Heat-pump calorimetry 2026-10-14
A motivating bench experiment asks whether a hot bath receives more heat than the electrical work supplied to a heat pump. PHYWE's authored operating instructions, section 4, describe a sealed vapor-compression refrigeration apparatus with its cold and hot coils immersed in separate water vessels. The quantitative procedure calls for insulated vessels, stirring to reduce temperature gradients, timed temperature readings, and a wattmeter for the input. It waits about five minutes for operating conditions to settle before measurements and explicitly includes the glass vessel and immersed copper pipe when calculating the heated body's heat capacity. One bath cooling while the other warms makes the origin of the delivered energy observable. [ref]
The following analysis extends that procedure to simultaneous measurements of both finite baths. Weigh the water in each vessel; determine the effective heat capacities and , including immersed hardware, by a separate known electrical-input calibration over the working temperature range. A matched pair of calibrated thermometers placed in stirred water measures bulk and ; keeping probes away from the coils avoids mistaking a local contact temperature for the whole bath. Record the room temperature and synchronise the electrical meter and thermometry. Use short intervals after startup, with no boiling or freezing and small enough temperature changes that constant effective heat capacities are a reasonable approximation. These are proposed measurement controls, rather than reported results from the manual.
Over a common interval , the uncorrected bath estimates and measured electrical input are
Here the wattmeter records real electrical power . The integral handles changing input, and the signs make both estimated useful heat quantities positive during heating and cooling. The effective heat capacity includes water mass times its specific heat capacity plus the response of the vessel and immersed hardware. Without that correction, equal temperature changes need not imply equal heat transfers.
Insulation reduces heat leaks but does not establish an isolated system. A warmer bath can lose heat to the room while a colder bath gains it. As a proposed control, measure pump-off drift at matched bath and room temperatures, with the same stirring and insulation. Writing for outward hot-bath leak power and for inward cold-bath leak power, the bath balances give
Integrate these rates over the same interval before dividing by . For example, in a pump-off hot-bath control, ; in the cold-bath control, . Check that a stopped machine does not create a different leak path from the running one; otherwise use an independent bath calibration and keep that discrepancy as an uncertainty. Repeat at the same starting temperatures to distinguish drift from a change in COP as the baths separate.
Figure 1.
Heat-pump calorimetry: bath temperature slopes, electrical work, and heat-leak corrections
. Original apparatus schematic and stipulated constant-rate calculation, not measured data and not a performance claim for the cited apparatus. Each finite bath has effective heat capacity . The hypothetical machine takes , extracts , and delivers ; hot-bath loss is and cold-bath gain is . Over the hot bath warms by and the cold bath cools by . Correcting the two bath estimates restores the machine's heating COP and cooling COP. Details.
In the illustrated toy model, , , , , , and are stipulated constants for . The slopes are and . Consequently,
The raw heating and cooling ratios are and , whose difference is . Correcting the bath leaks gives and , whose difference is one. The apparent failure of was incomplete measurement of heat, not a violation of the first law of thermodynamics.
Real apparatus can also release motor heat and pipe heat directly to the room, and store internal energy during startup. For a machine boundary that receives electrical work, an additional outward heat transfer gives . The residual therefore tests the boundary and corrections; do not force it to zero. Include any auxiliary electrical work in the input if quoting whole-apparatus COP, and account separately for stirring work deposited in a bath. Ignoring vessel heat capacity, sensor lag or room leaks can bias the inferred result. These limitations follow from extending the measured energy budget, and the manual's explicit hardware correction illustrates why such boundaries matter.
Finally, these vessels are finite bodies whose temperatures change; they are not fixed thermal reservoirs. Their ratio of integrated useful heat to integrated input work is an interval COP. A Carnot's theorem comparison using simultaneous bath temperatures is a local approximation only when the machine settles much faster than the baths change. An exact reversible comparison for changing bath temperatures instead uses the bath entropy changes, for example for constant heat capacities, negligible net machine entropy change and no room leaks. The experiment's beauty is that two thermometers, known heat capacities and an electrical-energy measurement expose both conservation of energy and the practical gap from reversibility.
Helmholtz free energy Created 2026-09-29 Updated 2026-10-14
The Helmholtz free energy is
where is internal energy, also denoted elsewhere here, is absolute temperature, and is entropy. For the closed system of fixed composition and only pressure-volume work described above, spontaneous relaxation at fixed temperature and volume cannot increase . Stable thermodynamic equilibrium is a global minimum over the permitted alternatives, including different thermodynamic phases. A rigid vessel fixes the total volume; it does not require each coexisting thermodynamic phase to retain its own volume.
For a constructed comparison at , suppose one candidate has more internal energy but more entropy than another at the same temperature and volume. Its difference is
Thus the candidate with higher internal energy has lower Helmholtz free energy. This is the practical reason for retaining the entropy term when selecting a state.
For equilibrium states within a smooth branch of fixed composition, the first law of thermodynamics gives , hence
These derivatives extract entropy and pressure; the differential is not an evolution equation for relaxation. If chemical species amounts vary, add , with molar chemical potentials . A closed system can still change chemical species amounts through chemical reactions; the fixed-composition simplification then needs the conserved-inventory constraints of chemical equilibrium. See MIT 5.60, Lecture 14, fundamental equations and chemical potentials.
The minimum principle has a short explanation. At fixed volume, no boundary work occurs, so . A thermal reservoir at positive temperature loses entropy . Consequently
by the Second law of thermodynamics. Here both endpoint states share the imposed temperature. See MIT 5.60, Lecture 12, criteria for spontaneous change.
The microscopic counterpart is in the canonical ensemble, with Boltzmann constant and partition function . The partition function examples compute the balance between internal energy and entropy; the chemical equilibrium example already displays an ideal-mixture Helmholtz free energy minimum.
A practical test compares three transformations involving sodium hydroxide: dissolution of the solid, neutralization of its chemical solution by hydrochloric acid, and reaction of the solid directly with hydrochloric acid. In the ideal compatible-state notation, the hydrogen ion and hydroxide give
The hydroxide cancels, whereas the newly formed water remains. Sodium and chloride spectator ions must still be included in the full material inventory when matching actual chemical solutions.
The teaching protocol Calorimetry: Heat of Neutralization Reaction & Hess's Law, Procedure A–C, uses nested insulated cups, a cardboard lid and a thermometer. It mixes each of hydrochloric acid and sodium hydroxide chemical solutions; separately it adds solid sodium hydroxide to diluted hydrochloric acid, or to water. It records initial temperatures, stirs, reads temperature every , repeats runs, and extrapolates cooling trends to mixing time. Rapid weighing limits moisture uptake. The Introduction neglects vessel heat capacity and room losses; these are approximations, not a measured calibration. The readout tests whether the inferred enthalpies add, rather than directly measuring a state function.[ref]
Matching the chemical equations alone is insufficient. The direct-acid recipe uses hydrochloric acid, whereas the solution-mixing recipe uses . The dissolution run also does not automatically reproduce the water inventory of the mixed-solution run. Exact closure requires matching initial and final chemical solution compositions and adding any necessary mixing or dilution enthalpies. Agreement after approximate corrections is a limited test, rather than proof of exact cancellation for those recipes.
To improve the calorimetry, use calorimeter calibration for the participating hardware, and estimate calorimeter heat-loss correction under comparable conditions. With an approximately constant total receiving heat capacity , common initial temperature, constant external pressure, only expansion work, negligible escaping material and negligible bulk kinetic energy and potential energy changes, the balance is
Here is the chemical solution mass, its specific heat capacity at constant pressure, the hardware heat capacity, and the signed outward heat loss. Different initial temperatures require separate warming terms. Calorimeter calibration and matching chemical solution compositions address different errors.
For an original constructed model, choose and . Neglect outward heat loss and other corrections. The receiving chemical solution and hardware gain , so . If the completed extent of reaction is for a unit-coefficient chemical equation, the inferred reaction enthalpy per mole of that reaction is
These chosen numbers illustrate the signs and normalization; they are not measured sodium hydroxide data and are independent of the abstract example.
Figure 1.
Stoichiometry cancels intermediate states; calorimetry converts warming into signed reaction enthalpy
. Left: original abstract chemical equations use to give and for one mole. Right: an independent chosen heat capacity model gives receiver gain and reaction change . Neither panel reports experimental measurements. Details.
A loss of heat to the room makes the uncorrected warming underestimate the magnitude of a heat-releasing reaction. Incomplete dissolution, incomplete reaction, uncertain amount of substance, or a slow thermometer response can also spoil closure. A useful experimental residual is after all changes have been placed on the same amount and state basis. Compare with the measurement uncertainty, accounting for shared calibration errors; averaging repeated runs reduces scatter without removing a common bias.
Hot-water heating Created 2026-10-31 Updated 2026-11-05
Hot-water heating carries thermal energy in circulating liquid water. The water gives up sensible heat in a radiator or pipe network and returns at a lower temperature. Under approximately constant pressure and specific heat capacity, the transferred power is
where is the mass flow rate.
The radiator's thermal conduction and its exchange with the room through convection and thermal radiation determine how large a temperature difference is needed. A circulation pump supplies mechanical work against fluid resistance, while the main heating source replenishes the lost enthalpy. Measuring inlet and outlet temperatures together with mass flow estimates the delivered thermal power. The circulating water can remain entirely within the liquid phase, making sensible cooling rather than a phase transition the principal transport budget.
I. A. Waitz Created 2026-10-14 Updated 2026-11-05
An engineering educator whose joint teaching text connects the first law of thermodynamics with practical energy bookkeeping. The change in stored internal energy differs from the heat and work transferred during a process. Separating a state function from transfers across a system boundary helps compare different ways of taking the same material between two states.[ref]
Ice-melting calorimetry 2026-10-14
Ice-melting calorimetry measures the fusion enthalpy through conservation of energy, rather than inferring it from the small temperature change of the melting ice itself. In a mixing experiment, warm water and the container cool while added ice melts and the resulting water warms. Their combined energy balance separates fusion from sensible heat.
Weigh the container and stirrer, then weigh the added warm water to obtain . Let the container and water reach a common initial temperature . Prepare ice in a clean ice-water mixture near , rather than assuming freezer ice is already at ; drain and blot its surface immediately before transfer. Add it without splashing, stir gently, and record temperature before addition and until after the last ice disappears. Reweigh to find the added mass . Wisconsin's laboratory guide specifies drying each piece, weighing before and after, and recording the temperature history; it places the initial and final temperatures on opposite sides of the jacket temperature to reduce net environmental exchange. [ref]
For fully melted, initially dry ice at , a common final temperature , constant water specific heat capacity at constant pressure , and apparatus heat capacity initially at ,
Here denotes net heat entering the apparatus from its surroundings. The left side supplies the fusion enthalpy and the sensible heat on the right; the water produced by melting must be warmed too. The balance extends Case Western's mixing demonstration by including the apparatus and environmental terms. If the specific heat capacities vary appreciably, replace their products with the corresponding temperature integrals. [ref]
Determine independently, for example by a warm/cold water mixing balance with no phase transition, including the wetted stirrer and thermometer in the apparatus. Check the thermometer calibration, keep the lid in place between transfers, and monitor the surroundings. Before/after drift and a comparable run without ice help bound ; such corrections assume comparable heat transfer conditions and are not automatically exact. Choosing a jacket temperature between and reduces exchange but does not prove . Ensure all the ice has melted and the contents are well mixed before using the measured : with residual ice, the weighed addition is not the mass that has undergone fusion. Repeat the experiment and report measurement errors in mass, temperature, and the leak correction.
An independent way to see the plateau is David Roundy's Oregon State experiment: heat an ice-water mixture slowly with a waterproof resistor in an insulated cup, stir, and record temperature, heater electric current , electric potential difference and elapsed time. Plot temperature against , recording changes in or . The initial solid mass must exclude accompanying water. For a boundary enclosing the heater as well as the sample, this input is electrical work, so under the constant-pressure assumptions above. Across an ideal isothermal plateau the apparatus has no sensible heat change, giving ; departures from uniform temperature require accounting for the heater and container response. [ref]
Ideal gas law Created 2026-10-14 Updated 2026-10-14
The ideal gas law is the equation of state relating the absolute pressure , volume , particle count , and absolute thermodynamic temperature of a classical ideal gas:
Here is the Boltzmann constant, is the Avogadro constant, is the molar gas constant, and is the amount of substance in moles. The distinct symbol avoids confusing amount of substance with the number density , measured in : the same law can be written . Use absolute pressure, relative to a vacuum, rather than an unconverted gauge pressure. [ref]
The exact SI definitions of the Avogadro constant and Boltzmann constant give
This numerical value is exact; the applicability of the ideal gas model to a real gas is an approximation. [ref] [ref]
For , , and , direct substitution gives
At fixed temperature and particle count, doubling the volume to halves the pressure to approximately : this is Boyle's law. At fixed volume and particle count, doubling the temperature from to doubles the pressure to approximately . These ratios use kelvins, not readings in the degree Celsius scale: is , and is .
Figure 1.
Ideal gas law: pressure versus volume at fixed temperatures, and pressure versus absolute temperature at fixed volume
. Details.
The curved slices show Boyle's law at three temperatures; the straight slice shows proportionality to absolute temperature at one fixed volume. They are relations among equilibrium states, rather than a prescribed process history. Compression could hold the temperature fixed through heat transfer, or raise it during adiabatic compression. To calculate a process's heat or work, supplement the ideal gas law with the process constraint, the first law of thermodynamics, and the relevant internal energy or heat capacity information. The graph's extrapolation toward is mathematical; it does not assert that a real gas remains classical or gaseous there.
For a stellar mixture with fixed mean molecular weight, the ideal gas law can be written , where is mass density and is the stellar gas constant. It assumes negligible interaction corrections and a nondegenerate gas; excitation or ionization can change the particle count and thermodynamic response.
In his 1834 memoir on the motive power of heat, Émile Clapeyron combined gas laws into a relation between pressure, volume, and temperature while analysing heat engines (printed page 164). The equation above uses modern SI notation and absolute thermodynamic temperature; it is not a transcription of his historical temperature convention. [ref]
Intensive and extensive thermodynamic quantities Created 2026-09-24 Updated 2026-10-14
An extensive quantity measures how much of a thermodynamic system there is; an intensive quantity describes its state independently of that amount. The clean comparison is replication: take independent identical systems at the same thermodynamic equilibrium, with the same substance, composition and thermodynamic phase. Combining their accounting boundaries multiplies internal energy , entropy , volume , mass and amount of substance , while leaving temperature , pressure and chemical potential unchanged.[ref]
For copies the ledger is
An integer counts actual copies; a positive real describes the corresponding bulk scaling law. Copies can remain separated: replication need not involve removing a partition, changing a container or performing any work. Distinguish , the amount of substance measured in moles, from , the dimensionless count of specified entities, where is the Avogadro constant in . For example, contains specified entities; tripling it gives and entities. This uses the exact Avogadro constant , which does not triple. In the differential , is a molar chemical potential in ; the coefficient of instead has units of energy per entity.
Figure 1.
Replication multiplies the extensive ledger while preserving the equilibrium state
. Details.
As a direct check, the ideal gas law gives
For a monatomic ideal gas with only translational degrees of freedom active, internal energy of an ideal gas gives , so at unchanged temperature. Here is the molar gas constant. The scaling distinction lets a small sample supply an equation of state for a much larger amount: material state and inventory become separate questions.
Replication does not mean multiplying every linear dimension by : a geometrically similar container then has times the volume. Nor does it mean combining samples at different states. Equal temperature establishes thermal equilibrium but does not imply equal pressure, composition or mass density. Density of a composite system makes this distinction concrete.
Internal energy Created 2026-09-24 Updated 2026-10-18
Internal energy is the energy stored in the microscopic degrees of freedom of a thermodynamic system: random molecular kinetic energy, internal molecular excitations, and interaction potential energy. Its value requires a specified system and an energy reference. In thermodynamic equilibrium, internal energy is a state function; it records the present state, without remembering whether a previous change came from stirring, compression or heat transfer. This makes internal energy the common accounting quantity connecting mechanical experiments to thermal ones.[ref]
Heat and work describe energy crossing the system boundary during a process. They are not separately stored contents of the system. For a closed system whose bulk kinetic energy and external potential energy have unchanged endpoint values, the first law of thermodynamics gives
Here is positive for incoming heat, and is positive for net work done on the system. Two routes between the same equilibrium states have the same , although their heat and work can differ. The infinitesimal notation distinguishes a state function change from path-dependent transfers.[ref]
For a calculated example, take one mole of a classical monatomic gas in the ideal gas regime, at fixed volume, from to . With stipulated molar gas constant , its internal energy change is . A purely thermal route supplies with . An ideal insulated stirring route supplies with , measured between equilibrium endpoints after the motion dies away. The same stored increment has two different transfer histories; neither history defines a separate kind of stored energy.
Figure 1.
Internal energy is stored; heat and work are ways to change it
. Original schematic and calculated ideal gas example with the stipulated parameters above; no experimental data. The two routes end at the same temperature and volume. Microscopic kinetic energy and interaction potential energy belong inside ; whole-body motion and external elevation are accounted for separately under the stated convention. Details.
The familiar internal energy of an ideal gas is a useful special case. Neglect intermolecular interactions, hold the amount of substance and composition fixed, and suppose that independent quadratic terms per particle obey the classical equipartition theorem. Choosing the constant reference contribution as zero gives
Here is absolute temperature, is the molar gas constant, and is the whole sample's heat capacity at constant volume. A monatomic gas has translational terms. A rigid diatomic molecule has when its two rotational terms are active and vibration is frozen; activating one harmonic oscillator vibration adds two terms, one kinetic energy and one potential energy, giving . The counting is of quadratic contributions, so a vibrational coordinate is not just one contribution.[ref]
This constant- formula applies only over a regime with those same active terms, classical translational statistics, nonrelativistic motion and unchanged composition. Quantum freezing, quantum degeneracy, chemical reactions or appreciable intermolecular interactions invalidate that simple counting. More generally, at fixed amount and composition an ideal gas can still have with varying heat capacity at constant volume, so use . A nonzero constant reference adds to without altering .
Constant temperature does not generally mean constant internal energy. At fixed amount and composition, within one smooth thermodynamic phase of a simple compressible system, an equilibrium identity is
The bracket vanishes for the ideal gas law . It need not vanish for a real gas: changing the volume changes the average interaction potential energy, even when the random-motion contribution stays at the same temperature. The identity follows by writing and using a Maxwell relation; it does not require the actual process to be reversible. [ref]
For a Van der Waals equation model with volume per mole , gives . Thus an isothermal expansion from to , at fixed amount of substance , increases internal energy by
Separating attractive particles makes their interaction potential energy less negative. For stipulated , , , , and , the calculated increase for this one-mole sample is , despite . This dilute stable-branch model example is not a measurement of a named gas. [ref]
A phase transition gives another reason. During equilibrium vaporization of a pure substance at fixed pressure, the temperature can stay at its phase coexistence value while the phase fraction changes. Molecular separation changes interaction potential energy.[ref] If the closed system does only pressure-volume work, its latent heat input obeys
Here is enthalpy. A vaporization latent heat therefore is not automatically the internal energy increment: some input supplies expansion work. A fixed temperature reading alone cannot tell how much internal energy changed.
Internal energy of an ideal gas Created 2026-09-24 Updated 2026-10-14
For a fixed amount of substance and composition of classical ideal gas, internal energy depends only on temperature: , with no independent volume dependence. This does not require constant heat capacity at constant volume. Between equilibrium endpoints,
Here is total heat capacity at constant volume, is molar heat capacity, and for sample mass is specific internal energy, whose change is . These endpoint relations apply even when the intervening process is not at fixed volume; they do not say that transferred heat always equals . With constant , the safe statement is . Writing additionally chooses the reference energy; it does not prove that a physical gas remains classical and ideal down to zero temperature. [ref]
A thermodynamic proof outline uses the first law of thermodynamics in equilibrium form, , and the Maxwell relation :
The last step uses the ideal gas law and the molar gas constant , with fixed composition and amount of substance. Microscopically, separated ideal gas particles store no intermolecular interaction energy; their translational kinetic energy and any internal molecular energy are determined by temperature. Negligible intermolecular interactions still allow a molecule to have internal vibrational potential energy.
An original analytical model makes the distinction between an ideal gas and constant heat capacity concrete. Assume classical translation, two fully active rotational terms, and one independent quantum harmonic oscillator per diatomic molecule, with fixed angular frequency . Define the vibrational temperature scale , using the reduced Planck constant and Boltzmann constant . Summing its thermal occupations gives
The constant includes the chosen reference and zero-point energy, whose temperature derivative vanishes. When the vibrational contribution to heat capacity is exponentially small; when it tends to per mole. This model assumes rotations remain active throughout the interval and excludes dissociation, electronic excitation and real gas corrections. Its low- edge is not a model of rotational freezing or of an actual substance at absolute zero.[ref]
Figure 1. . Analytical illustration with active rotations and one quantum harmonic oscillator per molecule. Left: Mayer's relation keeps the capacity difference equal to . Middle: heat capacity ratio falls from toward . Right: the internal energy change from is an integral of varying heat capacity at constant volume; the dashed line suppresses the vibrational contribution. No measured gas curve or material-specific vibrational frequency is implied. Details.
For this stipulated model, heating from to gives . Holding the frozen-vibration value throughout would give , underestimating the required internal energy increase by about of the model's full value. The integral measures all thermal storage activated along the interval. At fixed volume with only heat input, this is also the supplied heat; at fixed pressure, expansion work must be added through Mayer's relation.
Joule's free-expansion experiment tests the independence from volume in a different way: an insulated rigid vessel allows gas into an evacuated compartment, so no external work is delivered and ideally no heat is exchanged. The first law of thermodynamics gives ; positive then predicts equal initial and final equilibrium temperatures for the ideal gas. The apparatus, thermal-capacity controls and limitations of this inference are described in Joule's free-expansion experiment and temperature sensitivity of a free-expansion experiment. This endpoint test differs from the reversible adiabatic expansion of a loaded piston, which delivers work and lowers internal energy.
Inverse-square potential Created 2026-09-24 Updated 2026-10-14
Here inverse-square potential means the potential energy associated with an attractive inverse-square law for a central force: the potential energy itself varies as inverse distance. For and position with ,
The gradient makes the distinction explicit: doubling halves the magnitude of potential energy but quarters the magnitude of force. The neighboring repulsive inverse-square potential instead has potential energy proportional to and an inverse-cube force; the two naming conventions must be distinguished.
For Newton's law of universal gravitation, , where is the gravitational constant. Treat the attracting mass as an isolated fixed point mass, or as a body with spherical symmetry and exterior test particle of constant mass . The spherical shell theorem gives the same exterior Newtonian gravitational potential energy without requiring uniform mass density. Here is measured from the centre, and the formula applies outside the body; it does not describe its interior. The Newtonian gravitational potential per unit mass is .[ref]
The negative sign records binding relative to infinite separation, rather than a negative kinetic energy. Moving from to increases potential energy by
Because the force is a conservative force, this work depends on the endpoints, including for a curved path. Integrating with zero at infinity gives ; adding a constant leaves the force unchanged but shifts the numerical escape energy threshold by that same constant. This is why the reference at infinity must accompany an escape velocity calculation.
For an exact synthetic comparison, stipulate the standard gravitational parameter and a test particle. At , and the inward force has magnitude ; at , and the inward force has magnitude . The increase is of potential energy. These stipulated parameters are an arithmetic example, not observations of a particular body. The escape-energy balance diagram uses the same parameters.
Isothermal compressibility Created 2026-10-14 Updated 2026-10-20
Isothermal compressibility is the compressibility measured at fixed temperature, particle number and composition. A slowly compressed sealed piston in contact with a thermal reservoir can approach this constraint. For an ideal gas, differentiating at fixed temperature gives , hence . A small pressure increase therefore gives . For a finite change, use the whole isothermal process curve instead.
Take one mole of classical monatomic gas, with stipulated molar gas constant , initially at and . The initial volume is . Compressing to gives
The isothermal compression releases heat to the thermal reservoir; the reversible adiabatic compression retains the input work as internal energy. Reversing either curve gives the corresponding expansion, with adiabatic expansion cooling the gas.
Figure 1.
An ideal gas resists reversible adiabatic compression more strongly than isothermal compression
. Left: two pressure-volume curves through the same initial state; arrows point toward compression. Right: the corresponding temperature changes, including cooling during reversible adiabatic expansion. The dotted line marks half the initial volume. One mole of classical monatomic gas, with and the stipulated above. Details.
The equilibrium curvature of Gibbs free energy provides another measurement of isothermal compressibility:
This follows from , with denoting Gibbs free energy. Within a stable single thermodynamic phase, increasing pressure reduces volume and makes this curvature negative. At phase coexistence, changing the phase fraction can change the volume at constant temperature and pressure; the usual finite single-phase isothermal compressibility then need not exist. Near a critical point, it can diverge. [ref]
Isothermal compression Created 2026-10-05 Updated 2026-10-14
An isothermal compression reduces the volume of a thermodynamic system at fixed temperature. For a fixed-composition ideal gas compressed reversibly from to , isothermal ideal-gas work gives
The surroundings supply positive work , and the gas rejects the same amount of heat to the heat bath. Compressing at from to requires approximately of work input and rejects of heat. If that heat transfer cannot keep up, the temperature rises: the process departs from isothermal compression toward adiabatic compression. The finite-rate isothermal piston model explains the required small temperature excess.
Isothermal expansion Created 2026-10-05 Updated 2026-10-14
An isothermal expansion increases the volume of a thermodynamic system at fixed temperature. For a reversible ideal gas path, the gas delivers positive work and absorbs positive heat, with and . The heat bath supplies the expansion energy; without sufficient heat transfer, doing work instead lowers the temperature. This transfer is the hot step of a Carnot cycle, and it is why an expanding gas need not cool when its surroundings continuously replenish its internal energy.
An insulated free expansion into an evacuated compartment is different. In the idealization with no heat exchange, no external work and no container energy change, . Since internal energy of an ideal gas depends only on temperature, its final equilibrium temperature equals its initial value. The transient need not have a uniform temperature or pressure, so equal endpoint temperatures do not make it a reversible isotherm. Do not apply the logarithmic isothermal ideal-gas work formula to this irreversible process. Its entropy still increases by ; entropy change in free expansion evaluates this state function difference using a hypothetical reversible path.[ref]
For a real gas, even equal-temperature equilibrium endpoints need not have equal internal energy. An insulated free expansion can therefore change the temperature, while a controlled isothermal expansion must provide the heat needed for both its internal energy change and its delivered work. The internal energy at constant temperature article gives the Van der Waals equation example.[ref]
A gas syringe makes the distinction between a constant-temperature path and equal-temperature endpoints observable. The Ideal Gas Law Apparatus (TD-8596A), Instruction Manual 012-08420B by PASCO scientific specifies a TD-8596A gas syringe with a mechanical stop, an internal thermistor, an attached pressure sensor and data acquisition. Record the stop volume with the coupling disconnected, then set and seal the coupling. Record pressure and temperature, compress quickly to the stop near , and hold until both readouts settle. The manual gives roughly for settling and about for the thermistor response. Releasing the plunger gives a second transient and a final volume to record.[ref]
The manual describes a pressure overshoot and temperature rise during rapid compression, followed by cooling at the held volume; the settled pressure remains higher than initially. This illustrates approximately adiabatic compression followed by heat rejection, rather than an isothermal compression throughout. Use the settled initial and compressed readings to test Boyle's law, correcting both syringe readings by the unmarked tubing volume :
Use absolute pressure, not gauge pressure. A transient comparison of must pair simultaneous pressure and temperature readings: the sensor lag can separate their peaks.[ref]
For an additional test of the path, repeat with small compression steps, wait for stable temperature and pressure at every step, and repeat in reverse. Keep the sealed amount of substance, connecting tubing and ambient temperature fixed; prevent hand warming of the barrel and check for leaks by watching for continued pressure drift at held volume after thermal equilibrium. These are proposed controls, rather than reported measurements. Agreement of equilibrated points with Boyle's law supports the fixed-temperature ideal gas law approximation, but does not prove that the rapid segments between them were isothermal processes. A single local temperature sensor can miss temperature gradients; plunger friction can produce direction-dependent readings. The pressure and volume data estimate reversible work only insofar as the path approaches a reversible thermodynamic process; they do not directly measure heat. After motion stops, cooling at fixed volume has zero boundary work and , even though the complete equal-endpoint-temperature ideal gas run has .
Isothermal ideal-gas work Created 2026-10-14 Updated 2026-10-18
For a reversible thermodynamic process that changes the volume of a fixed amount of ideal gas at absolute temperature , the work done by the gas is
Here is the amount of substance, the molar gas constant, the pressure, and the natural logarithm. The ideal gas law supplies ; the integral is the signed area under the pressure-volume curve. Isothermal expansion gives positive work; isothermal compression gives negative work. A convention counting work done on the gas instead has .[ref]
For a calculated example, let , and , . The pressure falls from approximately to , and
Traversing the same isothermal process backward changes both transfer signs: . Neither direction changes internal energy of an ideal gas. These stipulated values illustrate the energy balance; they are not experimental measurements.
Figure 1.
Signed work along an ideal-gas isotherm: expansion and compression
. Details.
The shaded geometric area is positive in both panels; the direction of the integral supplies the sign of work. Knowing only the two equilibrium endpoints does not justify shading this curve for an irreversible process. For an idealized piston moving against a specified external resisting pressure, boundary work is . Substitution of the equilibrium gas pressure requires the appropriate reversible limit. In free expansion into vacuum, and this boundary work is zero.
Isothermal process Created 2026-09-29 Updated 2026-10-20
An isothermal process changes a thermodynamic system while keeping its temperature constant. It need not keep its pressure, volume or internal energy constant, and it does not generally have zero heat transfer. An adiabatic process specifies zero heat transfer instead. For example, a slowly expanding ideal gas can absorb heat from a heat bath and deliver the same amount of work while its temperature stays fixed.
For a fixed amount of substance and composition, internal energy of an ideal gas depends only on temperature. With for heat entering and for work delivered to the surroundings, the first law of thermodynamics therefore gives
This equality does not determine either transfer until the path is specified. Isothermal ideal-gas work gives the reversible value; an insulated free expansion has a different value even when its equilibrium endpoints have the same temperature. The closed system convention here excludes changes in bulk kinetic energy and external potential energy.
A conducting cylinder in a large, stirred heat bath approximates an isothermal process when the piston moves slowly enough for thermal conduction to keep up. A quasistatic process keeps the gas close to internal thermodynamic equilibrium; a reversible thermodynamic process additionally requires negligible friction and vanishing driving differences in pressure and temperature. Exact equality of the gas and heat bath temperatures gives no finite heat transfer rate through a finite thermal resistance. The reversible description is a limit of arbitrarily small differences and arbitrarily slow motion, illustrated by the finite-rate isothermal piston model.[ref]
For a real gas, internal energy at constant temperature can change with volume because molecular interaction potential energy changes. In one smooth thermodynamic phase of a fixed-composition simple compressible system,
Thus an isothermal process generally has , with no reason to set . The right-hand side vanishes for the ideal gas law, but not generally for another equation of state.[ref]
James Prescott Joule Created 2026-10-13 Updated 2026-10-31
James Prescott Joule was an English physicist whose private experiments connected electricity, heat and mechanical work. His family ran a brewery in Salford; he developed an experimental laboratory while working in the family business.[ref] The Royal Society's archive identifies his occupation as physicist and records his life as 1818–1889.[ref]
His electrical heating experiments placed metal coils in water and measured the temperature change. Later, he used the brewery cellar to reduce ambient temperature fluctuations during friction experiments and took particular care over thermometer calibration.[ref] These practical choices address a central difficulty of calorimetry: a small temperature rise can reflect both the intended input of energy and unwanted heat transfer to the surroundings. A quiet baseline and a known heat capacity are part of the measurement, not decoration around the formula.
The contrast between Joule heating and Joule's paddle-wheel experiment makes the connection especially clear. One supplies electrical work; the other supplies mechanical work. In each case, calorimetry tracks the resulting change of internal energy. Comparing quantitatively controlled inputs with thermal changes gives an experimental test of conservation of energy, rather than merely showing that stirring or an electric current can warm water. The paddle-wheel account describes the descending weights, vanes, thermal controls and corrections needed for that comparison.
Joule's free-expansion experiment asks a different question: how does a gas's temperature change during expansion without external work? The Joule-Thomson effect instead concerns flow through a restriction and a balance involving enthalpy. Keeping the two experiments distinct prevents a shared historical name from hiding different system boundaries and different conserved quantities.
Jarzynski equality Created 2026-10-14 Updated 2026-11-05
Integrating the Crooks fluctuation theorem over work and using normalization of the reverse probability distribution gives
Consequently Jensen's inequality gives . Some realizations may nevertheless have . This is a work relation with equilibrium endpoint preparations, while the integral fluctuation theorem for total entropy production uses the actual final forward distribution. [ref]