An RL circuit alternating for equal intervals between zero drive and constant drive has a unique attracting periodic electric current when . Write and . Sampling just before each zero-drive interval gives the recurrence relation . Subtracting its fixed point yields . Thus the limiting upper and lower electric currents are and ; convergence is geometric. This gives a simple exact ripple estimate for a periodically driven inductor.
Permanent magnet 2026-10-18
A permanent magnet retains magnetization after the applied magnetizing magnetic field is removed. Its working magnetic field persists without a driving electric current in a winding.
The persistent current in a superconducting loop experiment prepares a closed cold path, changes the externally linked magnetic flux, and follows the resulting superconducting persistent current through its magnetic field. The essential measurement starts after the inducing magnetic field has stopped changing.
Figure 1.
Preparing a closed superconducting loop and sensing its magnetic field with a compass
. Original schematic of field preparation and magnetic compensation. Shapes, winding counts and distances are illustrative. The magnetic compass senses a static magnetic field; it is not a stationary voltage-reading pickup coil. Details.
Heike Kamerlingh Onnes's 1914 Communication 140c describes a lead coil closed by fused lead connections. Cooling in a magnetic field and removing the inducing magnet prepared the circulation. A magnetic compass sensed it without contacts; a separate copper coil, cooled in liquid air, supplied a measured balancing magnetic field. Reversing orientation and compensation reduced offsets. [ref]
The liquid-helium cryostat maintained low temperature, but refilling temporarily warmed the apparatus and changed the inferred electric current. A later run used a smaller initial electric current. Apparatus deformation and needle-zero drift limited comparison; cutting the closure still left a smaller magnetic signal, warning against assigning every signal to circulation around the whole loop. The report consequently gave an upper bound under a residual Ohm's law hypothesis. [ref]
For interpreting such a measurement, write a locally calibrated magnetic signal as
Here is the coupling and gain, while contains the background magnetic field and readout offset. Only after establishing stable and correcting does a ratio of signals determine a ratio of electric currents. A gradual change in geometry, a drifting background magnetic field, or a change of temperature can imitate or conceal a decay. A pickup coil connected only to a voltage meter obeys Faraday's law of induction and detects a changing linked magnetic flux; its zero steady output cannot demonstrate that the underlying static magnetic field has disappeared. A magnetic compass can respond to that static magnetic field.
Opening the path, reversing the preparation, and repeating with smaller initial electric current are useful controls for separating whole-loop circulation, magnetic background and a superconducting critical current. Keep the temperature and external magnetic field stable during the interval used to infer electrical resistance. A sudden reduction near a superconducting critical current is not evidence for the constant- exponential decay model at smaller electric current. Persistence also does not replace the Field-cooled Meissner experiment: transport around a hole and magnetic expulsion from the material test different properties.
A superconducting persistent current is an electric current circulating in a closed path with superconductivity after the external driving electromotive force has ceased. Its associated magnetic field makes the circulation observable without inserting a measuring instrument into the path. This article concerns macroscopic circulation in superconductivity. A finite experiment establishes persistence over its observation window with a stated measurement error, rather than directly observing an infinite lifetime.
The ideal circuit connection is flux conservation in a superconducting loop. In a fixed loop with self-inductance , a single circulating electric current , and externally imposed linked magnetic flux , the lumped circuit approximation gives
When the electrical resistance vanishes, integrating gives . Removing the external linked magnetic flux therefore changes the circulating electric current; holding it fixed afterwards preserves that electric current. This is a circuit approximation with a specified self-inductance, not a derivation of the microscopic state of superconductivity. For thin paths where the carrier contribution matters, fluxoid and flux quantization in a superconductor supplies the more careful description.
For constant , constant effective electrical resistance , and no changing external linked magnetic flux, the RL circuit instead predicts
This exponential decay follows by separating . A long relaxation time tests extremely small electrical resistance without trying to resolve an equally tiny steady electric potential difference. The stored magnetic energy is , so the same model gives , exactly the Joule heating loss. The finite energy stored in the loop supplies the magnetic readout; the circulating electric current is not a continuously powered source of useful work.
Phenomenological paired-pulse recovery Created 2026-10-14 Updated 2026-11-05
Choose the following original phenomenological mathematical model solely to calculate a paired-pulse readout. It is not experimental data, a fit to the teaching experiment, or a simulation of the Hodgkin-Huxley model. The conditioning action potential defines time zero. Test pulses have fixed width ; denotes their inward electric current amplitude. Impose an absolute neuronal refractory period . No finite test amplitude generates an event for . After that hold, stipulate a required amplitude
with , dimensionless and recovery time constant . Emit a test event exactly when , including equality. For this ideal event readout, assign events to pulse onset and assume the conditioning event is successfully evoked. There is no calculated membrane potential, spike waveform, response latency or stochastic success probability.
Figure 1.
Chosen refractory hold and required test-pulse amplitude
. Original phenomenological paired-pulse recovery rule, not measured data or a channel simulation. The grey region forbids a second event; after , the required electric current falls from toward . Horizontal lines show two chosen test amplitudes; their intersections give different first permitted test delays. The dashed baseline is approached asymptotically. Details.
The exponential decay represents a combined recovery deficit; it is neither the Hodgkin-Huxley model's nor a measured potassium electrical conductance. One such curve cannot separately identify sodium recovery and persistent potassium electrical conductance. A constant pulse-width amplitude criterion also does not establish an equivalent rule for other pulse widths or voltage trajectories.
At the hold boundary, . At ,
at ,
Thus a test fails at because of the imposed hold, fails at because recovery is insufficient for that amplitude, and succeeds at . At , a pulse would instead succeed: failure of the weaker pulse was relative refractoriness.
Figure 2.
Paired pulses and event readout at three chosen delays
. Each row is a separately reset ideal trial with a conditioning pulse assumed to evoke an action potential at time zero and a test pulse. Rectangles are imposed electric current pulses; vertical ticks are abstract events, not voltage recordings. Only the test passes the stipulated recovery rule. The and failures have different causes. Details.
For , solve using the logarithm. The earliest allowed delay is
For this is ; for it is . If , the imposed hold is the only delay restriction. If , no finite delay passes the rule: the required amplitude stays strictly above its asymptotic baseline. A finite experimental “fully recovered” time therefore also needs a tolerance and a readout definition.
Photocurrent Created 2026-10-18 Updated 2026-10-20
Photocurrent is the electric current attributable to illumination. Here denotes its magnitude after background subtraction. For a photodiode without internal multiplication, let be incident radiant flux concentrated at wavelength , and let be external quantum efficiency, including losses before collection. An incident photon has energy , with Planck constant and speed of light . Dividing energy per second by energy per photon gives the incident photon rate . Multiplying by gives the mean collected pair rate ; multiplying by elementary charge gives
The factor is : each collected pair replenishes an electron at one contact and removes an electron at the other. A cross-section of the external wire therefore counts one transferred electron per completed pair, consistent with charge conservation. The electron and electron hole are two internal charge carriers, not two independent transfers through that wire.
Choose an illustrative photodiode model with , and . Using the exact SI values , and gives
A mean fraction does not require each fourth photon to be lost in a repeating pattern. It specifies a mean conversion rate. Nor does mean : quantum efficiency is dimensionless, whereas spectral responsivity compares electric current with power.
Figure 1. . Original model, not a measured device spectrum. Left: constant external quantum efficiency is assumed only on the displayed wavelength interval –. Right: the example at , with photon and collected-pair rates distinguished. Details.
At fixed radiant flux, longer-wavelength photons have less energy, so more arrive per second. This explains the rising straight line when quantum efficiency is held constant. A real photodiode's quantum efficiency varies with wavelength; the plotted assumption cannot be extended past its band gap absorption cutoff. At fixed incident photon rate instead, has no additional factor of .
Photodetector Created 2026-10-05 Updated 2026-11-05
A photodetector converts incident light into a measurable signal, often an electric current or stored electric charge. A photodiode collects carriers produced by absorbed photons; a photomultiplier first releases photoelectrons and then amplifies their number.
A basic characterization measures the output with the beam blocked and with several known incident radiant powers. The slope gives spectral responsivity at the chosen wavelength; the blocked signal reveals offsets and dark current. Quantum efficiency, response speed and noise-equivalent power describe different limitations. Detecting a weak signal requires separating its contribution from both background photons and the detector's own fluctuations.
Photodiode Created 2026-10-05 Updated 2026-10-20
The useful bridge is from light to a measurable electric current: spectral responsivity converts incident radiant flux into photocurrent, while quantum efficiency counts collected signal charge carriers per incident photon. These answer different questions: how much electric current one watt produces, and how many incident photons contribute to the signal. The worked calculation below makes their relationship explicit.
Figure 1.
Photon absorption, junction separation and an external electric current
. Original qualitative p-n junction schematic with an external resistor. Collecting one electron–electron hole pair transfers one elementary charge through the external electrical circuit, rather than twice that amount. Dimensions and paths are schematic. Details.
Thorlabs' silicon FDS100 experiment sends a collimated, fibre-coupled laser beam through a beam splitter to the photodiode and a reference optical power meter. Varying reverse bias and electrical load tests linearity. The measured readout approaches saturation when its electric potential difference consumes the bias.[ref] A separate low-signal setup uses a laser, polarizers, an optical chopper, optical attenuators and a lock-in amplifier inside a light-tight enclosure. These are actual apparatus choices, not the parameters of the constructed example above.[ref]
For an independent dark current check, Thorlabs enclosed a photodiode in nested metal boxes to exclude light and suppress interference, used a Keithley 6487 ammeter supplying reverse bias, and monitored temperature. Only one specimen per type was tested; the nearby temperature sensor did not directly contact the semiconductor. Such a dark control measures the baseline but cannot certify spectral responsivity.[ref]
To infer spectral responsivity at a selected wavelength, measure the radiant flux actually reaching the test photodiode, rather than just the reference arm. A reference-arm measurement calibration must include the beam splitter ratio and path losses. With signed readouts in one fixed orientation, subtract a beam-blocked baseline and use
for direct electric current readout or a calibrated linear transimpedance amplifier, respectively. Repeating at several incident powers tests whether this ratio is constant. Blocking only the source estimates ambient light plus electronic baseline; covering the detector excludes ambient light too. Their difference distinguishes optical background from dark current and readout offset. For a broad optical spectrum, the ratio is a spectrum-weighted response, not generally at one wavelength.
For independent Poisson distribution collection, the root mean square shot noise in a one-sided rectangular frequency band of signal bandwidth is
Here is reverse dark current; thermal electrical noise and amplifier noise also contribute.[ref]
Subtracting a mean dark current removes its offset, not its fluctuations: subtracting a constant leaves the variance unchanged. Consequently makes dark current shot noise small relative to signal shot noise; it does not by itself establish high signal-to-noise ratio. With an independently measured baseline, the variance of that estimate must also be propagated. Increasing the measurement signal bandwidth raises the admitted noise. Calibration of wavelength, reference radiant flux, readout gain and collection geometry, together with drift and linearity, determines what measurement uncertainty can be claimed.
Photometer Created 2026-10-14 Updated 2026-11-05
A photometer measures a visually weighted radiation quantity. A filtered photodiode approximates the photopic luminous efficiency function with its spectral responsivity; calibration establishes how its electric current maps to illuminance or collected luminous flux. NIST's detector-based realization traces the spectral responsivity to a cryogenic radiometer and the aperture area to length measurements. [ref]
Photometric calibration 2026-10-14
NIST Measurement Services: Photometric Calibrations by Yuqin Zong and others describes a concrete experiment (§§3.2–3.3.2): a silicon photodiode, multilayer photometric matching filter, precision aperture, temperature sensor and current-to-voltage converter. A monochromator supplies a small beam; scanning wavelength yields spectral responsivity, and scanning position checks response across the aperture. Finite spectral bandwidth requires correction. The average aperture response converts uniform illumination to illuminance.
On the photometric bench (§§3.4.1, 3.5.3, 3.5.6), align lamp and aperture reference planes, set the lamp's electric current, stabilize, and compare a check lamp with standard photometers. The readout is a voltmeter signal. Subtract the shutter-closed reading (§3.4.4), divide by converter gain to infer electric current, and use calibrated illuminance responsivity to obtain . For a distant source, luminous intensity follows from , with between reference planes.
Controls include repeat lamp operations, distance changes to test the inverse-square law, temperature corrections and gain linearity checks. Spectral mismatch correction is source dependent (§3.4.3). Unknown optical spectra, spatial nonuniformity and incorrect reference planes remain limitations.[ref]
Photovoltaic effect Created 2026-10-18 Updated 2026-10-20
The photovoltaic effect is the production of an electric potential difference or an electric current by photon absorption in a material or device. In a semiconductor, absorbed photons generate electrons and electron holes; asymmetric transport and selective collection can maintain distinct quasi-Fermi levels and deliver electrical power to a load. An equilibrium built-in potential alone supplies no continuous electrical power.
Physical quantity Created 2026-10-13 Updated 2026-11-05
A physical quantity is a property of a phenomenon, body or substance whose magnitude can be expressed by a number and a reference, usually a unit of measurement. For example, the length and the same length have different numerical values but represent the same physical quantity. The numerical value alone is therefore incomplete.
A physical quantity must also be specified physically. A gas's pressure at a stated position and temperature, the mass of a specimen, and a detector's electric current are different measurands even if an instrument happens to display the same number for them. Dimensional analysis constrains which quantities can be added or compared, while a system of units supplies a consistent convention for expressing them.[ref]
In a functional model, a driven piston reduces the chamber's volume. A surrounding cylinder carries the wall stress; a deformable packing or gasket closes a potential leak path around a moving or removable closure. A sample and a separate pressure sensor occupy the chamber. A transmitting liquid can load the sample on all sides; the sensor's signal passes to an external readout. This is a division of experimental functions, not a scale reconstruction of one historical machine.
For a slowly moving frictionless drive piston of effective surface area , external pressure and inward applied force , Newton's second law at mechanical balance gives
Thus on a stipulated effective surface area corresponds ideally to of differential pressure. These are illustrative numbers, not apparatus specifications. The effective surface area is the one associated with the pressure difference, not necessarily an external ram's visible cross section. With resisting friction during inward motion, the balance becomes . During outward motion the sign of the friction contribution reverses. An applied-load reading can therefore show hysteresis even when the chamber's actual pressure is well defined.
A resistance pressure gauge instead supplies an internal signal. In a generic electrical readout, measure an electric potential difference with a known electric current and use Ohm's law to infer electrical resistance . This describes a possible readout principle; the lecture passage cited below identifies the sensor material and calibration landmarks without specifying this wiring.
Planar London screening 2026-10-14
Take a homogeneous material exhibiting superconductivity in the half-space , bounded by vacuum at . Fix the temperature below the superconducting transition temperature, take a constant positive London penetration depth , and assume a local, linear London equation description in a Meissner state. The applied magnetic field is weak enough that it neither nucleates a quantum vortex nor appreciably suppresses the material's superconductivity. Assume translation invariance in , no imposed transport electric current, and a tangential boundary magnetic field . Here is the local surface magnetic flux density, equal to the imposed vacuum magnetic field in this infinite planar geometry; it is not automatically the far-away applied value for a finite object.
For a bounded, decaying interior solution, the London equations, magnetostatic Ampère law and Gauss's law for magnetism give
The solution is
At one London penetration depth the magnetic field is of its surface value; at five it is . There is no finite depth where this exponential is exactly zero. The differential equation excludes the growing exponential through the deep-interior boundary condition. A constant normal component cannot decay in this geometry: Gauss's law for magnetism would require . A perpendicular applied magnetic field on a finite plate therefore requires a different spatial problem.[ref]
Figure 1. . Original analytic profiles, not measured data. Left: the Planar London screening solution, with vacuum at and the material at . Right: Finite-thickness Meissner screening in a slab of thickness , with equal tangential surface magnetic fields. Blue is ; orange is the signed current density . Different horizontal origins are labeled explicitly. Details.
Platinum resistance thermometer Created 2026-10-14 Updated 2026-10-20
A resistance thermometer using platinum as its sensing material. It provides the temperature readout in the saturation-pressure equilibrium measurement.
Transduction turns electrical resistance into a temperature indication. A uniform wire has , with electrical resistivity , length , and cross-sectional area . Stronger atomic vibrations at higher temperature scatter electrons more, increasing platinum's electrical resistivity; impurities and crystallographic defects also contribute.[ref]
A small known electric current and measured electric potential difference give through Ohm's law; a resistance bridge can instead compare the element with a reference resistor. Inverting the calibrated response then gives the sensing element's temperature. This makes electrical readout convenient, but the platinum resistance thermometer lead resistance error, resistance thermometer self-heating, and resistance thermometer immersion test explain why a stable display alone is insufficient. The gas thermometer supplies a contrasting route through an equation of state; the saturation-pressure equilibrium measurement shows why a trustworthy independent temperature readout matters.
A 1998 NIST account describes stirred comparison baths, a reference SPRT with a 30 Hz resistance bridge, and a temperature-controlled reference resistor. The test IPRTs used a current source, a scanner, and an 8½-digit voltmeter. Separate current and voltage leads made each test probe a four-terminal device; readings in both current directions supplied voltage ratios to the reference resistor.[ref]
Its program bracketed test readings with reference SPRT readings and repeated a block if bath drift was excessive. A check SPRT monitored the system; reference water-triple point measurements checked stability. Sensor depths were matched, and unfamiliar probes underwent immersion tests. Ice and water-triple point readings before and after calibration checked changes. Additional points could test the fitted curve. Reported calibration uncertainties excluded transport, subsequent drift, hysteresis, and the user's measurement system. These are controls and limits of that documented apparatus, not an accuracy promise for all platinum resistance thermometers.[ref]
The inference is a pair: reference temperature from the SPRT, test electrical resistance from the electrical readout. Record measuring electric current, wiring, immersion, and the stated reference scale with each pair. A bath controller's setpoint is not itself the calibrated reference temperature. Returning to a reference and checking intermediate points distinguishes a fitted relation from a reproducible calibration over the intended interval.
With two shared leads of electrical resistances , Ohm's law gives
Using the Pt100 local linear model, positive lead electrical resistance appears as a warmer temperature. At true , stipulated leads of each change into , producing : a calculated bias of about . Subtracting a measured loop correction works only while that loop electrical resistance remains appropriate; cable temperature changes can invalidate it.
Figure 1.
Two shared leads include cable voltage drops; four leads sense at the platinum element
. Red paths carry electric current; blue paths connect an ideal high-input-electrical resistance voltmeter. In the two-wire circuit the sensing terminals span both leads and the element. In four-terminal sensing they span the element alone. The stated lead values and temperature bias are illustrative calculations. Source. Details.
Four-terminal sensing separates the electric current and electric potential difference leads. Negligible sensing electric current suppresses sensing-lead drops; force-lead drops lie outside the interval. Real voltmeter loading, connections, and leakage still need checking.[ref]
For a stable offset from the thermoelectric effect, two signed electric potential differences obey and . The current-reversal calculation in four-terminal sensing yields
At stipulated , and , and recover exactly in this model. If the offset or sample temperature changes between readings, cancellation is incomplete. Reversal also leaves Joule heating unchanged because it depends on .
P-n junction Created 2026-10-18 Updated 2026-10-20
A p-n junction joins regions of a semiconductor with opposite majority charge carriers: electrons on one side and electron holes on the other. It provides the junction modeled by the Shockley diode equation.
Near an unbiased p-n junction, diffusion leaves exposed ionized donor impurities and acceptor impurities. Their electric field opposes further majority-carrier transfer. At thermal equilibrium, drift and diffusion balance; there is no net electric current. Forward bias reduces the barrier to carrier injection; reverse bias enlarges it. A heavily doped tunnel diode adds a different transport route, quantum tunnelling, that the ordinary Shockley diode equation omits.
Power factor 2026-11-05
The power factor is the ratio of active power to the product of RMS terminal electric potential difference and electric current:
when the denominator is positive. It measures how effectively those terminal magnitudes support net average transfer. With signed absorbed power it can be negative for net generation; for an absorbing load the usual range is zero to one. The Cauchy-Schwarz inequality applied to the time average of proves .
For one-frequency sinusoids, , so the power factor is . The phase-only quantity for the fundamental is also called displacement power factor. The primary treatments make the RMS convention explicit.[ref][ref]
A distorted current can give another result. For purely sinusoidal supply potential difference and a zero-mean current with fundamental RMS magnitude ,
Only the in-phase fundamental contributes to the average product with that supply waveform. For an original example with fundamental3A in phase and third-harmonic RMS magnitude4A, total RMS current is5A. The displacement power factor is one but total power factor is0.6. Changing the fundamental phase alone would not remove this harmonic-current contribution.
At fixed absorbed active power and supply RMS potential difference, . For an original900W,230V example, changing the factor from0.6 to0.9 changes source current from about6.52 to4.35A. In the same upstream resistance, the corresponding loss becomes of its former value. This is a stipulated circuit calculation, not an assertion about a particular tariff or installation.
Power factor correction addresses the mechanism causing the low ratio. An ideal capacitive branch can offset inductive reactive power at one frequency; it does not automatically correct an arbitrary distorted current. State the waveform conditions and whether the reported number is total or displacement factor when comparing measurements.