Absolute pressure 2026-10-14
Pressure referred to an ideal vacuum, rather than to atmospheric pressure or another chosen reference. Gauge pressure instead subtracts atmospheric pressure.
Absolute pressure in Boyle's law 2026-10-14
With gauge pressure referenced to atmospheric pressure , absolute pressure is . Thus Boyle's law becomesEven a constant atmospheric pressure cannot be dropped from these sums. For the calculated to example, take . The initial and final gauge pressures are and ; their products with the respective volume are and , which are unequal. Zero gauge pressure means balance with the surroundings, not absence of gas or absence of wall momentum transfer.
For a frictionless piston of surface area with its outer face exposed to atmospheric pressure, the additional inward force required for mechanical equilibrium is . This is why a mechanical load can reveal a pressure difference even though the equation of state requires absolute pressure. With piston friction, the load alone no longer determines ; a direct pressure sensor avoids that particular inference.
Atmospheric pressure Created 2026-10-14 Updated 2026-10-18
Atmospheric pressure often provides the reference for a gauge pressure measurement: . The absolute pressure is measured relative to an ideal vacuum, whereas zero gauge pressure means equality to the chosen reference. The actual atmospheric pressure can vary; it must be specified when converting between these descriptions. [ref]
University Physics Volume 1 by William Moebs and others. OpenStax: atmospheric, gauge and absolute pressure
. Boyle's law 2026-10-14
Boyle's law states that a fixed amount of substance of an ideal gas at constant absolute thermodynamic temperature has constant product of absolute pressure and total volume:Here is measured relative to a vacuum, not relative to the surrounding atmospheric pressure. Halving the volume doubles the absolute pressure; doubling the volume halves it. A real gas approximately follows Boyle's law when it is sufficiently dilute and far from a phase transition.[ref]
2.1 Molecular Model of an Ideal Gas. §2.1, The Gas Laws, Figure 2.4, and equations 2.1–2.2: fixed temperature and particle number, absolute pressure, and the dilute-gas limit
For an original calculated example, suppose a sealed gas sample has and . After isothermal compression to ,Both states have , since . The volume ratio is dimensionless, so using the same volume units in numerator and denominator leaves the pressure in kilopascals. Although has units of energy, its value is not the compression work: isothermal ideal-gas work involves an integral along the path.
The blue curves show calculated ideal gas states, not measurements. The left panel is a branch of a hyperbola; the right panel plots exactly the same absolute pressures against . The dashed line uses gauge pressure with a stipulated constant atmospheric pressure of . Its displaced intercept explains why absolute pressure in Boyle's law matters.
A short derivation starts from the ideal gas law, . Fixing the amount of substance and absolute thermodynamic temperature fixes , where is the molar gas constant. The microscopic picture in ideal gas pressure from wall collisions is equally appealing: at fixed temperature, the typical particle kinetic energy stays fixed, while compression increases the number density and wall momentum transfer per unit surface area and time. A larger absolute pressure follows without requiring hotter particles.
The constraints are essential. A leak changes the amount of substance. Rapid compression can raise the temperature before heat transfer restores thermal equilibrium, as in the isothermal gas syringe experiment. Molecular size and interactions in a dense real gas invalidate the ideal gas approximation; the Van der Waals equation describes one such correction. Equal endpoint temperatures allow an endpoint test of Boyle's law, but do not establish that every intervening state lies on an isothermal process.
For fixed mass , mass density satisfies , so it is proportional to absolute pressure. In a dilute bubbly flow, Boyle's law allows bubble volume and its contribution to buoyancy to increase as hydrostatic pressure decreases, provided gas mass is retained, dissolution is negligible, and the temperature stays approximately fixed. Its usefulness is that a simple ratio predicts a substantial size change without tracking individual molecules.
Buoyancy Created 2026-09-28 Updated 2026-10-13
Buoyancy is the resultant upward force exerted by a resting fluid on an immersed body in a downward gravitational field. The fluid pushes inward, perpendicular to each part of the body's surface. Because hydrostatic pressure increases with depth, the upward push from below exceeds the downward push from above. Opposite horizontal forces cancel; buoyancy is the remaining vertical force. A sinking body still receives this upward support. [ref]
More precisely, pressure on a small surface of area with outward unit normal vector exertsThe minus sign says the force points into the body. On a cube, the top force points down, the bottom force points up, and the side forces cancel in pairs because opposite sides have the same pressure at each depth. This is a surface traction; gravity on the body itself is a separate body force. [ref]
For a computed example, stipulate resting water with constant mass density and uniform gravitational acceleration . A fully immersed cube has side , top depth and bottom depth . Each horizontal face has area , and its volume is . Its top and bottom gauge pressures areThe corresponding forces relative to the common surface pressure are downward and upward. Therefore the net buoyancy isUsing absolute pressures adds the same to both face forces, leaving their difference unchanged. Zero gauge pressure at the surface does not mean zero absolute pressure. Increasing both depths by the same amount also leaves buoyancy unchanged in this constant-mass density model.
Archimedes' principle turns this calculation into a rule for bodies of any shape. It explains floating equilibrium and lets us infer an immersed body's volume from the difference between its force from gravity and the force needed to support it. In particular, buoyancy depends on the displaced volume and the surrounding mass density, whereas the body's own mass density decides how its force from gravity compares with that support.
Clément-Desormes experiment 2026-10-14
The Clément-Desormes experiment infers heat capacity ratio from a rapid pressure reduction followed by reheating. Use a rigid bottle, pump or bellows, a quick release valve, and a U-tube pressure gauge connected to the bottle and atmosphere. First pump to a small excess pressure, close the inlet, and wait for thermal equilibrium with the room at . Measure , where is local absolute atmospheric pressure. Briefly open the release valve until the bottle reaches , then close it promptly. The remaining gas is cooler than the room. As it warms back to at fixed bottle volume, its pressure rises to , with in the ideal procedure. The rebound reveals cooling during release. [ref]
https://labs.physik.kit.edu/prakt-nebenfach-manuals/Experiments/Adiabatenexponent_Luft/. KIT Faculty of Physics, Clément-Desormes laboratory protocol: release and pressure rebound
The bottle loses molecules while the valve is open: it is not a closed system whose geometrical volume expands. In the well-mixed, negligible-heat transfer discharge model with constant heat capacity ratio, the remaining gas obeys . Subsequent heating at fixed amount of substance and bottle volume gives by the ideal gas law. Eliminating the unmeasured yieldsThe logarithms use absolute pressure ratios. The approximation requires small gauge pressures ; are signed height differences between the two liquid surfaces, and the final equality uses the same liquid in both readings with hydrostatic equilibrium. It is not an instruction to insert gauge readings into the logarithmic formula. For synthetic states generated with , and , the model gives . The logarithmic inference recovers ; the small-pressure approximation gives approximately .
Wait for a stable initial pressure after pumping and for a stable final pressure after closing. Vary the valve-open interval and initial excess pressure across repeated trials: closing too early leaves residual excess pressure, while leaving the valve open during warming lets more gas escape and suppresses the rebound. Test the seal for leaks, and read both liquid levels consistently. In a variant with incomplete release, measure the actual immediate post-release absolute pressure and replace the numerator by . The Zurich laboratory manual explains that liquid-column oscillations can delay this reading until appreciable reheating has occurred; its proposed backward extrapolation assumes an exponential thermal relaxation and must be checked against the recorded readings. These controls separate a thermal signal from valve timing, instrumental lag and leakage. [ref]
https://www.physik.uzh.ch/~matthias/espace-assistant/manuals/en/anleitung-cd_e.pdf. University of Zurich, PHY112/122 laboratory manual, §1.3 and Appendix 1.4(a): pressure readings and reheating correction
Gas thermometer 2026-10-14
A gas thermometer infers temperature from the equation of state of a confined gas. The constant-volume version measures pressure; the constant-pressure version measures volume. For an ideal gas,Here is particle number, is amount of substance, is the Boltzmann constant, and is the molar gas constant. The reference and test states must be internally equilibrated, and must be known on the absolute thermodynamic temperature scale. A pressure ratio cancels the unknown gas inventory and bulb size in the ideal constant-volume limit. This cancellation is the method's appeal: a mechanical measurement transfers an absolute temperature scale between different materials. The Maxwell speed distribution and ideal gas pressure from wall collisions explain the microscopic origin of the response. [ref]
2.1 Molecular Model of an Ideal Gas. OpenStax, University Physics Volume 2, section 2.1: the ideal gas law
For an original ideal-model calculation, choose , and . Then . These are stipulated values, not observations. Both pressures must be absolute, referred to vacuum. If the surrounding atmospheric pressure were , the corresponding gauge pressures would be and ; their ratio is three, which would incorrectly predict . A pressure difference from the atmosphere cannot be substituted into the ideal gas law.
An original schematic of constant-volume gas thermometer measurements
. Simplified measurement layout inspired by the NBS apparatus, not a replica or a report of measurements. A controlled bath thermalizes the bulb, separate calibrated thermometers check its surroundings, and a diaphragm transmits pressure to an absolute-pressure standard without mixing gases. The connecting tube contains gas at other temperatures; its inventory belongs in the gas thermometer dead-space correction. Details.A motivating calibration experiment. Schooley's account of the NBS/NIST work describes a helium-filled metal bulb, a capillary, a capacitance diaphragm gauge, and a mercury pressure standard. Circulating bath liquid passed the bulb assembly and platinum resistance thermometers. The resistance probes checked the surrounding temperature; gas pressure supplied the independent thermodynamic comparison. In an early series, workers filled the cleaned bulb with purified helium at the highest test temperature, checked pressure over three hours, then measured at lower temperatures and near a reference at . Six filling pressures allowed extrapolation toward zero pressure. [ref]
To turn this principle into a readable measurement protocol, use the following controls. These steps are a proposed teaching interpretation of the cited apparatus, rather than additional reported NBS observations:
- Define the inventory. Evacuate and clean the bulb and connecting tube, fill with a known pure gas, close the filling valve, and keep the same total particle number during each reference/test comparison. Monitor pressure at a maintained temperature before and after a run. A leak changes ; a slow drift can also reflect contamination, sorption, or imperfect equilibration, so pressure stability alone does not identify its cause.
- Measure the reference. Place the bulb in a controlled environment with independently assigned . Record after thermalization, together with the independent thermometer readings and their spread. A reference near has an absolute temperature near , rather than zero temperature. Repeating reference readings tests whether the inventory or apparatus changed during the comparison.
- Measure the test state. Change the bath setting, allow the bulb and gas to settle, and record only when the surrounding thermometers and pressure are sufficiently stable for the intended accuracy. Compare readings at different positions and vary the settling time. Record the connecting tube's temperature profile too: uniform bath temperature does not make the entire connecting tube isothermal.
- Measure the volume changes. Determine the bulb's thermal expansion and its deformation from differences between inside and outside pressure. Measure the connecting tube's dimensions and the effective valve volume. “Constant volume” is the ideal model; a real vessel expands. These measurements supply corrections, even though the absolute bulb volume cancels in the simplest pressure ratio.
- Measure pressure against a reference. For a liquid-column pressure standard, measure the vertical column height, the liquid's temperature-dependent mass density, local gravitational acceleration, and the reference-side pressure. Include diaphragm offsets and the gas's hydrostatic pressure change between the sensor and bulb. The pressure needed in the equation of state is at the bulb, not automatically at the remote gauge.
The corresponding NBS apparatus used a protective case with an outer helium layer held close to the bulb pressure, reducing pressure deformation while aiding thermal contact. Thermal expansion was measured on bulb material. Gas in the capillary remained partly outside the hot region, requiring a measured tube temperature profile and geometry. Pure working gas also mattered: contaminants that sorb on surfaces can change the freely circulating inventory. These are physical effects to measure or bound, rather than scatter that disappears by averaging. [ref]
https://nvlpubs.nist.gov/nistpubs/Legacy/IR/nbsir85-3133.pdf. J. F. Schooley, Thermometry, NBS IR 85-3133, chapter V, gas thermometry and the NBS experiment
For illustration, neglect connecting-tube gas and describe a uniform bulb with a dilute virial expansion in particle number density :Here and is the second virial coefficient in the particle-density convention; a molar-density convention rescales the coefficient. The second relation follows exactly from the definition of for the same , and shows the signs of the corrections. If the bulb expands by and at both states, the toy pressure ratio above gives , rather than . Improved expansion and helium nonideality models really changed the temperatures inferred from the 1989 NIST data: Harvey's 2023 reanalysis found expansion updates of roughly – and nonideality updates of roughly to . Those are reported corrections to historical results, not the much larger illustrative correction just calculated. [ref]
With equal corrected bulb volumes, negligible or corrected connecting-tube gas, fixed reference and test temperatures, and a dilute classical single gas phase,Repeat the reference/test comparison for several sealed fills, reducing between comparisons but keeping it fixed within each pair. The extrapolation varies density at fixed endpoint temperatures; it does not cool one fill toward zero temperature. Real gases have different finite-density corrections but share the classical ideal gas intercept under these assumptions. The limit does not remove leaks, bulb expansion, gauge offsets, or the gas thermometer dead-space correction. At very low pressures, thermal transpiration can make remote sensor and bulb pressures differ through a nonisothermal capillary; this must be corrected too. The NBS work measured this pressure correction using a separate comparison of narrow and wider tubes. [ref]
At , the toy models give and , respectively, although their imposed true test temperature is the same. Their zero-density intercept is the reusable quantity. Measuring several low-density points permits a fit and a check for curvature; asserting a straight line from one finite-density point supplies neither. A practical fit must balance nonideality against increasing relative pressure error and the capillary corrections, and keep its extrapolation uncertainty.
Gauge pressure Created 2026-10-14 Updated 2026-10-31
Gauge pressure is pressure measured relative to the surrounding atmospheric pressure; absolute pressure is instead measured relative to an ideal vacuum. At an open surface , so the constant-mass density hydrostatic pressure law becomes . Here at the surface denotes equality with atmospheric pressure. It does not denote zero absolute pressure. [ref]
Hydrostatic pressure Created 2026-09-24 Updated 2026-10-13
Hydrostatic pressure is the pressure distribution in a resting fluid, balancing its body force from gravity. Take upward and a uniform downward gravitational field , where . The vertical pressure gradient satisfiesA short derivation balances the vertical forces on a thin horizontal layer of area and thickness : . Dividing by and taking the derivative gives the equation above. For constant mass density , integration from the surface at givesThus pressure increases with depth . For the stipulated water with mass density and gravitational acceleration , depths and give gauge pressures and . Their difference produces the buoyancy of the example cube. [ref]
https://www.feynmanlectures.caltech.edu/II_40.html#Ch40-S1. Feynman Lectures, II §40-1: hydrostatics
Ideal gas law Created 2026-10-14 Updated 2026-10-14
The ideal gas law is the equation of state relating the absolute pressure , volume , particle count , and absolute thermodynamic temperature of a classical ideal gas:Here is the Boltzmann constant, is the Avogadro constant, is the molar gas constant, and is the amount of substance in moles. The distinct symbol avoids confusing amount of substance with the number density , measured in : the same law can be written . Use absolute pressure, relative to a vacuum, rather than an unconverted gauge pressure. [ref]
2.1 Molecular Model of an Ideal Gas. OpenStax University Physics Volume 2, section 2.1: particle-count and molar forms
The exact SI definitions of the Avogadro constant and Boltzmann constant giveThis numerical value is exact; the applicability of the ideal gas model to a real gas is an approximation. [ref] [ref]
https://www.bipm.org/en/si-base-units/kelvin. BIPM, exact Boltzmann constant in the kelvin definition
For , , and , direct substitution givesAt fixed temperature and particle count, doubling the volume to halves the pressure to approximately : this is Boyle's law. At fixed volume and particle count, doubling the temperature from to doubles the pressure to approximately . These ratios use kelvins, not readings in the degree Celsius scale: is , and is .
Ideal gas law: pressure versus volume at fixed temperatures, and pressure versus absolute temperature at fixed volume
. Details. The curved slices show Boyle's law at three temperatures; the straight slice shows proportionality to absolute temperature at one fixed volume. They are relations among equilibrium states, rather than a prescribed process history. Compression could hold the temperature fixed through heat transfer, or raise it during adiabatic compression. To calculate a process's heat or work, supplement the ideal gas law with the process constraint, the first law of thermodynamics, and the relevant internal energy or heat capacity information. The graph's extrapolation toward is mathematical; it does not assert that a real gas remains classical or gaseous there.
For a stellar mixture with fixed mean molecular weight, the ideal gas law can be written , where is mass density and is the stellar gas constant. It assumes negligible interaction corrections and a nondegenerate gas; excitation or ionization can change the particle count and thermodynamic response.
In his 1834 memoir on the motive power of heat, Émile Clapeyron combined gas laws into a relation between pressure, volume, and temperature while analysing heat engines (printed page 164). The equation above uses modern SI notation and absolute thermodynamic temperature; it is not a transcription of his historical temperature convention. [ref]
https://www3.nd.edu/~powers/ame.20231/clapeyron1834.pdf. Clapeyron, Mémoire sur la puissance motrice de la chaleur (1834), page 164
Isothermal gas syringe experiment 2026-10-14
A gas syringe makes the distinction between a constant-temperature path and equal-temperature endpoints observable. The Ideal Gas Law Apparatus (TD-8596A), Instruction Manual 012-08420B by PASCO scientific specifies a TD-8596A gas syringe with a mechanical stop, an internal thermistor, an attached pressure sensor and data acquisition. Record the stop volume with the coupling disconnected, then set and seal the coupling. Record pressure and temperature, compress quickly to the stop near , and hold until both readouts settle. The manual gives roughly for settling and about for the thermistor response. Releasing the plunger gives a second transient and a final volume to record.[ref]
Ideal Gas Law Apparatus (TD-8596A), Instruction Manual 012-08420B by PASCO scientific. pp. 2–4: Introduction, Equipment Setup and Experiment 1 procedure
The manual describes a pressure overshoot and temperature rise during rapid compression, followed by cooling at the held volume; the settled pressure remains higher than initially. This illustrates approximately adiabatic compression followed by heat rejection, rather than an isothermal compression throughout. Use the settled initial and compressed readings to test Boyle's law, correcting both syringe readings by the unmarked tubing volume :Use absolute pressure, not gauge pressure. A transient comparison of must pair simultaneous pressure and temperature readings: the sensor lag can separate their peaks.[ref]
Ideal Gas Law Apparatus (TD-8596A), Instruction Manual 012-08420B by PASCO scientific. pp. 4–5: Constant Temperature and Varying Temperature analyses; p. 8: answers 1–3
For an additional test of the path, repeat with small compression steps, wait for stable temperature and pressure at every step, and repeat in reverse. Keep the sealed amount of substance, connecting tubing and ambient temperature fixed; prevent hand warming of the barrel and check for leaks by watching for continued pressure drift at held volume after thermal equilibrium. These are proposed controls, rather than reported measurements. Agreement of equilibrated points with Boyle's law supports the fixed-temperature ideal gas law approximation, but does not prove that the rapid segments between them were isothermal processes. A single local temperature sensor can miss temperature gradients; plunger friction can produce direction-dependent readings. The pressure and volume data estimate reversible work only insofar as the path approaches a reversible thermodynamic process; they do not directly measure heat. After motion stops, cooling at fixed volume has zero boundary work and , even though the complete equal-endpoint-temperature ideal gas run has .
Linearized plot of Boyle's law 2026-10-14
A linearized plot of Boyle's law uses as its horizontal coordinate. Then absolute pressure satisfies : the slope is , and the ideal intercept is zero. In the figure, and in and become and in and . Both give slope . The Vernier experiment includes this reciprocal-volume graph as its extension.[ref]
6 Boyle's Law: Pressure-Volume Relationship in Gases. Experiment 6, printed p. 6–4, Extension steps 1–2: pressure versus reciprocal total volume and linear fitting
At constant atmospheric pressure, gauge pressure instead obeys , with the same slope and a negative intercept. A nonzero fitted intercept in an intended absolute pressure plot can indicate sensor bias or an incorrect volume correction, but is not by itself a unique diagnosis. The origin is a mathematical extrapolation to infinite volume and zero absolute pressure, rather than an additional measured state.
Measuring the Joule-Thomson effect with a porous plug Created 2026-10-14 Updated 2026-10-20
The 2024 operating instructions for PHYWE's 04361-00 apparatus describe a frit between two gas chambers. A copper-coil heat exchanger first restores inlet temperature after cylinder-regulator cooling. An upstream manometer reads gauge pressure; the outlet stays at atmospheric pressure. Two NiCr-Ni thermocouples sit 2–5 mm from the frit. The digital meter's difference mode reports precision.[ref]
Joule-Thomson apparatus 04361-00, Operating instructions by PHYWE. PHYWE, Joule-Thomson apparatus 04361-00, 2024 instructions, pp. 1–2, sections 2–4: heat exchanger, frit, manometer, probes and difference readout
The manual specifies at most gauge pressure, one hour of apparatus acclimatization, half an hour of meter warm-up, about 1.5 minutes of flow settling, then 10–15 seconds of averaging. It identifies a stagnant-gas zeroing error that shifts the fitted intercept and notes a maximum nitrogen signal of about in this setup.[ref]
Joule-Thomson apparatus 04361-00, Operating instructions by PHYWE. PHYWE, 2024 instructions, p. 2, section 4, Hints concerning the experiment: pressure limit, equilibration, drift, averaging and zero compensation
Define and , so a small-drop fit is . A constant sensor offset leaves the slope unchanged; pressure-dependent drift or heat leakage can alter it. The observable is the differential temperature slope across the plug, not cylinder-outlet cooling. Display precision alone does not establish accuracy or prove an isenthalpic process. This low-pressure, room-temperature apparatus does not map a complete Joule-Thomson inversion curve.
Since the velocity tends to zero at infinity, the steady Bernoulli equation gives . On the wall, therefore,Compute the finite source-induced force relative to uniform ambient pressure, or equivalently use the gauge pressure . The fluid lies above the wall, so its normal force on the wall points in for positive pressure. The negative excess pressure found here consequently produces an upward net force, toward the source.
For the symmetric section , there is no horizontal force, and the vertical excess force isHere an antiderivative of is . Taking gives the hydrodynamic attraction of a wall to a line source:Each component is a force per unit length of the wall in the out-of-plane direction. The result is an attraction even though the source sends fluid away from itself: faster tangential motion reduces the wall pressure.
The excess torque about the origin isfor every , because the integrand is odd. Thus the symmetric infinite-wall limit isThe symmetric prescription matters: the separate one-sided moment integrals grow logarithmically, whereas the force integral converges absolutely. This is the appropriate Cauchy principal value for the infinite wall. If one instead asks for the absolute fluid force with no balancing ambient pressure on the other side, its vertical component on a finite section is and has no finite infinite-wall limit. The finite answer above is the flow-induced load.
Boyle's law states that a fixed amount of gas at constant temperature obeyswhere is absolute pressure, not gauge pressure. It is an ideal-gas approximation, not an assertion that every rapid compression is isothermal.
The respiratory diaphragm contracts and lowers its dome, increasing thoracic volume. Mechanical coupling through the pleural space lets the lungs expand; alveolar pressure falls below atmospheric pressure, drawing air in. Relaxation of the respiratory diaphragm and elastic recoil drive quiet expiration. These respiratory mechanics describe ventilation, distinct from chemical cellular respiration.[ref]
How the Lungs Work: What Breathing Does for the Body by the National Heart, Lung, and Blood Institute. “Breathing in” and “Breathing out”: diaphragm movement, expansion and recoil
For the stipulated sealed, isothermal gas, initial and final volumes are and . Expressing absolute pressure as a mercury-column height,The mercury column is supported by the atmospheric-to-internal pressure difference:The specified volume change belongs to the historical idealized model. Real tube-air volume, temperature changes and limits on respiratory-muscle pressure generation would affect a physical apparatus.
Past exam of the theoretical mechanics course of the University of Oxford 2009 A 3 b Solution 2026-10-31
For spatially uniform , the gap equation gives . Axisymmetry makes this . Its general integral contains a logarithmic term; bounded pressure at zero excludes that term. Enforcing zero gauge pressure at the outer edge gives the axisymmetric squeeze-film pressureThe mean radial velocity is . At the material edge this must equal , so . HenceIt is conservation of the incompressible liquid volume, with no liquid entering or leaving through the plates or material rim.
When the plates approach, and inside: the liquid is squeezed outward. Separation requires relative to the surrounding air and inward motion of the edge. Negative gauge pressure is not intrinsically impossible, but substantial suction can cause cavitation or air invasion, and an advancing air/liquid interface can develop fingering. Thus the same mathematical lubrication solution exists for either sign, whereas its opening-plate physical validity needs these additional failure mechanisms to be absent; the closing case is the robust regime of the stated idealized model.
Past exam of the topics in fluid mechanics course of the University of Oxford 2023 3 b Solution 2026-10-25
Let be a representative mixing-layer thickness, , and let be a representative streamwise velocity. Use the scalesThe transverse velocity scale follows from the mean continuity equation. For turbulent transverse transport to balance streamwise inertia,With this ordering, the streamwise eddy-stress divergence isThus transverse diffusion of streamwise linear momentum is retained, while streamwise diffusion and the contribution are smaller by . Molecular kinematic viscosity is neglected where .
In the transverse momentum equation, inertia and the retained eddy stresses have scale , whereas an order-one transverse pressure variation would give . Hence the leading dimensionless transverse balance is . The resulting boundary layer equations for the planar turbulent mixing layer areHere . The effective pressure convention explained in part (a)(iii) is understood. Since the outer streams have uniform horizontal velocities, their inviscid momentum equations give ; choosing the far-field gauge pressure as zero gives throughout this approximation.
Write and below, so the subscript in this shorthand denotes differentiation with respect to . The vertical pressure gradient is zero, and integrating twice with the wall boundary conditions gives the local Couette-Poiseuille flowThe first two terms are Couette flow due to the moving plate; the last is plane Poiseuille flow due to the local pressure gradient. Its dimensionless volume flux per unit width isUse the Leibniz integral rule and incompressibility to differentiate this integral:Thus the flux is independent of position, andThe normal velocity can then be recovered from incompressibility:It is zero on the lower wall by construction and zero on the upper wall by the constant-volume flux relation. This supplies the remaining component of the leading velocity field.
The endpoint pressure conditions are matching conditions, not an assertion that the lubrication approximation remains valid at arbitrarily large dimensional . On an outer length scale , a viscous pressure variation is of size , whose ratio to the inner scale isEven an outer inertial pressure scale has ratioConsequently the gap pressure is much larger than either outer variation. Matching to fluid with a common ambient pressure at both ends, and taking that value as zero gauge pressure, givesThere is an overlap region before the wall slope becomes order one. The infinite endpoints represent matching through that region. An externally imposed pressure difference comparable with would change these conditions and the resulting volume flux; equal outer pressure is the physical assumption being used here.
Pressure Created 2026-09-24 Updated 2026-10-14
Pressure measures the compressive normal force per unit surface area. For a uniformly loaded surface of surface area , the magnitude of the perpendicular force is ; locally, . Its SI unit is the pascal, . A force parallel to the surface contributes shear stress, rather than pressure. In a gas, the steady macroscopic force comes from many microscopic transfers of momentum; ideal gas pressure from wall collisions makes this connection explicit. [ref]
https://openstax.org/books/university-physics-volume-2/pages/2-2-pressure-temperature-and-rms-speed. OpenStax University Physics Volume 2, section 2.2: microscopic origin of pressure
For a computed example, a uniform pressure of on a piston of surface area produces a perpendicular force of . If the opposite face is exposed to , the net force due to the two pressures is , directed toward the lower-pressure side. The pressure on either face and their difference answer different questions; gauge pressure records a difference from a specified reference, often atmospheric pressure.
A scalar pressure describes a gas in local thermodynamic equilibrium with isotropic random particle velocities: every wall orientation has the same normal force per unit surface area. Without this isotropy, normal stresses can depend on direction and a pressure tensor is needed. Thus random motion alone is insufficient; a preferred direction matters.
In thermodynamics, pressure and volume are thermodynamic conjugate variables. For a simple compressible system, reversible expansion delivers work , and at fixed entropy and particle number ,where is internal energy. A large pressure therefore means both a large wall force per unit surface area and a large amount of expansion work per small increase in volume.
Pressure coefficient 2026-10-14
The pressure coefficient compares a local static pressure with a specified reference static pressure, using a fixed reference dynamic pressure:The reference mass density, speed, pressure and location must accompany a reported pressure coefficient. Keep the same throughout a surface distribution; replacing it by local gives a different quantity and fails at a stagnation point. The definition itself requires neither Bernoulli's principle nor inviscid flow, so measured pressure coefficients remain useful when those idealizations fail.
Positive means ; zero means equality; negative means . This sign does not specify the sign of absolute pressure. For stipulated absolute and , gives absolute: a deficit, not a negative absolute pressure. If equals atmospheric pressure, the numerator is gauge pressure; a wind tunnel's reference static pressure need not equal room atmospheric pressure. These are reference choices, not different physical pressures.
Under the same-height, constant-mass density , steady flow and inviscid flow hypotheses of the Bernoulli equation, and with the same Bernoulli invariant as the reference flow,The first equation immediately gives the second after subtracting and dividing by . Hence a lossless stagnation point has , gives , and gives . For a real stationary wall satisfying the no-slip boundary condition, the velocity is zero right at the wall; inserting that wall velocity in an inviscid flow formula would wrongly predict everywhere. A thin boundary layer approximation instead relates the wall pressure to the outer-flow pressure and uses the outer velocity where the Bernoulli equation applies.
For potential flow around a circular cylinder of radius with zero circulation, the surface tangential velocity is . Thus the ideal pressure coefficient isWith measured from downstream, the front is at and the rear at zero. Both ideal stagnation points have , while the sides at and have . This exact model is a useful baseline for an experiment; its complete rearward pressure recovery is not a prediction for a separated real fluid wake.
To obtain a force, integrate the local normal contributions with their directions. Let be the unit normal vector from the body into the fluid and its closed wetted surface:The second equality removes a spatially constant reference pressure over the closed surface; it does not remove a nonuniform hydrostatic pressure or justify dropping the reference term on an isolated open patch. A negative contributes outwards in this reference-subtracted calculation, even though a positive absolute pressure still pushes inwards locally. Project onto the chosen lift and drag directions. The full fluid force also includes from the viscous stress tensor, including tangential shear stress; pressure tapping does not measure that contribution.
For the two-dimensional cylinder, force per unit span follows from per unit span and :Substituting gives : opposite points have identical pressure coefficients and opposite normal vectors. Zero drag despite a strongly nonuniform pressure is the instructive limitation associated with D'Alembert's paradox. Real viscosity, boundary layers and flow separation can prevent rearward pressure recovery and produce pressure drag. Thus the measured curve can test the model and estimate a pressure contribution without proving that total drag vanishes.
A motivating measurement is Virginia Tech's Experiment 3 - Flow Past a Circular Cylinder. Its open-jet wind tunnel holds a rotatable cylinder with end plates and 36 one-millimetre pressure tappings at ten-degree intervals. A reference Pitot-static tube supplies and to one digital manometer; a second connects a surface tap to its positive port and reference static pressure to its negative port. Record the simultaneous readings:The last equality uses the low-speed stopping relation, not an additional speed measurement. The manual requires sealed tubing and zero checks with the tunnel off. It also provides a water-column manometer: read each column's initial level, and subtract the independently measured reference-to-atmosphere pressure from the tap-to-atmosphere readings. Record tap positions after rotation. Compare the two instruments and check spanwise uniformity; end plates do not guarantee two-dimensional flow. [ref]
The inferential distinction matters: the manometers directly measure pressure differences, the ratio estimates pressure coefficients, and a surface integral estimates the pressure part of the force. Measure room absolute pressure and temperature if converting the reference difference to speed through the ideal gas law ; use thermodynamic temperature in that equation. For force reconstruction, cover the complete contour and use numerical integration with the tap geometry. Finite tap spacing, unknown end effects, probe disturbance and fluctuations limit that reconstruction; tubes and instruments also average or delay changing pressures. Retain the signed raw readings, repeated readings and reference readings so that these limits can be assessed. A small reference difference magnifies any zero error in the ratio, rather than producing a reliably large .






