Free-slip convection neutral curve 2026-10-06
For nonrotating convection under the Boussinesq approximation with fixed-temperature impermeable stress-free boundary conditions, a marginal vertical mode has neutral Rayleigh number . The minimum is at , , and . These follow by eliminating the temperature from and , then minimizing over .
Freezing 2026-10-07
Freezing is a liquid-to-solid phase transition. A moving phase boundary releases latent heat; its rate is constrained by heat transport and the Stefan condition. In a solution the liquidus depends on composition, so being below the pure solvent's melting temperature does not imply that the solution must freeze.
Freezing density increase 2026-10-14
Freezing-point depression Created 2026-09-24 Updated 2026-09-24
Freezing-point depression is the lowering of a solvent's equilibrium freezing temperature by dissolved solute. In the linear dilute approximation, with .
At a fixed temperature, the frequency maximizing and the wavelength maximizing in Planck's law are different ways of describing the same black-body radiation. They are maxima of different densities. The inverse relation between frequency and wavelength does not preserve those maxima because its interval-conversion factor varies with wavelength.
For a short extremum calculation, substitute into , leaving a positive constant times . Substitute into , leaving a positive constant times . Setting each derivative to zero givesSelect the positive roots: and . Although solves the rearranged equations, it is an endpoint, not either peak. For or , the function initially increases, its derivative crosses zero once, and it eventually becomes negative. Hence it has exactly one positive root. Each density tends to zero at both ends, so this root gives its unique maximum. The bisection method brackets these roots in and .
Consequently Wien's displacement law givesThese constants agree with NIST's tabulation. [ref] The wavelength peaks at , , and are , , and . Doubling temperature halves this peak wavelength.
At , the frequency peak is , whose converted wavelength is . The peak is instead , corresponding to about . These numbers are compatible: equal-width frequency bins and equal-width wavelength bins partition the same radiation differently. Specifying which density was maximized is essential when using a peak to infer temperature.
One spectrum, two spectral-coordinate peaks
. The left panel plots the Planck function per terahertz and marks its frequency peak. The right panel plots the same black-body radiation per micrometre and marks its wavelength peak. Dashed lines locate the counterpart of the other panel's peak. The converted frequency peak at is not the peak at : the change of variables changes density heights through . Details. Frictional heating 2026-10-14
In the ideal kinetic friction model, the loss of combined mechanical energy of the two contacting bodies becomes thermal internal energy at rateThis follows by adding the contact powers and from Newton's third law. It depends on contact slip, whereas the contact power received by either body separately depends on its velocity in the chosen inertial frame. Thus positive frictional heating is compatible with increasing kinetic energy of one body, as the moving-belt example in kinetic friction shows. Heating need not cause a temperature rise if thermal energy escapes by heat transfer.
Frost 2026-10-31
Frost point 2026-10-31
The temperature at which the given water-vapour partial pressure equals saturation vapour pressure over ice. It is defined relative to ice, whereas the usual dew point uses equilibrium with liquid water.
Frozen dew 2026-10-31
Retaining perturbation thermal diffusion givesThe displaced interfacial temperature and Stefan condition give the displayed implicit dispersion relation. In the capillarity-free limit its growing branch is exactly , which checks the limits of the small-wavelength thermal approximation for a solidification front.
Gadolinium sulfate octahydrate 2026-10-18
The hydrated chemical compound used in the Giauque-MacDougall magnetic cooling experiment.
GaN blue LED development Created 2026-10-18 Updated 2026-10-31
Isamu Akasaki chose metalorganic vapour-phase epitaxy and sapphire, which tolerated hot Ammonia. Large lattice mismatch hindered film growth. In 1985, Hiroshi Amano and Isamu Akasaki's group developed an aluminium nitride epitaxial buffer layer deposited before hotter gallium nitride growth. The buffer supplies initial growth sites; growing islands join into a continuous film. It improves growth without making the lattices identical. Isamu Akasaki separately credits Shuji Nakamura's 1991 gallium nitride buffer.[ref]
https://www.kapitza.ras.ru/~glazkov/students/Nobel-akasaki-lecture.pdf. §§3.2–4.2, pp. 13–20, Figs. 7–11: growth sequence and microscopy
The 1986 growth paper reports a vertical metalorganic vapour-phase epitaxy reactor, source gases mixed just before delivery to slanted sapphire, and subsequent gallium nitride growth over aluminium nitride. Surface microscopy, X-ray diffraction and photoluminescence tested different aspects of the result. Buffered and unbuffered growth supplied the comparison. Its reported preparation conditions should not be replaced by one universal buffer temperature: the later lecture's account of the first low-temperature trial and the paper's parameter study describe different stages of development.[ref]
Metalorganic vapor phase epitaxial growth of a high quality GaN film using an AlN buffer layer by Hiroshi Amano and others. p. 353, Fig. 1 and Table I; p. 354, surface and diffraction comparisons
A magnesium acceptor impurity in gallium nitride accepts a valence band electron, leaving a mobile electron hole. Hydrogen passivation of magnesium acceptors in gallium nitride suppresses this activity. Post-growth activation restores existing impurities' electrical activity; ionization of active acceptor impurities supplies mobile electron holes. Adding magnesium alone is insufficient.[ref]
In 1989, Hiroshi Amano, Masahiro Kito, Kazumasa Hiramatsu and Isamu Akasaki obtained p-type semiconductor conduction in magnesium-doped gallium nitride using low-energy electron beam irradiation. Hiroshi Amano recalls initially resistive samples, an unreliable hot-probe test, then Masahiro Kito's Hall effect confirmation. A junction irradiated in an M-shaped region emitted in that pattern. Treatment, carrier-sign measurement and spatial emission supplied different evidence. Earlier zinc-doped samples brightened under irradiation without becoming p-type semiconductors: brightness alone did not prove the carrier sign.[ref]
https://journals.aps.org/rmp/pdf/10.1103/RevModPhys.87.1133. §V, pp. 1135–1136, Figs. 3–4; authors of 1989 paper in references, p. 1137
Shuji Nakamura, Takashi Mukai, Masayuki Senoh and Naruhito Iwasa's 1992 thermal-activation paper established another route using annealing in nitrogen. Their companion study compared already conducting films heated in nitrogen with films heated in Ammonia: the latter became highly resistive again. Its hydrogen-complex explanation connected this atmosphere dependence to acceptor impurity passivation. These are thermal treatment and atmosphere comparisons, distinct from the earlier electron beam intervention; neither observation alone images an individual hydrogen complex.[ref]
Thermal Annealing Effects on P-Type Mg-Doped GaN Films by Shuji Nakamura and others. abstract: nitrogen annealing above 700 degrees Celsius and electrical changes
[ref]Hole Compensation Mechanism of P-Type GaN Films by Shuji Nakamura and others. p. 1258, abstract and §§1–2: activation history, nitrogen/ammonia comparison and hydrogenation hypothesis
Shuji Nakamura's two-flow metalorganic chemical vapor deposition reactor, gallium nitride buffer, thermal activation and improved indium gallium nitride growth enabled bright double heterostructure devices, followed by thinner quantum well devices.[ref]
The original 1994 paper by Shuji Nakamura, Takashi Mukai and Masayuki Senoh describes a sapphire/gallium nitride-buffer device with zinc-doped indium gallium nitride emission and aluminium gallium nitride confinement. After nitrogen annealing, etching exposed the n-side for a separate contact. At room temperature, varied forward electric currents gave emission spectra and optical-output curves alongside a current-voltage characteristic. At : output, peak wavelength, terminal electric potential difference. The peak stayed nearly fixed across the current series. The paper omits a complete calibrated collection apparatus; comparisons with earlier devices change several preparation details.[ref]
Candela-class high-brightness InGaN⁄AlGaN double-heterostructure blue-light-emitting diodes by Shuji Nakamura and others. pp. 1687–1688, fabrication and Figs. 1–4
Gas 2026-10-14
Gas clumping factor 2026-10-14
The gas clumping factor measures unresolved inhomogeneity in electron number density within a chosen averaging region. Nonnegative variance gives . At a uniform temperature, thermal bremsstrahlung scales with , whereas the thermal Sunyaev-Zeldovich effect scales with . A homogeneous cluster distance from Sunyaev-Zeldovich and X-ray signals inference then gives , for the same geometry and temperature. Correlated temperature variations require additional weighted averages.
Gas density 2026-10-18
Gas density is the mass density of a gas. For an ideal gas sample of amount of substance , mean molar mass and volume , combine with to obtainHere is the molar gas constant. For a homogeneous mixture of gases, , where is the mole fraction of component . Equal pressure and temperature make a gas density ratio equal to a mean-molar mass ratio within this model. Argon detection in a binary gas mixture gives a worked example and a temperature confounder.
Gas-globe density measurement 2026-10-18
In an idealized rigid-globe weighing, subtract the evacuated assembly's mass from the filled assembly's mass, then divide by the internal volume. Establish the filled sample's pressure and temperature before comparing gases. The subtraction removes the container's large contribution, making a small gas contribution accessible. A real measurement also needs corrections or bounds for residual gas, adsorption, leaks, changing vessel volume and external buoyancy. The Rayleigh-Ramsay argon density experiment shows why preparation-route comparisons add information beyond repeated weighings of one sample.
Gas law 2026-10-31
A relation among quantities such as pressure, volume, temperature and amount of substance for a gas.
Gas-syringe dead-volume correction 2026-10-14
In a gas syringe, the graduated barrel reading can omit a fixed connecting or sensor volume occupied by the same sealed gas. For a common settled temperature, Boyle's law then requiresThis second relation makes a graph of against a straight line; its intercept divided by its slope estimates under these assumptions. Unequal temperatures between the barrel and the connecting space require a different correction, analogous to the gas thermometer dead-space correction.
For an original hypothetical example, let , , , and . The total volume falls from to , so , rather than . The omitted volume makes the total compression ratio smaller than the barrel-reading ratio. These stipulated values are independent of the manufacturer's apparatus correction and are not measurements.
Gas-syringe measurement of Boyle's law 2026-10-14
Vernier's Boyle's Law: Pressure-Volume Relationship in Gases, Experiment 6 of Chemistry with Vernier, specifies a graduated gas syringe, a pressure sensor and LabQuest recording. Preparation sets the disconnected syringe to , then attaches it to the sensor valve, trapping air. At several plunger positions, the displayed pressure is allowed to stabilize before the paired volume is entered. The analysis tests whether remains approximately constant. It adds to each syringe reading for the sensor's internal volume; this correction belongs to that apparatus. The worksheet supplies a blank results table, not measured values.[ref]
The manufacturer's Gas Pressure Sensor User Manual identifies a membrane sensor referenced to a vacuum, hence an absolute pressure readout. Its calibration section uses the actual local atmospheric pressure for an ambient reference check, rather than a weather report corrected to sea level. Checking this baseline before sealing separates an offset error from compression physics.[ref]
Gas Pressure Sensor User Manual, Vernier. Calibration: station pressure and one-point reference; Specifications: internal volume; How the Sensor Works: vacuum-referenced membrane and absolute-pressure output
Additional controls follow from the model. Keep the same sealed amount of substance, connections and ambient temperature; persistent pressure drift at held volume after thermal equilibrium suggests a leak or another uncontrolled change. Small steps and waiting allow compression heat to leave the gas; avoiding hand warming reduces another heat input. Compare settled readings when approaching a position from either direction. Plunger friction affects the applied force and position reproducibility, so record actual pressure and actual plunger position rather than predicting pressure from the load. These are proposed controls, not reported experimental outcomes.
Nearly constant across equilibrated states supports the fixed-temperature ideal gas approximation over the tested range. Stabilized pressure alone does not demonstrate uniform or constant temperature; the isothermal gas syringe experiment adds a temperature readout to distinguish a transient from its cooled endpoint. Finite sensor accuracy and uncertain total volume limit the test, and agreement does not establish ideal behavior at every pressure. Repeated return readings help distinguish reproducibility from a one-direction trend.
Gas thermometer 2026-10-14
A gas thermometer infers temperature from the equation of state of a confined gas. The constant-volume version measures pressure; the constant-pressure version measures volume. For an ideal gas,Here is particle number, is amount of substance, is the Boltzmann constant, and is the molar gas constant. The reference and test states must be internally equilibrated, and must be known on the absolute thermodynamic temperature scale. A pressure ratio cancels the unknown gas inventory and bulb size in the ideal constant-volume limit. This cancellation is the method's appeal: a mechanical measurement transfers an absolute temperature scale between different materials. The Maxwell speed distribution and ideal gas pressure from wall collisions explain the microscopic origin of the response. [ref]
2.1 Molecular Model of an Ideal Gas. OpenStax, University Physics Volume 2, section 2.1: the ideal gas law
For an original ideal-model calculation, choose , and . Then . These are stipulated values, not observations. Both pressures must be absolute, referred to vacuum. If the surrounding atmospheric pressure were , the corresponding gauge pressures would be and ; their ratio is three, which would incorrectly predict . A pressure difference from the atmosphere cannot be substituted into the ideal gas law.
An original schematic of constant-volume gas thermometer measurements
. Simplified measurement layout inspired by the NBS apparatus, not a replica or a report of measurements. A controlled bath thermalizes the bulb, separate calibrated thermometers check its surroundings, and a diaphragm transmits pressure to an absolute-pressure standard without mixing gases. The connecting tube contains gas at other temperatures; its inventory belongs in the gas thermometer dead-space correction. Details.A motivating calibration experiment. Schooley's account of the NBS/NIST work describes a helium-filled metal bulb, a capillary, a capacitance diaphragm gauge, and a mercury pressure standard. Circulating bath liquid passed the bulb assembly and platinum resistance thermometers. The resistance probes checked the surrounding temperature; gas pressure supplied the independent thermodynamic comparison. In an early series, workers filled the cleaned bulb with purified helium at the highest test temperature, checked pressure over three hours, then measured at lower temperatures and near a reference at . Six filling pressures allowed extrapolation toward zero pressure. [ref]
To turn this principle into a readable measurement protocol, use the following controls. These steps are a proposed teaching interpretation of the cited apparatus, rather than additional reported NBS observations:
- Define the inventory. Evacuate and clean the bulb and connecting tube, fill with a known pure gas, close the filling valve, and keep the same total particle number during each reference/test comparison. Monitor pressure at a maintained temperature before and after a run. A leak changes ; a slow drift can also reflect contamination, sorption, or imperfect equilibration, so pressure stability alone does not identify its cause.
- Measure the reference. Place the bulb in a controlled environment with independently assigned . Record after thermalization, together with the independent thermometer readings and their spread. A reference near has an absolute temperature near , rather than zero temperature. Repeating reference readings tests whether the inventory or apparatus changed during the comparison.
- Measure the test state. Change the bath setting, allow the bulb and gas to settle, and record only when the surrounding thermometers and pressure are sufficiently stable for the intended accuracy. Compare readings at different positions and vary the settling time. Record the connecting tube's temperature profile too: uniform bath temperature does not make the entire connecting tube isothermal.
- Measure the volume changes. Determine the bulb's thermal expansion and its deformation from differences between inside and outside pressure. Measure the connecting tube's dimensions and the effective valve volume. “Constant volume” is the ideal model; a real vessel expands. These measurements supply corrections, even though the absolute bulb volume cancels in the simplest pressure ratio.
- Measure pressure against a reference. For a liquid-column pressure standard, measure the vertical column height, the liquid's temperature-dependent mass density, local gravitational acceleration, and the reference-side pressure. Include diaphragm offsets and the gas's hydrostatic pressure change between the sensor and bulb. The pressure needed in the equation of state is at the bulb, not automatically at the remote gauge.
The corresponding NBS apparatus used a protective case with an outer helium layer held close to the bulb pressure, reducing pressure deformation while aiding thermal contact. Thermal expansion was measured on bulb material. Gas in the capillary remained partly outside the hot region, requiring a measured tube temperature profile and geometry. Pure working gas also mattered: contaminants that sorb on surfaces can change the freely circulating inventory. These are physical effects to measure or bound, rather than scatter that disappears by averaging. [ref]
https://nvlpubs.nist.gov/nistpubs/Legacy/IR/nbsir85-3133.pdf. J. F. Schooley, Thermometry, NBS IR 85-3133, chapter V, gas thermometry and the NBS experiment
For illustration, neglect connecting-tube gas and describe a uniform bulb with a dilute virial expansion in particle number density :Here and is the second virial coefficient in the particle-density convention; a molar-density convention rescales the coefficient. The second relation follows exactly from the definition of for the same , and shows the signs of the corrections. If the bulb expands by and at both states, the toy pressure ratio above gives , rather than . Improved expansion and helium nonideality models really changed the temperatures inferred from the 1989 NIST data: Harvey's 2023 reanalysis found expansion updates of roughly – and nonideality updates of roughly to . Those are reported corrections to historical results, not the much larger illustrative correction just calculated. [ref]
With equal corrected bulb volumes, negligible or corrected connecting-tube gas, fixed reference and test temperatures, and a dilute classical single gas phase,Repeat the reference/test comparison for several sealed fills, reducing between comparisons but keeping it fixed within each pair. The extrapolation varies density at fixed endpoint temperatures; it does not cool one fill toward zero temperature. Real gases have different finite-density corrections but share the classical ideal gas intercept under these assumptions. The limit does not remove leaks, bulb expansion, gauge offsets, or the gas thermometer dead-space correction. At very low pressures, thermal transpiration can make remote sensor and bulb pressures differ through a nonisothermal capillary; this must be corrected too. The NBS work measured this pressure correction using a separate comparison of narrow and wider tubes. [ref]
At , the toy models give and , respectively, although their imposed true test temperature is the same. Their zero-density intercept is the reusable quantity. Measuring several low-density points permits a fit and a check for curvature; asserting a straight line from one finite-density point supplies neither. A practical fit must balance nonideality against increasing relative pressure error and the capillary corrections, and keep its extrapolation uncertainty.



