . Substitution into the cantilever mass gives
The material performance index to maximize is therefore
Changing the free geometric variable changes the strength exponent. A stiffness-limited problem would instead involve Young's modulus; the supplied modulus does not enter this strength-only index.
The rectangular second moment of area is . The maximum bending stress in the cantilever is . If is fixed, choose the minimum allowed width . Its mass is then
Thus the material performance index to maximize is
For unequal tensile and compressive strengths, use the smaller permissible stress in this symmetric section. The derivation assumes bending-strength control, not a separate deflection, shear, buckling or existing-crack constraint.
Both outer faces of the cantilever are loaded, so use the pine's lower longitudinal compressive strength, , rather than its tensile strength. Consistent SI units give values for steel, aluminium and pine respectively. Their values are . Hence Design A: steel best, pine worst; Design B: pine best, steel worst. Aluminium is intermediate in both cases. These rankings concern the stated strength constraint; a toughness or deflection constraint could change feasibility.
In contact-mode atomic force microscopy, a sharp tip on a compliant cantilever remains in contact with the surface while a feedback loop controls its normal deflection. A laser reflected from the cantilever onto a position-sensitive detector converts bending into an electrical error signal. With calibrated deflection and stiffness , the normal force is approximately under the stated contact convention.
A piezoelectric scanner raster-scans the sample or tip. The controller adjusts vertical position to maintain the chosen deflection setpoint; that vertical displacement is recorded as topography. The right panel of the microscopy diagram identifies tip, cantilever, optical deflection readout, controller and vertical actuator.
Figure 1.
Original instrument diagrams; the right panel shows contact-mode atomic force microscopy and the recorded vertical feedback signal
. Details.
The result is a probe-accessible surface, affected by finite tip shape, sample deformation, adhesion and feedback bandwidth. Lateral friction can distort or damage a soft surface, so constant deflection is not a guarantee of unchanged specimen structure.
The measured potential is a spatially weighted electrostatic signal, not an infinitesimal-point voltage. The apex, cone and cantilever all contribute capacitance; increasing lift height weakens the signal and broadens its lateral averaging. Small pits can consequently show reduced or distorted contrast.
Imperfect topographic tracking changes the true separation on the lift pass, producing crosstalk correlated with pit geometry. Tip contamination or work-function drift changes the reference. Adsorbed water, mobile charges, illumination, atmosphere and surface states can change the specimen during measurement; semiconductor band bending can also depend on the applied electrical excitation.
There is a tradeoff between small lift height and suppressing short-range/topographic interference. Repeat at several heights, check trace/retrace and an electrical reference, and compare with independently measured electrical response. A stable potential contrast is evidence of a local electrostatic difference, not by itself proof of a particular dislocation's conductivity.
The controlled quantity is the maximum repulsive force in each tapping cycle. In PeakForce quantitative nanomechanical mapping, a vertical modulation below the cantilever resonance produces a force-distance cycle at each pixel. The optical lever measures deflection; the calibrated cantilever stiffness converts this to force. Subtract the measured peak force from the setpoint and use proportional-integral control to adjust the mean vertical scanner position. Excessive peak force makes the controller withdraw the sample or tip; insufficient force makes it approach. The scanner position required to maintain the setpoint supplies the topography signal. This controls the top of the repulsive branch in the preceding curve, rather than the oscillation amplitude used in conventional tapping mode.[ref]
Figure 1.
Peak-force feedback with an independent vertical modulation
. Source. Details.
The atomic force microscope measures the cantilever deflection , converted to normal force by after calibration. On approach, the force is initially almost zero, then attractive; a sufficiently steep attractive force gradient causes jump-to-contact. After contact, further approach bends the cantilever and indents the sample, giving a rising repulsive branch. With tip-surface separation increasing to the right, approach proceeds from right to left. The instability occurs when the magnitude of the attractive force gradient exceeds the cantilever stiffness. The contact-branch slope depends on both cantilever and sample compliance.
Figure 1.
Approach force curve and the additional retraction information used in mechanical mapping
. Source. Details.
friction force microscopy records cantilever torsion caused by lateral tip–sample force. Choose the detector sign so that friction on a flat region contributes positively during a left-to-right trace. On a common laboratory-position axis, write the two signals schematically as
Here is an offset and represents a slope-related lateral-force or detector cross-talk contribution. A reversed electronic convention reverses all plotted signs without changing the physical conclusions.
At the same controlled normal contact force , the flat-region contribution is approximately . The lower perovskite structure plateaux give a larger trace–retrace separation than the upper inorganic surface, because their coefficient of kinetic friction is larger. Genuine friction reverses sign with scan direction, so a larger friction magnitude means a higher positive trace and a more negative retrace in this convention.
Sloped edges produce additional peaks or shoulders. The normal reaction to a slope has a lateral component, and the finite tip flank can contact the rim, so edge contrast is not a direct map of the local friction coefficient. With the two scans registered and load control maintained,
The half-difference therefore emphasizes friction while the half-sum diagnoses common geometrical contributions. Real contact mechanics and imperfect registration can prevent exact cancellation. The bottom panel of the figure shows these two kinds of contrast separately.
The recorded profile has approximately 30 nm deep, flat-bottomed depressions with broadened, sloping edges. The full-height regions are the inorganic surface; the lower plateaux are the infiltrated perovskite structure. With adequate feedback control, imaging trace and retrace should agree when plotted against the same laboratory position.
The tip has a finite opening angle. Its flank contacts a pore rim before its apex can follow an ideal vertical wall, so tip convolution makes a narrow depression appear narrower and its sidewall less steep. For a flank angle measured from the vertical, the horizontal distance needed to descend a height is . Along the cantilever, the nominal front and back values give
Thus the two edges can have different apparent slopes. Across the cantilever, the nominal side angle instead gives about nm on each side. The scan orientation relative to the cantilever must therefore be specified before choosing the relevant pair of slopes.
Even a 100 nm wide pore leaves ample room for the tip apex to reach a flat bottom: in the along-cantilever idealization its bottom plateau is about nm wide. A 200 nm pore gives about 178 nm. The tip height is far greater than the 30 nm step, so it does not prevent access. An unreported finite apex radius adds edge rounding; its size cannot be inferred from the listed flank angles.
Figure 1.
Original model of tip convolution, feedback lag and friction-loop contrast
. Details.
Use a force-tilted activation barrier for a generic one-coordinate escape model. A tensile force tilts a free-energy landscape to . If is the initial minimum and the saddle, define and the zero-force barrier . The relevant activation barrier is measured relative to the initial minimum, not an arbitrary absolute transition-state energy. To first order in force,
The attempt prefactor is assumed force independent, and the saddle displacement is approximated as constant. This single-barrier approximation need not survive large landscape deformations or a change of escape pathway. TODO: user-deferred biomolecular implementation; the protein-specific shear-force arrangement and attachment chemistry are not solved.
For a generic force-ramp measurement, a calibrated atomic force microscope cantilever pulls an attachment while a stage changes separation. cantilever deflection gives force , and the cantilever compliance correction subtracts from imposed separation to give tether extension, with further corrections for other apparatus compliance. A prescribed stage speed does not automatically produce constant force loading rate when tether stiffness varies; the assumed must be measured or controlled. The original schematic below illustrates a generic nonbiological bond/tether geometry. Protein-specific construction is deferred.
The no-rebinding approximation is a physical escape assumption: after the link breaks, separation removes the partners from the capture region before appreciable reassociation. It requires the integrated return/association probability during that separation to be small. If the separated state stays nearby or the force is held without separation, setting can fail. Under the stipulated approximation, with the link initially intact,
This survival under a linear force ramp differs from the probability density for rupture. It starts at one, decreases to zero and shifts to larger force as the loading rate increases. For small , ; a slow ramp allows much escape before substantial force is reached.
The median rupture force under a linear ramp satisfies , giving
At high loading rate, the slope of against tends to ; it is not exactly for a slow ramp. Fit the full expression across loading rates to recover and . The rupture-force probability density is , whose most probable rupture force under a linear ramp is if , and zero otherwise. median and mode are different observables and must not be interchanged when fitting.
At known temperature, yields the activation distance ; gives the zero-force mean lifetime in this exponential model. Absolute barrier height additionally requires the attempt prefactor, and neither equilibrium binding free energy nor an association rate follows from dissociation-only data. A loading-rate-dependent rupture force is a kinetic readout, not an intrinsic equilibrium strength. TODO: user-deferred biological conclusions. The protein-specific stability interpretation and cell-adhesion strategy are left unsolved.
Figure 1.
Original generic pulling arrangement, tilted energy landscape and calculated survival under a force ramp
. Details.
A force-curve mapping mode that controls the peak repulsive normal force during repeated low-frequency tip-surface contacts. Calibrated force curves provide topography and spatial maps of adhesion, deformation, modulus and dissipated energy. Mechanical-property accuracy depends on tip geometry, cantilever calibration and the contact model.