A Michelson interferometer uses a beam splitter to divide collimated input light between a fixed mirror and a moving mirror. Returning beams recombine at the beam splitter; one output reaches a photodetector. A compensator plate can equalize the glass traversed in the two arms. Moving one mirror by changes the optical path difference by .
The usual nominal Fourier transform spectroscopy specification is , where is the maximum optical path difference measured from equal arms. It gives
This is a nominal spectral resolution element convention.[ref] A symmetric rectangular scan from to instead has a sinc instrumental line shape with first-zero offset . Quoting that first-zero criterion would give cm; interferogram apodization and other linewidth criteria change the numerical factor. An exact mathematical minimum requires such a criterion.
In the measurement chain, the Michelson interferometer is the spectral analyzer or modulator, between the collection optics and the photodetector. It encodes the incoming radiation spectrum into an interferogram; detector electronics digitize that signal and a Fourier transform reconstructs the radiation spectrum. It does not itself turn radiant power into an electrical signal.
Let the two returning monochromatic fields at the chosen detector port be
Here is wavenumber, rather than temporal frequency. An optical-cycle average uses and . In units absorbing the detector's field-to-power conversion,
The choice of output port can reverse the cosine sign. For mutually incoherent components of the radiation spectrum, add powers rather than instantaneous fields, giving
Subtracting the constant term leaves the Fourier cosine transform of the detected spectral weighting , including throughput and detector response.
The printed prefactor requires a normalization convention. In the derivation, is the amplitude of each returning field at the detector, not an entrance amplitude with input power . For a lossless balanced beam splitter and literal input power , conservation of energy instead gives , so the source formula identifies . Taking both and the printed formula literally would give at zero optical path difference, a concrete normalization contradiction.
Outcome determinism means that a complete hidden state , together with the chosen measurement settings, fixes each outcome. The observed probabilities can still arise from an ensemble of unknown . This differs from deterministic unitary evolution of a wavefunction, which does not by itself specify a unique measurement outcome. Quantum nonlocality means failure of a Bell-local account of the correlations, under assumptions including measurement independence; it is stronger than the existence of entanglement and does not imply controllable faster-than-light signaling.
Use the first ket for the Positron and the second for the electron. Before either final beam splitter, the first pair of splitters and annihilation produce
The photon state is orthogonal to the surviving charged-particle sector. Its probability is , while the remaining sector has squared norm . The states below retain that photon amplitude and are normalized over all trials, without conditioning on nonannihilation. For compactness write , and similarly for .
Take equal arm lengths, no plate, and coherent illumination. At a coated beam splitter, choose the real-amplitude convention where transmission has amplitude and reflection has amplitude from one side and from the other; which reflection carries the minus sign is a basis convention. The two paths interfere constructively at detector 1 and destructively at detector 2 in the diagram's labeling. Thus
For example, writing the beam splitter as and the two equal-reflection mirrors as , one obtains . The first output component propagates downwards to detector 2; the nonzero second component propagates upwards to detector 1. The overall minus sign is an unobservable global phase. The redistribution is interference of amplitudes, rather than two successive independent coin tosses.
For a monochromatic photon of vacuum wavelength , a plate of thickness replaces air by glass and adds phase . Neglect losses and reflections, or assume antireflection coatings. Between the first beam splitter and mirror 1, the affected arm is the second basis component, so . The successive quantum states, omitting common propagation phases, are
The first output component feeds detector 2 and the second feeds detector 1, giving
The imbalance gives . Thus the measured probabilities determine only up to , and hence only a corresponding family of thicknesses. There is no unique absolute thickness from this one measurement. The zero-plate limit recovers , .
Bosonic creation operators commute, so the cross terms cancel:
Thus
The square root of two in is essential. Destructive interference removes the one-particle-per-output amplitude; the particles do not independently choose ports.
For photons this is the ideal Hong-Ou-Mandel effect. Its laboratory test combines one photon from each input at a balanced beam splitter and scans relative arrival delay while measuring output coincidences. Spatial, polarization and spectral indistinguishability must be controlled; distinguishable photons restore a nonzero coincidence baseline. The original experiment used down-converted photons, a translated beam splitter and coincidence counting, with an independently estimated accidental-coincidence background.[ref]
Thorlabs' silicon FDS100 experiment sends a collimated, fibre-coupled laser beam through a beam splitter to the photodiode and a reference optical power meter. Varying reverse bias and electrical load tests linearity. The measured readout approaches saturation when its electric potential difference consumes the bias.[ref] A separate low-signal setup uses a laser, polarizers, an optical chopper, optical attenuators and a lock-in amplifier inside a light-tight enclosure. These are actual apparatus choices, not the parameters of the constructed example above.[ref]
For an independent dark current check, Thorlabs enclosed a photodiode in nested metal boxes to exclude light and suppress interference, used a Keithley 6487 ammeter supplying reverse bias, and monitored temperature. Only one specimen per type was tested; the nearby temperature sensor did not directly contact the semiconductor. Such a dark control measures the baseline but cannot certify spectral responsivity.[ref]
To infer spectral responsivity at a selected wavelength, measure the radiant flux actually reaching the test photodiode, rather than just the reference arm. A reference-arm measurement calibration must include the beam splitter ratio and path losses. With signed readouts in one fixed orientation, subtract a beam-blocked baseline and use
for direct electric current readout or a calibrated linear transimpedance amplifier, respectively. Repeating at several incident powers tests whether this ratio is constant. Blocking only the source estimates ambient light plus electronic baseline; covering the detector excludes ambient light too. Their difference distinguishes optical background from dark current and readout offset. For a broad optical spectrum, the ratio is a spectrum-weighted response, not generally at one wavelength.
For independent Poisson distribution collection, the root mean square shot noise in a one-sided rectangular frequency band of signal bandwidth is
Here is reverse dark current; thermal electrical noise and amplifier noise also contribute.[ref]
Subtracting a mean dark current removes its offset, not its fluctuations: subtracting a constant leaves the variance unchanged. Consequently makes dark current shot noise small relative to signal shot noise; it does not by itself establish high signal-to-noise ratio. With an independently measured baseline, the variance of that estimate must also be propagated. Increasing the measurement signal bandwidth raises the admitted noise. Calibration of wavelength, reference radiant flux, readout gain and collection geometry, together with drift and linearity, determines what measurement uncertainty can be claimed.
Photon counting 2026-10-18
Photon counting registers discrete detector events and their times. A beam splitter sends the selected light to two photodetectors; a coincidence circuit or time-recording system compares the outputs. With a vacuum input at the unused optical port, splitter fractions and independent quantum efficiencies , a selected input photon reaches the two recorded channels with probabilities and . Conditional on input occupation , each pair of distinct photons contributes to the cross-channel pair moment. Therefore
Their normalized ratio recovers . This assumes number-resolving counts, or sufficiently dilute gates for threshold detectors to approximate them. One photon cannot create both channel counts in this ideal model.
Figure 1.
Functional two-detector photon-correlation measurement and its normalization
. Original schematic: optical selection and a beam splitter feed independent detectors; all cross-channel time differences populate a histogram. A single photon produces at most one ideal detection. The empty splitter input is vacuum. Details.
A postselection scheme in which two polarization-entangled photon pairs pass through polarizing beam splitters, a half-wave plate and a nonpolarizing beam splitter. One photon at each of a trigger and three output detectors selects a GHZ state in the three output polarizations, provided optical mode matching removes distinguishable pair labels. All four detections are needed for the selected event; the trigger alone does not guarantee the three other output occupations.
Precision displacement measurement with a Michelson interferometer Created 2026-10-18 Updated 2026-10-20
A Michelson interferometer turns displacement into a count of interference fringes: a moving mirror changes both the outward and return paths. This lets a wavelength act as a ruler whose divisions need no engraving. Assume normal incidence, a fixed reference arm, stable refractive index along the moving arm, and fixed vacuum wavelength . Write for the wavelength in that medium and take positive away from the beam splitter. The change in optical path difference, the accumulated phase difference, and the signed number of complete interference fringe cycles are
The factor two comes from propagation geometry: moving the mirror by adds before reflection and afterward. One bright-to-bright cycle means mirror travel ; bright-to-dark means , half a cycle. All lengths must use compatible units, and is dimensionless. A constant reflection phase difference changes the initial fringe position but cancels from this change formula.
For an original hypothetical example, suppose in the actual surrounding medium and a continuous scan gives cycles, including the final fractional cycle. Then . With the wavelength treated as exact, a fringe-reading measurement uncertainty of cycle contributes to the displacement measurement uncertainty. An uncertainty of in a local phase difference instead corresponds to . This illustrates measurement resolution, not a complete claim of accuracy: wavelength calibration, air changes, mirror tilt and missed cycles can dominate.
To separate the oscillating readout from its optical coherence envelope, use the local model
Here is detected light intensity, is the mean level, is a fixed bias wave phase, and is interferometric visibility. A fine scan of measures the carrier phase difference; a coarse scan can map the temporal optical coherence envelope through . The envelope maximum locates matched delays only under the source and apparatus assumptions; it is not a bright fringe identifying an arbitrary mirror position. Even with known , one intensity reading leaves the cosine sign and whole-cycle ambiguities. Continuous motion of known direction, or an independently obtained pair of signals in phase quadrature, is needed to recover signed travel through reversals.
Figure 2.
Proposed Michelson interferometer readout and hypothetical optical coherence and refractive index controls
. Details.
The figure's original control example stipulates equal arm light intensities, matching linear polarizations, and a Lorentzian optical spectrum giving with . In vacuum, arm mismatches and give and . Matching the arms improves the phase difference readout without changing its travel per cycle. A fine mirror scan of one cycle changes by only for the stipulated source, much less than . These are calculated values, not historical data.
A proposed measurement sequence follows from this model. Overlap the returning beams, match their linear polarizations, find a high-contrast delay and use a small calibrated mirror scan to determine the sign and fringe period. Read a fixed spatial sample of the interference fringes with a photodiode; record each arm separately and the dark signal, avoid detector saturation, and retain enough time resolution to count every cycle. Approach mechanical settings consistently and repeat the return scan to expose lost motion or drift. For very small changes, bias near phase quadrature, where is largest; an intensity extremum has zero first-order sensitivity. A known drive voltage alone does not provide measurement calibration for mirror travel: compare the actual travel with the fringe count using an independently known wavelength.
Control temperature, pressure, vibration and air currents because the instrument measures optical path difference, not mirror motion alone. For a fixed reference arm and fixed , the first-order change of the moving arm's round-trip path is . Thus an unrecognized mimics . A hypothetical , , and give apparent travel. Holding the mirror still while varying or monitoring the air conditions tests this confounder. Separate checks of alignment, wavelength stability and thermal expansion are necessary before sub-fringe measurement resolution becomes a defensible measurement uncertainty.
In 1977, H. Jeff Kimble, Mario Dagenais, and Leonard Mandel excited a dilute sodium atomic beam with a resonant dye laser. Apertures restricted the observed atoms; a microscope objective collected fluorescence perpendicular to both beams. A beam splitter sent the collected light to two photomultipliers. Shaped pulses drove the start and stop inputs of a time-to-digital converter; its delay histogram rose away from zero delay, indicating photon antibunching. Turning the atomic beam off measured scattered-laser backgrounds. More than one contributing atom and scattered light limited the comparison with an ideal isolated two-level system; the paper explicitly treated corrections.[ref]
A simple reason for suppressed immediate re-emission is visible in an ideal two-level system. After spontaneous emission, the atom is in . Its emission operator obeys , so two instantaneous emissions are impossible. Excitation must rebuild the excited-state population before another emission. This argument explains the zero-delay idealization; the actual photon counting record additionally depends on collection, background, timing, and the number of contributing atoms.
Sagnac effect 2026-10-18
The Sagnac effect is a difference in return times, or an associated interference fringe displacement, for light sent in opposite directions around a closed optical path in a rotating apparatus. The light leaves one beam splitter, travels around the same loop in opposite senses, and returns to the rotating beam splitter and readout. For a slowly rotating apparatus in flat spacetime, the leading laboratory arrival-time difference is
Here is the vacuum speed of light, is the apparatus's angular velocity relative to an inertial frame, and is the vector area oriented by the chosen positive direction around the closed curve . With positive rotation about the upward normal vector, the counterclockwise beam is labelled and arrives later. Reversing rotation, or reversing the orientation convention alone, changes the sign of .
The surprise is that a small loop can reveal its rotation by comparing two beams locally, even though each local vacuum speed of light measurement gives . A rotating reference frame is not a single inertial frame: a beam must catch the moving receiver after a full circuit. The Sagnac area law makes this difference a practical way to measure angular velocity, while the Relativistic Sagnac delay on a circular ring distinguishes laboratory coordinate time from the receiver's proper time.
Sagnac interferometer 2026-10-18
A Sagnac interferometer uses a beam splitter to send light with optical coherence around a common closed optical path in opposite directions, then recombines the beams to read their phase difference. Mirrors or a guiding path define the loop; the source, beam splitter, loop and readout rotate together in the basic experiment. A small relative beam tilt makes spatial interference fringes whose displacement reveals a change of phase difference. At one fixed output, a photodetector can instead record the corresponding change in light intensity.
Figure 1.
Counterpropagating light around a rotating closed loop, with a shared beam splitter and readout
. Original rectangular teaching layout. The vector area points out of the page for the positive counterclockwise path. The source and readout rotate with the mirrors. Arrow offsets separate the two directions visually; they do not represent different physical paths. The drawing is not a reconstruction of the historical mirror arrangement. Details.
In 1913 Georges Sagnac mounted a lamp, collimator, beam splitter, four mirrors and photographic plate on a horizontal rotating platform. The mirrors defined a closed polygon; the two beams traversed it in opposite senses. A slight beam splitter adjustment produced visible interference fringes. A motor brought the platform to steady rotation, and switching the lamp exposed the plate. Comparing opposite senses of rotation revealed a reversing displacement. Changing the beam splitter adjustment supplied a further check against optical deformation. The report compared measured and calculated displacements. Georges Sagnac interpreted this as evidence for a luminiferous aether; the relativistic derivation below reproduces the delay without that hypothesis.[ref]
The readout measures a relative wave phase, not two stopwatch readings separated by an optical transit time. Reversal isolates a contribution odd in angular velocity: if a signed interference fringe position is to leading order, then cancels the fixed offset . Changes of alignment or temperature correlated with reversal can survive this subtraction. Thus the experimental inference concerns a reproducible rotating-loop phase difference, with controls for disturbances; it is not by itself a unique test of an underlying propagation hypothesis.
Arthur Leonard Schawlow's 1981 lecture describes an external-cell arrangement: a beam splitter makes opposed beams, an optical chopper interrupts the pump, and a lock-in amplifier reads the probe detector with the chopper reference. Pump-on reduces absorption and raises the detector signal; scanning the laser finds the narrow resonance. Its first application used iodine; pressure and laser jitter limited the resolution.[ref]
Figure 1.
Counterpropagating pump and probe with synchronous transmission readout
. Original functional drawing of the cited external-cell principle. Beam separation is exaggerated; this is not a reconstruction of the pulsed sodium apparatus. Details.
For an original detector model, let be the probe power reaching a photodetector without sample attenuation and let be the sample's probe optical depths. After subtracting a stable detector background, the transmitted powers are
A positive synchronous change means less probe photon absorption, provided incident probe power is stable and scattered pump light is excluded. For small optical-depth changes the fractional signal is approximately . A lock-in amplifier's voltage also depends on gain, reference phase and chopping waveform; it is not automatically this dimensionless ratio.
Useful controls are blocking the pump, detuning far from the feature, reversing or adjusting beam alignment, and repeating the scan at several pump light intensities. Frequency calibration, sample pressure, transit through the beams and laser frequency fluctuations must be assessed separately. Nearly opposed beams leave a residual Doppler shift: for wave vectors , the simultaneous-resonance condition gives . Exact opposition removes this transverse sensitivity in the first-order model.
Ultraviolet Created 2026-10-05 Updated 2026-10-20
Ultraviolet is electromagnetic radiation with wavelengths shorter than the violet end of ordinarily visible light. The CIE convention uses –, subdivided into UV-A at –, UV-B at – and UV-C at –. These divisions name intervals; they do not mark universal changes in every material's response. The boundary with visible light is not sharp, and other fields use near-, far-, vacuum- and extreme-ultraviolet divisions. [ref]. For example, NASA's Solar Dynamics Observatory observes extreme-ultraviolet bands below ; the CIE interval is therefore not the full extent of astronomical ultraviolet usage. [ref].
The CIE ultraviolet interval corresponds to frequencies about – and photon energies –, using vacuum wavelength and with Planck constant and speed of light . Shorter wavelength means larger energy per photon; larger irradiance means more incident power per area. These are different comparisons. At the same incident power, a ultraviolet beam has fewer photons per second than a infrared beam, even though each ultraviolet photon has more energy.
A common ultraviolet mechanism is photon absorption promoting an electron from a ground state to an excited state of a chemical atom or molecule. By contrast, many infrared absorption bands involve stretching or bending chemical bonds. For either mechanism, both the energy difference and the coupling to the incident electromagnetic field matter. One concrete example is the absorption discussed for butadiene: an electron moves between molecular orbitals, and the position of the absorption reveals their energy separation. Ultraviolet absorption can therefore probe molecular orbitals, while infrared spectroscopy probes many molecular motions. [ref].
Subsequent chemical reactions depend on the absorbing chemical substance and its environment; the label ultraviolet alone does not predict a reaction rate. At sufficiently short wavelength, a photon can cause photoionization if its energy exceeds the relevant ionization energy. There is no single wavelength that separates all ionizing from all nonionizing behavior for every chemical substance. Likewise, eventual heating after photon absorption is possible in both ultraviolet and infrared. The distinction is useful because different optical spectra expose different processes, not because the two ranges obey different laws of electromagnetism.
Ritter's 1801 observation used a chemical readout rather than a thermometer. In Caltech/IPAC's account, sunlight passed through a glass optical prism and silver chloride was placed in different parts of the optical spectrum. Darkening was weak toward red, stronger toward violet, and pronounced beyond the visible violet edge. The measured evidence was a change in the chemical substance's appearance at a spatial location with no visible color, establishing radiation beyond the eye's ordinary range. It was not a calibrated measurement of a universal ultraviolet optical spectrum. [ref].
A controlled modern interpretation of such a chemical test would compare equal exposure durations and matched fresh samples, keep an unexposed sample in darkness, monitor temperature, and distinguish scattered visible light from the intended short-wavelength beam with characterized filtering. These are proposed controls, not an assertion that Ritter performed them. Unequal sample preparation or heating can change a reaction rate. The optical prism, terrestrial atmosphere and sample response restrict which wavelengths the test detects. Darkening is a detector-weighted response, so its spatial maximum is not automatically the maximum of incoming radiant flux or of photon energy.
Thomas Larason and Jeanne Houston's NIST report documents a quantitative ultraviolet detector comparison. An argon source feeds a double-diffraction grating spectrometer; a shutter and mirrors deliver the selected band to a working-standard or test photodiode, successively at the same position. A beam splitter supplies a monitor photodiode. Amplifiers and voltmeters record detector and monitor simultaneously; shutter-closed readings provide dark baselines. The – scan, typically with bandpass, measures spectral responsivity. Monitor normalization suppresses source drift; known spectral lines calibrate wavelength. The report assesses temperature effects and unwanted light outside the band, weighted by detector response. The latter need not cancel between different spectral responsivities. [ref].
This determines electrical response without visual perception. Bandpass and spatial nonuniformity limit the calibration; aging and changed temperature can affect later use. The scan includes visible light and ultraviolet; no response below follows from it.
For a narrow beam, the idealized substitution calculation is
Here is spectral responsivity; and are dark-subtracted detector and monitor electric currents. This follows by writing for each detector and using the monitor to track changes in radiant flux. It assumes linear response, unchanged beam splitting and collection geometry, and a sufficiently narrow band. For an illustrative standard with , ratios and give . These invented arithmetic inputs illustrate the method; they are not NIST measurements.
An independent photon-counting model connects spectral responsivity to quantum efficiency. If is the number of collected electrons per incident photon, there is no internal multiplication, and is the elementary charge, then
At illustrative constant and , this gives about at and at . The ultraviolet readout is smaller in this model because there are fewer incident photons, not because their individual energies are smaller. Real quantum efficiency depends on wavelength, windows, absorption and carrier collection; a detector's band gap and construction determine its usable range. Constant across these two wavelengths is an analytical comparison, not a claimed specification of a real photodiode.
Figure 1.
Photon energy across conventional ultraviolet and infrared ranges, with a schematic detector-substitution experiment
. Details.
The curve is the calculated relation , and the shaded ranges use the stated CIE conventions. The lower diagram abstracts the detector comparison rather than reproducing an instrument drawing. Together with the infrared heat-balance illustration, it shows why a calibrated detector response, rather than what looks bright to an observer, is the bridge between invisible radiation and a quantitative measurement.