Serial-section electron microscopy reconstructs a volume from an ordered image stack of thin physical slices. In the transmission electron microscopy implementation, an ultramicrotome cuts successive slices, which are collected on a support and imaged through the specimen. Image registration aligns corresponding features; image segmentation identifies structures to follow between slices. JEOL's glossary describes sections of 40–100 nm and illustrates the separate registration and segmentation stages.[ref]
The motivation is geometrical: one isolated slice cannot show whether a feature continues, divides, or ends outside that slice. An ordered image stack provides observations along the third coordinate, although finite section-thickness anisotropy and missing material limit what can be inferred. A change between neighboring images may come from specimen structure, a slice deformation, or imperfect image registration.
For a chosen numerical example, suppose each of consecutive sections has uniform thickness , with lateral pixel pitch . Ignoring lost material and thickness variation, the sections cover
The voxel sampling is 12.5 times coarser along the cutting direction. These are chosen dimensions, not measurements from either cited experiment. The first and last section centers are only apart: center-to-center extent differs from total represented thickness. Neither spacing is automatically the achieved spatial resolution.
Image resampling onto a cubic grid can make an image stack easier to display, but interpolation supplies no additional independent observations. Likewise, smooth image registration cannot resolve missing-section ambiguity in volume reconstruction: a small connecting feature could lie wholly inside lost material. Checking a reconstructed connection therefore needs evidence in the acquired slices, not just a visually smooth rendered surface.
A historically important automation change illustrates the distinction between section and surface imaging. In 2004, Winfried Denk and Heinz Horstmann placed an ultramicrotome with a diamond knife inside a scanning electron microscope. Their serial block-face scanning electron microscopy imaged the exposed block surface between cuts instead of collecting the removed sections. They reported stacks several hundred layers deep, 50–70 nm sectioning, and typically less than 10 nm lateral position jitter. These observations tested an apparatus for acquiring closely aligned serial images; they did not remove the finite spacing between observations.[ref]
A mechanical detail makes the measurement chain concrete. Their specimen-advance lever reduced the motorized micrometer's motion by 0.11. Thus its stated position measurement uncertainty of 50 nm contributed after that reduction; other sources of section-thickness error still required assessment.[ref]
Sharron Anderson 2026-11-05
Sharron Anderson is an author of research on analytical sampling and laboratory measurement of food. The primary byline of the2002 collaborative sampling study identifies her with Michael Thompson, Paul Willetts, Paul Brereton and Roger Wood.[ref]
The team had multiple samplers collect portions of wheat and green coffee from bulk material. Its findings separated variation under different collection conditions, showing that the variability observed for one food could not simply be assumed for the other.[ref]
An original example illustrates the experimental logic. Repeated readings of one extract might give10.0,10.1 and9.9 concentration units, while a second independently collected portion gives14.0,14.1 and13.9. Each extract is measured consistently, but the collection-to-collection difference is much larger. A study replicating only instrument readings would miss that component of measurement uncertainty.
Sharron Anderson's documented coauthorship concerns this joint experimental investigation. The scientific lesson is to replicate the stage whose variation is at issue: additional measurements of the same material cannot substitute for independent collections when assessing analytical sampling uncertainty.
Siegert relation 2026-10-18
For one selected polarization of waves, write the detected complex fields as and their light intensities as , absorbing a common conversion constant. With stationary ensemble averages and nonzero , define
Here second-order optical coherence concerns distinct detector channels, with the corresponding normal ordering in the quantum formulation. Independent detection has . Subtracting the means gives the normalized covariance
Thus multiplying raw photocurrents contains a large mean-product contribution; multiplying their fluctuations isolates the small excess. Independent detector shot noise does not supply a mean cross-channel excess, although it increases measurement uncertainty.
The displayed Siegert relation requires a zero-mean, jointly circular complex Gaussian random field, the usual chaotic thermal light model. Circular means invariance under a common arbitrary complex phase rotation; in particular . Apply the Gaussian moment pairing theorem to the real and imaginary parts:
The remaining pairing vanishes by circularity. Division proves the Siegert relation. Gaussianity is an assumption about fourth moments, not a consequence of merely measuring first-order optical coherence; a coherent state or photon antibunching cannot be interpreted by this thermal formula.
For equal-intensity beams combined in a common optical mode with matched wave polarizations, interferometric visibility is . Consequently ideal thermal stellar intensity interferometry measures , while the full includes the independent-count reference one. Unpolarized thermal light with two independent equally populated wave polarizations has half this excess: adding their covariances doubles the numerator while multiplying total means quadruples the denominator. Finite detector response further suppresses the peak and must be calibrated.[ref]
SI prefix 2026-11-05
An SI prefix supplies an exact decimal factor when attached to a unit of measurement. For example, milli means and micro means ; a millimetre and a micrometre are correspondingly scaled metres. The prefix symbol and unit symbol form one unit that can itself be raised to a power.[ref]
For an original conversion example,
The area conversion squares the factor as well as the unit. Applying just one factor of to that area would give , a value 1000 times too large. Scientific notation makes the power explicit.
Prefix symbols are case-sensitive and cannot be stacked. For mass, prefixes attach to gram, because the base-unit name kilogram already contains a prefix. SI prefixes denote powers of ten, even in a computing context; they do not silently change into powers of two.[ref] An exact unit conversion scales any stated measurement uncertainty with the value; it does not make an uncertain measurement exact.
A calibration mark can be tested against an independent mass measurement. The Buffalo State-hosted Laboratory Experiment 1. Calibration of Volumetric Glassware uses a pipette, filler, conical flask, weighing bottle, balance, and thermometer. It specifies clean, freely draining glassware, bubble-free filling, an eye-level liquid meniscus reading, gravity drainage, and retaining the residual liquid allowed for by the pipette's calibration. Degassed water is delivered to a tared bottle; ten weighings and initial/final temperature readings give repeated delivered volumes using the water mass density. Their mean and standard deviation assess delivery and repeatability. The balance and thermometer must themselves be checked before claiming accuracy. The same handout describes weighing an empty volumetric flask and then weighing it filled to its mark.[ref]
The independent inference is . Agreement between repeats constrains scatter; agreement with the nominal delivery additionally needs trustworthy mass density, temperature, and balance readings. This experiment tests delivered or contained volume under its conditions. It does not prove that a subsequent solute transfer is complete or a prepared chemical solution is homogeneous.
The University of Wisconsin–Madison teaching module supplies the complementary preparation demonstration. Dissolved solute and vessel rinsings reach the volumetric flask; the contents reach room temperature before a dropper adjusts the eye-level liquid meniscus to the calibration mark. The module excludes a stirring bar during filling because its displaced volume is absent from the calibration, and requires mixing the neck liquid with the bulk chemical solution. Its self-check identifies incomplete dissolution and transfer as a cause of lower-than-intended molar concentration.[ref][ref][ref]
The observable endpoints are a liquid boundary at a mark and a delivered mass, not a direct count of solute particles. An overfilled flask has excessive final volume; removing some well-mixed chemical solution also removes solute, so it cannot recover the intended inventory merely by restoring the mark. Measurement uncertainty, contamination, incomplete transfer, and sampling before full mixing limit the inferred molar concentration. The apparatus demonstrates why the words “dilute to” refer to final chemical solution volume.
A spectroscopic response factor relates a specified measured signal in spectroscopy to an amount of material under fixed measurement conditions. In the convention
is the signal after blank correction, is the amount of substance, and has units of signal per amount. Specify whether the signal is a peak height or an integrated feature, and retain the same acquisition conditions during measurement calibration and unknown-sample measurement. Some literature instead defines a response factor as amount per signal: it is the reciprocal of in this simple model.[ref]
For an original calibration, two standards with amounts and give raw integrated signals 25 and 45 in fixed instrument units. Fitting gives
An unknown with raw signal 35 therefore has
Ratios of uncorrected signal to amount would give 2.5 and 2.25 for the standards, falsely suggesting a changing response. Here the cause is an additive background, not a changing .
Two standards fit this illustrative straight line exactly. Additional standards and repeats are needed to test the useful linear range. The inferred amount inherits measurement uncertainty in the signal, background and response factor. Samples calibrated using the same uncertain factor can have correlated inferred amounts; repeating the unknown measurement does not independently average away that shared calibration contribution.[ref]
Standard uncertainty Created 2026-10-14 Updated 2026-11-05
A standard uncertainty expresses a component of measurement uncertainty as an estimated standard deviation. It has the same unit of measurement as the measurand: describes a spread in a length, rather than a percentage or a known measurement error. A Type A evaluation of uncertainty uses statistical analysis of observations; a Type B evaluation of uncertainty uses other information, such as a reference certificate or instrument specification. These classify the evaluation method, not whether a physical effect is random or systematic.[ref]
For independent repeated readings with common variance, the sample mean has estimated standard uncertainty , where is their sample standard deviation. This is the standard error of the mean; the scatter instead describes individual readings. Repetition reduces this contribution, but a shared measurement calibration contribution remains. A quoted tolerance supplies a different example: assigning a continuous uniform distribution within those bounds gives . That assignment is a model of the available information, not a consequence of repeating measurements.[ref]
Stephen L. R. Ellison 2026-11-05
Steven L. R. Ellison is the name printed for an analytical chemistry coauthor of the recovery-information report, with an affiliation at the Laboratory of the Government Chemist in Teddington. Eurachem's later organizer biography identifies the same report among the works of Stephen L. R. Ellison, an LGC Fellow specializing in analytical statistics and measurement uncertainty.[ref][ref]
The Eurachem biography also identifies him as a principal author of Quantifying Uncertainty in Analytical Measurement and a coauthor of method-validation and laboratory-proficiency-testing guidance. This work develops the statistical interpretation needed to compare analytical results under stated conditions.[ref]
For an original illustration, two independent uncertainty contributions of and combine to in quadrature. If their correlation coefficient is 0.5 and both enter the model with positive sensitivity, the combined standard uncertainty is instead
The model determines which inputs contribute and whether covariance matters. Listing more decimal places does not resolve an omitted source of uncertainty. This calculation illustrates the subject of the guidance; it is not experimental data attributed to Ellison.
Steve R. Domen 2026-11-05
Steve R. Domen developed physical water calorimetry methods at the National Institute of Standards and Technology. His 1994 sealed-water instrument measured absorbed dose from a minute radiation-induced temperature rise, while investigating how dissolved gases change the fraction of deposited energy that appears as heat.[ref]
The apparatus enclosed water in a thin glass vessel immersed in a larger water tank. Two calibrated thermistors in glass capillaries formed part of a Wheatstone bridge that detected the temperature response. The vessel also limited convection. Comparing different dissolved-gas conditions tested the heat defect, rather than assuming that all absorbed energy immediately becomes heat.[ref]
For an original ideal calculation, neglect the heat defect and heat transfer during a reading. With water specific heat capacity and measured rise ,
This illustrates why sub-millikelvin thermometry matters. It is not a reported run: the real measurement model must include the instrument's corrections and measurement uncertainty.
Consider an original ideal model: of hydrochloric acid at , titrated with sodium hydroxide at . Treat these numbers as exact model inputs. Assume complete dissociation of an ionic compound, complete strong acid ionization, additive volumes, uniform mixing, and activity coefficients equal to one. Use stipulated , approximately appropriate near . These are calculated pH values, not measurements of these materials.
The chemical equation for neutralization is
Initial acid inventory is , so the equivalence point is . Added water does not change this inventory. But the final volume , including the titrant, determines all concentrations.
Keep self-ionization of water even at the equivalence point. With , define the dimensionless acid excess , hydronium level , and hydroxide level by
The sodium cation and chloride anion inventories fix the electric charge balance , while chemical equilibrium gives . Eliminating yields a quadratic equation with exactly one positive root. For stable calculation on either side, write
The second expression follows by multiplying by ; it avoids subtracting two nearly equal large quantities on the alkaline side. A logarithm acts on the dimensionless , as required by the pH definition.
Figure 1. . Original analytic model for of hydrochloric acid and sodium hydroxide, with . Both panels use the complete electric charge balance and ideal chemical equilibrium. The right panel resolves microlitre additions around the equivalence point. The marked points are calculated, not measured. Details.
At , the excess is in , giving and pH . At , excess hydroxide is in , giving and pH . Neglecting dilution would give the wrong values even far from the equivalence point.
Exactly at , and , giving pH in this model. At and , the same smooth expression gives pH and . The familiar excess-only formulas would predict zero hydronium or zero hydroxide at the equivalence point; retaining self-ionization of water removes that failure. A steep curve is still continuous.
The sensitivity is especially revealing. Differentiating gives
when . At this model's equivalence point, the derivative is about pH units per millilitre. This is a local derivative, not a linear prediction over a whole drop. Real endpoint accuracy also depends on dosing, mixing, measurement uncertainty, temperature, extra reacting chemical species, and nonideal thermodynamic activities.
A substitution measurement method compares an unknown with a reference by placing them successively in the same relevant position of a measuring system. The comparison seeks a match or measures a difference while retaining as much of the system response as possible. The VIM lists substitution among measurement methods; the detailed sequence belongs to the measurement procedure.[ref]
For a physical application, NIST's double substitution weighing procedure compares a standard and an unknown of the same nominal mass twice, using a weighing instrument as a comparator. It also determines response with a known sensitivity mass and includes measurement-assurance checks.[ref] The measurement standard supplies the reference value; the comparator need not establish the whole mass independently from its own zero.
An original simplified drift model shows why sequence matters. Let a comparator indication at equally spaced time index be
with fixed response coefficient , additive offset and linear instrumental drift . Observe standard S at , unknown X at , and S again at . The mean X indication and mean S indication have the same mean time , giving
Both the constant offset and the stipulated linear drift cancel. For , , and , an unknown gives indications . Their paired means differ by two, recovering .
This teaching sequence isolates one cancellation mechanism; it is not the full NIST procedure. Nonlinear drift, unequal timing, changing response or different positioning can defeat it. NIST's procedure separately addresses buoyancy and uncertainty contributions.[ref] A close match does not remove the reference's measurement uncertainty or establish that the substitution left every relevant condition unchanged.
Systematic error Created 2026-10-14 Updated 2026-11-05
A systematic error is a component of measurement error that remains constant or changes predictably when measurements are repeated under specified conditions. A ruler with an incorrect scale, a thermometer with a measurement zero offset, and an unaccounted instrumental drift can all produce such effects. Closely agreeing readings can therefore have poor measurement accuracy: repetition tests their scatter, not the correctness of the shared reference.
An estimated systematic error should be corrected when possible, with the correction's measurement uncertainty included in the measurement uncertainty budget. Suppose a balance reads and a reference gives an estimate of its offset. The corrected result is ; uncertainty in remains even if the same unknown offset affected every weighing. Changing the reference, using a blank, or testing systematic-error cancellation by reversal can reveal effects that a long sequence of identical readings misses. The Type A evaluation of uncertainty and Type B evaluation of uncertainty distinction concerns how uncertainty is evaluated, so neither label is synonymous with systematic error.[ref]
Tangent stiffness 2026-11-05
The tangent stiffness at a point on a differentiable force-displacement curve is its local slope,
It governs the first-order response to a small displacement change:
The graphical interpretation is the slope of the tangent line. Structural stiffness measured from force-deflection behaviour includes geometry as well as material response.[ref]
For the illustrative model , choose and . At , the tangent stiffness is . An additional therefore predicts a force increase of to first order. Direct substitution gives ; the difference is the higher-order contribution omitted by the tangent approximation. The secant stiffness from the origin is only at this point.
In a displacement-controlled experiment, one can estimate the local slope with a central finite difference,
Making smaller reduces the approximation error for a smooth exact curve, but measured differences can become dominated by measurement uncertainty. The force channel, displacement channel and loading branch must therefore be specified. Crosshead movement can include machine mechanical compliance, while an extensometer records a selected specimen extension.[ref]
For a curve with hysteresis, loading and unloading can have different local slopes at the same displacement. At a sharp corner a unique derivative may not exist, and a finite-interval slope should be reported as such.
T. E. Lutz Created 2026-10-20 Updated 2026-11-05
T. E. Lutz and D. H. Kelker analyzed systematic effects in luminosity calibration from noisy stellar parallaxes. Their 1973 treatment assumes a spatial distribution of stars and a model for measurement uncertainty, then examines the inferred absolute magnitudes.
The resulting Lutz-Kelker bias illustrates why an unbiased measurement of one quantity does not automatically yield unbiased inference after selection or a nonlinear transformation. The calculation's population and measurement assumptions are essential to interpreting the proposed corrections.[ref]
A ternary tie-line consistency residual quantifies how far an overall three-component composition is from the mass balance predicted by specified endpoints of a ternary liquid-liquid tie line. It is a geometric consistency check on a proposed two-phase decomposition. A thermodynamic tie line connects coexisting equilibrium compositions; the calculation below checks the lever rule for those endpoints, rather than independently proving that they are equilibrium phases.[ref]
Use mole fraction vectors for the endpoint liquids and for the overall composition, all normalized to sum to one. Put . For a proposed mole-based phase fraction , define
Exact mass balance requires every component of to vanish. Closure gives automatically, so this residual has only two independent components. A zero sum alone does not imply that the component balances hold.
When the phase fraction is unknown and the endpoints are distinct, an unweighted least squares tie-line phase-fraction fit minimizes the squared Euclidean norm of . Differentiating gives
This is an orthogonal projection onto the line through the endpoints. A physically admissible phase fraction also lies in ; if the unconstrained fit lies outside that interval, the closest admissible composition is an endpoint of the line segment. Coincident endpoints do not determine a phase fraction through this formula.
For the hypothetical endpoints and , the composition gives and zero residual. Changing the overall composition to preserves closure but gives
No adjustment of a single phase fraction can eliminate that discrepancy under the stipulated exact endpoints. For experimental use, its significance depends on measurement uncertainty in both the overall composition and endpoint compositions. Unweighted distance is a chosen metric, not an automatic statistical rejection threshold; correlated composition errors require their covariance matrix on the independent composition coordinates. Even a tiny residual establishes only compatibility with this mass balance, not equality of chemical potentials or stability of the proposed liquid-liquid equilibrium.
An electronic comparator registers a pulse only if its input reaches the chosen threshold. For an initially uncharged capacitor driven through an electrical resistance , a short rectangular pulse may end before that happens. The resulting threshold-crossing time of a finite rectangular pulse through an RC load depends on threshold and pulse duration, as well as the time constant ; it is not determined by group delay alone.
Take a positive input electric potential difference for , zero input outside that interval, and output across a capacitance . Include the source's effective series electrical resistance in . The initial condition is . Kirchhoff's voltage law and the capacitor current relation give the first-order linear differential equation
The output is a continuous function and reaches its maximum at the pulse's falling edge:
The input jumps there, but an ideal capacitor's stored electric charge cannot change discontinuously under the finite currents of this model.
Choose a fixed threshold with . Solving the rising exponential function using a natural logarithm gives
If , the output never reaches the threshold. At , it touches the threshold only at its peak; the ideal time spent strictly above threshold is zero. A nonzero high interval in an ideal non-inverting electronic comparator requires . This is RC pulse-width discrimination: different input durations can give a missing pulse or a finite output pulse even when their input heights are equal.
For , solving the falling exponential decay gives
Here is the time above threshold for an RC-filtered rectangular pulse, before any electronic comparator switching dynamics. It is not generally the input duration . Just above , differentiation gives and hence to first order; a half-height threshold gives slope two at the detection boundary.
For an original numerical example, take , , , and , so . A input peaks at only and is rejected. A input peaks at and gives
These are model calculations, not recorded instrument traces.
Figure 1.
Pulse duration determines whether an RC circuit reaches a threshold and how long it stays above it
. Original calculation for and a half-height threshold. The short pulse is rejected, the boundary-duration pulse just touches the threshold, and the longer pulse yields two crossings. The right panel shows the corresponding ideal above-threshold duration. Details.
A threshold near the asymptotic input height is also especially sensitive to noise and threshold measurement uncertainty. Holding the absolute input height and fixed, the threshold timing sensitivity of an RC step response is
For the same , model, a effective threshold perturbation gives approximately of timing change at , but at , by first-order uncertainty propagation. The rectangular pulse must still last long enough to reach the selected threshold; changing together with is a different sensitivity calculation.
A bench test uses a pulse generator and an oscilloscope to record both the applied electric potential difference and the capacitor output. Sweep while keeping and the actual threshold fixed, and allow enough time between pulses for the measured initial output to be negligible at the desired precision. Include probe and comparator-input capacitance in . MIT's laboratory measures RC decay with a capacitor, a resistor and square-wave excitation; that gives a directly measurable component-value time constant.[ref] The pulse-duration sweep proposed here extends that decay measurement to a threshold observation. A real electronic comparator's input offset voltage, hysteresis and finite switching time must be separated from the ideal RC crossing times; a Schmitt trigger uses different rising and falling thresholds.
A tie-line phase-fraction fit estimates how much of each liquid phase is needed to reproduce an overall composition from specified liquid-liquid tie line endpoints. If are endpoint mole fraction vectors and is the mole-based phase fraction of phase , the mass balance predicts
The endpoints must be supplied by equilibrium measurements or a suitable thermodynamic model; fitting does not establish that an arbitrary pair of compositions is an equilibrium tie line.[ref]
For a ternary liquid-liquid tie line, use two independent composition coordinates, such as the first two mole fractions, since the third is fixed by their sum. Let be these two measured coordinates, the corresponding phase- coordinates, the difference between phase- and phase- coordinates, and a positive-definite covariance matrix for . With exact endpoints, generalized least squares minimizes
Differentiation yields the unconstrained estimate
The second expression imposes the physical bounds. It follows because is a convex quadratic when the endpoints differ in the measured coordinates. The ternary tie-line consistency residual then records the remaining failure of the mass balance.
For an original hypothetical example, take endpoint compositions and , with overall measurement . Suppose the first two measured mole fractions have independent standard measurement uncertainties and , so . Then
The predicted composition is , giving residual . The separate first- and second-component balances would suggest and ; the fit combines them according to their stated measurement uncertainties, rather than averaging those two numbers without regard to precision.
With Gaussian composition errors, fixed endpoints and an interior unconstrained fit, the tie-line phase-fraction uncertainty follows directly from the linear estimator:
for this example. Endpoint uncertainty, a fit on a boundary, or a nearly vanishing endpoint difference needs additional treatment. In particular, all three normalized mole fractions must not be treated as independent observations with an invertible three-coordinate covariance matrix: their exact closure creates dependence. A small fitted residual is evidence of consistency with the specified mass balance and error model, not a substitute for testing chemical potential equality or phase stability.
The uncertainty of a gravimetric dilution describes how uncertain mass measurements propagate into the retained stock fraction . Write , where is stock mass and is added diluent mass. The first-order propagation of uncertainty is
The negative covariance coefficient follows from the opposite signs of and . This is the covariance form of the uncertainty law applied to this specific measurement model.[ref]
For an original sequential-weighing example, let the empty vessel reading be , the reading after adding stock , and the final reading . Then , and . Suppose the three readings have independent standard uncertainties of , with other effects neglected. Both net masses have variance , but
Thus , a relative standard uncertainty of about . Omitting this shared-reading covariance would give and underestimate the uncertainty. Equivalently, propagate the three independent readings directly through .
A common multiplicative balance scale error instead cancels exactly from , since multiplying both masses by the same positive number leaves their ratio unchanged. Identify which errors are shared before assigning independent terms. If the desired result is final mass fraction , the stock mass fraction and its measurement uncertainty also belong in the measurement model.
For reference-assisted spectral offset separation, the corrected estimate depends on the measured sample change , reference change and sensitivity ratio . Its measurement uncertainty must include their joint variation, including uncertainty in a fitted sensitivity ratio. Subtracting a reference can reduce shared noise while adding independent noise.
The gradient of the measurement model with respect to is . If is their covariance matrix, first-order uncertainty propagation gives , or
The signs follow from the sensitivity coefficients, rather than a rule that all correlation reduces uncertainty. This is an application of the correlated-input propagation law in the primary GUM.[ref]
For an original independent-input example, take , , , , and . Then and
Treating as exact would give only . A large reference response amplifies uncertainty in its conversion ratio, even when that response is measured precisely.
For mutually independent random inputs, the exact product variance also includes :
where on the right denote the input means. It follows by expanding . This term is omitted by linearization. In the example it changes the standard uncertainty to about .
Experimentally, use paired sample/reference spectra to estimate shared covariance, and independent sensitivity measurements to test the proposed value of . If the same spectra estimate both and the correction, independence cannot simply be assumed. Drift between acquisitions or a nonlinear reference response is a measurement model problem as well as a precision problem.
Unit of time 2026-11-05
A unit of time is a reference duration used to express a time interval. Writing changes the numerical representation, not the elapsed time in physics. A date labels a point on a time scale; a duration measures separation between two such points.
The SI base unit is the second, defined by fixing the unperturbed ground-state hyperfine transition frequency of the caesium-133 atom to exactly . A clock realizes that definition by comparing an oscillator with the atomic response; practical imperfections still create measurement uncertainty.[ref]
A minute is exactly 60 seconds, an hour 3600, and a day used as a duration unit 86400. A calendar day need not coincide with a uniform 86400-second interval on every civil time scale. Specifying the time convention matters when combining clock measurements, astronomical observations and Julian years.