The photopic luminous efficiency function is dimensionless and describes standardized relative sensitivity for photopic vision. It is normalized to and tabulated over –. The wavelength coordinate is in standard air, rather than vacuum; changing coordinates requires converting the spectral radiant flux density as well. The official prescription uses linear interpolation between the specified 1 nm entries.[ref]
Write for the standard-air wavelength at . With spectral radiant flux in and in nanometres, the luminous flux is
The integral has units of watts, and converts them to lumens. For radiation at a single wavelength with total radiant flux , this reduces to . For a sufficiently narrow finite band, this is an approximation requiring nearly constant across the band.[ref]
Planck's law 2026-10-13
Planck's law gives the spectral radiance of an ideal blackbody in thermal equilibrium at absolute temperature . It predicts how much power is carried in each small frequency or wavelength interval, per projected area and per solid angle. The whole curve is determined by : no material-specific parameter enters this ideal Planck function.
Writing for frequency in hertz and for wavelength in metres, the two forms of the Planck function are
Here is the Planck constant, the Boltzmann constant, and the speed of light in vacuum. Their exact values in the International System of Units are , , and . [ref]
The units matter: is measured in , whereas is measured in . Both are spectral radiance, rather than radiant exitance or energy density. To report per micrometre, multiply its per-metre value by ; a micrometre-wide interval is metres wide.
For a computed example, at and , the exponent is . Substitution into Planck's law gives , or per micrometre. At the same wavelength, the illustrative temperatures and give and in the latter units. These are ideal predictions for stipulated temperatures.
Figure 1. . The vertical axis is spectral radiance per micrometre, including the conversion from per metre. Increasing temperature raises the curve at every fixed wavelength and moves its marked peak toward shorter wavelengths. The peaks are approximately , , and for , , and , respectively; only to of each spectrum is shown. Details.
The key change of variables is . Corresponding positive intervals carry the same power, so and
Equal-width wavelength intervals have unequal widths in frequency; the factor compensates for this. Merely replacing the horizontal-axis labels of by does not produce . This explains the frequency and wavelength peaks of Planck's law.
A short derivation uses the Boltzmann distribution for a single electromagnetic mode with thermal energies , . Put . The geometric series gives and mean thermal energy . The electromagnetic mode density in a box, including both independent polarization of an electromagnetic wave states of a photon, is per volume per unit frequency. Multiplication gives spectral energy density ; isotropy gives . This counts thermal excitation energy, excluding zero-point energy. [ref]
Planck's law connects quantum mechanics to a measurable radiation curve. Its small- limit is the Rayleigh-Jeans law, ; its large- limit is the Wien approximation, . The suppression at high frequency makes the total energy density finite. Integrating the curve gives the Stefan–Boltzmann law, while locating its peak gives Wien's displacement law. Thus one formula predicts both the spectral shape and the total emission.
Plane electromagnetic wave Created 2026-09-24 Updated 2026-10-14
A monochromatic vacuum plane electromagnetic wave has phase , where is the wavevector, the angular frequency, and . Its field amplitudes obey and .
For the linear polarization and orientation in the visualization, choose a right-handed Cartesian coordinate system and put
Constant phase means , so propagation is along . The cross product fixes the positive Poynting vector. Both fields reverse together every half-period and are in phase. Reversing the propagation direction with the same electric orientation reverses the magnetic orientation.
These fields are spatially uniform across every transverse plane. An infinite plane electromagnetic wave is an idealization with infinite total electromagnetic energy; it is useful as a local approximation and as a building block for superposition. For frequency , and wavelength , independently of .
For blackbody star and planet observed at one common wavelength sufficiently long for the Rayleigh-Jeans law to hold for both,
This follows by substituting in the thermal eclipse depth. A radiative flux ratio alone does not determine the radius without a planetary brightness temperature and a stellar temperature. Spectral peaks interpreted with Wien's displacement law must refer to , not or .
Point spread function Created 2026-10-05 Updated 2026-10-14
A point spread function is an imaging system's response to an optical point source. Here it means the distribution of image light intensity: an ideal point object becomes a finite spot, often with faint rings, because of diffraction. An unresolved star and a sufficiently small bead emitting fluorescence are practical approximations to an optical point source. Their image tells us how nearby object details will overlap, and reveals blurring from optical aberrations, defocus, and, for an optical telescope, astronomical seeing.
A useful convention separates shape from throughput. At a fixed image plane, let be the point spread function with . Its units are inverse area. A point delivering detected radiant flux at image position produces light intensity ; this point spread function normalization makes the integral over the image equal to . A profile divided by its own peak is convenient for comparing widths, but is a different point spread function normalization and does not by itself compare total radiant flux.
In incoherent imaging, different object points have negligible mutual spatial optical coherence over the measurement, so their image light intensities add. If the response also obeys translation invariance over the chosen field of view, the expected continuous image is the convolution
Here is the object brightness distribution expressed in image-scaled coordinates, with magnification already included, and is constant transmission. For , where is the two-dimensional Dirac delta function, this gives . A changing response across the field of view instead requires a position-dependent integral kernel ; one measured point spread function is then only a local calibration.
In coherent imaging, the superposition principle applies to the complex wave amplitudes. With amplitude spread function , the corresponding ideal model is
Thus generally differs from : coherent contributions produce wave interference terms that depend on their relative wave phases. Their time average vanishes for mutually incoherent emitters, which explains the intensity convolution. Partial optical coherence requires the correlations between object points as well.[ref]
Figure 1.
Ideal circular-pupil point responses and the overlap of two mutually incoherent sources
. Details.
The figure is an original diffraction model, not measured data. All panels have the same total radiant flux: each member of a pair contributes half the single-source radiant flux. Dashed curves show the separate light intensity contributions; the solid curve is their sum. The point spread functions have tails, so even sources satisfying the Rayleigh criterion overlap. As separation decreases, the image changes continuously; there is no abrupt disappearance of all information at one separation.
A motivating measured example is Measurement of point-spread function (PSF) for confocal fluorescence microscopy by InCheon Song and others (2005). A laser, delivered through an optical fiber, excited fluorescent beads on a slide through a , numerical aperture oil-immersion objective. A moving stage scanned the bead; emitted light returned through the objective, an emission filter and a pinhole to a photodiode. The readout was fluorescence light intensity against scanned position. Under the same setup, nominal bead diameters , , and produced progressively broader measured spots. The size comparison controls for the finite object: the measured response is the convolution of bead emission with the effective point spread function, rather than a pure point response. Inverting this convolution model gave a response compared with the theoretical one. The smaller bead gave a weaker signal relative to noise; uncertainty in actual bead diameter also affected the restored width. These are limits of the calibration. In confocal microscopy, excitation and pinhole detection jointly determine the effective point spread function; this experiment does not measure the ideal widefield Airy disk plotted above.[ref]
For a new calibration, useful additional controls are to measure detector background, avoid saturating the photodetector, repeat the scan to check stability, and compare isolated beads at several positions in the field of view. Record wavelength, numerical aperture, pinhole setting, sampling interval and optical focus. These are practical recommendations, not extra measurements claimed for the published experiment.
Polarimetry Created 2026-10-14 Updated 2026-10-18
Polarimetry measures optical rotation with a polarimeter. The vertical Vernier Go Direct instrument uses a light-emitting diode source, fixed polarizer, sample cell, motor-driven polarimeter analyzer, photodetector and optical rotary encoder. Calibrate using the same solvent at fill height; rinse with sample, fill to , and read the meniscus to . The readout plots illumination against analyzer angle; a maximum near zero or a cosine-squared fit locates .[ref]
An actual teaching experiment in the Truman State University-hosted Optical Activity Laboratory handout, revised July 2013, uses sucrose chemical solutions. Weigh the sucrose, dissolve completely without heating, and measure final solution volume for concentration. Use one polarimeter throughout; measure the water blank's extinction reading, then each sample's extinction reading and actual liquid path length near . Subtract the blank and plot corrected optical rotation against concentration. The attached apparatus description specifies a vertical light path, an incandescent or tungsten source with yellow filter, sheet polarizers and a rotating eyepiece polarimeter analyzer. Its coarse demonstration permits temperature control to the nearest , path length to the nearest millimeter and approximately angular precision. These are design tolerances, not measured results.[ref]
For a quantitative comparison, equilibrate blank, reference and sample at the same recorded temperature; use a thermostatted cell when needed. This is a recommended control, not a claim that the teaching instruments provide a thermostat. A yellow filter also passes a band of wavelengths, so its result need not match a narrow-source reference exactly. The laboratory's concentration series tests the linear model rather than establishing it in advance; a second series varying only path length tests the other factor.
For our original ideal model, Malus's law gives
where is light intensity immediately before the ideal polarimeter analyzer, after sample transmission. It is not the intensity of the original unpolarized source. The angle difference is evaluated in consistent units; the plotted degree values are converted to radians before computing the cosine. Assume transmitted light retains linear polarization, with negligible scattering and no polarization-dependent absorption. The pure positive model, mixture and racemate shift the maxima to , and , respectively. Minima shift by the same amount, to , and . Thus an extinction measurement and a maximum measurement recover the same shift after their respective blanks are subtracted.
Figure 1.
Computed optical rotations and polarimeter analyzer intensity for a stipulated enantiomer pair
. Original model, not experimental data: , and total pair concentration . Left: additive optical rotation versus positive-member mole fraction. Right: normalized Malus's law curves for , with a close view of their extinction minima. Normalization by each trace's pre-analyzer compares orientations rather than sample transmission losses. Details.
Repeat blanks and use a reference to check drift and scale. Bubbles, particles, stressed windows and sample absorption can spoil the ideal curve. Vernier specifies accuracy; the manual's fit reproducibility does not establish that accuracy. It warns against interpreting its illumination readout quantitatively and requires a transparent homogeneous sample.[ref] Neither a zero reading nor a matching specific rotation uniquely identifies a molecule.
A Michelson interferometer turns displacement into a count of interference fringes: a moving mirror changes both the outward and return paths. This lets a wavelength act as a ruler whose divisions need no engraving. Assume normal incidence, a fixed reference arm, stable refractive index along the moving arm, and fixed vacuum wavelength . Write for the wavelength in that medium and take positive away from the beam splitter. The change in optical path difference, the accumulated phase difference, and the signed number of complete interference fringe cycles are
The factor two comes from propagation geometry: moving the mirror by adds before reflection and afterward. One bright-to-bright cycle means mirror travel ; bright-to-dark means , half a cycle. All lengths must use compatible units, and is dimensionless. A constant reflection phase difference changes the initial fringe position but cancels from this change formula.
For an original hypothetical example, suppose in the actual surrounding medium and a continuous scan gives cycles, including the final fractional cycle. Then . With the wavelength treated as exact, a fringe-reading measurement uncertainty of cycle contributes to the displacement measurement uncertainty. An uncertainty of in a local phase difference instead corresponds to . This illustrates measurement resolution, not a complete claim of accuracy: wavelength calibration, air changes, mirror tilt and missed cycles can dominate.
To separate the oscillating readout from its optical coherence envelope, use the local model
Here is detected light intensity, is the mean level, is a fixed bias wave phase, and is interferometric visibility. A fine scan of measures the carrier phase difference; a coarse scan can map the temporal optical coherence envelope through . The envelope maximum locates matched delays only under the source and apparatus assumptions; it is not a bright fringe identifying an arbitrary mirror position. Even with known , one intensity reading leaves the cosine sign and whole-cycle ambiguities. Continuous motion of known direction, or an independently obtained pair of signals in phase quadrature, is needed to recover signed travel through reversals.
Figure 2.
Proposed Michelson interferometer readout and hypothetical optical coherence and refractive index controls
. Details.
The figure's original control example stipulates equal arm light intensities, matching linear polarizations, and a Lorentzian optical spectrum giving with . In vacuum, arm mismatches and give and . Matching the arms improves the phase difference readout without changing its travel per cycle. A fine mirror scan of one cycle changes by only for the stipulated source, much less than . These are calculated values, not historical data.
A proposed measurement sequence follows from this model. Overlap the returning beams, match their linear polarizations, find a high-contrast delay and use a small calibrated mirror scan to determine the sign and fringe period. Read a fixed spatial sample of the interference fringes with a photodiode; record each arm separately and the dark signal, avoid detector saturation, and retain enough time resolution to count every cycle. Approach mechanical settings consistently and repeat the return scan to expose lost motion or drift. For very small changes, bias near phase quadrature, where is largest; an intensity extremum has zero first-order sensitivity. A known drive voltage alone does not provide measurement calibration for mirror travel: compare the actual travel with the fringe count using an independently known wavelength.
Control temperature, pressure, vibration and air currents because the instrument measures optical path difference, not mirror motion alone. For a fixed reference arm and fixed , the first-order change of the moving arm's round-trip path is . Thus an unrecognized mimics . A hypothetical , , and give apparent travel. Holding the mirror still while varying or monitoring the air conditions tests this confounder. Separate checks of alignment, wavelength stability and thermal expansion are necessary before sub-fringe measurement resolution becomes a defensible measurement uncertainty.
Consider a chosen primitive cubic lattice with and one identical point scatterer per primitive unit cell. Retain the incident wavelength . This is an ideal point model. Its reciprocal lattice vectors are for integer triples other than . Write . The associated interplanar spacing and first-order Bragg angle satisfy
The reflection labels use the Miller index convention, keeping unreduced triples to encode diffraction order. With that convention, set in Bragg's law for each label. The reflection is equivalently the second diffraction order of : does not imply extra scatterers halfway between the occupied planes. Counting both descriptions as distinct reflections would count the same direction twice.
For , , and , respectively, and the calculated interplanar spacings are , , and . Their calculated deflections are , , and . Equal steps in give equal steps in , not equal steps in . This regularity makes an angular pattern a test of a proposed crystal structure.
Figure 1.
Calculated angular positions for four reflections of an ideal primitive cubic point lattice
. Separate strips show positions only, for and . Vertical row placement and marker length encode no wave intensity; this is not a simulated intensity spectrum or measured data. The four labels are a selected subset. Details.
A real crystal structure has a Bravais lattice together with an atomic basis of a crystal. The crystal lattice structure factor selects the periodic directions; the crystal basis structure factor combines the scattering amplitudes of atoms within a unit cell and can cancel a reflection. Consequently Bragg's law alone predicts neither allowed reflections nor their wave intensities. [ref] Systematic absences further restrict observed families; for example, face-centered cubic lattice centering permits only indices all even or all odd in conventional cubic indexing. The four strips above apply to the stated primitive cubic lattice, not to sodium chloride or a face-centered cubic lattice. [ref]
A prism minimum-deviation index measurement infers a transparent sample's relative refractive index from two angles: its known apex angle and the smallest deviation of a selected spectral line. A spectrometer combines a source and slit, a collimator, a rotating optical prism table, and a viewing optical telescope with an angular readout. The collimator supplies a nearly parallel beam; the optical telescope locates its emergent direction. This turns an otherwise difficult measurement of the phase velocity of light into a measurement of angles.
For the symmetric path through a uniform optical prism, with the same external medium at both faces and positive refractive indices,
The angles are measured from face normals. This is the minimum deviation relation, not a formula for an arbitrary prism orientation. An absolute refractive index additionally requires .
Figure 1.
Prism minimum-deviation index measurement: symmetric ray geometry and an independently calculated deviation minimum
. The original model chooses and at one fixed wavelength. The left panel marks face normals, equal internal angles, apex angle and deviation from the undeflected incident direction. The right panel calculates deviation versus incidence from two applications of Snell's law; the marked minimum is the only orientation used for the index inference. No points are experimental observations. Details.
For example, stipulate and . Then and . Substitution recovers . These synthetic values illustrate the geometry; they are not readings for the prism in the laboratory protocol below.
A Michigan State University teaching experiment uses a mercury lamp, slit, collimator, prism table and optical telescope, with a subsidiary angular scale read to one arcminute. First focus the optical telescope on a distant object; with the prism removed, align its crosshair to a narrow vertical slit image, focus the collimator, and establish the direct-beam zero. Insert and centre the prism. Select the green mercury spectral line, tabulated at , then rotate the prism while tracking its image with the optical telescope. Locate the turning point where continued prism rotation reverses the image motion, and read the deviation. Re-establish the minimum separately for each other spectral line. The protocol assumes ; that assumed apex angle is not an independently measured result. [ref]
For an uncertainty-aware implementation, repeat the turning-point search from both prism-rotation directions, recheck the direct-beam reference, and record individual readings before averaging. Independently determine or include its calibration uncertainty; record the chosen spectral line, sample identity, temperature and surrounding-air conditions. These are proposed controls deduced from the inference, not claimed additional instructions or results of that manual. Finite slit width, unresolved spectral lines, imperfect collimation, table tilt, angular calibration and sample nonuniformity can limit the inferred refractive index. One arcminute readout resolution alone does not establish one arcminute total measurement uncertainty. A single selected wavelength gives one index, not a measured optical dispersion curve.
A short geometric derivation uses and . Reversibility exchanges entrance and exit. At the symmetric stationary path, and ; applying Snell's law gives the displayed index relation. The plotted calculation checks that this stationary point is a minimum on the chosen transmitting branch. To assess local sensitivity, differentiate the inferred index using angles in radians:
This first-order relation exposes both angular contributions; it is not an instruction to add signed systematic errors as independent random errors. In the synthetic example, a deviation error of arcminute at fixed changes the inferred relative refractive index by approximately .