Photodetachment threshold Created 2026-10-14 Updated 2026-10-20
In the published chlorine experiment, negative ions were accelerated and mass-selected, then overlapped with a tunable laser beam. A downstream electric field deflected charged ions; neutral chemical atoms continued to a plate, whose secondary electrons were counted. Scanning the laser frequency produced an onset in this neutral-atom signal. The apparatus therefore inferred an electron affinity from a threshold, rather than measuring the heat of attachment.[ref]
Reversing the laser direction and averaging the two thresholds corrected the first-order Doppler shift. A Fabry–Pérot interferometer and reference spectral lines calibrated the frequency. Finite laser bandwidth, unresolved quantum state splitting and shifts of the calibration lines limited the inference; the authors included these effects in fitting and the error estimate. The level diagram shows only the extracted gap, not a measured signal curve.[ref]
Thorlabs' silicon FDS100 experiment sends a collimated, fibre-coupled laser beam through a beam splitter to the photodiode and a reference optical power meter. Varying reverse bias and electrical load tests linearity. The measured readout approaches saturation when its electric potential difference consumes the bias.[ref] A separate low-signal setup uses a laser, polarizers, an optical chopper, optical attenuators and a lock-in amplifier inside a light-tight enclosure. These are actual apparatus choices, not the parameters of the constructed example above.[ref]
For an independent dark current check, Thorlabs enclosed a photodiode in nested metal boxes to exclude light and suppress interference, used a Keithley 6487 ammeter supplying reverse bias, and monitored temperature. Only one specimen per type was tested; the nearby temperature sensor did not directly contact the semiconductor. Such a dark control measures the baseline but cannot certify spectral responsivity.[ref]
To infer spectral responsivity at a selected wavelength, measure the radiant flux actually reaching the test photodiode, rather than just the reference arm. A reference-arm measurement calibration must include the beam splitter ratio and path losses. With signed readouts in one fixed orientation, subtract a beam-blocked baseline and use
for direct electric current readout or a calibrated linear transimpedance amplifier, respectively. Repeating at several incident powers tests whether this ratio is constant. Blocking only the source estimates ambient light plus electronic baseline; covering the detector excludes ambient light too. Their difference distinguishes optical background from dark current and readout offset. For a broad optical spectrum, the ratio is a spectrum-weighted response, not generally at one wavelength.
For independent Poisson distribution collection, the root mean square shot noise in a one-sided rectangular frequency band of signal bandwidth is
Here is reverse dark current; thermal electrical noise and amplifier noise also contribute.[ref]
Subtracting a mean dark current removes its offset, not its fluctuations: subtracting a constant leaves the variance unchanged. Consequently makes dark current shot noise small relative to signal shot noise; it does not by itself establish high signal-to-noise ratio. With an independently measured baseline, the variance of that estimate must also be propagated. Increasing the measurement signal bandwidth raises the admitted noise. Calibration of wavelength, reference radiant flux, readout gain and collection geometry, together with drift and linearity, determines what measurement uncertainty can be claimed.
Pierre Agostini Created 2026-10-18 Updated 2026-10-31
A physicist researching high harmonic generation and interactions between lasers and matter.[ref]
Point spread function Created 2026-10-05 Updated 2026-10-14
A point spread function is an imaging system's response to an optical point source. Here it means the distribution of image light intensity: an ideal point object becomes a finite spot, often with faint rings, because of diffraction. An unresolved star and a sufficiently small bead emitting fluorescence are practical approximations to an optical point source. Their image tells us how nearby object details will overlap, and reveals blurring from optical aberrations, defocus, and, for an optical telescope, astronomical seeing.
A useful convention separates shape from throughput. At a fixed image plane, let be the point spread function with . Its units are inverse area. A point delivering detected radiant flux at image position produces light intensity ; this point spread function normalization makes the integral over the image equal to . A profile divided by its own peak is convenient for comparing widths, but is a different point spread function normalization and does not by itself compare total radiant flux.
In incoherent imaging, different object points have negligible mutual spatial optical coherence over the measurement, so their image light intensities add. If the response also obeys translation invariance over the chosen field of view, the expected continuous image is the convolution
Here is the object brightness distribution expressed in image-scaled coordinates, with magnification already included, and is constant transmission. For , where is the two-dimensional Dirac delta function, this gives . A changing response across the field of view instead requires a position-dependent integral kernel ; one measured point spread function is then only a local calibration.
In coherent imaging, the superposition principle applies to the complex wave amplitudes. With amplitude spread function , the corresponding ideal model is
Thus generally differs from : coherent contributions produce wave interference terms that depend on their relative wave phases. Their time average vanishes for mutually incoherent emitters, which explains the intensity convolution. Partial optical coherence requires the correlations between object points as well.[ref]
Figure 1.
Ideal circular-pupil point responses and the overlap of two mutually incoherent sources
. Details.
The figure is an original diffraction model, not measured data. All panels have the same total radiant flux: each member of a pair contributes half the single-source radiant flux. Dashed curves show the separate light intensity contributions; the solid curve is their sum. The point spread functions have tails, so even sources satisfying the Rayleigh criterion overlap. As separation decreases, the image changes continuously; there is no abrupt disappearance of all information at one separation.
A motivating measured example is Measurement of point-spread function (PSF) for confocal fluorescence microscopy by InCheon Song and others (2005). A laser, delivered through an optical fiber, excited fluorescent beads on a slide through a , numerical aperture oil-immersion objective. A moving stage scanned the bead; emitted light returned through the objective, an emission filter and a pinhole to a photodiode. The readout was fluorescence light intensity against scanned position. Under the same setup, nominal bead diameters , , and produced progressively broader measured spots. The size comparison controls for the finite object: the measured response is the convolution of bead emission with the effective point spread function, rather than a pure point response. Inverting this convolution model gave a response compared with the theoretical one. The smaller bead gave a weaker signal relative to noise; uncertainty in actual bead diameter also affected the restored width. These are limits of the calibration. In confocal microscopy, excitation and pinhole detection jointly determine the effective point spread function; this experiment does not measure the ideal widefield Airy disk plotted above.[ref]
For a new calibration, useful additional controls are to measure detector background, avoid saturating the photodetector, repeat the scan to check stability, and compare isolated beads at several positions in the field of view. Record wavelength, numerical aperture, pinhole setting, sampling interval and optical focus. These are practical recommendations, not extra measurements claimed for the published experiment.
A 2018 photonic experiment makes the distinction between populations and quantum coherence controllable. A 400-nm laser pumped a crystal to generate paired photons, filtered near 800 nm. One photon triggered collection; the other entered the apparatus. A half-wave plate and polarizing beam splitter selected horizontal photon polarization; another half-wave plate prepared balanced quantum superpositions. Separated polarization paths and diffraction gratings let a spatial light modulator control their frequency-dependent quantum phases and spectral weights. Recombined paths passed through quartz plates whose adjustable thickness controlled polarization–frequency coupling. Quarter- and half-wave plates plus a polarizing beam splitter selected tomography bases; jointly registered photodetector events supplied the readout. Quantum state tomography inferred polarization density matrices at each thickness. The dephasing model preserves diagonal populations while changing off-diagonal quantum coherence. Here the unobserved photon frequency acts as the environment for the polarization qubit. Finite modulator resolution and imperfect spectral shaping limited agreement with the intended dynamics; this experiment does not establish exact preparation or the ideal probabilities above. [ref]
Population inversion Created 2026-10-14 Updated 2026-10-20
A population inversion puts more emitters in the upper than in the lower quantum state of a transition in the nondegenerate two-level model. Let be their number densities. Equal stimulated-transition strengths make the emission rate proportional to and the competing photon absorption rate proportional to , so net optical gain is proportional to . For and , the difference is positive; swapping them makes the medium absorb. Equal populations give zero net stimulated contribution. With degenerate levels, the relevant comparison uses populations per statistical weight of an atomic level, rather than simply .
Why pump through extra levels? Ordinary steady resonant pumping of a nondegenerate two-level system drives both photon absorption and stimulated emission. With equal induced rate and upper-state decay rate , its elementary population equation is
at a stationary population. Increasing this pump saturates towards ; it cannot sustain population inversion in this model. This is a steady rate-equation conclusion, not a universal prohibition on all driven two-level dynamics or all possible laser gain mechanisms. [ref].
In a familiar three-level path, pumping excites a higher level which relaxes into the upper lasing level; the lower lasing level is the ground state, so substantial ground-state depletion is needed. A four-level path also provides rapid emptying of the lower lasing level into a separate ground state. Suitable lifetimes then allow accumulation above and depletion below the lasing transition. Pumping still supplies the energy: extra levels provide a route to population inversion, not free energy. [ref].
A 2013 experiment by Paul A. Williams and collaborators mounted a 25-mm dielectric mirror on a commercial force-restoration balance. Removing its pan-weight compensation spring allowed horizontal force sensing with a horizontal laser path. They measured mirror optical reflectance and varied incidence from about to . A 1071-nm laser, capable of 530 W, was switched at 0.1 Hz. Digitized mass-equivalent readings were converted to force using ; quadrature demodulation isolated the switched signal from drift. [ref]
Measurement calibration used a 20-g mass, with checks down to 1 mg. The optical forces lay below that checked range, limiting the accuracy claim. An angle scan tested the predicted cosine dependence at fixed intercepted power. Replacing the mirror with a carbon nanotube coated absorber increased absorbed power while reducing the predicted optical force; no modulation peak was detected. This bounded gas-heating artifacts under those conditions, rather than proving their universal absence. The fitted slope was about 7% below . Inferring laser power requires calibrated force, optical reflectance, radiometric absorptance and angle of incidence, with measurement uncertainty for each. [ref]
Rainer Weiss Created 2026-10-18 Updated 2026-10-31
Figure 1.
Portrait of Rainer Weiss
. Michael Hauser; required Commons attribution Dhs at English Wikipedia; existing crop by Neegzistuoja and level adjustment by WikiPedant. Creative Commons Attribution-ShareAlike 3.0 Unported (selected from Commons GFDL migration licensing)., via Wikimedia Commons. Source.
Eliminating a common far-detuned intermediate state with single-photon couplings and gives effective Rabi angular frequency of magnitude . The complex laser phases set the coupling phase, and differential AC Stark shifts change the two-photon resonance.
Raman spectroscopy 2026-10-18
Raman spectroscopy measures shifted scattering of light to infer internal energy spacings and the molecular response that makes them visible. In a modern illustrative arrangement, narrowband laser illumination replaces the historical arc, an optical filter rejects the strong excitation line, and a spectrograph resolves the weak shifted spectral lines. The illumination, rejection and spectral dispersion are separate jobs. Increasing rejection can conceal small Raman shifts close to the excitation line.
A proposed measurement should record the excitation wavelength, a dark signal, a blank container and the liquid, then compare calibrated line positions and integrated signals under the same collection conditions. Measure optical filter transmission and detector response before using the Thermal Raman line ratio for temperature. Repeat at lower incident power to look for heating or altered sample response. These are suggested checks, not an assertion that the 1928 apparatus implemented modern calibration procedures.
Its appeal is the conversion of internal molecular energy differences into visible spectral lines: changing the excitation can move the whole pattern while preserving characteristic Raman shifts. Combined with vibrational infrared activity, Raman vibrational activity supplies complementary information about molecular vibration. A line pattern can constrain a material's identity or environment, but overlapping bands, fluorescence backgrounds and the finite spectral resolution element prevent an unqualified identification from one peak alone.
Rayleigh criterion 2026-10-14
The Rayleigh criterion calls two equal, mutually incoherent optical point sources just resolved when one circular-pupil Airy disk's centre coincides with the other's first zero. It is a specified discrimination convention, distinct from full width at half maximum and from precision in estimating one source's position.
For this circular optical pupil, the Rayleigh criterion uses angular separation , or in the paraxial approximation to the focal plane. The point spread functions still overlap.[ref]
For the Airy disk design example with , and , the angular convention gives , or about arcseconds, and the image-plane separation is . Two equally bright, mutually incoherent optical point sources at separations , and times this scale would have image separations , and respectively. These calculated cases are compared in incoherent two-source Airy profile.
The Rayleigh criterion is useful because it ties an easily recognized two-peak convention to a predictable optical scale. It does not assert that all information vanishes below that separation. Inferring two unknown positions also depends on photon shot noise, background, unequal brightness and uncertainty in the point spread function; locating a single isolated source is a different task. A theoretical analysis shows that alternative optical measurements can estimate the separation of two incoherent optical point sources below this scale.[ref]
The mutual optical coherence condition is essential. In incoherent imaging, time-averaged light intensities add. In coherent imaging, wave amplitudes add before squaring, leaving a wave interference term controlled by relative wave phase. Two holes illuminated by the same coherent laser therefore do not implement the equal-incoherent-source convention merely because their individual patterns resemble Airy disks. The coherent single-aperture laboratory test establishes the response of each aperture; it does not establish the pair's mutual incoherence.
R. E. Drullinger Created 2026-10-18 Updated 2026-10-24
Rick Trebino Created 2026-10-20 Updated 2026-10-31
Arthur Leonard Schawlow's 1981 lecture describes an external-cell arrangement: a beam splitter makes opposed beams, an optical chopper interrupts the pump, and a lock-in amplifier reads the probe detector with the chopper reference. Pump-on reduces absorption and raises the detector signal; scanning the laser finds the narrow resonance. Its first application used iodine; pressure and laser jitter limited the resolution.[ref]
Figure 1.
Counterpropagating pump and probe with synchronous transmission readout
. Original functional drawing of the cited external-cell principle. Beam separation is exaggerated; this is not a reconstruction of the pulsed sodium apparatus. Details.
For an original detector model, let be the probe power reaching a photodetector without sample attenuation and let be the sample's probe optical depths. After subtracting a stable detector background, the transmitted powers are
A positive synchronous change means less probe photon absorption, provided incident probe power is stable and scattered pump light is excluded. For small optical-depth changes the fractional signal is approximately . A lock-in amplifier's voltage also depends on gain, reference phase and chopping waveform; it is not automatically this dimensionless ratio.
Useful controls are blocking the pump, detuning far from the feature, reversing or adjusting beam alignment, and repeating the scan at several pump light intensities. Frequency calibration, sample pressure, transit through the beams and laser frequency fluctuations must be assessed separately. Nearly opposed beams leave a residual Doppler shift: for wave vectors , the simultaneous-resonance condition gives . Exact opposition removes this transverse sensitivity in the first-order model.
An actual teaching experiment in LM-0300 Fabry Perot Spectrum Analyser by Walter Luhs, §§3.1.1–3.1.2, prepares a green laser beam by aligning it near and far along a rail, then centring back-reflected light from adjustable curved mirrors. An actuator using piezoelectricity moves one mirror. A screen first displays interference fringes; a focusing optical lens and photodiode then replace the viewing arrangement. The oscilloscope records the photodiode signal and the piezo drive reference on separate channels, using the latter as its trigger. Alignment and mirror separation are adjusted to suppress extra transverse normal modes; source temperature and drive electric current are controlled. Repeated peaks calibrate the scan using the appropriate free spectral range, and their spacing divided by their full width at half maximum estimates finesse.[ref]
This experiment uses a confocal curved-mirror cavity, whose repeated peak families can have spacing in air; it must not be calibrated blindly with the parallel-mirror . Extra transverse normal modes can mimic source features. A finite measured peak width bounds source sharpness only after instrumental broadening is considered; it does not determine the source's intrinsic width by itself.[ref]
For an independent ideal scan calculation, hold a narrow laser wavelength fixed and vary the parallel-mirror gap at normal incidence. Advancing one resonance order requires . If changes uniformly over the measured interval, the time between repeated peaks divided by their full width at half maximum in time equals finesse. The cancellation of scan rate follows from : both peak spacing and peak width are divided by the same . It fails if the scan rate changes appreciably over that interval. A practical control in the University of Washington's motor-driven flat-mirror experiment is to check that successive peaks are equally spaced before using the time axis this way; its handout also warns that table strain and aggressive low-pass filtering alter the trace.[ref]