Peter Rostron 2026-11-05
Peter Rostron is a co-editor, with Michael H. Ramsey and Stephen L. R. Ellison, of the 2019 guide to measurement uncertainty arising from analytical sampling. Its editorial page identifies him with the University of Sussex. The guide connects laboratory analysis with the earlier act of collecting material.[ref]
Measurement uncertainty arising from sampling by Michael H. Ramsey and others. Editorial page, PDF p. 3, and Summary
An illuminating experimental design question is where to place replication. Repeating an analysis of one collected sample reveals a different source of variation from collecting another sample from the target material. A nested design can distinguish the corresponding variance components. The guide also describes an unbalanced sampling-uncertainty design with two analyses of one field sample and one analysis of its field duplicate.[ref]
An original counting example makes the design concrete. At each of ten targets, two collected samples each analysed twice require analytical results. Analysing the first sample twice and its field duplicate once instead requires . The design retains a within-sample analytical contrast at each target and a contrast between separately collected samples, with fewer analytical results. The information content and the appropriate analysis of variance must nevertheless follow the actual replication pattern; thirty results do not automatically have the same precision as forty.
This distinction matters when deciding whether to spend effort on another instrument reading or on another collection. Rostron's verified contribution here is editorial participation in the shared guidance, rather than sole authorship of every method it discusses. Repeated collection can reveal variability without, by itself, proving the absence of a common collection bias.
Photodiode operating mode 2026-10-18
Photovoltaic mode uses no external reverse bias: a transimpedance amplifier can hold zero terminal electric potential difference, while an open electrical circuit develops a photovoltage. Photoconductive mode biases the diode cathode positive relative to the diode anode, generally reducing junction capacitance and increasing speed and dark current.[ref]
A simple calculation separates the readout from the conversion. If a linear transimpedance amplifier has feedback electrical resistance , the model above produces an output change of magnitude . The output sign depends on the connection. Increasing increases that electric potential difference, but does not create more collected charge carriers.
For a reverse-biased photodiode feeding an electrical load , Ohm's law gives a readout magnitude while the response remains linear. With negligible dark current and other series drops, the remaining reverse bias is , where is the supply magnitude. Thus the chosen electrical load can consume the available bias: a larger readout does not automatically reduce measurement uncertainty. For the same constructed example, and leave about ; this algebra alone does not establish a real device's linear range.
Photodiode responsivity measurement 2026-10-18
Thorlabs' silicon FDS100 experiment sends a collimated, fibre-coupled laser beam through a beam splitter to the photodiode and a reference optical power meter. Varying reverse bias and electrical load tests linearity. The measured readout approaches saturation when its electric potential difference consumes the bias.[ref]
Photodiodes by Thorlabs. Photodiodes, Lab Facts, Photodiode Saturation Limit and Noise Floor, opening experiment description and Figs. 1–3: reference power sensor, bias/load comparisons and saturation
A separate low-signal setup uses a laser, polarizers, an optical chopper, optical attenuators and a lock-in amplifier inside a light-tight enclosure. These are actual apparatus choices, not the parameters of the constructed example above.[ref]Photodiode Saturation and Noise Floor by Thorlabs. Photodiode Saturation and Noise Floor, p. 7 of 12, Experimental Design — Noise Floor: 637 nm source, polarizers, chopper, attenuation filters, lock-in readout and light-tight enclosure
For an independent dark current check, Thorlabs enclosed a photodiode in nested metal boxes to exclude light and suppress interference, used a Keithley 6487 ammeter supplying reverse bias, and monitored temperature. Only one specimen per type was tested; the nearby temperature sensor did not directly contact the semiconductor. Such a dark control measures the baseline but cannot certify spectral responsivity.[ref]
Photodiodes by Thorlabs. Photodiodes, Lab Facts, Dark Current as a Function of Temperature, Experimental Setup and Experimental Limitations: nested enclosure, ammeter and specimen/temperature limitations
To infer spectral responsivity at a selected wavelength, measure the radiant flux actually reaching the test photodiode, rather than just the reference arm. A reference-arm measurement calibration must include the beam splitter ratio and path losses. With signed readouts in one fixed orientation, subtract a beam-blocked baseline and usefor direct electric current readout or a calibrated linear transimpedance amplifier, respectively. Repeating at several incident powers tests whether this ratio is constant. Blocking only the source estimates ambient light plus electronic baseline; covering the detector excludes ambient light too. Their difference distinguishes optical background from dark current and readout offset. For a broad optical spectrum, the ratio is a spectrum-weighted response, not generally at one wavelength.
For independent Poisson distribution collection, the root mean square shot noise in a one-sided rectangular frequency band of signal bandwidth isHere is reverse dark current; thermal electrical noise and amplifier noise also contribute.[ref]
Si photodiodes by Hamamatsu Photonics K.K.. Si photodiodes, p. 6, §2-4, equations (2-4)–(2-7): independent thermal, dark-current and photocurrent noise terms
Subtracting a mean dark current removes its offset, not its fluctuations: subtracting a constant leaves the variance unchanged. Consequently makes dark current shot noise small relative to signal shot noise; it does not by itself establish high signal-to-noise ratio. With an independently measured baseline, the variance of that estimate must also be propagated. Increasing the measurement signal bandwidth raises the admitted noise. Calibration of wavelength, reference radiant flux, readout gain and collection geometry, together with drift and linearity, determines what measurement uncertainty can be claimed.
Physical section thickness 2026-11-05
Physical section thickness is the distance through a recovered physical section, measured normal to its faces. For a flat section with parallel faces, write this distance as ; a nonuniform section instead requires a local thickness or a stated averaging convention. It is distinct from the nominal section feed commanded to the microtome, the spacing of sampled planes in an image stack, and the spatial resolution of the resulting image.
In ultramicrotomy, a controlled advance presents material to a diamond knife on each cutting stroke. JEOL describes preset feed increments for thin sections and notes that actual thickness depends on the material and temperature.[ref]
Ultramicrotomy by JEOL. Glossary, preset specimen feed and following paragraph on actual section thickness
Section compression and other deformation can further complicate the relation between the machine setting and the recovered section. A feed setting alone is not an independent thickness measurement.An original geometric example shows why finite thickness matters. Consider a straight thin feature whose direction makes angle with the section's normal vector. Choose coordinates so the section occupies and the feature lies alongAn image combining contributions through the section can therefore contain feature positions spanning the lateral distanceFor stipulated and , this range is . It is a geometric range, not a universal formula for the measured image width: image contrast, the imaging response and the feature's own width also matter. Replacing the slab by one mathematical plane discards this depth variation.
For an ideal undeformed series of contiguous sections, imaging every fifth section gives a centre-to-centre sampling interval of , while each imaged section remains thick. Record omitted sections and the image stack axis calibration separately. To test a feed setting experimentally, measure recovered thickness with an independent method and stated measurement uncertainty, rather than using the same feed display as both command and verification.
Physik Instrumente 2026-11-05
Physik Instrumente makes precision motion equipment used in scientific apparatus. Its archived M-222–M-226 motorized micrometer family illustrates a concrete measurement issue: motor-shaft encoder counts and screw geometry give a design resolution, while actual motion also depends on mechanics and loading.[ref]
High Precision Actuators with Closed-Loop Feedback Motor by Physik Instrumente. M-221–M-226 description: micrometer screw, gearhead and motor-shaft encoder
For a worked reconstruction from the manufacturer's M-222.20 specifications, use screw pitch per revolution, gear reduction and 60 encoder counts per motor revolution. The nominal linear increment isThis agrees with the listed design resolution of . The same table separately lists minimum incremental motion of and backlash of for that model.[ref]
High Precision Actuators with Closed-Loop Feedback Motor by Physik Instrumente. Technical-data table, M-222.20 column; design resolution, minimum incremental motion, backlash, encoder resolution, drive screw pitch and gear ratio
Thus a command corresponding to one encoder count is not evidence for an independently resolved 59 nm movement of the load. The listed backlash corresponds to roughly 84 nominal counts, illustrating why reversing direction is a different test from continuing in one direction. These are calculations from a legacy specification, not new performance measurements. In an assembled instrument, compare commanded and independently measured displacement, including reversals, to establish the usable response and its measurement uncertainty.
Pion decay at rest in emulsion 2026-10-18
Cecil Frank Powell's 1950 lecture illustrates the successive decays of pions and muons through their tracks and explains why the nearly fixed secondary muon range, allowing for range straggling, supports a single neutral decay partner. He describes its identification as a neutrino tentatively. The neutral particle itself leaves no recorded track. He contrasts this decay with the nuclear reactions caused by stopped negative pions, connecting the latter with the nuclear force.[ref]
Here is a modern special relativity derivation for the dominant leptonic pion decay, assuming a free pion at rest, a massless muon neutrino, and no emitted photon. Write and for the pion and muon masses. By conservation of momentum the daughter momenta are opposite with equal magnitude . By conservation of energy and the energy–momentum relation,Substitute , square, and cancel . This givesThe resulting is fixed, although the muon's direction is not set by the incoming stopping track. Thus the kink angle relative to that earlier track is not the angle between the two daughters: in the pion rest frame, the daughters are back to back. A neutral-partner arrow in a diagram represents inferred momentum, not an observed string of grains.
For an original calculation, adopt rounded input values and . They give , and . Their sum satisfies . These are rounded calculation inputs and outputs, not a fit to the 1947 plates.
At fixed material and particle identity, a fixed initial kinetic energy suggests a narrow particle range distribution broadened by range straggling and measurement uncertainty. The synthetic example convolves a single chosen range with a normal distribution; a second example superposes two such chosen ranges. It explains why repeated comparable stopping distances can be informative, without pretending to extract a historical spectrum.
The decay equations do not identify the neutral partner as a neutrino on their own: a massless photon would satisfy the same energy and momentum equations. Nor does any isolated kink establish the pion's identity. The inference gains force from calibrated particle range, track grain density, repeated decay patterns and independent particle properties. Historical particle names should not obscure the distinction: a muon is a lepton, whereas a pion is a meson.
Positioning repeatability 2026-11-05
Positioning repeatability describes the scatter of positions reached when a motion system repeatedly returns to the same commanded position under specified conditions. It is an application of measurement repeatability, a form of measurement precision. A system may return to almost the same wrong position each time, so small scatter does not establish measurement accuracy.[ref]
International vocabulary of metrology by JCGM. VIM3 entries2.15,2.20 and2.21: precision and repeatability conditions
The approach direction belongs in the test description. Unidirectional positioning repeatability keeps it fixed; bidirectional positioning repeatability includes returns from both directions. To measure the assembled stage, independently observe its final position rather than only the motor's rotary encoder count. Physik Instrumente describes a static test that lets the platform stop before reading it, using an interferometer for displacement and an autocollimator for angular error, with two reflecting mirrors on the stage.[ref]
How to Consistently Verify the Performance of Nanopositioning Motion Control Equipment by Physik Instrumente. §§3.1–3.2, static readout, direction convention and paired-mirror apparatus
For an original numerical example, repeatedly command . Suppose an independent position readout with known measurement calibration gives the errors , in micrometres,Each direction has the same sample standard deviation,Yet their arithmetic means are 0 and . The magnitude of their mean difference, a positioning reversal error at this target, is 2.0 micrometres, far larger than the within-direction scatter. Pooling all eight values gives a sample standard deviation of ; that statistic mixes the directional shift with the smaller return-to-return variation.
Two tightly repeated return positions separated by approach direction
. Original chosen errors at one target. Four trials per direction; small bars show mean plus or minus one sample standard deviation. The figure is not measured product performance. Details.The calculation uses a declared sample standard deviation, not an unspecified manufacturer's tolerance or a universal repeatability convention. A positioning specification must state its approach directions, target positions, settling/readout procedure, load, environmental conditions, statistic and measurement uncertainty. A single target does not establish performance over the travel range. The manufacturer's method explicitly separates directional means from scatter and combines information from multiple targets when forming axis-level specifications.[ref]
Backlash is one possible cause of direction-dependent behavior, while thermal expansion and instrumental drift can move the apparent target over time. A repeatability experiment must distinguish those effects from random scatter. Measuring small returns to one target also answers a different question from minimum incremental motion, which asks whether a small commanded change can actually be executed.
Precision displacement measurement with a Michelson interferometer Created 2026-10-18 Updated 2026-10-20
A Michelson interferometer turns displacement into a count of interference fringes: a moving mirror changes both the outward and return paths. This lets a wavelength act as a ruler whose divisions need no engraving. Assume normal incidence, a fixed reference arm, stable refractive index along the moving arm, and fixed vacuum wavelength . Write for the wavelength in that medium and take positive away from the beam splitter. The change in optical path difference, the accumulated phase difference, and the signed number of complete interference fringe cycles areThe factor two comes from propagation geometry: moving the mirror by adds before reflection and afterward. One bright-to-bright cycle means mirror travel ; bright-to-dark means , half a cycle. All lengths must use compatible units, and is dimensionless. A constant reflection phase difference changes the initial fringe position but cancels from this change formula.
Mirror displacement, doubled optical path difference, and counted interference fringes for a hypothetical Michelson interferometer
. Details. For an original hypothetical example, suppose in the actual surrounding medium and a continuous scan gives cycles, including the final fractional cycle. Then . With the wavelength treated as exact, a fringe-reading measurement uncertainty of cycle contributes to the displacement measurement uncertainty. An uncertainty of in a local phase difference instead corresponds to . This illustrates measurement resolution, not a complete claim of accuracy: wavelength calibration, air changes, mirror tilt and missed cycles can dominate.
To separate the oscillating readout from its optical coherence envelope, use the local modelHere is detected light intensity, is the mean level, is a fixed bias wave phase, and is interferometric visibility. A fine scan of measures the carrier phase difference; a coarse scan can map the temporal optical coherence envelope through . The envelope maximum locates matched delays only under the source and apparatus assumptions; it is not a bright fringe identifying an arbitrary mirror position. Even with known , one intensity reading leaves the cosine sign and whole-cycle ambiguities. Continuous motion of known direction, or an independently obtained pair of signals in phase quadrature, is needed to recover signed travel through reversals.
Proposed Michelson interferometer readout and hypothetical optical coherence and refractive index controls
. Details. The figure's original control example stipulates equal arm light intensities, matching linear polarizations, and a Lorentzian optical spectrum giving with . In vacuum, arm mismatches and give and . Matching the arms improves the phase difference readout without changing its travel per cycle. A fine mirror scan of one cycle changes by only for the stipulated source, much less than . These are calculated values, not historical data.
A proposed measurement sequence follows from this model. Overlap the returning beams, match their linear polarizations, find a high-contrast delay and use a small calibrated mirror scan to determine the sign and fringe period. Read a fixed spatial sample of the interference fringes with a photodiode; record each arm separately and the dark signal, avoid detector saturation, and retain enough time resolution to count every cycle. Approach mechanical settings consistently and repeat the return scan to expose lost motion or drift. For very small changes, bias near phase quadrature, where is largest; an intensity extremum has zero first-order sensitivity. A known drive voltage alone does not provide measurement calibration for mirror travel: compare the actual travel with the fringe count using an independently known wavelength.
Control temperature, pressure, vibration and air currents because the instrument measures optical path difference, not mirror motion alone. For a fixed reference arm and fixed , the first-order change of the moving arm's round-trip path is . Thus an unrecognized mimics . A hypothetical , , and give apparent travel. Holding the mirror still while varying or monitoring the air conditions tests this confounder. Separate checks of alignment, wavelength stability and thermal expansion are necessary before sub-fringe measurement resolution becomes a defensible measurement uncertainty.
Preset tare value 2026-11-05
A preset tare value is a numerical tare mass entered or recalled for use in subsequent weighings without individually determining each container's tare. The pinned OIML terminology denotes it by and calls the measured gross or net indication minus this value a calculated net value. It can be supplied through a preset tare device rather than obtained by weighing the current empty container.[ref]
For an original example, a stored container tare is , but the actual empty container has tare mass . Adding of material gives gross mass . Applying the stored value givesalthough the actual net mass is . Their difference isAn underestimated preset tare therefore overestimates the contents. Repeating the same discrepancy across ten containers produces a total discrepancy; more displayed digits do not remove it.
A useful physical check independently weighs representative empty containers, including the same lids and attachments used during the loaded measurement, and compares them with the stored value. Record which load components the preset represents. If it is subtracted from an indication that has already had those components removed by a tare device, the same tare mass is deducted twice. The stored number's provenance and variability belong in the measurement model and its measurement uncertainty.
Pressure gauge temperature error 2026-11-05
Pressure gauge temperature error is the change in a gauge's indication error associated with a change from its specified temperature conditions, while the pressure comparison is otherwise controlled. It belongs to the complete Bourdon tube pressure gauge response, not merely the number of scale divisions. Distinguish an instrument-temperature effect from a real change in the pressure being measured.
A concrete manufacturer specification is WIKA's June2025 US data sheet, which gives an additional error bound of of span per departure from the reference.[ref] This is a version-specific bound; it is not a measured signed correction for an individual gauge.
For an original gauge with span at , the specified form givesAt a pressure reading of , that bound is of the reading. The percentage is referenced to span, so the same absolute bound represents a larger relative contribution near the bottom of the range. Without the sign and an appropriate measurement calibration, subtracting would not be justified.
For an original controlled comparison, hold reference pressure at and allow the gauge assembly to equilibrate first at and then at . Suppose indications are and . Their errors relative to the reference are and , so the observed temperature-associated change is , or a local average . These are illustrative readings, not manufacturer test results. Repeat at other pressures and on return to the reference temperature to check reversibility and hysteresis.
Record gauge and reference conditions, equilibration and uncertainty in the pressure comparison. Heating a sealed gas can change its actual pressure through the ideal gas law; an uncontrolled change of that kind cannot be assigned entirely to instrument error. The bound or fitted correction remains one contribution to the final measurement uncertainty.
Pressure measurement 2026-10-18
A pressure measurement determines pressure relative to a specified reference using an instrument such as a manometer or pressure transducer. Gauge pressure, absolute pressure and differential pressure require different references; measurement calibration and measurement uncertainty determine how the reading can be interpreted.
Primary measurement standard 2026-11-05
A primary measurement standard is a measurement standard established by a primary reference measurement procedure, or an artifact selected by convention. A primary reference measurement procedure obtains its result without calibration against a standard of the same kind of quantity; it may still require measurements of other kinds of quantity.[ref]
For an original ideal pressure realization, independently established force acts on an effective piston area . The measurement model givesIf the inputs have independent standard uncertainties and , first-order propagation of uncertainty givesso . This is an illustrative model, not a complete piston-gauge uncertainty budget: a physical realization must also account for effects such as piston deformation and friction.[ref]
The essential distinction is how the pressure value is established. Merely comparing a gauge with another pressure gauge does not make it a primary measurement standard. Primary status also does not mean zero measurement uncertainty.
Primary reference measurement procedure 2026-11-05
A primary reference measurement procedure determines a result without calibration against a measurement standard for a quantity of the same kind. It can still depend on calibrated measurements of other quantities. The VIM example determines delivered liquid volume from a mass difference and liquid mass density, rather than comparing it with another calibrated delivered volume.[ref]
For an original ideal calculation, let the receiving vessel's tare mass be and its gross mass after delivery be . Suppose the liquid's mass density at the measurement conditions is . ThenWith independent standard uncertainties and ,so . These are stipulated values, not a recommended liquid or an actual calibration result.[ref]
The physical procedure would weigh the receiver before and after delivery, establish the liquid temperature and density, and include significant corrections in its measurement model. Evaporation, buoyancy and correlated mass readings can invalidate the simplified expression above. Primary describes the reference relation, not freedom from measurement uncertainty or the absence of any instrument calibration. Its primary status is therefore a claim about the complete procedure, not about this equation alone.
Prism minimum-deviation index measurement 2026-10-18
A prism minimum-deviation index measurement infers a transparent sample's relative refractive index from two angles: its known apex angle and the smallest deviation of a selected spectral line. A spectrometer combines a source and slit, a collimator, a rotating optical prism table, and a viewing optical telescope with an angular readout. The collimator supplies a nearly parallel beam; the optical telescope locates its emergent direction. This turns an otherwise difficult measurement of the phase velocity of light into a measurement of angles.
For the symmetric path through a uniform optical prism, with the same external medium at both faces and positive refractive indices,The angles are measured from face normals. This is the minimum deviation relation, not a formula for an arbitrary prism orientation. An absolute refractive index additionally requires .
Prism minimum-deviation index measurement: symmetric ray geometry and an independently calculated deviation minimum
. The original model chooses and at one fixed wavelength. The left panel marks face normals, equal internal angles, apex angle and deviation from the undeflected incident direction. The right panel calculates deviation versus incidence from two applications of Snell's law; the marked minimum is the only orientation used for the index inference. No points are experimental observations. Details.For example, stipulate and . Then and . Substitution recovers . These synthetic values illustrate the geometry; they are not readings for the prism in the laboratory protocol below.
A Michigan State University teaching experiment uses a mercury lamp, slit, collimator, prism table and optical telescope, with a subsidiary angular scale read to one arcminute. First focus the optical telescope on a distant object; with the prism removed, align its crosshair to a narrow vertical slit image, focus the collimator, and establish the direct-beam zero. Insert and centre the prism. Select the green mercury spectral line, tabulated at , then rotate the prism while tracking its image with the optical telescope. Locate the turning point where continued prism rotation reverses the image motion, and read the deviation. Re-establish the minimum separately for each other spectral line. The protocol assumes ; that assumed apex angle is not an independently measured result. [ref]
The Prism Spectrometer. introductory-physics-experiments-physics-252-version-4-0-michigan-state-university; Michigan State University, The Prism Spectrometer, Physics 252 v4.0: pp. 2, 5–9; apparatus, vernier readout, alignment, turning point and wavelength table
For an uncertainty-aware implementation, repeat the turning-point search from both prism-rotation directions, recheck the direct-beam reference, and record individual readings before averaging. Independently determine or include its calibration uncertainty; record the chosen spectral line, sample identity, temperature and surrounding-air conditions. These are proposed controls deduced from the inference, not claimed additional instructions or results of that manual. Finite slit width, unresolved spectral lines, imperfect collimation, table tilt, angular calibration and sample nonuniformity can limit the inferred refractive index. One arcminute readout resolution alone does not establish one arcminute total measurement uncertainty. A single selected wavelength gives one index, not a measured optical dispersion curve.
A short geometric derivation uses and . Reversibility exchanges entrance and exit. At the symmetric stationary path, and ; applying Snell's law gives the displayed index relation. The plotted calculation checks that this stationary point is a minimum on the chosen transmitting branch. To assess local sensitivity, differentiate the inferred index using angles in radians:This first-order relation exposes both angular contributions; it is not an instruction to add signed systematic errors as independent random errors. In the synthetic example, a deviation error of arcminute at fixed changes the inferred relative refractive index by approximately .
Proficiency testing 2026-11-05
Proficiency testing evaluates participants' performance against established criteria through an interlaboratory comparison. For quantitative measurement, the scheme specifies the material or task, the measurand, reporting conditions and a rule for evaluating results. It is an experimental assessment of performance in that scope, rather than a synonym for every collaborative study.[ref]
Selection, Use and Interpretation of Proficiency Testing (PT) Schemes by Brian Brookman and Ian Mann. §3, printed p. 4, definition; §§4.1–4.2, role and scheme types
A typical round distributes comparable proficiency-test items to laboratories, which analyse them using their routine measurement procedures. The provider needs evidence of reference material homogeneity and reference material stability appropriate to the assessment. Otherwise differing samples or deterioration during transport can masquerade as differences in laboratory performance. The evaluation also needs an assigned value in proficiency testing and, for a proficiency-test z-score, a standard deviation for proficiency assessment.[ref]
Selection, Use and Interpretation of Proficiency Testing (PT) Schemes by Brian Brookman and Ian Mann. §§7.2.1–7.2.4 and Annex D, assigned values, uncertainty and scoring
An original two-level experiment distinguishes possible response patterns. Suppose independent assigned values are and in the same units. One laboratory reports and ; another reports and . The first pattern fits a multiplicative factor of 1.10 at both levels, whereas the second fits an additive offset of 1. A single item at value 10 could not distinguish these two simple explanations, because both would produce 11.
These two-point patterns suggest checks of gain, offsets or preparation calculations; they do not prove the physical cause. Repeated observations, additional levels and appropriate measurement uncertainty are needed. If the scheme uses assessment scale 0.5 at each level, the first pattern has scores and the second , illustrating how the experimental design and scoring rule work together.
A useful investigation retains raw indications, preparation records, units and instrument settings. Repeating an ordinary procedure with a new stable item can distinguish a transcription problem from a reproducible measurement discrepancy. Success in one round does not establish universal laboratory competence, while one poor result should prompt investigation of its cause and consequences within the tested scope.[ref]
Selection, Use and Interpretation of Proficiency Testing (PT) Schemes by Brian Brookman and Ian Mann. §8.2, laboratory evaluation; §8.3, investigating unsatisfactory performance
Proficiency-test z-score 2026-11-05
A proficiency-test z-score expresses a laboratory result's deviation from the assigned value in proficiency testing in units of the scheme's standard deviation for proficiency assessment:The score is dimensionless because numerator and denominator describe the same kind of quantity in the same units. Its sign gives the direction of the deviation. The denominator is an assessment scale, not automatically the participant's own measurement uncertainty.[ref]
For an original resistance example, take , and a reported value . The score is . Under the guide's commonly used interpretation, is satisfactory, raises a warning, and raises an action signal. Thus this example is questionable; a result of gives and lies exactly at the satisfactory boundary, whereas gives . The scheme must state its interpretation.[ref]
Selection, Use and Interpretation of Proficiency Testing (PT) Schemes by Brian Brookman and Ian Mann. §7.2.4, printed p. 17, score interpretation
The word z does not by itself make this a hypothesis test with a universally fixed error probability. If scores actually followed a standard normal distribution, the stated thresholds would have the corresponding normal-tail probabilities. That distribution is an additional model assumption; the numerical scoring rule alone does not establish it.
Uncertainty in the benchmark can change the comparison. The proficiency-test z-prime score usesWith the same deviation and , an assigned-value standard uncertainty changes the denominator to and the modified score to 2. A smaller modified score may reflect a less certain benchmark, not improved laboratory measurement.[ref]
Selection, Use and Interpretation of Proficiency Testing (PT) Schemes by Brian Brookman and Ian Mann. §7.2.4(b), printed pp. 17–18, modified score and caution
A score should therefore travel with the assigned-value method, assessment scale and relevant uncertainty information. Comparing scores from schemes with different criteria can confuse a change in the benchmark with a change in performance.
Radiation-pressure laser power measurement 2026-10-14
A 2013 experiment by Paul A. Williams and collaborators mounted a 25-mm dielectric mirror on a commercial force-restoration balance. Removing its pan-weight compensation spring allowed horizontal force sensing with a horizontal laser path. They measured mirror optical reflectance and varied incidence from about to . A 1071-nm laser, capable of 530 W, was switched at 0.1 Hz. Digitized mass-equivalent readings were converted to force using ; quadrature demodulation isolated the switched signal from drift. [ref]
Measurement calibration used a 20-g mass, with checks down to 1 mg. The optical forces lay below that checked range, limiting the accuracy claim. An angle scan tested the predicted cosine dependence at fixed intercepted power. Replacing the mirror with a carbon nanotube coated absorber increased absorbed power while reducing the predicted optical force; no modulation peak was detected. This bounded gas-heating artifacts under those conditions, rather than proving their universal absence. The fitted slope was about 7% below . Inferring laser power requires calibrated force, optical reflectance, radiometric absorptance and angle of incidence, with measurement uncertainty for each. [ref]
Randall P. Wagner 2026-11-05
Randall P. Wagner is a physicist at the US National Institute of Standards and Technology whose research includes acoustic measurement. In 2011, Wagner and colleagues described improved microphone reciprocity calibration for laboratory-standard microphones.
The method measures pairs of microphones coupled through air-filled cavities, using reciprocity to determine pressure sensitivity. Comparing long and short coupling cavities tests corrections rather than assuming one apparatus geometry is exact. The reported service covered 31.5 hertz to 20 kilohertz and reduced stated measurement uncertainty relative to the previous service. This supports a traceable conversion from electric potential difference to sound pressure.[ref]
Institutional profile: www.nist.gov/people/randall-p-wagner.
Range-overlap calibration 2026-11-05
Range-overlap calibration establishes or checks the relationship between instrument settings using inputs that both can measure validly. In gain ranging, it tests whether changing sensitivity preserves the inferred measurand. It can determine relative scale and offset between ranges, but agreement between ranges alone does not establish their absolute accuracy or independent metrological traceability.
For an original two-range measurement model, let the same stable electric current produceEliminating givesAn overlap experiment therefore needs enough distinct, stable input levels to distinguish relative gain from offset. One common input gives one equation for and and cannot identify both without additional information.
For example, suppose , , and . Currents of , and give output pairs , and . They satisfy . Dividing raw outputs by their nominal gains without removing the offsets would not give consistent current estimates.
A practical test can drive a switchable transimpedance amplifier from a stable current source, record multiple overlap levels, and alternate the setting order to reveal instrumental drift. Avoid amplifier clipping, allow settling time, and retain the bandwidth and temperature conditions. The manufacturer tables show why range-dependent noise and gain drift must be included rather than assumed equal.[ref]
Datasheet DLPCA-200 by FEMTO. p. 2, gain accuracy, gain drift, input limits and gain-dependent performance
At each level, compare corrected estimates and through their difference . The measurement uncertainty of that difference depends on covariance:A common reference can move both estimates together and disappear from this difference. Thus good overlap agreement checks a relative relationship; it cannot by itself exclude a shared calibration error. The general uncertainty-combination law explicitly retains correlated contributions.[ref]