Range straggling 2026-10-18
Range straggling is the event-to-event spread in particle range for otherwise similar incident particles, caused by fluctuations in their energy loss and transport through matter. It must be distinguished from measurement uncertainty and from the incomplete recording of particles that escape.
Ratiometric measurement 2026-11-05
A ratiometric measurement infers a quantity from a ratio of two measured signals chosen to share an unwanted multiplicative variation. The useful question is which inputs actually share that variation: a reference resistor error or a change in the sensor itself generally survives the cancellation.
In a shared-excitation resistance ratio measurement, a current source passes the same electric current through a sensor of electrical resistance and a reference resistor . Their electric potential differences are and . An ideal normalized analog-to-digital converter reading is thereforeThe current source need not have a perfectly constant output for this cancellation, provided both measurements sample the same electric current, the denominator is nonzero, and the converter stays within its operating ranges. This is a standard arrangement for a resistance thermometer.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. §2.1.3, pp. 10–11, shared-current resistance measurement
For an original ideal example, take , and . Then , and . Increasing by gives and : the ratio remains . If the converter instead uses an independent fixed reference and the calculation assumes the original , the inferred electrical resistance becomes .
Shared excitation cancels in a resistance ratio; a fixed independent reference retains the excitation error. Original ideal calculation, not measured data.
Details. The reference-resistor error in ratiometric resistance measurement remains even with perfect shared excitation. If the actual reference resistor is but the calculation uses its nominal value, thenThus a reference error makes the nominal estimate of a true sensor about . Reference accuracy and its variation with temperature still matter.[ref]
Nor does the ratio remove sensor self-heating in a ratiometric resistance measurement. Joule heating dissipates , which is in the example. A temperature rise changes the actual electrical resistance of a resistance thermometer; the ratio then measures that changed resistance. Reducing the electric current or comparing several currents can reveal this physical effect.[ref]
The law of propagation of uncertainty makes the role of covariance explicit. For small perturbations of nonzero mean signals, the first-order measurement uncertainty of obeysThe contribution from a shared excitation fluctuation cancels through the covariance term. Treating the two signals as independent random variables would discard that cancellation. Other errors need their own inputs in the complete model.[ref]
For time-varying excitation, ratiometric cancellation with matched signal and reference filters requires matching dynamics as well as shared wiring. With constant resistances and identical linear time-invariant systems of impulse response , the filtered signals are and , so their ratio is still wherever defined. Here denotes convolution. Different filters or different sampling times can leave an excitation-dependent error.[ref]
RTD Ratiometric Measurements and Filtering Using the ADS1148 and ADS1248 Family of Devices by Luis Chioye. First-page abstract and low-pass filter design discussion, matching input and reference filters
A simple proposed bench test first replaces the sensor by a stable calibrated resistor, sweeps the electric current within the instrument limits, and records both electric potential differences and their ratio. That separates electrical cancellation from changing sensor temperature. Repeating with a resistance thermometer in a stable bath then tests Joule heating; changing the reference resistance or one filter tests the predicted failure modes. These are proposed checks of the model, rather than reported experimental results.
Reference material stability 2026-11-05
Reference material stability concerns whether the specified properties of a reference material remain suitable over a stated time and under stated storage, transport and use conditions. Stability is a requirement relative to an intended measurement use, rather than a promise that every property remains exactly constant.[ref]
International vocabulary of metrology by JCGM. VIM3entry5.13, definition
Reference material homogeneity asks a different question: whether different portions adequately represent the same material.A stability investigation must distinguish material change from a changing measuring system. Repeatedly measuring one bottle with an instrument that drifts does not identify which changed. Keep the property definition and temperature conditions consistent, use suitable comparison references, record handling and examine the uncertainty of any fitted trend. A stable comparison reference helps separate the two effects; it also has its own measurement uncertainty.
For an original optical example, suppose a trend model for refractive index at fixed wavelength and temperature iswhere is measured in months. A stipulated predicts a change after three months. If the intended use permits a change of at most , that model reaches the limit after months. This is a teaching calculation under an assumed trend, not an observed shelf life. A real decision also accounts for uncertainty in , changes between transport and storage conditions, and the reliability of the trend model.
The archived NIST SRM1922 certificate gives concrete handling conditions: keep the oil in its original bottle, tightly recapped, at approximately –, protected from excessive moisture and light. Its certification conditions exclude damaged, contaminated or modified material.[ref]
Certificate for Standard Reference Material 1922 by National Institute of Standards and Technology. 2009certificate, p1, Expiration of Certification; p3, Storage and Handling
Those conditions belong alongside the assigned value; an unchanged label does not establish reference material stability. Reference resistor 2026-11-05
A reference resistor is a resistor whose assigned electrical resistance supplies the comparison value in a resistance-measurement circuit. Its value may come from measurement calibration rather than its nominal marking. The useful reference is its value under the stated temperature, loading and connection conditions, together with the associated measurement uncertainty.
In a shared-excitation resistance ratio measurement, the sensor and reference resistor carry the same electric current. Measuring their electric potential differences gives , so . A reference value error therefore sets a scale error even when excitation cancels perfectly. Reference-resistor error in ratiometric resistance measurement derives the exact relation and distinguishes a correction from its remaining measurement uncertainty.[ref]
The connections define which resistance is compared. In four-terminal sensing, two leads carry the electric current while a separate high-impedance pair senses the electric potential difference at the reference terminals. This avoids including the current-lead drops in the assigned reference value, within the sensing instrument's limitations.
For an original ideal example, a reference resistor carries through two current leads, each of electrical resistance . The reference element has a potential difference of , while a reading outside both leads is . If this latter reading is used as the denominator while the calculation assumes , a true sensor is inferred asSensing directly at the reference terminals removes this idealized lead contribution. It does not correct a wrong reference value. The circuit guide explicitly shows a Kelvin connection at the reference terminals for this purpose.[ref]
A small temperature change shifts the value approximately by , with temperature coefficient of resistance . Excitation also dissipates electric power ; a reference with a different temperature or load from its measurement calibration may require a correction or additional measurement uncertainty. A proposed check compares the reference before and after the measurement sequence under the same connection and loading conditions. Repeating the comparison at a lower electric current tests for a heating contribution. Stability of the observed ratio alone cannot certify the reference's assigned value.
Reference-resistor error in ratiometric resistance measurement is the error left when an instrument converts a measured resistance ratio using an inaccurate value for its reference resistor. Shared excitation removes one nuisance input; it does not establish the denominator's true electrical resistance.
Let be the actual reference resistor value and the value used in calculation. For an ideal simultaneous reading , the inferred sensor resistance isWriting gives the exact fractional error . The familiar is its first-order approximation. If a deterministic bound is , the inferred fractional error lies inA symmetric reference tolerance therefore produces a slightly asymmetric resistance-error interval.
For an original example, suppose the instrument uses while the actual reference resistor is . A true sensor is reported as , an error near . Multiplying the shared excitation by any common positive number leaves this error unchanged in the ideal model. Using a measured and corrected instead of its nominal marking can remove the known part, while the corrected value retains its own measurement uncertainty.
Temperature variation can produce the remaining error after measurement calibration. Locally, , where is the temperature coefficient of resistance. For example, a -parts-per-million-per-kelvin coefficient over contributes roughly parts per million; it survives excitation cancellation. The primary circuit guide explicitly requires reference accuracy and low drift, and separates these requirements from shared-current cancellation.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. §2.1.3, printed pp. 10–11, reference accuracy and drift
For small random errors, applying the law of propagation of uncertainty to givesThe covariance term must follow how the reference value and reading were obtained; it cannot simply be assumed zero when inputs are shared.[ref]
Regulated calibration bath 2026-10-18
A regulated bath holds its temperature near a chosen setting through feedback control, while stirring reduces internal temperature gradients. Its finite heat capacity buffers disturbances; external heating and cooling replace the ideal heat bath's unlimited capacity. This is useful for thermometer calibration and for maintaining a blackbody calibration source at a known temperature.
Joel B. Fowler's A Third Generation Water Bath Based Blackbody Source reports a NIST test of a stirred water bath containing a copper blackbody cavity. A bottom plate supplied heating and cooling; a double stirrer circulated water. Electronic feedback control regulated heating. An electronic thermometer with a thermistor probe provided the readout. The preparation was to approach a selected setting, wait for stabilization, adjust against the measured temperature and then log continuously.[ref]
A Third Generation Water Bath Based Blackbody Source by Joel B. Fowler. A Third Generation Water Bath Based Blackbody Source, §§2–3, printed pp. 592–593: bath construction, controller and setpoint procedure
To test spatial uniformity, one thermistor stayed near the cavity tip while another visited ten locations at three heights. The simultaneous reference distinguished spatial differences from temporal drift. An insulated dome reduced thermal conduction along probe casings. Table 1 reports nonzero departures at several settings: circulation did not make every location identical. The measurement uncertainty analysis separately included uniformity, stability, setting, immersion and readout contributions. These tests supported a bounded temperature assignment around the cavity, rather than infinite heat capacity or perfect thermal equilibrium throughout the apparatus.[ref]
A Third Generation Water Bath Based Blackbody Source by Joel B. Fowler. A Third Generation Water Bath Based Blackbody Source, §4.1 and Tables 1–2, printed pp. 594–595: fixed and movable probes, immersion control, measured nonuniformity and uncertainty components
René Dybkær 2026-11-05
René Dybkær worked on measurement concepts and terminology in laboratory medicine. The IUPAC traceability report identifies him with the Department of Standardization in Laboratory Medicine at Frederiksberg Hospital, Denmark, and names him as a coauthor with Paul De Bièvre, Aleš Fajgelj and David Brynn Hibbert.[ref]
Metrological traceability of measurement results in chemistry: Concepts and implementation by Paul De Bièvre and others. Publisher byline, affiliation2 and Abstract.
The report develops a shared account of metrological traceability for chemical measurement. Such terminology matters because a result must identify the measurand and reference relationship before numerical comparisons become meaningful. A chemical name, number and unit together specify more than an instrument indication alone.
For an original conversion example, a stipulated compound with molar mass at has mass concentration . The numbers2 and200 describe equivalent content under that specification, rather than disagreement between laboratories. Conversely, equal numbers attached to different quantity definitions need not be equivalent. Uncertainty in the inputs also belongs in the conversion's measurement uncertainty.
René Dybkær's coauthored standards work addresses this need to make the meaning of a measurement communicable. The example is an original illustration of the problem, not data attributed to his experiments or a statement of a clinical decision threshold.
Resistance pressure gauge 2026-10-18
A resistance pressure gauge is a pressure sensor whose electrical resistance is calibrated against pressure. Even a simple local response needs both pressure and temperature. For an original illustrative model definewhere is the reference electrical resistance at , , and zero history offset. The derivatives and are local sensitivities with units and , respectively; is a dimensionless history offset. Constant sensitivities are an approximation over a chosen range, not material constants valid at every pressure.
Choose synthetic coefficients and , with and . A measured ratio then gives . Ignoring the temperature change gives , a error. This model is deliberately stipulated; these numbers are not measured manganin coefficients or Bridgman's data.
Pressure calibration: a synthetic electrical resistance response with temperature and hysteresis offsets
. Original calculated example. The blue reference line has , ; the orange line has , . Dashed lines illustrate two histories with , not observed data. A ratio 1.022 maps to 1.00 GPa after correcting temperature, or 1.10 GPa if read against the blue line. The lower chain distinguishes the specimen, the internal sensor, the electrical readout and the calibrated pressure. Details.If two histories differ by , their outputs at the same pressure differ by ; interpreting both with a single history-free curve gives an apparent pressure separation . The half separation is . Reversing the drive can additionally cause mechanical friction hysteresis, independently of this sensor model. Record both the electrical calibration history and the loading direction.
For independent small standard measurement uncertainties and exact stipulated , first-order propagation givesFor , , and , this is . Real calibration also includes coefficient measurement uncertainties, correlations, reference-point measurement uncertainty, and any unresolved nonhydrostatic loading. A sensor calibrates its local environment; it cannot by itself certify the absence of deviatoric stress throughout the sample.
Resistor tolerance Created 2026-10-14 Updated 2026-10-25
Resistor tolerance specifies the permissible manufacturing deviation of a resistor's electrical resistance from its nominal value under stated conditions. For nominal electrical resistance and fractional resistor tolerance , the specified range is . This differs from measurement uncertainty: resistor tolerance describes a component specification, while measurement uncertainty describes knowledge of a measured value; the tolerance alone is not a probability distribution for that value.
Two nominally equal resistors can have different actual electrical resistances, changing their voltage divider ratio even without an external electrical load. In a loaded voltage divider, use the actual electrical resistances to calculate both the unloaded output and the Thévenin resistance governing voltage divider loading error. A mismatch from resistor tolerance and a change from voltmeter loading are distinct effects that can occur together.
RNA melting baseline 2026-11-05
An RNA melting baseline is the predicted absorbance of a specified folded or unfolded RNA population as a function of temperature, before mixing those populations in an RNA optical melting model. It differs from a solvent blank: a folded RNA sample still absorbs light, and that response can change with temperature even without the modeled transition.
For a two-state RNA melting model with unfolded fraction , writeThe folded and unfolded RNA melting baselines might be modeled as and . Their continuation into the transition region is RNA baseline extrapolation. Solving the optical mixture givesFor original stipulated readings , and , the inferred fraction is 0.50. Changing only the folded RNA melting baseline to 0.45 gives 0.40; changing only the unfolded RNA melting baseline to 0.80 gives 0.375. The same measured absorbance can therefore imply different populations under different baseline models.
Sensitivity is explicit. With ,A small baseline separation amplifies measurement uncertainty. Correlated fitted baselines require covariance terms in uncertainty propagation, rather than independent error bars by assumption.[ref]
Evaluation of measurement data — Guide to the expression of uncertainty in measurement by JCGM. §§5.1–5.2, sensitivities and correlated inputs.
Optical-melting studies infer thermodynamics by fitting measured curves, rather than directly counting unfolded molecules.[ref]
Thermodynamic characterization of naturally occurring RNA tetraloops by Justin P. Sheehy and others. Materials and Methods, individual-curve fitting.
Inspect the pre-transition and post-transition regions and test the chosen RNA baseline extrapolation. An inferred fraction outside signals noise or incompatibility with the assumed optical model; it is not a negative molecular population. Rodney W. Leonhardt 2026-11-05
Rodney W. Leonhardt is the author of a2004 National Institute of Standards and Technology report on measurement calibration of low-level pulsed-laser radiometers at a wavelength of1.06micrometres. It treats laser pulse energy and peak power as separate measurement quantities and describes the apparatus and measurement uncertainty needed to transfer a laboratory reference to a test instrument.[ref]
Calibration Service for Low-level Pulsed-Laser Radiometers at 1.06 µm: Pulse Energy and Peak Power by Rodney W. Leonhardt. NIST SP250-64, title/byline and §§3–5
The pulse-energy procedure uses a modulated laser, a calibrated beam splitter arrangement, a relatively slow thermal reference and a fast test radiometer. The reference accumulates energy over many pulses. Pulse repetition rate and the illumination interval determine the pulse count; the splitter ratio transfers the reference energy to the lower-energy test beam. The thermal detector need not resolve individual pulses.[ref]
For an original idealized transfer calculation, let the reference receive a total over equal pulses. Define as the reference-to-test beam energy ratio. The average energy per test pulse isIf the calibrated radiometer's selected output measure is0.20volts, the corresponding energy-per-output coefficient would be50picojoules per volt. These invented values explain the transfer arithmetic and do not reproduce the report's calibration results.
The output measure must be specified together with its coefficient. Leonhardt's report discusses offsets in the integrating-amplifier waveform and use of peak-to-peak electric potential difference as a readout. Its peak-power procedure separately compares continuous and pulsed conditions.[ref]
Calibration Service for Low-level Pulsed-Laser Radiometers at 1.06 µm: Pulse Energy and Peak Power by Rodney W. Leonhardt. §4, waveform/offset discussion; §5.1, peak-power comparison
The work illustrates how a measurement becomes reproducible by recording the optical transfer and electrical readout conventions, rather than only stating a detector sensitivity. Ruby calibration against sodium chloride 2026-10-18
The 1975 experiment mixed powdered sodium chloride and 0.5%-chromium ruby in a 4:1 volume ratio. An Inconel gasket, initially thick with a hole, held the powder and 4:1 methanol–ethanol. At , X-ray diffraction rings on film away gave salt compression; a reference equation of state assigned pressure. After decreasing load, they waited at least five hours and bracketed diffraction exposures with readings to check drift.[ref]
Calibration of the pressure dependence of the R1 ruby fluorescence line to 195 kbar by Gasper J. Piermarini and others. pp. 2774–2775, Experimental Procedures
The recommended linear slope was , with a fit confidence interval of , to . Reference-scale systematic measurement uncertainty remained separate. Above about the medium was only quasihydrostatic; separate range fits checked consistency.[ref]
Calibration of the pressure dependence of the R1 ruby fluorescence line to 195 kbar by Gasper J. Piermarini and others. pp. 2775–2776, Table I and recommended calibration; p. 2778, pressure-distribution discussion
The logical chain is compression to reference pressure, paired with optical shift, then a fitted response. X-ray diffraction supplies the reference comparison during calibration; subsequent ruby readings transfer that scale to other samples. The reference equation of state is therefore part of the meaning of the reported pressure. Changing it can move all calibrated readings together, even if the optical measurements repeat perfectly. A regression residual tests agreement with the chosen reference within the sampled conditions; it cannot independently certify that reference.
Ruby pressure uncertainty worked example 2026-10-18
Construct a synthetic local calibration to show the calculation independently of the historical data. Choose reference pairs , with shifts in and pressures in . The reference zero is already subtracted. Treat the shifts as exact for this fit and give the three reference pressures equal weights. Minimizing the ordinary least squares objectiveand setting its derivative to zero givesThe regression residuals are . Their weighted moment vanishes, as required by the least-squares normal equations. An unknown reference offset requires fitting an intercept instead; forcing it to zero can bias the slope. If shift errors matter, this elementary fit is insufficient.
Synthetic ruby pressure scale: fitting the response and propagating measurement uncertainty
. Original stipulated reference points and calculations, not measured historical calibration. The left panel shows the fitted slope and a chosen reading. The right panel separates independent contributions to standard uncertainty; bar lengths are standard contributions, which combine by sums of squares. The assigned slope standard uncertainty includes a hypothetical reference-scale contribution and is not estimated from the three residuals. Details.Use the local measurement modelHere denotes pressure relative to the reference, is calibrated sensitivity, and is a local wavelength sensitivity to temperature at fixed pressure. Choose , and a reading at . The inferred pressure is . These coefficients and readings are stipulated, not measured ruby constants. With the same pressure, a increase in temperature would add to the shift in this model; neglecting it would overestimate pressure by .
Assign independent standard uncertainties , and for the temperature difference. Here is an independently assigned calibration uncertainty, not an estimate from the regression residuals above; treat as exact in this example. First-order propagation of uncertainty follows by differentiating the measurement model:The four standard contributions are , giving . This is a standard uncertainty, not automatically a interval. Correlated reference and loaded wavelength readings require the corresponding covariance term; common wavelength-axis drift may cancel in their difference. Uncertain , nonzero , model inadequacy and unresolved deviatoric stress require a broader budget. Taking more optical readings reduces some random errors while leaving a common reference-scale error.
Sagitta momentum uncertainty Created 2026-10-24 Updated 2026-11-05
A particle track projected onto the plane perpendicular to a uniform magnetic field is a circular arc with radius . Here is the momentum component perpendicular to the magnetic field; it is the collider transverse momentum when the field follows the beam axis. For chord length and sagitta , circle geometry givesFor a short arc with , this yieldsThe measurement uncertainty follows by differentiating the inverse relation, treating and as known. A fixed spatial measurement uncertainty therefore makes fractional momentum uncertainty grow with momentum, whereas increasing the lever arm improves it quadratically in this idealization. Multiple Coulomb scattering, field uncertainties and alignment add other contributions; a helical particle track also requires its pitch to recover the full momentum.[ref]
Review of Particle Physics by Particle Data Group (2022). Particle detectors at accelerators, section 35.12.1, equation 35.49: sagitta and magnetic momentum analysis
Scalar rescaling invariance of a tie-line weight matrix means that multiplying every fit weight by the same positive number leaves the fitted phase fraction unchanged. Let endpoint compositions and overall composition be expressed in independent mole fraction coordinates. With , a fixed symmetric positive-definite matrix defines the tie-line phase-fraction fit by minimizingFor , . Multiplication by a positive constant preserves every comparison between objective values, so the sets of minimizers agree. The proof also applies when the same constraint is imposed in both fits.
For distinct endpoints, the unconstrained formula makes the cancellation visible:The differential in sensitivity of a tie-line fit to its weight matrix agrees: for a perturbation , its numerator is by the fit's normal equation. Thus the scalar direction is insensitive even when the residual is nonzero.
In an original example, take , , and . The remaining composition coordinate is fixed by closure. These values giveReplacing by gives the same fraction, although the minimum objective increases from0.00072to0.00504. Changing only the second diagonal weight from4to7 instead changes the fitted fraction to0.640625. Relative weights determine the geometry; their common multiplier does not.
Point-estimate invariance does not by itself establish invariance of physical measurement uncertainty. Merely rewriting the same objective with leaves propagation through the same measured-input covariance matrix unchanged. If instead represents absolute precision and asserts a genuinely different covariance , that is a different noise model. With exact endpoints, the linear-model variance then decreases by a factor . The measurement model and covariance interpretation must be stated.[ref]
Guide to the expression of uncertainty in measurement by JCGM. Sections 5.1.3 and 5.2.2, model sensitivities and covariance propagation; tie-line result derived here
Scientific instrument Created 2026-10-13 Updated 2026-11-05
A scientific instrument converts a physical interaction into an observable indication under a stated measurement model. An ionization chamber converts deposited energy into collected electric charge; a telescope collects incoming electromagnetic radiation; a force probe converts a mechanical load into a calibrated response.
An indication is not yet the desired measurand. The instrument's response, measurement resolution, measurement calibration and measurement uncertainty determine what can be inferred. Changing a controlled input and comparing with a reference is often more revealing than reading the same display repeatedly: it tests whether the instrument is responding to the intended quantity or to an unwanted influence.
Scientific reproducibility Created 2026-10-20 Updated 2026-11-05
Scientific reproducibility concerns whether the basis of a scientific result can be reconstructed and checked. It requires enough information about the measurement, data handling and reasoning to distinguish repeating a procedure from merely obtaining a similar-looking answer.
Terminology varies between fields. The 2019 National Academies report distinguishes computational reproducibility, using the same data and computational procedure, from scientific replication, which collects new data to address the same scientific question. These test different links in the argument. Repeating a calculation can expose a software or transcription error; new observations can test whether the finding persists beyond the original sample.[ref]
Reproducibility and Replicability in Science by National Academies of Sciences, Engineering, and Medicine. Summary, pp. 6–9: Definitions, Reproducibility and the distinction from new-data replication
Consider an original toy measurement of three lengths, , and units. Recomputing their arithmetic mean as checks the recorded calculation. Measuring three new objects tests a different question: whether the estimated mean describes further observations, allowing for measurement uncertainty and sampling variation. Neither exercise alone establishes that the instrument was correctly calibrated.
Scott M. Ransom Created 2026-10-20 Updated 2026-11-05
Scott M. Ransom works in radio astronomy and explains its measurement physics with James J. Condon in their textbook Essential Radio Astronomy by James J. Condon and Scott M. Ransom. Their treatment connects received radio waves to measurable antenna signals and the physical properties of astronomical sources.
For example, the radiometer analysis relates measurement uncertainty to receiver noise, observing bandwidth and integration time. Increasing either the independent frequency range or observing duration improves sensitivity through statistical averaging. This supplies the physical reasoning needed to design observations, rather than treating telescope sensitivity as a single unexplained specification.[ref]
Essential Radio Astronomy by James J. Condon and Scott M. Ransom. Radiometers chapter: radiometer equation, system temperature and continuum sensitivity
Portrait of Scott M. Ransom
. External portrait hosted by National Radio Astronomy Observatory, astronomer homepage. Source. Second Created 2026-10-18 Updated 2026-11-05
The second, symbol , is the SI base unit of time. Its definition fixes the unperturbed frequency of the ground-state atomic transition between levels split by hyperfine structure in caesium with mass number to exactly . One second therefore corresponds to that many cycles of the specified electromagnetic radiation.[ref]
The International System of Units (SI), ninth edition, SI Brochure. §2.3.1: definition of the second through the caesium transition frequency
A practical atomic clock realizes the reference by interrogating atoms and controlling a reference oscillation. Magnetic fields, motion and environmental effects require corrections: an exact unit definition does not make every physical realization free of measurement uncertainty.
Sensitivity mass 2026-11-05
A sensitivity mass is a small calibrated mass added to a loaded weighing instrument to determine its response to a known mass increment. NIST's double-substitution procedure calls this a sensitivity weight and uses it to establish sensitivity under the conditions of the comparison.[ref]
Weighing by Double Substitution by National Institute of Standards and Technology. SOP4–2019, p3, Summary and apparatus; pp6–7, selection and use of sensitivity weight
It supplies a local input step rather than a new determination of the entire load from zero.Let an original comparator give indications before and after adding mass . For a stable linear response over that increment,where is another indication difference near that working load. A constant additive zero cancels from each difference; instrumental drift or load-dependent response need not cancel.
For original indications , and assigned sensitivity mass , the local response is about indication increments per milligram. A difference increments gives . Using only a nominal label instead would give . The assigned mass and its measurement uncertainty, rather than the nominal label alone, enter the comparison.
The added step must be large enough to resolve reliably but remain within the response region relevant to the unknown/reference comparison. NIST specifies selection criteria and regularly spaced observations; it also requires the standards to have valid calibration information.[ref]
Weighing by Double Substitution by National Institute of Standards and Technology. p1, prerequisites; p3, Summary; pp6–8, sensitivity selection and observations
Keep handling, settling and environmental conditions consistent. A sensitivity mass does not by itself cancel buoyancy, temperature effects or uncertainty in the main measurement standard. Those contributions remain in the full double substitution weighing model.